K211941 · Energetic Designs, Inc. · DSA · Jan 27, 2022 · Cardiovascular
Device Facts
Record ID
K211941
Device Name
9Line
Applicant
Energetic Designs, Inc.
Product Code
DSA · Cardiovascular
Decision Date
Jan 27, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2900
Device Class
Class 2
Indications for Use
9Line patient leadwires are intended to be used to connect electrodes placed at appropriate sites on the patient to an ECG or EKG monitoring device for general monitoring and/or diagnostic evaluation by a health care professional. The device is intended for single-patient use.
Device Story
9Line consists of multiple patient leadwires with snap or pinch terminations; connects patient-affixed electrodes to external trunk cables or directly to ECG/EKG monitors. Transmits ECG signals from patient to monitor for diagnostic/monitoring purposes. Used in healthcare facilities by healthcare professionals. Color-coded per AAMI recommendations. Single-patient use; non-sterile. Benefits include reliable signal transmission for cardiac monitoring.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-5:2009, ISO 10993-10:2010) confirming non-cytotoxic, non-sensitizing, and non-irritating properties; electrical safety (IEC 60601-1:2005); and mechanical safety (ANSI/AAMI EC 53:2013).
Technological Characteristics
Medical-grade PVC jacket; shielded lead wires; snap or pinch terminations; 3, 5, 6, or 10 lead configurations; 30" or 50" lengths. Standards: ISO 10993-5:2009, ISO 10993-10:2010 (biocompatibility), IEC 60601-1:2005 (electrical safety), ANSI/AAMI EC 53:2013 (mechanical safety). Passive device; no power source.
Indications for Use
Indicated for adult patients requiring ECG or EKG monitoring or diagnostic evaluation in a healthcare facility. Intended for single-patient use.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 27, 2022
Energetic Designs, Inc. % John Gillespy President FDA 510k Consultants, LLC 1100 Del Lago Circle #104 Palm Beach Gardens, Florida 33410
Re: K211941
Trade/Device Name: 9Line Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer And Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: December 22, 2021 Received: December 27, 2021
Dear John Gillespy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211941
Device Name 9Line
#### Indications for Use (Describe)
9Line patient leadwires are intended to be used to connect electrodes placed at appropriate sites on ECG or EKG monitoring device for general monitoring and/or diagnostic evaluation by a healthcare professional. The device is intended for single-patient use.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|--|
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |
| <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| 1. 510(k) Submitter: | Energetic Designs, Inc.<br>1100 Del Lago Cir, Palm Beach Gardens, FL<br>33410 Phone: 386-243-4332<br>Email: john@fda510kconsultants.com |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Company Contact: | John F. Gillespy, MBA |
| 3. Date of Submission: | June 17, 2021 |
| 4. 510(k) Preparer: | John F. Gillespy, MBA<br>FDA 510k Consultants, LLC<br>Palm Beach Gardens, FL 33410<br>Phone: 386-243-4332<br>Email: john@fda510kconsultants.com |
| 5. Device Classification: | Trade name: 9Line<br>Common name: ECG Disposable Lead Wires<br>Device: Cable, Transducer & Electrode, Patient,<br>(Including Connector)<br>Class: II<br>Regulation #: 870.2900<br>Product Code: DSA |
| 6. Predicate: | Applicant: APK Technology Co.<br>Device: ECG Disposable Lead Wires<br>510(k) Number: K170536 |
### 7. Indications For Use
9Line patient leadwires are intended to be used to connect electrodes placed at appropriate sites on the patient to an ECG or EKG monitoring device for general monitoring and/or diagnostic evaluation by a health care professional. The device is intended for single-patient use.
### 8. Device Description
9Line is comprised of multiple patient leadwires that connect electrodes affixed to the patient's body either to an external trunk cable or directly into an electrocardiogram (ECG) monitor. Each
Image /page/3/Picture/7 description: In the image, a coiled white cable with multiple colored clips is displayed. The clips are brown, red, white, green, and black. The clips are attached to the end of the cable. The cable is coiled in a circular shape.
leadwire connects to a separate patient electrode (see photo of product at right). Leadwires and electrode placement are color-coded in accordance with Association for the Advancement of Medical Instrumentation (AAMI) recommendations.
The device transmits ECG signals from the patient for both diagnostic and monitoring purposes.
{4}------------------------------------------------
9Line is for prescription-use only. The product is sold non-sterile for single-patient use.
