K211953 · Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd. · KGO · May 25, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211953
Device Name
Disposable Sterilized Latex Surgical Gloves
Applicant
Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
May 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Device Story
Disposable sterilized surgical glove; manufactured from natural rubber latex; intended for use by operating room personnel. Device acts as a physical barrier to protect surgical wounds from contamination during procedures. Provided EO sterilized with a 3-year shelf life. Available in multiple sizes (6 through 8.5). Performance verified through bench testing against ASTM standards for physical properties, watertightness, and biocompatibility.
Clinical Evidence
No clinical data. Bench testing only, including physical dimensions, tensile strength, ultimate elongation, watertightness (ASTM D5151), residual powder (ASTM D6124), protein content (ASTM D5712), biocompatibility (ISO 10993-10/11), pyrogenicity (USP <151>), and bacterial endotoxin (USP <85>).
Technological Characteristics
Natural rubber latex; powder-free; EO sterilized (SAL 10^6). Complies with ASTM D3577-09(15) (Type 1 gloves). Biocompatibility per ISO 10993-1, 10, 11. Residuals per ISO 10993-7. Dimensions: 265mm length; 0.10mm min thickness. No software or electronic components.
Indications for Use
Indicated for use by operating room personnel to protect a surgical wound from contamination. No specific patient population or contraindications described.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable.
May 25, 2022
Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd % Alice Huang RA Manager Shanghai Mind-Link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District Shanghai. 200040 China
Re: K211953
Trade/Device Name: Disposable Sterilized Latex Surgical Gloves Regulation Number: 21 CFR 878.4460 Regulation Name: Non-Powdered Surgeon's Glove Regulatory Class: Class I, reserved Product Code: KGO Dated: April 15, 2022 Received: April 25, 2022
Dear Alice Huang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K211953
Device Name Disposable Sterilized Latex Surgical Gloves
Indications for Use (Describe)
The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
## K211953
#### I. SUBMITTER:
Jiangxi Kemei Medical Apparatus & Instruments Group Co., Ltd. No. 316 National road, Xinju, Changshan Town, Jinxian County, Nanchang City, 331724, Jiangxi Province, China. Contact Person: Jenny Huang
TITLE: International Trade Supervisor
TEL. : +86 17702094798
Email: sales@jxkmjt.com
Submission Correspondent: Alice Huang Email: alice.huang(@mind-link.net Tel:+86 15618536177 Shanghai Mind-link Business Consulting Co., Ltd. Room A08, Floor 14th, No 699, Jiaozhou Road, Jingan District, Shanghai
Summary prepared: 05/23/2022
II. DEVICE 510(k) Number: K211953 Name of Device: Disposable Sterilized Latex Surgical Gloves Regulation Number: 21 CFR878.4460 Common Name: Surgeon's Gloves Classification Name: Non-powdered surgeon's glove Regulatory Class: I Product Code: KGO
III. PREDICATE DEVICE
510(k) Number: K171550 Product Name: Sterile Latex Surgical Gloves, Powder Free Manufacture: Sanrea Healthcare Pvt Ltd
{4}------------------------------------------------
## IV. DEVICE DESCRIPTION
The proposed device, Disposable Sterilized Latex Surgical Gloves is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination.
The proposed device is made of natural rubber latex, per standard ASTM D3577-09(15), the rubber surgical gloves classification is:
"Type 1-gloves compounded primarily from nature rubber latex." The gloves are powder-free and available in white in sizes 6, 6.5, 7, 7.5, 8, and 8.5.
The proposed device is provided EO sterilized to achieve the sterility Assurance Level (SAL) of 106. The disposable sterilized latex surgical glove shelf life is 3 years.
