The Aveir™ Introducer is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
Device Story
Aveir Introducer is a guiding sheath system used to provide venous access for diagnostic or interventional devices, specifically for the implantation or retrieval of leadless pacemakers via the femoral vein. The device consists of an introducer sheath with a flush port, three-way stopcock, and hemostasis valve, paired with a dilator. The sheath features a hydrophilic coating for lubricity and a radiopaque tip marker for visualization. The dilator and sheath utilize a twist-locking mechanism to maintain connection during insertion. The device is operated by clinicians in a clinical setting. By providing a stable conduit and maintaining hemostasis, the device facilitates the delivery of interventional tools while minimizing air introduction and vessel trauma. It is supplied sterile for single-procedure use.
Clinical Evidence
No human clinical data provided. Evidence consists of biocompatibility testing per ISO 10993-1, bench testing (physical/dimensional, functional, lubricity/durability, packaging/shelf-life), and an acute GLP animal usability study. Usability validation confirmed acceptable performance and residual risk for the intended femoral vein access workflow.
Technological Characteristics
25 Fr inner diameter introducer sheath and dilator. Features include hydrophilic coating, radiopaque tip marker, hemostasis valve, and twist-locking mechanism. Supplied sterile via ethylene oxide. Biocompatibility evaluated per ISO 10993-1. Particulate matter testing per USP <788>.
Indications for Use
Indicated for providing a conduit into the venous system for insertion of diagnostic and interventional devices in patients requiring such access. Not specifically tested for pediatric use.
Regulatory Classification
Identification
A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.
Predicate Devices
Nanostim™ Introducer Kit 30 cm, model S1S18F (K160716)
Nanostim™ Introducer Kit 50 cm, model LSN18501 (K161102)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 27, 2021
Abbott Medical Bijal Patel-Jain Sr. Manager, Regulatory Affairs 15900 Valley View Court Sylmar, CA 91342
Re: K211989
Trade/Device Name: Aveir Introducer Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: June 25, 2021 Received: June 28, 2021
#### Dear Bijal Patel-Jain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K211989
Device Name Aveir Introducer
Indications for Use (Describe)
The Aveir™ Introducer is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Abbott. The logo consists of a stylized, rounded square shape in blue, positioned above the word "Abbott" in a bold, black font. The square shape is open at the bottom, creating a frame-like appearance.
#### Submitter
Abbott Medical 15900 Valley View Court Sylmar, CA 91342 USA Phone: (818) 362-6822 Date Prepared: June 25, 2021
## Contact Person
Allison Kamiya Regulatory Affairs Specialist Phone: (818) 493-2702 allison.kamiya@abbott.com
Bijal Patel-Jain Sr. Manager, Regulatory Affairs Phone: (818) 493-2476 bijal.jain(@abbott.com
## Device
| Name of Device: | Aveir™ Introducer |
|-----------------------|---------------------------------------|
| Model Number: | LSN25301 (30cm) and LSN25501 (50cm) |
| Common or Usual Name: | Introducer, Catheter |
| Classification Name: | Catheter Introducer (21 CFR 870.1340) |
| Regulatory Class: | II |
| Code: | DYB |
## Pediatric Use
The Aveir™ Introducer has not been specifically tested for use in pediatric patients.
#### Predicate Device
The modified Aveir™ Introducer is substantially equivalent in intended use and method of operation to the predicate devices manufactured by Abbott:
- Nanostim™ Introducer Kit 30 cm, model S1S18F (K160716) ●
- Nanostim™ Introducer Kit 50 cm, model LSN18501 (K161102) ●
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Abbott. The logo consists of a stylized letter 'A' in blue, positioned above the word "Abbott" in black, sans-serif font. The 'A' is designed with rounded corners and a continuous line, giving it a modern and clean appearance.
## Device Description
The Aveir™ Introducer is designed to perform as a guiding sheath for introduction of diagnostic or interventional devices. It is intended to provide a conduit into the venous system for insertion of diagnostic or interventional devices. The Aveir™ Introducer has a 25 French (Fr) inner diameter and comes in two lengths - Model LSN25301, 30 cm. and Model LSN25501, 50 cm. The only difference between the two models is the length.
The Aveir™ Introducer comprises the following two components:
- 1) Introducer sheath with a flush port and a three-way stopcock
The introducer sheath is coated with a hydrophilic coating and is fitted with a suture loop, a sideport with a three-way stopcock, and a hemostasis valve, which minimizes air introduced and maintains hemostasis during insertion and/or exchange. The sheath contains a radiopaque tip marker incorporated within the sheath material to identify the location of the distal tip sheath. The dilator contains a locking mechanism that locks the dilator into the introducer sheath.
