The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
Device Story
Ortho MI System consists of orthodontic anchor screws inserted into jaw or palatal bone; provides fixed anchorage point for orthodontic appliances; facilitates controlled tooth movement; prevents unwanted tooth movement. Used in clinical dental settings by orthodontists. Device is a machined titanium alloy (Ti-6Al-4V ELI) screw; available in various head types (Double, Button, Wing, Slot) and dimensions. Benefits patient by enabling precise orthodontic correction. Output is mechanical stability for appliance attachment.
Clinical Evidence
No clinical data were necessary for the demonstration of substantial equivalence. Bench testing only.
Technological Characteristics
Material: Ti-6Al-4V ELI titanium alloy (ASTM F 136-13). Form: Orthodontic anchor screw with various head types (Double, Button, Wing, Slot). Dimensions: Diameter Ø1.4mm–Ø2.0mm; length 6.0mm–16.0mm. Sterilization: Radiation (sterile) or moist heat (non-sterile). Standards: ISO 19023 (mechanical), ASTM F543 (bone screws), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for use as a fixed anchorage point for orthodontic appliances to facilitate tooth movement; intended for temporary use and removal post-treatment; single use only.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 15, 2021
Osteonic Co., Ltd. Lee Jisun Manager 303Ho, 405Ho, 505-2Ho, 505-3Ho, 902Ho, 1004Ho, 1201Ho 1202Ho, 1206Ho, 38 Digital-ro 29-gil Guro-gu, Seoul 08381 REPUBLIC OF KOREA
Re: K211992
Trade/Device Name: Ortho MI System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: OAT Dated: July 14, 2021 Received: July 19, 2021
Dear Lee Jisun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211992
Device Name ORTHO MI SYSTEM
Indications for Use (Describe)
The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Osteonic, a medical company. The logo features a blue geometric shape resembling a stylized diamond or cube, positioned to the left of the company name. The text "OSTEONIC" is written in a bold, sans-serif font, with the tagline "The Medical Explorer" appearing in a smaller font size beneath it.
# 510(k) Summary
The assigned 510(k) Number: K211992
## 01. Date of Submission: 2021.10.15
#### 02. Applicant
LEE JISUN OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-2Ho, 505-3Ho, 902Ho, 1004Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8411 FAX: +82-2-6902-8401 Email: jisun@osteonic.com
#### 03. Submission Correspondent
LEE JISUN OSTEONIC Co., Ltd. 303Ho, 405Ho, 505-2Ho, 505-3Ho, 902Ho, 1004Ho, 1202Ho, 1206Ho, 38 Digital-ro 29-gil, Guro-gu, Seoul, Korea TEL: +82-2-6902-8411 FAX: +82-2-6902-8401 Email: jisun(@osteonic.com
#### 04. Subject Device Identification
Trade Name: ORTHO MI SYSTEM Common Name: Non sterile, sterile orthodontic anchor screw Classification Name: Implant, Endosseous, Orthodontic Product Code: OAT Panel: Dental Regulation Number: 21 CFR 872.3640 Device Class: II
#### 05. Indication for use
The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
# 06. Predicate devices
Predicate device 510(k) Number: K103105 Device Name: Orthodontic screw Manufacturer: OSSTEM Implant Co., Ltd.
Reference device 510(k) Number: K161335 Device Name: Dual Top Screw System Manufacturer: Jeil Medical Corporation
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for a company called OSTEONIC. The logo consists of a blue geometric shape on the left and the company name "OSTEONIC" in bold, dark font on the right. Below the company name is the tagline "The Medical Explorer" in a smaller, lighter font.
# 07. Device Description
The ORTHO MI SYSTEM is indicated for use as a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and is removed after orthodontic treatment has been completed. It is intended for single use only.