# V. INDICATIONS FOR USE
The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Disposable Sterilized Latex Surgical Gloves are compared with the predicate device (Sterile Latex Surgical Gloves, Powder Free, K171550). The results are shown below in the Technological Characteristics Comparison Table:
| DEVICE | Subject Device | Primary Predicate Device<br>(K171550) | Comparison |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Indications For Use | The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. | The Sterile Latex Surgical Gloves, Powder free, is a disposable device made of natural rubber, intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination. | Same |
| Intended Use | The Disposable Sterilized Latex Surgical Glove is a device made of nature rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. | Intended for medical purposes that is worn by operating room personnel to protect a surgical wound from contamination. | Same |
| Classification<br>Product Code | KGO | KGO | Same |
| Regulation No. | 21 CFR 878.4460 | 21 CFR 878.4460 | Same |
| Class | I | I | Same |
| Powdered or<br>Powdered free | Powdered free | Powdered free | Same |
| Material | Natural Rubber Latex | Natural Rubber Latex | Same |
| Dimensions | Length:<br>Size 6: 265mm, min<br>Size 6.5: 265mm, min<br>Size 7: 265mm, min<br>Size 7.5: 265mm, min<br>Size 8: 265mm, min<br>Size 8.5: 265mm, min<br>Width:<br>Size 6: 76 ±6 mm<br>Size 6.5: 83 ± 6 mm<br>Size 7: 89 ± 6 mm<br>Size 7.5: 95 ± 6 mm<br>Size 8: 102 ± 6 mm<br>Size 8.5: 108 ± 6 mm<br>Palm Thickness: 0.10mm, min<br>Finger Thickness: 0.10mm, min<br>Cuff Thickness: 0.10mm, min | size 7.5 :<br>Length : 300mm<br>size 7.5 :<br>Width : 95mm<br>Palm Thickness: 0.19mm<br>Finger Thickness: 0.22mm<br>Cuff Thickness: 0.14mm | Similar<br>Meeting<br>requirement<br>of ASTM<br>D 3577 |
| Biocompatibility | Under the conditions of the study, not a sensitizer.<br>Under the conditions of the study, not an irritant.<br>Under the conditions of the study, no evidence of acute systemic toxicity. | Non-sensitizing<br>Non-irritating | Similar |
| Tensile strength | Before Aging 24MPa, min<br>After Aging 18MPa, min | Before Aging 31.22MPa<br>After Aging 24.43MPa | Similar<br>Meeting |
| Ultimate<br>Elongation | Before Aging 750%, min<br>After Aging 560%,min | Before Aging 843.09%<br>After Aging 762.28% | requirement<br>of ASTM<br>D 3577 |
| Freedom from | Meets ASTM D5151-06(2015) | Meets ASTM D 5151 -06 | Same |
| Holes | AQL 1.5 | (2011) | |
| Protein Content | Meets ASTM D5712-2015 | Meets ASTM D5712 | Same |
| Powdered residue | Meets ASTM D6124-06(Reapproved 2017) | Meets ASTM D 6124-06(2011) | Same |
| Sterility | EO | EO | Same |
| EtO and ECH residuals | Meets ISO 10993-7:2008 | Meets ISO 10993-7:2008 | Same |
| Shelf life | 3 Years | 3 Years | Same |
| Single Use | YES | YES | Same |
| Label and labeling | Meet FDA's Requirement | Meet FDA's Requirement | Same |
{5}------------------------------------------------
{6}------------------------------------------------
## VII. SUMMARY OF NON-CLINICAL TESTING
The gloves have the same technological characteristics compared to ASTM or equivalent standards as given below:
- ASTM D5151-06(Reapproved 2015) Standard Test Method for Detection of Holes in Medical Gloves
- ASTM D6124-06 (Reapproved 2017) Standard Test Method for Residual Powder on Medical Gloves
- ASTM D3577-09 (Reapproved 2015) Standard Specification for Rubber Surgical Gloves
- . ASTM D5712-15 Standard Test Method for Analysis of Aqueous Extractable Protein in Natural Rubber in Latex, Natural Rubber, and Elastomeric Products Using the Modified Lowry Method
- ISO 10993-1 : 2009 Biological evaluation of medical devices - Part1: Evaluation and testing within a risk management process
- ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- USP 43-NF 38<151> Pyrogen Test
- USP 43<85> Bacterial Endotoxin Test
- ASTM F88/F88M-15 Standard Test Method for Seal Strength of Flexible Barrier Materials
{7}------------------------------------------------
- ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- . ASTM D7160 Standard Practice for Determination of Expiration Dating for medical Gloves