- 2) Dilator
The dilator contains a locking mechanism that locks the dilator into the introducer sheath.
The introducer sheath and dilator are supplied sterile (via ethylene oxide) in one package and are intended for single procedure.
## Indications for Use/ Intended Use
The Aveir™ Introducer is intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
## Comparison of Technological Characteristics with Predicate Device
The subject and predicate devices have the same intended use, and similar technological characteristics and performance. Both the subject device and predicate devices are intended to provide a conduit into the venous system for insertion of diagnostic and other interventional devices.
At a high-level, the subject and predicate devices are based on the following technological elements that are the same:
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Abbott, a healthcare company. The logo features a stylized, rounded "A" shape in blue, positioned above the company name "Abbott" in a bold, black font. The "A" symbol is a prominent and recognizable element of the Abbott brand identity.
- Same intended use statement ●
- Radiopaque filler and radiopaque marker band in the distal tip to view its location during . a procedure
- Length options: 30 cm and 50 cm .
- A hermetic seal included in the Introducer's hub/seal to provide hemostasis ●
- . Smooth introducer tip transition and retention when the introducer sheath and dilator are connected, preventing movement of the introducer beyond the dilator tip transition, thus mitigating vessel trauma during insertion.
- The extrusions of the Introducer sheath and dilator are the same. ●
The following technological differences exist between the subject and predicate devices:
- a 25 French (Fr) inner diameter to enable the Aveir™ leadless pacemaker to remain fully . covered during venous access to simplify the procedure and reduce risk of procedural error
- a twist locking mechanism to the dilator rather than a snap fit .
- a hydrophilic coating to the sheath for ease of insertion. .
These differences were evaluated through performance testing, which demonstrated that the A veir™ Introducer performs in a substantially equivalent manner to currently marketed predicate devices.
#### Performance Data
The Aveir™ Introducer is substantially equivalent to the predicate devices based on comparisons of the intended use, device functionality, and technological characteristics. The following performance testing was conducted to demonstrate that the device meets its design specifications and is substantially equivalent to the predicate devices:
#### Biocompatibility Testing
The biocompatibility evaluation for the Aveir™ Introducer was conducted in accordance with ISO 10993-1 for biological evaluation and FDA Guidance Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" issued September 4, 2020. For biological testing in animal models, testing was conducted in accordance with FDA Good Laboratory Practice (GLP) Regulations 21 CFR 58.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for Abbott Laboratories. The logo features a stylized letter 'a' in blue, positioned above the word "Abbott" in a bold, sans-serif font. The blue color of the 'a' provides a visual contrast against the white background, making the logo easily recognizable.
The Aveir™ Introducer is considered an external communicating device in contact with circulating blood for less than 24 hours per ISO 10993-1 guidelines.
The testing included the following:
- Cytotoxicity ●
- Sensitization ●
- Irritation (Intracutaneous Reactivity) ●
- Acute Systemic Toxicity ●
- Materials Mediated Pyrogenicity ●
- Hemocompatibility
- Direct contact and extract hemolysis O
- Complement activation assay SC5b-9 O
- In-vivo thrombogenicity Anticoagulated Venous Implant (AVI) and O Non-Anticoagulated Venous Implant (NAVI)
- Particulate Matter per USP < 788> ●
#### Bench Testing
- Design Verification
- o Physical and Dimensional Characteristics
- Functional Characteristics о
- Hydrophilic Coating Lubricity and Durability O
- Packaging and Shelf Life Testing ●
- Nominal T=0 о
- Accelerated Aging 12-month o
#### Animal Testing
Usability validation testing including the Aveir™ Introducer was conducted through an acute GLP study. This GLP study were performed as defined by 21 CFR Part 58 "Good Laboratory Practice for Nonclinical Laboratory Studies" at a USDA registered facility, which is also accredited by AAALAC International.
The usability study was conducted for the intended use workflows where the Introducer is utilized to implant or retrieve a leadless pacemaker via femoral vein.
Based on the results of usability validation testing, the Aveir™ Introducer possesses acceptable usability and acceptable residual usability risk based on the observed difficulties and use errors of
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Abbott, a healthcare company. The logo consists of a stylized letter "a" in blue, with rounded corners and a three-dimensional appearance. Below the symbol is the company name, "Abbott," in a bold, sans-serif font.
the participants. All acceptance criteria were met under the conditions of the study. There were no observed use errors that could result in serious patient harm.
## Conclusions
The resulting evidence obtained from the design verification and validation testing demonstrated that the subject device is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.