K211969 · Exact Medical Manufacturing · ITX · Aug 13, 2021 · Radiology
Device Facts
Record ID
K211969
Device Name
EMM Ultrasonic Transducer Cover
Applicant
Exact Medical Manufacturing
Product Code
ITX · Radiology
Decision Date
Aug 13, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Exact Medical Manufacturing Ultrasonic Transducer Cover can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe. This product is not intended for use of probes with endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic Transducer Cover are also sold as bulk non-sterile, single use items, to repackager/ relabeler establishments for further packaging and ethylene oxide sterilization.
Device Story
Device is a 0.05mm thin, 50 GSM, translucent polyurethane film tube (nominally 48" x 6") with a heat-sealed distal end. Applied over ultrasound transducer probes to minimize patient-to-probe contamination during external scanning; facilitates probe cleaning/disinfection. Used in clinical settings by healthcare professionals. Device is single-use; provided either sterile or as bulk non-sterile items for further packaging/sterilization by third parties. Output is a physical barrier; does not affect ultrasound image quality or signal transmission. Benefits include reduced cross-contamination risk and simplified probe maintenance.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via AAMI PB70 Level 3 barrier testing, ISO 527-3 tensile/elongation testing, ISO 10993-5 cytotoxicity, ISO 10993-10 sensitization/irritation, NEMA UD 2-2004 acoustic output, and ISO 10993-7 ethylene oxide residuals.
Technological Characteristics
Material: Polyurethane film (50 GSM). Form factor: Tubular, heat-sealed distal end. Energy source: None (passive barrier). Sterilization: Ethylene oxide (ISO 11135-1). Connectivity: None. Software: None.
Indications for Use
Indicated for use as a protective barrier to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures on external intact skin. Not indicated for endocavity or intraoperative use.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, and the full name of the agency, "U.S. Food & Drug Administration," is in a smaller font size below the acronym.
Exact Medical Manufacturing % Abdel Halim, PharmD, MSc, PhD, DABCC President, Global Quality and Regulatory Services Global Quality and Regulatory Services 10 Scenic Way MONROE, NEW JERSEY 08831
August 13, 2021
Re: K211969
Trade/Device Name: EMM Ultrasonic Transducer Cover Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: Class II Product Code: ITX Dated: August 10, 2021 Received: August 12, 2021
Dear Dr. Halim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211969
Device Name EMM Ultrasonic Transducer Cover
#### Indications for Use (Describe)
Exact Medical Manufacturing Ultrasonic Transducer Cover can be used to minimize contamination between patient and ultrasound probe during ultrasound scanning procedures for external intact skin. This may help with easier cleaning and disinfection of the probe. This product is not intended for use of probes with endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic Transducer Cover are also sold as bulk non-sterile, single use items, to repackager/ relabeler establishments for further packaging and ethylene oxide sterilization.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary for Exact Medical Manufacturing Inc., EMM Ultrasonic Transducer Cover
| Date Summary was Prepared | March 31, 2021 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter | Ryan Power, Vice President<br>Exact Medical Manufacturing Inc.<br>4917 William Street, Suite C<br>Lancaster, NY 14086<br>kpower@exactmm.com<br>(p)716-681-0866, (f) 716-681-4110 |
