K211940 · Shenzhen Taikang Medical Equipment Co., Ltd. · KNG · Jan 7, 2022 · Obstetrics/Gynecology
Device Facts
Record ID
K211940
Device Name
Fetal Doppler
Applicant
Shenzhen Taikang Medical Equipment Co., Ltd.
Product Code
KNG · Obstetrics/Gynecology
Decision Date
Jan 7, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2660
Device Class
Class 2
Indications for Use
The Fetal Doppler is used to detect the fetal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in hospital, clinic, community and home. The device is intended for use at or after 16 weeks gestation.
Device Story
Hand-held, battery-powered Doppler ultrasound device; detects fetal heart rate (FHR). System comprises main unit (main board, power module, battery, speaker, LCD) and probe (ultrasonic transducers). Probe transmits continuous wave ultrasound at 2.0, 2.5, or 3.0 MHz; receives reflected signals. Used by clinicians or patients in clinical/home settings. Device processes reflected signals to calculate FHR; displays rate on LCD; provides audio output via speaker. Assists clinicians in monitoring fetal well-being; enables non-invasive assessment of fetal heart activity.
Clinical Evidence
Bench testing only. Includes biocompatibility (ISO 10993-5, -10), electrical safety (IEC 60601-1, -1-11), electromagnetic compatibility (IEC 60601-1-2), battery safety (IEC 62133), and acoustic output (NEMA UD 2-2004). FHR accuracy verified per IEC 60601-2-37; measuring range 50-210 bpm with ±2 bpm accuracy and 1 bpm resolution.
Indicated for detection of fetal heart rate in patients at or after 16 weeks gestation. For use by healthcare professionals including nurses, midwives, and specialized technicians in hospital, clinic, community, or home settings.
Regulatory Classification
Identification
A fetal ultrasonic monitor is a device designed to transmit and receive ultrasonic energy into and from the pregnant woman, usually by means of continuous wave (doppler) echoscopy. The device is used to represent some physiological condition or characteristic in a measured value over a period of time (e.g., perinatal monitoring during labor) or in an immediately perceptible form (e.g., use of the ultrasonic stethoscope). This generic type of device may include the following accessories: signal analysis and display equipment, electronic interfaces for other equipment, patient and equipment supports, and component parts. This generic type of device does not include devices used to image some relatively unchanging physiological structure or interpret a physiological condition, but does include devices which may be set to alarm automatically at a predetermined threshold value.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
January 7, 2022
Shenzhen Taikang Medical Equipment Co., Ltd. Shigui Du General Manager 3F East, Building 4, Lanzhu East Road 8, Grand Industrial Park, Pingshan District Shenzhen, Guangdong 518015 CHINA
Re: K211940
> Trade/Device Name: Fetal Doppler Regulation Number: 21 CFR 884.2660 Regulation Name: Fetal Ultrasonic Monitor And Accessories Regulatory Class: II Product Code: KNG Dated: December 6, 2021 Received: December 6, 2021
Dear Shigui Du:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K211940
Device Name Fetal Doppler
Indications for Use (Describe)
The Fetal Doppler is used to detect the feal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in hospital, clinic, community and home. The device is intended for use at or after 16 weeks gestation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K211940
Prepared in accordance with the requirements of SMDA and 21 CFR §807.92
## 1.0 Information of Submitter and Correspondent
#### Submitter's information:
Company Name: Shenzhen Taikang Medical Equipment Co., Ltd.
| Street Address: | 3F East, Building 4, Lanzhu East Road 8, Grand Industrial Park,<br>Pingshan District |
|------------------|--------------------------------------------------------------------------------------|
| City: | Shenzhen |
| State/ Province: | Guangdong |
| Country: | China |
| Telephone: | +86(755)-89664986 |
| Fax: | +86(755)-89664986 |
| Contact Person: | Shigui Du |
| Contact Title: | General Manager |
| Contact Email: | dushigui@taikang-medical.com |
#### Submission correspondent's information:
| Name: | Shenzhen Reanny Medical Devices Management Consulting Co., Ltd |
|-----------------|---------------------------------------------------------------------------------------------------------------------|
| Address: | Room 2012#, Gebu Commercial Building, Hongxing Community<br>Songgang Street, Baoan District, Shenzhen 518000, China |
| Contact Person: | Reanny Wang |
| E-mail: | reanny@reanny.com |
#### Date Summary Prepared: January 5, 2022
#### 2.0 Device Information
| Trade/Device Name | Fetal Doppler |
|--------------------|------------------------------------------|
| Model: | TK-T802CR |
| Regulation Name: | Fetal ultrasonic monitor and accessories |
| Requlation Number: | 21 CFR 884.2660 |
| Common Name: | Fetal Doppler |
| Product Code: | KNG (monitor, ultrasonic, fetal) |
| Regulatory Class: | Class II |
| Review Panel: | Obstetrics/Gynecology |
{4}------------------------------------------------
## 3.0 Predicate Device Information
| Sponsor: | Vcomin Technology Limited Company |
|----------------|-----------------------------------------------------------------------|
| Device: | Fetal Doppler, Model: FD-200D |
| 510(K) Number: | K182526 |
| | The predicate device has not been subject to a design-related recall. |
## 4.0 Device Description
The Fetal Doppler, model TK-T802CR, is a hand-held, battery powered audio Doppler device used for detecting fetal heart beats and sounds. The Fetal Doppler is used for non-invasive detection and display of the fetal heart rate (FHR) utilizing Doppler ultrasound. It has two hand-held components, a main unit and a probe. The main unit consists of the main board, power module, battery, speaker, and LCD screen. The probe consists of the ultrasonic transducers for transmission and for signal reception. The ultrasonic signal is continuously transmitted at a frequency of 2.0, 2.5, or 3.0 MHz, depending on the probe settings selected.