The screw diameters and lengths vary as shown below. It is made of a machined piece of Ti-6Al-4V ELI titanium alloy (ASTM F 136-13).
| Structure (head type) | Thread diameter and Length | Remark |
|-----------------------|-------------------------------|------------------------------------------------|
| Double head | - Ø1.4mm, 6.0/8.0mm | Exceptionally S5T (Double head) and S5R (Slot) |
| Button head | - Ø1.6mm, 6.0/8.0/10.0mm | are Ø1.8mm, 10.0 to 16.0mm per 1.0mm. |
| Wing | - Ø1.8mm, 6.0/8.0/10.0mm | |
| Slot | - Ø2.0mm, 6.0/8.0/10.0/12.0mm | |
These size combinations are the same across all model types (Double Head, Wing, Slot) with exception case.
# 08. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the subject device complies with the following standards:
#### ▪ Mechanical performance
- ISO 19023: 2018 Dentistry Orthodontic anchor screws
- ASTM F543: 2013 Standard specification and test methods for metallic medical bone screws
### ▪ Sterilization, shelf-life and packaging for sterile product
- ISO11137-1:2006/Amd.2:2018 Sterilization of health care products - Radiation - Part 1: Requirement, validation and routine control of a sterilization process for medical devices [Including: Amendment 1 (2013) and Amendment 2 (2018)]
- ISO11137-2:2013Sterilization of health care products -Requirements for validation and routine control - Radiation -Part 2: Establishing the sterilization dose
- ISO11137-3:2017Sterilization of health care products -Requirements for validation and routine control - Radiation -Part 3: Guidance on dosimetric aspects
- ISO11737-1:2018 Sterilization of medical devices - Microbiological methods- Part 1: Estimation of microorganisms on products
- ISO 11737-2:2009 Sterilization of medical devices - Microbiological methods- Part 2. Tests of sterility performed in the validation of a sterilization process
- ISO 11607-1:2006/AMD1:2014 Packaging for terminally sterilized medical devices - part 1: requirements for materials, sterile barrier systems and packaging system
- ISO 11607-2:2006/AMD1:2014 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980:2016 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM F88/F88M:2015 Standard test method for seal strength of flexible barrier materials.
- ASTM F1140/F1140M:2013 Standard test methods for internal pressurization failure resistance of unrestrained packages for medical applications
- ASTM F2096:2011 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- ASTM F1929:2015 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration - ASTM D882:2012 Standard test method for tensile properties of thin plastic sheeting.
- ASTM F1886F1886M:2016 Standard Test Method for Determining Integrity of Seals for flexible packaging by visual inspection
#### · Sterilization for non-sterile product
- ANSI/AAMI ST79:2017 Comprehensive guide to steam sterility assurance in health care facilities - ISO 17665-1:2006 Sterilization of health care products -- moist heat -- part 1: requirement, validation, and routine control of a sterilization process for medical devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Osteonic, a medical company. The logo features a stylized blue geometric shape resembling a stylized letter "O" or a series of interconnected rectangles. To the right of the geometric shape is the company name "OSTEONIC" in bold, dark gray letters. Below the company name is the tagline "The Medical Explorer" in a smaller, lighter font.
- ISO TS 17665-2:2009 Sterilization of health care products - moist heat - part 2: guidance on the application of iso 17665-1.
- ISO 11138-1:2017 Sterilization of health care products - biological indicators - part 1: general requirements
▪ Biological safety assessment
Biological assessment has been performed according to ISO 10993-1:2018, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process," and to the FDA Guidance document, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process', Guidance for Industry and Food and Drug Administration Staff", for the subject devices.
# 09. Clinical Test Summary
No clinical data were necessary for the demonstration of substantial equivalence.
