Omeza Collagen Matrix
K211972 · Omeza, LLC · FRO · Sep 1, 2021 · SU
Device Facts
| Record ID | K211972 |
| Device Name | Omeza Collagen Matrix |
| Applicant | Omeza, LLC |
| Product Code | FRO · SU |
| Decision Date | Sep 1, 2021 |
| Decision | SESE |
| Submission Type | Traditional |
| Device Class | Class U |
| Attributes | Therapeutic |
Indications for Use
Omeza® Collagen Matrix is indicated for the management of wounds including: Partial and full-thickness wounds. Pressure ulcers. Venous ulcers. Diabetic ulcers. Chronic vascular ulcers. Tunneled/undermined wounds. Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence). Trauma wounds (abrasions, lacerations, superficial partial thickness burns, skin tears). Draining wounds. The device is intended for one-time use.
Device Story
Omeza Collagen Matrix is a sterile, single-use wound dressing composed of hydrolyzed whitefish collagen (Types I, II, III, IV) infused with cod liver oil, plant-derived oils, and waxes. Supplied in a 1.6g vial, the matrix has a gel-like consistency at ambient temperature. It is applied topically by a clinician to the wound bed, where it conforms to the surface to provide intimate contact. The matrix acts as an anhydrous skin protectant, maintains a moist wound environment, and is naturally incorporated into the wound tissue over time as part of the body's natural healing process. It is intended for use in clinical settings for various chronic and acute wound types.
Clinical Evidence
Clinical evidence includes a 25-subject skin prick study showing no immediate allergic reactions and a 58-subject Human Repeat Insult Patch Test (HRIPT) demonstrating no irritation or sensitization. Animal studies (swine) compared the device to saline and Hollister Endoform Dermal Template, showing no impairment of wound healing and no adverse biological reactions. Bench testing confirmed sterilization (ANSI/AAMI/ISO 11137-2), viral inactivation (ISO 22442), shelf life, and physical properties (extrusion, coverage, rheology).
Technological Characteristics
Hydrolyzed whitefish skin collagen (Types I, II, III, IV) with cod liver oil, plant oils, and waxes. Gel-like consistency; applied topically. Sterile (E-beam). Single-use. Shelf-stable at room temperature. Biocompatibility per ISO 10993-1. Packaging extractables/leachables per ISO 10993 and USP 1663.
Indications for Use
Indicated for management of partial/full-thickness wounds, pressure/venous/diabetic/chronic vascular ulcers, tunneled/undermined wounds, surgical wounds (donor sites, grafts, post-Moh's/laser, podiatric, dehiscence), trauma wounds (abrasions, lacerations, superficial partial thickness burns, skin tears), and draining wounds.
Predicate Devices
Reference Devices
- INTEGRA Flowable Wound Matrix (K072113)
- Kerecis MariGen Wound Dressing (K132343)
- Southwest Technologies Stimulen Collagen (K030774)
- Hollister Endoform Dermal Template (K092096)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services. The FDA logo features a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below. To the left of the FDA logo is the symbol of the Department of Health & Human Services, which consists of a stylized human figure.