| Primary Contact for this 510(k) Submission | Ryan Power, Vice President<br>Exact Medical Manufacturing Inc.<br>4917 William Street, Suite C<br>Lancaster, NY 14086<br>kpower@exactmm.com<br>(p)716-681-0866, (f) 716-681-4110 |
| Device Common Name | Cover, Ultrasonic Transducer Cover |
| Trade Name | Cover, Disposable, |
| Device Product Codes and<br>Classification Name | Diagnostic ultrasonic transducer ITX, 21CFR892.1570, Class II |
| Predicate Device | MedXPress Pro, EZ-Cover K 191491 |
| Device Description | A 0.05mm thin, 50 GSM (Grams per Square Meter), translucent high strength<br>polyurethane film tube shape, nominally 48" x 6" with heat sealed distal end to be<br>applied over a transducer probe to provide a Transducer Cover that can be used to<br>minimize contamination between patient and ultrasound probe during ultrasound<br>scanning procedures for external intact skin. This may help with easier cleaning<br>and disinfection of the probe. This product is not intended for use of probes with<br>endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic<br>Transducer Cover are also sold as bulk non-sterile, single use items, to<br>repackager/relabeler/kit packers establishments for further packaging and ethylene<br>oxide sterilization. |
| Indications for Use | Exact Medical Manufacturing Ultrasonic Transducer Cover can be used to<br>minimize contamination between patient and ultrasound probe during ultrasound<br>scanning procedures for external intact skin. This may help with easier cleaning<br>and disinfection of the probe. This product is not intended for use of probes with<br>endocavity or intraoperative use. The Exact Medical Manufacturing Ultrasonic<br>Transducer Cover are also sold as bulk non-sterile, single use items, to<br>repackager/relabeler establishments for further packaging and ethylene oxide<br>sterilization. |
| Technological Characteristics | Exact Medical Ultrasonic Transducer Cover has the same design, material, and<br>performance characteristics of the predicate device. |
| Summary of Non-Clinical Testing | Exact Medical Ultrasonic Transducer Cover is substantially equivalent and meets<br>the same acceptance criteria as the predicate device as in K191491. Non-clinical<br>performance testing includes;<br>- barrier properties- AAMI PB70, Level 3,<br>- tensile<br>- elongation<br>- flammability<br>- linting,<br>- sterility<br>- biocompatibility<br>- acoustics performance.<br>- ethylene oxide residuals<br>All results of the testing met acceptance criteria. See following (below) Summary<br>Tables. |
| Clinical Testing | None, not applicable |
| Substantial Equivalence Conclusion | The transducer cover described in this 510(k) submission are substantially<br>equivalent in all specifications and performance compared to the predicate device<br>identified in K191491 except for minor variations in the widths and lengths. There<br>are no significant differences among the submission device and the predicate<br>device which present concerns concerning safety and effectiveness. |
{4}------------------------------------------------
### Bench Testing Summary
| Test Performed<br>& Description of<br>test | Device Description / Sample Size | Reference<br>Standard | FDA<br>Recognition<br>number or Not<br>Recognized<br>(NR) | Acceptance<br>Criteria | Unexpected<br>Results/<br>Significant<br>Deviations | Result Summary | Conclusion | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Water Resistance:<br>Impact<br>Penetration Test,<br>Level 3 | Final Finished Device (both heat seal and TPU<br>film). Worst case, largest size in the submission.<br>Sample Size N= 32 with C=0<br>AAMI PB70 Sample Size AQL of 4<br>%/RQL of 20 %.<br>Test sample exceeds the reference standard<br>requirements (N=32, C=3) | AATCC<br>42:2017<br>(AAMI<br>PB70:2003<br>/(R)2009) | NR | blotter weight gain of<br>no more than 1.0 g | None | Accept<br>32 samples =/< 1.0g<br>(all samples = 0.0 g) | Ability of a protective product to<br>resist the penetration of liquids<br>and liquid borne<br>Microorganisms accordingly to<br>AAMI PB70, Level 3<br>Substantially equivalent barrier<br>property compared to the<br>predicate device | |