### 9. Device Models
9Line models are shown below:
| PART NUMBER | DESCRIPTION (SNAP TERMINATION) |
|-------------|-------------------------------------------------------------------------------|
| 3SS-030 | 30" 3 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 3SS-050 | 50" 3 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5SS-030 | 30" 5 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5SS-050 | 50" 5 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5VSS-030 | 30" 5 V Leads Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5VSS-050 | 50" 5 V Leads Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-3SS-030 | 30" 3 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-3SS-050 | 50" 3 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-6SS-030 | 30" 6 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-6SS-050 | 50" 6 Lead Snap Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| PART NUMBER | DESCRIPTION (PINCH TERMINATION) |
|-------------|--------------------------------------------------------------------------------|
| 3PS-030 | 30" 3 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 3PS-050 | 50" 3 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5PS-030 | 30" 5 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5PS-050 | 50" 5 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5VPS-030 | 30" 5 V Leads Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| 5VPS-050 | 50" 5 V Leads Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-3PS-030 | 30" 3 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-3PS-050 | 50" 3 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-6PS-030 | 30" 6 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
| NK-6PS-050 | 50" 6 Lead Pinch Single Patient Use AAMI<br>w/Shielded Systems Bonded Cable |
{5}------------------------------------------------
## 11. Comparison To Predicate Device
The following table compares 9Line with the predicate device:
# Comparison Table
| Characteristics | Device | Predicate | Comparison |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Name | 9Line | ECG Disposable Lead<br>Wires | NA |
| Applicant | Energetic Designs Inc | APK Technology Co | NA |
| 510k Number | Applied For | K170536 | NA |
| Device Photo | Image: 9Line device photo | Image: ECG Disposable Lead Wires device photo | NA |
| Classification | II | II | Same |
| Regulation Number | 870.2900 | 870.2900 | Same |
| Product Code | DSA | DSA | Same |
| Intended Use | Connect patient<br>electrodes with ECG for<br>monitoring or<br>diagnostic evaluation. | Connect patient<br>electrodes with ECG for<br>monitoring or<br>diagnostic evaluation. | Same |
| Indications For Use | 9Line patient leadwires<br>are intended to be used<br>to connect electrodes<br>placed at appropriate<br>sites on the patient to<br>an ECG or EKG<br>monitoring device for<br>general monitoring<br>and/or diagnostic<br>evaluation by a health<br>care professional. The<br>device is intended for<br>single-patient use. | The ECG Disposable<br>Lead Wires are<br>intended to be used<br>with ECG. The lead wire<br>is used to connect<br>electrodes placed at<br>appropriate sites on the<br>patient to ECG for<br>general monitoring<br>and/or diagnostic<br>evaluation by health<br>care professional. | Same |
| Target Population | Adults | Adults | Same |
| Anatomical Site | Upper body | Upper body | Same |
| Where Used | Health care facility | Health care facility | Same |
| Reuse | Single patient only | Single patient only | Same |
| Rx/OTC/Both | Rx | Rx | Same |
{6}------------------------------------------------
| System Description | ECG lead wires with<br>terminal connectors | ECG lead wires with<br>terminal connectors | Same |
|--------------------------------------|-----------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|------------------|
| Design Concept | Lead wires transmit ECG<br>signals from patient to<br>trunk cable or<br>monitoring device | Lead wires transmit ECG<br>signals from patient to<br>trunk cable or<br>monitoring device | Same |
| Lengths | 30", 50" | 50" (per brochure) | Minor difference |
| Lead Sets | 3, 5, 6, 10 (by<br>combining two sets of<br>5 leads: one limb lead<br>set and one set of<br>vleads) | 3, 5, 6, 10 (brochure) | Same |
| Termination | Snap, Pinch | Snap, Pinch | Same |
| Multiple Models<br>Based on Features | Yes | Yes | Same |
| Peelable Ribbon | Yes | Yes | Same |
| Material<br>Composition | Medical-grade PVC<br>jacket | PVC | Same |
| Powered | No | No | Same |
| Components | | | |
| Shielded Lead Wires | Yes | Yes | Same |
| Patient Termination<br>Connector | Yes (Snap & Pinch) | Yes (Snap & Pinch) | Same |
| Trunk/Monitor<br>Connector | Yes | Yes | Same |
| Trunk Cable | No | No | Same |
| Accessories | None | None | Same |
| S&E Testing | | | |
| Biocompatibility | ISO 10993-5:2009<br>ISO 10993-10:2010 | ISO 10993-5:2009<br>ISO 10993-10:2010 | Same |
| Electrical Safety | IEC 60601-1:2005<br>MOD (Part 8.5.2.3) | IEC 60601-1:2005<br>MOD (Part 8.5.2.3) | Same |
| Mechanical Safety | ANSI/AAMI EC 53:<br>2013 | ANSI/AAMI EC 53:<br>2013 | Same |
Physical Characteristics
The subject device is comparable to the predicate in terms of design, function, materials, and performance. None of the differences raised different issues of safety and effectiveness.
### Non-Clinical Testing
9Line passed the following non-clinical tests, all of which were performed to current FDArecognized standards:
- . Biocompatibility, ISO 10993-5:2009 and ISO 10993-10:2010... Device is non-cytotoxic, non-sensitizing, and non-irritating.
- Electrical & Mechanical Safety, IEC 60601-1 (Part 8.5.2.3) and ANSI/AAMI EC 53:2013... 510k Summary – Page 4
{7}------------------------------------------------
Device meets electrical and mechanical safety standards.
#### Substantial Equivalence
9Line patient leadwires successfully followed the pathway to Substantial Equivalence in the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications" (2014). The steps are summarized below:
- The predicate device is legally marketed and was found substantially equivalent through 510(k) premarket submission.
- The subject and predicate devices have the same intended use (and indications for use).
- Technological differences between the subject and predicate were evaluated; none of the differences raised different issues of safety and effectiveness.
- . The following methods for evaluation of the effects of different characteristics on safety and effectiveness were deemed acceptable—testing for biocompatibility, electrical safety, and mechanical performance. Evaluation methods were conducted to FDArecognized standards where applicable.
- . Data from these tests demonstrated equivalence and support the indications for use.
In summary, all necessary testing has been performed and the results support the conclusion that 9Line patient leadwires is substantially equivalent to the legally marketed predicate, based on both (a) comparison of intended use, materials, technology, and design and (b) testing to FDA-recognized standards, and the device thus does not raise any concerns of safety or effectiveness.
Based on the information contained within this submission, the applicant concludes that 9Line is substantially equivalent to the identified predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.