September 1, 2021
Omeza, LLC % Randy Prebula Partner Hogan Lovells US LLP 555 13th Street NW Washington, District of Columbia 20004
Re: K211972
Trade/Device Name: Omeza Collagen Matrix Regulatory Class: Unclassified Product Code: FRO Dated: June 24, 2021 Received: June 24, 2021
Dear Randy Prebula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lixin Liu, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
#### K211972
Device Name
#### Omeza® Collagen Matrix
Indications for Use (Describe)
Omeza® Collagen Matrix is indicated for the management of wounds including:
- · Partial and full-thickness wounds
- · Pressure ulcers
- Venous ulcers
- · Diabetic ulcers
- · Chronic vascular ulcers
- · Tunneled/undermined wounds
- · Surgical wounds (donor sites/grafts, post-Moh's surgery, podiatric, wound dehiscence)
- · Trauma wounds (abrasions, lacerations, superficial thickness burns, skin tears)
- · Draining wounds
The device is intended for one-time use.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# Omeza® Collagen Matrix
# K211972
# Submitter's name and address:
Omeza, LLC 25 South Osprey Avenue Sarasota, Florida 34231
## Contact person and telephone number:
Thomas Gardner Chief Executive Officer Telephone: 941-780-5274 Email: tgardner@omeza.com
## Date Summary was prepared:
August 23, 2021
## Product Name:
| Proprietary Name: | Omeza® Collagen Matrix |
|-----------------------------|------------------------|
| Common Name: | Wound Dressing |
| Device Classification Name: | Dressing, Wound, Drug |
| Regulatory Classification: | Unclassified |
| Product Code: | FRO |
| Predicate Device: | SweetBio Apis (K182725) |
|--------------------|----------------------------------------------------|
| Reference Devices: | INTEGRATM Flowable Wound Matrix (K072113) |
| | Kerecis MariGen Wound Dressing (K132343) |
| | Southwest Technologies Stimulen Collagen (K030774) |
#### Product Description:
Omeza® Collagen Matrix (OCM) is a wound care matrix comprised of hydrolyzed fish collagen infused with cod liver oil, which acts as an anhydrous skin protectant, and other plant-derived oils and waxes. When applied to a wound surface, the matrix is naturally incorporated into the wound over time. Omeza® Collagen Matrix is designed for intimate contact with both regular wound beds to providea conducive environment for the patient's natural wound healing process.
# Intended Use/Indications for Use:
Omeza® Collagen Matrix is indicated for the management of wounds including:
- Partial and full-thickness wounds .
- Pressure ulcers .
- Venous ulcers .
- Diabetic ulcers .
- Chronic vascular ulcers .
{4}------------------------------------------------
- Tunneled/undermined wounds
- · Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- · Trauma wounds (abrasions, lacerations, superficial partial thickness burns, skin tears)
- Draining wounds
The device is intended for one-time use.
#### Comparison with the Predicate Device:
Omeza® Collagen Matrix and the predicate device SweetBio Apis (K182725) have the same intended use, namely to manage wounds by providing an animal-derived collagen product that is biodegradable and incorporates into the surrounding tissue during the body's natural wound healing processes. Both supplement the collagen constituent with additional biocompatible materials to achieve a final product that covers and protects the wound, assists in managing wound exudate, and maintains a moist wound environment. Both products are supplied sterile and for single use, and have very similar specific indications for use.
As can be seen from the substantial equivalence table below, the comparison to the predicate device shows no new questions of safety or effectiveness. Both products rely primarily on collagen to achieve the intended use, and based on the function of the additional materials in each product, the intended use of the Omeza Collagen Matrix is the same compared to that of the predicate. Additionally, performance testing data shows that there are no different questions of safety or effectiveness.