# 10. Substantially Equivalent Conclusion
| | Table 1: Substantial Equivalence Comparison | |
|--|---------------------------------------------|--|
| | | |
| Product Name | SUBJECT Device<br>ORTHO MI SYTEM<br>(K211992) | PREDICATE Device<br>Orthodontic screw (K103105) | REFERENCE Device<br>Dual Top Screw System<br>(K161335) |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | OSTEONIC Co., Ltd. | OSSTEM Implant Co., Ltd. | Jeil Medical Corporation |
| Product code | OAT | OAT | OAT |
| Regulatory<br>class | Class II | Class II | Class II |
| Regulation<br>Number | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 |
| Indication for<br>use | The ORTHO MI SYSTEM is<br>indicated for use as a fixed<br>anchorage point for attachment<br>of orthodontic appliances to<br>facilitate the orthodontic<br>movement of teeth. It is used<br>temporarily and is removed<br>after orthodontic treatment has<br>been completed. It is intended<br>for single use only. | The Orthodontic Screw is<br>indicated for use as a fixed<br>anchorage point for attachment<br>of orthodontic appliances to<br>facilitate the orthodontic<br>movement of teeth. It is used<br>temporarily and is removed<br>after orthodontic treatment has<br>been completed. Screws are<br>intended for single use only. | The Dual Top Screw System is<br>intended for use as a temporary<br>anchor for orthodontic<br>treatment for use in patients<br>aged 12 and older. |
| Operating<br>Principles | Orthodontic anchorage screw is<br>inserted into jaw and palatal to<br>help the orthodontist move the<br>correct teeth and stop the<br>wrong teeth from moving in the<br>wrong direction. | Orthodontic anchorage screw is<br>inserted into jaw and palatal to<br>help the orthodontist move the<br>correct teeth and stop the wrong<br>teeth from moving in the wrong<br>direction. | Orthodontic anchorage screw is<br>inserted into jaw and palatal to<br>help the orthodontist move the<br>correct teeth and stop the wrong<br>teeth from moving in the wrong<br>direction. |
| Material | Titanium alloy (Ti6Al4V)-<br>ASTM F136 | Titanium alloy (Ti6Al4V)-<br>ASTM F136 | Titanium alloy (Ti6Al4V)-<br>ASTM F136 |
| Form | Orthodontic Anchorage Screw | Orthodontic Anchorage Screw | Orthodontic Anchorage Screw |
| Structure | Double head<br>Button head<br>Wing<br>Slot | Through hole: wire hole<br>Simple head: no hole<br>Tapered body | Double head<br>Button head<br>Wing<br>Slot |
| Color | Silver | Silver | Silver |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Osteonic, a medical company. The logo features a blue diamond shape with a stylized "O" inside, next to the company name in bold, dark gray letters. Below the name is the tagline "The Medical Explorer" in a smaller, lighter font.
| Product Size | Diameter: Ø1.4 to 2.0<br>Thread Length: 6.0mm to<br>16.0mm (included an<br>unthreaded length under head)<br>Diameter by thread length<br>Ø1.4mm- 6.0mm to 8.0mm<br>Ø1.6mm and Ø1.8mm- 6.0mm<br>to 10.0mm<br>Ø2.0mm- 6.0mm to 12.0mm<br>Exception -S5T and S5R-:<br>Ø1.8mm- 10.0mm to 16.0mm) | Diameter: Ø1.4 to 1.8<br>Thread Length: 6mm to 10mm | Diameter: Ø1.3 to 2.5<br>Thread Length: 5mm to 16mm<br>(included an unthreaded length<br>under head) |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|------------------------------------------------------------------------------------------------------|
| Sterilization | Non sterile (user sterilization)<br>or<br>Sterile (Radiation) | Sterile (Radiation) | Non sterile (user sterilization)<br>or<br>Sterile (Radiation) |
| Single Use/<br>Reuse | Single use | Single use | Single use |
| Packaging | 1 or 10ea / Pack | 1ea / Pack | - |
| Shelf-life | Non sterile product- N/A,<br>Sterile product-3 Years | 5 Years | Non sterile product- N/A,<br>Sterile product-5 Years |
There are slightly different diameters and thread length between the subject and predicate device was added to support the difference diameters and lengths of the size of the subject device is included in the range of the predicate and reference device. The result of the comparative bench testing in line with ISO 19023 demonstrate that all screws are similar performance with predicate and/or reference device.
The technological characteristics of subject device is same materials and has similar dimension as the predicate and/or reference device.
The Indication for use of subject device is same as the predicate device.
Based on above, the subject device, ORTHO MI SYSTEM, is determined to be Substantially Equivalent (SE) to the predicate devices and reference device, Orthodontic screw (K103105) and Dual Top Screw System (K161335).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.