## 5.0 Indications for Use
The Fetal Doppler is used to detect the fetal heart rate. The device should be used by health care professionals including nurses, midwives, and specialized technicians in hospital, clinic, community and home. The device is intended for use at or after 16 weeks gestation.
## 6.0 Comparison to the Predicate Device
The following table compares the subject device to the predicate device with respect to the intended use and technological characteristics:
| Device & Predicate<br>Device(s): | Subject Device | Predicate Device | Comparison |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | K211940 | K182526 | |
| | Fetal Doppler<br>Model: TK-T802CR | Fetal Doppler<br>Model: FD-200D | |
| Indications for Use | The Fetal Doppler is used to<br>detect the fetal heart rate. The<br>device should be used by health<br>care professionals including<br>nurses, midwives, and<br>specialized technicians in<br>hospital, clinic, community and<br>home. The device is intended for<br>use at or after 16 weeks<br>gestation. | The device is used to detect the<br>fetal heart rate. The device<br>should be used by health care<br>professionals including nurses,<br>midwives, and specialized<br>technicians in hospital, clinic,<br>community and home. | Same |
| Design | A main unit and a probe. The<br>main unit displays FHR | A main unit and a probe. The<br>main unit displays FHR | Same |
| Mode of action | Doppler ultrasound, continuous<br>wave | Doppler ultrasound, continuous<br>wave | Same |
| Ultrasound<br>frequency | 2.0 MHz, 2.5 MHz, and 3.0 MHz | 2.0 MHz, 2.5 MHz, and 3.0 MHz | Same |
{5}------------------------------------------------
| | FHR Specifications | FHR Measuring Range: 50-210 bpm | FHR Measuring Range: 50-210 bpm | Same |
|------------------------------------------------|--------------------|---------------------------------|---------------------------------|-----------|
| | | Accuracy: ±2 bpm | Accuracy: ±2 bpm | |
| | | Resolution: 1 bpm | Resolution: 1 bpm | |
| | | | | |
| Acoustic output | | 2.0MHz - ISATA: 6.01mW/cm² | 2.0MHz - ISATA: 17.24 mW/cm² | Different |
| | | 2.5MHz - ISATA: 6.67mW/cm² | 2.5MHz - ISATA: 18.57 mW/cm² | |
| | | 3.0MHz - ISATA: 6.87mW/cm² | 3.0MHz - ISATA: 11.496 mW/cm² | |
| Material - Patient<br>Contacting<br>Components | | ABS | ABS, Silicone, Colorants | Same |
The subject and predicate device have the same intended use (i.e., to detect the fetal heartbeat). The subject and predicate device have different technological characteristics, including different acoustic output specifications, physical dimensions, and weight. The differences in technological characteristics do not raise new questions of safety and effectiveness.
# 7.0 Performance Summary
The following performance data were provided in support of the substantial equivalence determination:
- a. Biocompatibility
Biocompatibility testing on the patient-contacting components of the subject device in accordance with the FDA guidance "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" dated September 4, 2020. Testing included the following assessments:
- Cytotoxicity per ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: ● Tests for in vitro cytotoxicity
- . Sensitization per ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
- Irritation per ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: ● Tests for irritation and skin sensitization
The patient-contacting materials were shown to be non-cytotoxic, non-irritating, and nonsensitizing.
- b. Electrical Safety, Electromagnetic Compatibility, and Wireless Technology
- IEC 60601-1:2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical electrical . equipment - Part 1: General requirements for basic safety, and essential performance with US deviations per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012.
- . IEC 60601-1-11:2015, Medical electrical equipment - Part 1-11: General requirements for basic safety, and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2:2014, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility
- IEC 62133:2012 - Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
- c. Software Verification
{6}------------------------------------------------
Software verification and validation was conducted and software documentation was provided in accordance with the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" dated May 11, 2005 for a moderate software level of concern.
- d. Performance Testing
The following testing was provided to support the safety and effectiveness of the subject device:
- i. Use-Life Testing
- ii. Fetal Heart Rate Accuracy Testing
- Testing per IEC 60601-2-37 Edition 2.1 2015 Medical electrical equipment Part iii. 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- iv. Acoustic output testing per NEMA UD 2-2004(R2009) Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment and the acoustic output measurement methodology as recommended in FDA guidance document "Marketing Clearance of Diagnostic Ultrasound Systems and Transducers" dated June 27, 2019 was followed for Track 1 devices.
## 9. Conclusion
The non-clinical performance data described above demonstrate that the Fetal Doppler is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.