| Test Method for<br>Water Resistance:<br>Hydrostatic<br>Pressure Level 3 | Final Finished Device (both heat seal and TPU<br>film). Worst case, largest size in the submission.<br>Sample Size N= 32 with C=0<br>AAMI PB70 Sample Size AQL of 4<br>%/RQL of 20 %.<br>Test sample exceeds the reference standard<br>requirements (N=32, C=3) | AATCC<br>127:2017<br>(2018)e<br>(AAMI<br>PB70:2003<br>/(R)2009) | NR | hydrostatic<br>resistance of at least<br>50 cm | None | Accept<br>32 samples > 50cm<br>(all samples =/><br>180cm) | Ability of a protective product to<br>resist the penetration of liquids<br>and liquid borne<br>Microorganisms accordingly to<br>AAMI PB70, Level 3<br>Substantially equivalent barrier<br>property compared to the<br>predicate device | |
| Flammability of<br>Clothing Textiles | Polyurethane Film<br>Sample Size per 16CFR1610, 5 replicates are to<br>be tested if no flame spread is observed upon<br>preliminary testing.<br>Material evaluation only for flammability, not<br>relevant to device configuration. | 16CFR1610 | NR | Burn Time >/= 3.5<br>seconds | None | Accept<br>5 samples DNI (Did<br>Not Ignite) | Able to meet and claim<br>flammability Class 1 rating in<br>accordance with 16CFR1610<br>Exceeds predicate device based<br>upon predicate device has not<br>been tested for this attribute. | |
| Determination of<br>tensile properties<br>— Part 3: Test<br>conditions for<br>films and sheets<br>(Tensile and<br>Elongation) | Polyurethane Film<br>Material evaluation only, not relevant to device<br>configuration.<br>Sample size per statistical sampling plan ISO2859-<br>1, based upon lot population quantity | ISO 527-3 | NR | Tensile strength<br>minimum 40N in<br>machine direction<br>Tensile strength<br>minimum 34N in<br>cross direction<br>Elongation 108 %<br>minimum in machine<br>direction<br>Elongation 119%<br>minimum in cross<br>direction | None | Accept<br>All samples must<br>pass acceptance<br>criteria in<br>accordance with<br>statistical sampling<br>plan per ISO2859-1,<br>otherwise, material<br>is rejected (not used<br>in production) | Specification confirmation of<br>tensile and elongation properties<br>of plastic films or sheets less than<br>1 mm thick to assure compliant<br>device barrier performance.<br>Substantially equivalent tensile<br>and elongation property<br>compared to material specification | |
| Lint and other<br>particles<br>generation in the<br>dry state | Polyurethane film<br>Material evaluation only, not relevant to device<br>configuration.<br>Sample size = 10<br>Standard indicates 5 samples | ISO 9073-10 | NR | IPM =/< 3.5<br>(Index for Particle<br>Matter) | None | Accept<br>IPM 1.3 (outside)<br>IPM 0.72 (inside) | Material does not significantly<br>generate lint or particles that may<br>shed during an ultrasonic scan<br>procedure.<br>Exceeds predicate device based<br>upon predicate device has not | |
| Acoustic Output<br>Measurement<br>Standard for<br>Diagnostic<br>Ultrasound | Final Finished Device<br>Worst case test configuration on distal sealed end<br>where ultrasonic probe is positioned. Polyurethane<br>50gsm tested. | NEMA UD 2-<br>2004<br>(R2009) | 12-105 | negligible effect on<br>the transmission of<br>ultrasound signal<br>and does not<br>degrade signal or<br>image | None | Accept<br>Speed of sound =<br>1698 m/sec<br>Acoustic Attenuation<br>= 0.51 dB | been tested for this attribute.<br>Based on the acoustic testing<br>results it was concluded that<br>EMM's transducer/probe cover<br>has negligible effect on the<br>transmission of ultrasound signal<br>and that EMM's transducer/probe | |