| Test<br>Method | Standard | Test Purpose | Criteria | Result |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Dimension | ASTM D3577-09(Reapproved<br>2015) Standard Specification for<br>Rubber Surgical Gloves | To determine the length<br>of the gloves | 265mm, min | Pass |
| | ASTM D3577-09(Reapproved<br>2015)Standard Specification for<br>Rubber Surgical Gloves | To determine the width<br>of the gloves | Size 6: 76 ±6 mm<br>Size 6.5: 83 ± 6 mm<br>Size 7: 89 ± 6 mm<br>Size 7.5: 95 ± 6 mm<br>Size 8: 102 ± 6 mm<br>Size 8.5: 108 ± 6 mm | Pass |
| | ASTM D3577-09(Reapproved<br>2015) Standard Specification for<br>Rubber Surgical Gloves | To determine the<br>thickness of the gloves | Palm: 0.10mm, min<br>Finger: 0.10mm, min<br>Cuff: 0.10mm, min | Pass |
| Physical<br>Properties | ASTM D3577-09(Reapproved<br>2015) Standard Specification for<br>Rubber Surgical Gloves | To Determine the<br>physical properties<br>Tensile strength | Before Ageing<br>Tensile Strength<br>24Mpa Minimal for all<br>sizes<br>After Ageing<br>Tensile Strength<br>18Mpa Minimal for all<br>sizes | Pass |
| | ASTM D3577-09(Reapproved<br>2015) Standard Specification for<br>Rubber Surgical Gloves | To Determine the<br>physical properties<br>Ultimate Elongation | Before Ageing<br>Ultimate Elongation<br>750% Min for all sizes<br>After Ageing<br>Ultimate Elongation<br>560% Min for all sizes | Pass |
| Watertight<br>test | ASTM D5151-19 Standard Test<br>Method for Detection of Holes in<br>Medical Gloves | To determine the<br>watertightness of the test<br>gloves | Sample size: 200 pcs<br>Inspection level : GI<br>AQL 1.5<br>Criteria:≤8 pieces | Pass |
| Residual<br>powder | ASTM D6124-06 (Reapproved<br>2017) Standard Test Method for | To determine the<br>residual powder in the | 2 mg per glove or less | Pass |
| | Residual Powder on Medical<br>Gloves | gloves | | |
| Protein<br>content | ASTM D5712 - 15, Standard Test<br>Method for Analysis of Aqueous<br>Extractable Protein in Natural<br>Rubber. | To determine the<br>extractable protein in the<br>gloves. | $200 \mu g/dm^2$ Max for<br>all sizes | Pass |
| Skin<br>Sensitization | ISO 10993-10<br>Biological Evaluation of<br>Medical Devices Test for<br>Irritation and Skin<br>Sensitization. Test done<br>for irritation | The test was designed to<br>evaluate the potential of<br>a test article to cause<br>skin sensitization. | Under the conditions<br>of the study not a<br>sensitizer | Pass |
| Irritation | ISO 10993-10<br>Biological Evaluation of<br>Medical Devices Test for<br>Irritation and Skin<br>Sensitization. Test done<br>for irritation | To evaluate the potential<br>skin irritation caused by<br>test article contact with<br>the skin surface of<br>rabbits. | Under the condition of<br>study not an irritant | Pass |
| Systemic<br>toxicity | ISO 10993-11:2017 biological<br>evaluation of medical devices - part<br>11, tests for systemic toxicity | The test article was<br>evaluated to determine<br>whether leachables<br>extracted from the test<br>article would cause acute<br>systemic toxicity<br>following injection into<br>mice. | Under the conditions<br>of study the device<br>extracts do not pose a<br>systemic toxicity<br>concern | Pass |
| Pyrogen | USP 43-NF 38<151> Pyrogen Test | The test article was<br>evaluated for the risks of<br>febrile reaction in the<br>rabbit to the<br>administration by<br>injection. | Under the condition of<br>study not a pyrogen | Pass |
| Bacterial<br>Endotoxin | USP 43 <85> Bacterial<br>EndotoxinTest | Bacterial EndotoxinTest | <20 EU/ pair of gloves | Pass |
| EO Residue | ISO 10993-7:2008 | Determine if the<br>Ethylene Oxide residues<br>of test article is within<br>the requirements. | ≤10ug/cm² | Pass |
| ECH<br>Residue | ISO 10993-7:2008 | Determine if the<br>Ethylene Chlorohydrin | ≤5mg/cm² | Pass |
| | | residues of test article is<br>within the requirements. | | |
{8}------------------------------------------------
{9}------------------------------------------------
#### VIII. SUMMARY OF CLINICAL TESTING No clinical testing is included in this submission.
# IX. CONCLUSION
The conclusions drawn from the non-clinical performance data demonstrate that the subject device, Disposable Sterilized Latex Surgical Glove, is as safe, as effective, and performs as well as or better than the legally marketed predicate device K171550.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.