K211995 · Sonendo, Inc. · KIF · Aug 27, 2021 · Dental
Device Facts
Record ID
K211995
Device Name
Sonendo Filling Material 5C
Applicant
Sonendo, Inc.
Product Code
KIF · Dental
Decision Date
Aug 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sonendo Filling Material 5C is intended for permanent obturation of the root canal following root canal treatment.
Device Story
Sonendo Filling Material 5C is a two-part, water-soluble, radiopaque acrylate resin used for permanent root canal obturation. The material is supplied in plastic syringes with a mixer and dispensing tip. It utilizes a chemical curing system to harden within the root canal. The device is intended for use by dental professionals trained in endodontics within a clinical setting. By providing a permanent seal, the material supports the success of root canal therapy. The device is applied directly into the canal; the clinician manages the placement and setting process. No electronic or software components are involved.
Clinical Evidence
No clinical data was provided. Substantial equivalence is supported by bench testing, including biocompatibility testing per ISO 10993-1 and ISO 7405:2018 (cytotoxicity, sensitization, reactivity, toxicity, genotoxicity, implantation, pyrogenicity) and physical/mechanical performance testing per ISO 6876:2012 (flow, working time, setting time, film thickness, solubility, radio-opacity).
Technological Characteristics
Two-part root canal filling resin; chemical curing system. Composition: Poly(ethylene glycol) diacrylate, water, 5-acrylamido-2,4,6-triiodoisophthalic acid, triethanolamine, sodium hydroxide, potassium persulfate. Form factor: Plastic syringes with mixer and dispensing tip. Performance standards: ISO 6876:2012 (flow, working time, setting time, film thickness, solubility, radio-opacity). Biocompatibility standards: ISO 10993-1, ISO 7405:2018.
Indications for Use
Indicated for permanent obturation of the root canal following root canal treatment in patients undergoing endodontic procedures.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 27, 2021
Sonendo, Inc. Steve Ziemba Vice President, Regulatory Affairs & Quality Assurance 26061 Merit Circle, Suite 102 Laguna Hills, California 92653
Re: K211995
Trade/Device Name: Sonendo Filling Material 5C Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: June 24, 2021 Received: June 28, 2021
## Dear Steve Ziemba:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211995
Device Name Sonendo Filling Material 5C
Indications for Use (Describe)
Sonendo Filling Material 5C is intended for permanent obturation of the root canal following root canal treatment.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K211995 510(k) Summary
- 1. Submitter:
Sonendo, Inc. 26061 Merit Circle, Suite 102 Laguna Hills, California 92653
| Contact Person: | Steven L. Ziémba |
|-------------------|------------------|
| Telephone Number: | (949) 667-8151 |
| Fax Number: | (949) 305-5201 |
June 24, 2021 Date Prepared:
- 2. Device:
| Name of Device: | Sonendo Filling Material 5C |
|----------------------|-------------------------------------------------|
| Common Name: | Root Canal Filling Resin |
| Classification Name: | Root Canal Filling Resin, per 21 CFR § 872.3820 |
| Device Class: | II |
| Product Code: | KIF |
- 3. Predicate Device:
| Primary Predicate: | EndoREZ Dual Cure, Ultradent Products, Inc., K071106, Product<br>Code KIF |
|--------------------|--------------------------------------------------------------------------------------|
| Reference Device: | Pulp Canal Sealer EWT, Sybron Dental Specialties, Inc.,<br>K152956, Product Code KIF |
#### 4. Device Description
The Sonendo Filling Material 5C is a two-part filling material developed for permanent root canal filling. It is a water-soluble acrylate which is radiopaque and hardens with a chemical curing system. The Sonendo Filling Material 5C is packaged in plastic syringes with a syringe mixer and a dispensing tip. The syringe and dispensing tips are the only patient contacting devices packaged with the Sonendo Filling Material 5C and are commercially available.
- 5. Statement of Intended Use:
The Sonendo Filling Material 5C is intended for permanent obturation of the root canal following root canal treatment.