| Equipment,<br>Revision 3 | Sample size<br>3 ea. Speed of Sound<br>3 ea. Acoustic Attenuation<br>3 ea. Acoustic Impedance | | | | | | Acoustic Impedance<br>= 1.63 MRayl | cover does not degrade signal or<br>image.<br>Exceeds predicate device based<br>upon predicate device has not<br>been tested for this attribute. |
| Standard<br>Practices for<br>Evaluating the<br>Resistance of<br>Plastics to<br>Chemical<br>Reagents.<br>Compatibility of<br>ultrasonic<br>coupling gel with<br>polyurethane | Polyurethane Probe Cover Samples<br>Sample Size N=45<br>Tested by Manufacturer of coupling gel, FDA 510K<br>cleared K130581. This element is not an Exact<br>Medical Manufacturing product. Bench Testing is<br>provided to confirm compatibility with polyurethane<br>film | ASTM D543-<br>20 | NR | Evaluated for Color,<br>Dimension, Weight,<br>Coordinates,<br>Mechanical<br>Properties.<br>Acceptance Criteria<br>Not Defined | None | Accept<br>Color N=9, Pass,<br>Dimension N=9<br>Pass, Weight N=9<br>Pass, Coordinates<br>N=9 Pass,<br>Mechanical<br>Properties N=9 Pass | Confirmation that device<br>polyurethane material is not<br>negatively affected by the<br>acoustic coupling gel during an<br>ultrasonic scanning procedure.<br>Exceeds predicate device based<br>upon predicate device does not<br>provide compatibility evaluation of<br>acoustic coupling gel with their<br>device. | |
| Ethylene Oxide<br>Residuals | Polyurethane film<br>Worst case, Run the sterilization cycle at the cycle<br>parameters which provide for the highest EO<br>exposure, i.e. high gas induction time and pressure<br>(most pounds of EO), highest length of EO dwell<br>time, least time for post cycle vacuums, lowest<br>amount of aeration time within aeration chamber.<br>Sample Size per sterilization batch N=15 X 3<br>consecutive sterilization batches = 45 samples | ISO 10993-<br>7:2008, | 14-408 | tolerable contact<br>limit (TCL) as<br>expressed in units of<br>micrograms per<br>square centimeter<br>for EO at 10ug/cm2<br>and<br>milligrams per<br>square centimeters<br>for ECH at 5mg/cm2 | None | Accept<br>Actual results for EO<br>Residuals at Time 0<br>hours = 8.67ug/cm2.<br>Actual results for<br>ECH residuals at<br>Time 0 hours = Not<br>detected (within the<br>limits of detection of<br>3.7ug/g). | Confirmation of maximum<br>allowable residues for ethylene<br>oxide (EO) within the submission<br>device<br>sterilized with EO and<br>to ensure that the levels of<br>residual EO, ethylene<br>chlorohydrin<br>(ECH) and ethylene glycol (EG)<br>pose a minimal risk to the patient<br>in normal product use.<br>Substantially equivalent to<br>predicate device based upon<br>acceptable testing results that<br>comply with testing standard. | |
| In Vitro<br>Cytotoxicity Test | The test article. Probe Cover polyurethane, EO<br>Exposed, was evaluated to determine the biological<br>reactivity of a mammalian cell culture (mouse<br>fibroblast L929 cells) in response to the test article.<br>Sample Size = 4 test article extracts | ISO 10993-<br>5;2009 | 2-245 | The lower the<br>Viab.% value, the<br>higher the cytotoxic<br>potential of the test<br>article is. If viability is<br>reduced <70% of the<br>blank, it has<br>cytotoxic potential.<br>The Viab.% of 100%<br>extract of the test<br>article is the final<br>result | None | 100% Test article<br>extract 76.5%<br>viability.<br>No Test article<br>extracts are <70%<br>viability. | Testing Conducted on final<br>sterilized device. Study<br>conclusion, "Under the conditions<br>of this study, the test article<br>extract did not show potential<br>toxicity to L929 cells | |