| | Omeza® Collagen Matrix<br>(Subject product) | SweetBio Apis (predicate)<br>(K182725) |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Indications for Use | |
| Product Code | FRO | FRO |
| Indications for Use | Indicated for the management of<br>wounds including: partial and full-<br>thickness wounds, pressure ulcers,<br>venous ulcers, diabetic ulcers, chronic<br>vascular ulcers, tunneled/undermined<br>wounds, surgical wounds (donor<br>sites/grafts, post- Moh's surgery, post-<br>laser surgery, podiatric, wound<br>dehiscence), trauma wounds (abrasions,<br>lacerations, superficial partial thickness<br>burns, and skin tears), draining wounds.<br>The device is intended for one-time use. | Indicated for use in management<br>of wounds including: full and<br>partial thickness wounds,<br>pressure ulcers (stages I-IV),<br>venous stasis ulcers, diabetic<br>ulcers, abrasions, surface<br>wounds, traumatic wounds<br>(healing by secondary intention),<br>donor site wounds, surgical<br>wounds |
| Rx | Yes | Yes |
| Single use | Yes | Yes |
| Technological Characteristics | | |
| Sizes | Single-use vial of 1.6g (2mL); volume<br>covers 3cm x 6cm area (18 cm²) | Sheet form, available in sizes 16<br>x16mm, 25x 25mm, 50 x 50mm |
| Intimate contact<br>with wound bed | Yes | Yes |
| Technological<br>Features | Collagen and other components<br>manufactured to gel-like consistency,<br>soft at ambient temperature, and applied<br>topically | Collagen derivative and other<br>components manufactured to<br>sheet, softened with saline, and<br>applied topically |
{5}------------------------------------------------
| Collagen Source | Whitefish skin collagen Types I, II, III, IV | Porcine skin collagen derivative(gelatin) |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Collagen product particle size | Collagen particles of ≤1000µm in diameter | Collagen particle derivative (gelatin) |
| Applied hydrated or flowable | Yes | Yes |
| Contains salt when applied | Yes (Sea salt) | Yes (Final product softened with saline before topical application) |
| Additional materials of composition | Cod liver oil, plant-derived oils and waxes, and process aids | Manuka honey and hydroxyapatite |
| Absorbable | Fully absorbable | Fully absorbable |
| Moist environment | Maintains moist wound environment | Maintains moist wound environment |
| Wound protection | Protects the wound tissue | Protects the wound tissue |
| Supplied sterile? | Yes (E-beam) | Yes (Gamma) |
| Shelf life | Shelf stable at room temperature | Shelf stable at room temperature |
| Additional performance testing | Porosity, mechanical, endotoxin, chemical composition, and distribution testing; sterilization validation | Porosity, mechanical, endotoxin, chemical composition, and distribution testing; sterilization validation |
The main differences between the Omeza Collagen Matrix and the predicate are the specific collagen source - Omeza uses whitefish skin-derived collagen and SweetBio Apis uses porcine skin-derived collagen – and the specific identity of the supplements, which serve the same fundamental purpose in enabling each wound dressing to achieve the shared intended use. The key questions of safety of the animal-derived material and effectiveness of achieving the intended use remain the same for both products, and are fully addressed for the Omeza Collagen Matrix through submitted performance testing. They are additionally supported by clearance of the reference devices with the same intended use and similar indications:
- MariGen Wound Dressing (K132343), which is also derived from fish collagen;
- Stimulen® Collagen Gel (K030774) which also contains collagen delivered in a gel-like . consistency; and
- INTEGRA Flowable Wound Matrix (K072113), which is also gel-like. In particular, Omeza® Collagen Matrix and the INTEGRA Flowable Wound Matrix (K072113) manage wounds by providing collagen particles of similar diameter in a gel-like consistency applied topically in a sterile, single-use application. Both products are designed to conform to the wound bed, achieving intimate contact with the surface, and are ultimately incorporated into the wound during the natural healing process.
# Performance Data:
The following testing was performed to demonstrate substantial equivalence:
# Biocompatibility Testing
All biocompatibility endpoints required according to ISO 10993-1 and FDA's corresponding 2016guidance were evaluated with favorable results. Testing was performed in the following categories:
- . Cytotoxicity
- Sensitization ●
- Irritation
- Acute systemic toxicity ●
- Genotoxicity ●
{6}------------------------------------------------
- . Material-mediated pyrogenicity
## Clinical Studies
A Skin Prick Study was performed in the US in accordance with Good Clinical Practice Standards. The single-center, monadic, one-day enrolled 25 subjects. The study evaluated Collagen Matrix for the potential of an immediate allergic reaction following a one-day Skin Prick Method Study. The test article, a positive control (1.0 mg/mL histamine base), and a negative control (aqueous negative control) were tested on subjects using the "Allergy Diagnostic Testing: An Updated Practice Parameter" skin prick test method. During this study, no immediate allergic reaction to the test product was observed for any subjects. Therefore, no potential allergenicity of the test product was observedunder the conditions of this study.