- 6. Summary of Technological Characteristics with the Predicate Device
The technological characteristics of the subject Sonendo Filling Material 5C are similar to the predicate device, EndoREZ Dual Cure (K071106). There are no substantial technical or functional differences between the Sonendo Filling Material 5C and the predicate device in
{4}------------------------------------------------
terms of chemical composition, function and intended use. Both are root canal filling resins consisting of a catalyst and base packaged in individual syringes. See Table 1 below for technological characteristics and comparisons of the root canal filling resins.
| Element | Sonendo Filling<br>Material 5C<br>(Subject Device) | EndoREZ Dual Cure<br>(Predicate Device) | Comparison |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Company | Sonendo, Inc. | Ultradent Products, Inc. | N/A |
| 510(k) | K211995 | K071106 | N/A |
| Indications for Use | The Sonendo Filling<br>Material 5C is intended<br>for permanent<br>obturation of the root<br>canal following root<br>canal treatment. | EndoRez is designed to be<br>used with EndoREZ Points<br>and/or gutta percha for the<br>filling of cleaned and shaped<br>root canals. EndoREZ, in<br>conjunction with a master cone<br>and accessory cones (as<br>needed), provides optimum<br>sealing. Although EndoREZ<br>Points are recommended,<br>EndoREZ may be used with all<br>conventional endodontic<br>obturation techniques. | Indicated for same<br>purpose - filling of root<br>canals. |
| Target Users | Dental Professionals<br>trained in endodontics. | Dental Professionals trained in<br>endodontics. | Same |
| Basic Chemical<br>Composition | Poly(ethylene glycol)<br>diacrylate, water, 5-<br>acrylamido-2,4,6-<br>triiodoisophthalic acid,<br>triethanolamine, sodium<br>hydroxide, potassium<br>persulfate | Diurethane dimethacrylate,<br>triethylene glycol<br>dimethacrylate,<br>organophosphine oxide,<br>benzoyl peroxide, radiopaque<br>agent | Similar - The predicate<br>device employs acrylate<br>monomers catalyzed by<br>a peroxide as does the<br>subject device. |
| Material<br>Compatibility | Biocompatibility profile<br>is similar to currently<br>marketed reference<br>dental filling material<br>and demonstrates no<br>increased risk with<br>respect to long-term<br>biological safety. | Meets Biocompatibility<br>Requirements of ISO 10993-1 | Same |
| Element | Sonendo Filling<br>Material 5C<br>(Subject Device) | EndoREZ Dual Cure<br>(Predicate Device) | Comparison |
| Flow<br>(ISO 6876: 2012) | 33.45 mm<br>≥ 17mm | 17.31 mm<br>≥ 17mm | Same - Passed ISO<br>6876 test requirements |
| Working Time<br>(ISO 6876: 2012) | 32.98 mm<br>≥ 17mm | 17.38 mm<br>≥ 17mm | Same - Passed ISO<br>6876 test requirements |
| Setting Time<br>(ISO 6876: 2012) | 60 minutes, 7 seconds<br>Claim of 1 - 24 hours<br>Between 1 - 24 hours | 25 minutes, 46 seconds<br>Claim of 30 minutes<br>No more than 10% longer | Same - Passed ISO<br>6876 test requirements |
| Film Thickness<br>(ISO 6876: 2012) | 5 μm<br>≤ 50 μm | 5 μm<br>≤ 50 μm | Same - Passed ISO<br>6876 test requirements |
| Solubility &<br>Disintegration<br>(ISO 6876:2012) | 2.03%<br>No disintegration<br>≤ 3.0% by mass<br>No Disintegration | 0.77%<br>No disintegration<br>≤ 3.0% by mass<br>No Disintegration | Same - Passed ISO<br>6876 test requirements |
| Radio-opacity<br>(ISO 6876:2012) | 3.10 mm Al<br>≥ 3 mm of Aluminum | 6.6 mm Al<br>≥ 3 mm of Aluminum | Same - Passed ISO<br>6876 test requirements |
Table 1: Comparison of Subject and Predicate Devices
{5}------------------------------------------------
## 7. Performance Data
## Biocompatibility Testing
The biocompatibility evaluation for the Sonendo Filling Material 5C was conducted in accordance with ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry, Annex A, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA. The biocompatibility testing included the following tests:
- 1. Cytotoxicity
- 2. Sensitization
- 3. Intracutaneous Reactivity
{6}------------------------------------------------
- 4. Acute Systemic Toxicity
- 5. Subacute Toxicity
- 6. Genotoxicity
- 7. Intramuscular Implantation
- 8. Material mediated pyrogenicity
The biocompatibility testing conducted demonstrates adequate biocompatibility for the Sonendo Filling Material 5C.
## ISO 6876 Testing
Testing according to ISO 6876:2012 Dentistry - Root canal sealing materials was performed on the Sonendo Filling Material 5C and as compared to the predicate device, it was substantially equivalent to the device and met the physical/mechanical properties of the standard.
### Clinical Studies
No human clinical testing was conducted to support substantial equivalence.
#### 8. Conclusion as to Substantial Equivalence
The similarities in chemical composition, function and intended use of the Sonendo Filling Material 5C with the legally marketed predicate device, EndoREZ Dual Cure (K071106), support substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.