| Test protocol<br>description. | Results<br>Summary | Analysis | Conclusion | Objective of test | Test Article | Test method and<br>procedure | Study endpoint | pre-defined<br>acceptance or<br>pass/fail criteria |
| ISO 10993-<br>10:2010:2010<br>Closed patch<br>Sensitization<br>Study in Guinea<br>Pigs | No skin<br>sensitization<br>reaction was<br>found in the<br>Guinea pigs<br>using extracts of<br>the test article<br>(0.9% sodium<br>chloride, sesame<br>oil). The positive<br>rate of<br>sensitization was<br>0%. | Dermal Reactions<br>Challenge<br>24 Hour Score<br>Ten test Guinea<br>pigs, 0 (zero)<br>reactions<br>48 Hour Score<br>Ten test Guinea<br>pigs, 0 (zero)<br>reactions | Under the conditions<br>of the study, the test<br>article extracts<br>showed no<br>significant evidence<br>of causing skin<br>sensitization in the<br>Guinea pigs. | The purpose of this<br>study was to<br>evaluate the<br>potential of the test<br>article to cause<br>dermal contact<br>sensitization<br>following repeated<br>occlusive patching<br>in the guinea pig. | Probe cover<br>Polyurethane | This study was<br>conducted to<br>evaluate skin<br>sensitization<br>using Guinea pig<br>maximization test<br>based on the ISO<br>10993-<br>10:2010:2010,<br>Biological<br>evaluation of<br>medical devices.<br>Part 10: Tests for<br>irritation and skin<br>sensitization. | The test article.<br>Probe Cover<br>polyurethane,<br>EO Exposed,<br>was evaluated<br>for the, was<br>evaluated for<br>the potential<br>skin<br>sensitization in<br>the Guinea pig. | No significant<br>evidence of<br>causing skin<br>sensitization in<br>the guinea pig.<br>Applying grading<br>scales as<br>follows,<br>No visible change<br>= 0<br>Discrete or patchy<br>erythema = 1<br>Moderate and<br>confluent<br>erythema =2<br>Intense erythema<br>and swelling = 3<br><br>Substantially<br>equivalent to the<br>predicate device<br>based upon<br>meeting the<br>acceptance<br>criteria of the test<br>standard to<br>evaluate the<br>potential of the<br>test article to<br>cause dermal<br>contact<br>sensitization<br>following repeated<br>occlusive<br>patching in the<br>guinea pig |
| ISO 10993-<br>10:2010:2010<br>Intracutaneous<br>Reactivity Test | All animals<br>were survived<br>and no<br>abnormal signs<br>were observed<br>during the<br>study.<br>According to<br>what observed | Summary<br>Analysis Test<br>Results of<br>Dermal<br>Observations in<br>Test Article<br>Extract SC<br>24 hr = 0 | The test result<br>showed that the<br>polar and non-<br>polar extract of the<br>final test sample<br>score is less 1.0,<br>the requirements<br>of the test are met. | To evaluate the<br>potential<br>intracutaneous<br>reactivity caused<br>by<br>intracutaneously<br>inject the extract<br>to rabbit and<br>extrapolating the | Probe cover<br>Polyurethane | To evaluate the<br>potential<br>intracutaneous<br>reactivity<br>caused by<br>intracutaneously<br>inject the extract<br>to rabbit in<br>accordance with<br>Reactivity Test | The test<br>article. Probe<br>Cover<br>polyurethane,<br>EO Exposed,<br>was evaluated<br>for the<br>Intracutaneous<br>Reactivity Test | The<br>requirements of<br>the test are met<br>if the final test<br>article score is<br>1.0 or less.<br><br>Substantially<br>equivalent to |
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Testing – Animal Summary Table:
{7}------------------------------------------------
| the skin<br>reaction of<br>polar and non-<br>polar extract on<br>testing side did<br>not exceed that<br>on the control<br>side. Thus, the<br>final test article<br>score was<br>calculated to be<br>0. See table 3. | 48 hr = 0<br>72 hr = 0<br><br>Extract SO<br>24 hr = 0<br>48 hr = 0<br>72 hr = 0 | results to<br>humans, but it<br>does not<br>establish the<br>actual risk of<br>irritation. | ISO I 0993-<br>10:2010<br>Biological<br>evaluation of<br>medical devices<br>- Part 10: Tests<br>for irritation and<br>skin<br>sensitization. | in New<br>Zealand white<br>rabbits. | the predicate<br>device based<br>upon meeting<br>the acceptance<br>criteria of the<br>test standard to<br>evaluate the<br>potential<br>intracutaneous<br>reactivity |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