A Human Repeat Insult Patch Test was performed in the US to Good Clinical Practice Standards. The study design was based on a modified Human Repeat Insult Patch Test. The study was single site, evaluator-blinded using a within-subject randomization design. The study had (58) subjects enrolled. Omeza® Collagen Matrix was evaluated for the potential to cause irritation and sensitization compared to that of a negative control (0.9% aqueous sodium chloride) based on a modified human repeat insult patch test. Under the conditions of this study, the test product showed to be safe for use with no side effects.
## Animal Studies
Two wound healing studies were conducted to assess the implantation and the wound healing endpoints of Omeza® Collagen Matrix on porcine wounds. The first swine study compared the subject product to saline. The two products were applied to 10 full-thickness circular wound sites (approximately 3.0 cm in diameter) on four animals, with 14 and 25 day evaluations followed by pathology and independent expert review. The second swine study compared an exaggerated dose of Omeza Collagen Matrix to Hollister Endoform Dermal Template (K092096), with the products applied to 10 full-thickness circular wound sites (approximately 2.0 cm in diameter) on two animals. The final wound evaluation occurred at 24 days, followed by pathology and independent expert review. Both studies showed no evidence of impairment of wound healing with the use of Omeza® Collagen Matrix, and demonstrated that the subject product was safe and did not trigger adverse biological reactions inthe animals.
#### Additional Testing (Bench)
- . Extrusion: Omeza performed testing on representative lots to quantify the amount of producta typical user can extrude from a vial.
- Final Product Coverage: The time required for the product to cover 18 sq. cm - target coveragefor one vial - was calculated in a simulated model.
- Product Characterization under Environmental Conditions: Surface morphology of Omeza ● Collagen Matrix was imaged at varying magnifications using Scanning Electron Microscopy (SEM), to observe andconfirm microscopic and macroscopic stability of the OCM structure.
- . Interaction with Wound Exudate: SEM was employed to compare the surface structure/morphology of sterile, dry OCM to OCM immersed in solutions of varying salinity concentrations, simulating moist wound exudate conditions.
- Sterilization/endotoxins: OCM vials are sterilized using a process validated per . ANSI/AAMI/ISO 11137-2. Biologic testing for sterilization validation was also
{7}------------------------------------------------
performed. Endotoxin testing confirmed an acceptable level of endotoxins in the final product.
- Viral Inactivation: Three viral inactivation studies were performed on the OCM utilizing a . panel of model viruses per ISO 22442 Part 3 (Validation of the elimination and/or inactivation of viruses and transmissible spongiform encephalopathy (TSE) agents).
- Shelf Life: OCM meets requirements for a 9-month shelf life, based on product tested at ● theend of this period passing all finished product specifications.
- Collagen Characterization: Standard ELISA techniques were used to determine the types . ofcollagen in OCM.
- Lot-to-Lot Consistency: Testing of samples from representative lots demonstrated lot-to-lot ● consistency of all OCM raw materials.
- Cod Liver Oil Identification and Quantification: Fatty Acid Methyl Ester analysis following AOAC 996.06 was conducted on representative lots to quantify the Cod Liver Oil (CLO) in thefinal product.
- . Elastic Modulus: The product's rheological properties were evaluated by analyzing elastic modulus, with results showing that this feature of the final product is consistent across lots.
- Extractables and Leachables: Extractables testing was completed on the OCM primary ● and secondary packaging system, following ISO 10993 and USP 1663. No extractable compounds were detected in the product primary packaging above the analytical evaluationthreshold.
- Final Product Specifications: Representative samples of final, sterilized OCM were . analyzed for conformance to pre-defined specifications for quantity (%w/w) of collagen, pH, specific gravity, heavy metals, product structure, and bioburden.
#### Conclusion:
Omeza® Collagen Matrix functions as intended, and is as safe and effective as SweetBio Apis (K182725). As explained above, OCM has the same intended use and similar indications for use, and similar key technological and design characteristics and mechanism of action. The minor differences between the subject and predicate products do not raise different questions of safety or effectiveness, as supported by performance data on the subject product and prior clearance of the reference devices. Thus, Omeza® Collagen Matrix is substantially equivalent with the predicate device.