{8}------------------------------------------------
# 510K Summary: Submission Device to Predicate Device Comparison
| Exact Medical Manufacturing -<br>Transducer Cover | Substantially<br>Equivalent | MedXPress.Pro EZ-Cover K191491<br>PREDICATE DEVICE |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: Exact Medical Ultrasonic Transducer<br>Cover can be used to minimize contamination between<br>patient and ultrasound probe during ultrasound scanning<br>procedures for external intact skin. This may help with<br>easier cleaning and disinfection of the probe. This product<br>is not intended for use of probes with endocavity or<br>intraoperative use. The Exact Medical Manufacturing<br>Ultrasonic Transducer Cover are also sold as bulk non-<br>sterile, single use items, to repackager/relabeler<br>establishments for further packaging and ethylene oxide<br>sterilization. | Substantially<br>Equivalent | Indications for Use: Probe cover can be used to<br>minimize contamination between patient and<br>ultrasound probe during ultrasound scanning<br>procedures for external intact skin. This may help<br>with easier cleaning and disinfection of the probe.<br>This product is not intended for use of probes with<br>endocavity or intraoperative use. |
| Classification & Code Classification & Code Diagnostic<br>ultrasonic transducer ITX, 21CFR892.1570, Class II | Substantially<br>Equivalent | Classification & Code Diagnostic ultrasonic<br>transducer ITX, 21CFR892.1570, Class II |
| Barrier properties - AATCC 42:2007, AATCC 127:2008:<br>Liquid Barrier Performance and Classification of Protective<br>Apparel and Drapes intended for Use in Health Care<br>Facilities, AAMI PB70:2003 /(R)2009, Level 3, AATCC 42<br>≤ 1.0 g, AATCC 127 ≥ 50 cm | Substantially<br>Equivalent | Water Resistance Polyethylene film can withstand<br>xx hydrostatic pressure time should not be less than<br>XX.<br>Standards not indicated |
| Materials & Construction: POLYMER, Polyurethane,<br>tubular, sealed | Substantially<br>Equivalent | Materials & Construction: POLYMER, Polyethylene,<br>tubular, sealed |
| Sterile (via EO Gas) ISO 11135-1:2014, Sterilization of<br>health care products - Ethylene oxide - Part 1. SAL 10-6 | Substantially<br>Equivalent | Sterilization Validation (Re-validation of Report of<br>Ethylene Oxide Sterilizer Performance) |
| Sterile Barrier Packaging: Chevron peel pouch (coated<br>paper (73gsm), PET12/PE40 film construction), individual<br>internal wrap. | Substantially<br>Equivalent | Aseptic packaging seal test |
| Disposable, Single Use Only | Substantially<br>Equivalent | Disposable, Single Use Only |
| Non-Sterile, Bulk Pack | | Not offered |
| Tensile strength Machine Direction / Cross Direction -<br>ISO 527- 3 Plastics - Determination of tensile properties<br>- Part 3: Test conditions for films and sheets Machine<br>Direction / Cross Direction: 40N per 2.5cm / 34N per<br>2.5cm | Substantially<br>Equivalent | The maximum tensile strength of polyethylene film<br>for manufacture protective sleeve should be no less<br>than XX, and the maximum tensile force of traverse<br>fracture is no more than XX<br>Test Standard not Indicated |
| Elongation - - ISO 527- 3 Plastics - Determination of<br>tensile properties - Part 3: Test conditions for films and<br>sheets Machine Direction / Cross Direction: 108% / 119%…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.