The ORIO-Ti cervical intervertebral body fusion cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six (6) weeks of non-operative treatment with an intervertebral cage. ORIO-Ti cervical intervertebral body fusion implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C-3 to C-7 disc levels using autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. ORIO-Ti cervical intervertebral body fusion implant is to be used with supplemental fixation that has been cleared for use in the cervical spine. The ORIO-Ti lumbar intervertebral body fusion cage is indicated for intervertebral body fusion procedures in sketally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six (6) months of non-operative treatment. ORIO-Ti lumbar intervertebral body fusion implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via a transforaminal, open posterior, anterior/anterolateral, or lateral approach. The ORIO-Ti lumbar intervertebral body fusion implants are to be used with supplemental fixation that have been cleared for use in the lumbosacral spine.
Device Story
ORIO-Ti Intervertebral Body Fusion Cage System; box/trapezoidal-shaped titanium alloy implants; designed for cervical (ACDF) or lumbar (PLIF, TLIF, ALIF, lateral) spinal fusion. Implants feature internal windows for bone graft placement and external surface teeth to prevent migration. Used by surgeons in clinical settings to facilitate fusion in patients with DDD. Requires supplemental fixation cleared for spinal use. Benefits include stabilization of spinal segments and promotion of bone fusion.
Clinical Evidence
No clinical studies were performed.
Technological Characteristics
Titanium alloy Ti-6Al-4V ELI (ASTM F136). Box/trapezoidal form factor with surface teeth. Mechanical testing per ASTM F2077-18 (compression, shear, torsion), ASTM F2267-04 (subsidence), and MED-SPN-EXP (expulsion).
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (one level, C3-C7) with radicular symptoms, or lumbar spine (one or two levels, L2-S1) with discogenic back pain. Lumbar patients may have up to Grade 1 spondylolisthesis or retrolisthesis and prior non-fusion surgery. Requires 6 weeks (cervical) or 6 months (lumbar) of non-operative treatment. Must be used with supplemental fixation and bone graft.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image contains the logos of the Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
September 30, 2021
SpineCraft, LLC Ami Akallal-Asaad Vice President of Regulatory Affairs & Quality Assurance 777 Oakmont Lane, Suite 200 Westmont, Illinois 60559
Re: K211935
Trade/Device Name: ORIO-Ti Intervertebral Body Fusion Cage System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: ODP, MAX Dated: August 27, 2021 Received: September 2, 2021
Dear Ms. Akallal-Asaad:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211935
### Device Name
ORIO-Ti Intervertebral Body Fusion Cage System
### Indications for Use (Describe)
The ORIO-Ti cervical intervertebral body fusion cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six (6) weeks of non-operative treatment with an intervertebral cage. ORIO-Ti cervical intervertebral body fusion implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C-3 to C-7 disc levels using autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. ORIO-Ti cervical intervertebral body fusion implant is to be used with supplemental fixation that has been cleared for use in the cervical spine.
The ORIO-Ti lumbar intervertebral body fusion cage is indicated for intervertebral body fusion procedures in sketally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six (6) months of non-operative treatment. ORIO-Ti lumbar intervertebral body fusion implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via a transforaminal, open posterior, anterior/anterolateral, or lateral approach. The ORIO-Ti lumbar intervertebral body fusion implants are to be used with supplemental fixation that have been cleared for use in the lumbosacral spine.
Type of Use (Select one or both, as applicable)
| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# SpineCraft ORIO-Ti Intervertebral Body Fusion Cage System
In accordance with 21 CFR 807.92 of the Federal Code of Regulations the following 510(k) summary is submitted for the ORIO-Ti Cage System.
Date Prepared: August 27, 2021
| 1. Submitter: | Contact Person: |
|------------------------|-----------------------------|
| SpineCraft, LLC | Ami Akallal-Asaad |
| 777 Oakmont Lane | VP, Regulatory Affairs & QA |
| Westmont, IL 60559 USA | SpineCraft, LLC |
| Tel: 1 630-920-7300. | a.asaad@spinecraft.com |
| Fax: 1 630-920-7310 | |
- ORIO-Ti Intervertebral Body Fusion Cage System 2. Trade name:
| Common Name: | Intervertebral body fusion device |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name: | Intervertebral body fusion device - Cervical per ODP 21 CFR 888.3080<br>Intervertebral body fusion device - Lumbar per MAX 21 CFR 888.3080<br>Class II |
### 3. Predicate or legally marketed devices which are substantially equivalent:
## Primary
- ORIO Intervertebral Body Fusion Cages (K090887) SpineCraft, LLC .
# Additional predicate devices
- . BAK/C Vista Interbody Fusion (P980048) Zimmer
- . Shoreline ACS Anterior Cervical Standalone (K170569) SeaSpine
- IVA Cage Ti (ACIF, PLIF, TLIF, DLIF, and ALIF) (K173080) Huvexel Co., Ltd ●
- Brigade Hyperlordotic ALIF Lumbar Cage (K123045) Nuvasive ●
- APEX-DL Spine System (K143683) SpineCraft, LLC
- ASTRA Spine System (K150417) SpineCraft, LLC ●
### 4. Description of the device:
The subject devices are intervertebral body fusion devices for use in cervical lumbar spinal surgery. They may also be referred to as interbody fusion devices or interbody cages. The devices are generally boxshaped or trapezoidal shaped with windows or pockets in their design to allow for the placement of autograft or allogenic bone graft. The ORIO-Ti Cage System is a Titanium alloy interbody fusion system comprised of various device configurations based on surgical approach and patient anatomy, and may be implanted via one of the following approaches: as a single device via an anterior cervical (ACDF), bilaterally in pairs via a posterior (PLIF) approach; as a single device via a transforaminal (TLIF) approach; or as a single device via an anterior (ALIF) approach. The exterior of the devices has "teeth" on the superior and inferior surfaces to help prevent the devices from migrating once they are implanted.
{4}------------------------------------------------
## Materials:
Titanium alloy Ti-6Al-4V ELI per ASTM F136.
### Summary of Similarities and Differences in Technological Characteristics and Performance 5.
ORIO-Ti Intervertebral Body Fusion Cage System is substantially equivalent to the ORIO cervical caqe (K090887), the the BAK/C Vista (P980048), the Shoreline ACS (K170569), the ORIO lumbar cages (K090887), IVA Cage Ti (ACIF, PLIF, TLIF, DLIF, and ALIF) (K173080) , the Brigade Hyperlordotic ALIF Lumbar Cage (K123045) in terms of intended use, design, materials used, mechanical safety and/or performances.
### 6. Indications for Use:
The ORIO-Ti cervical intervertebral body fusion cage is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one disc level. Degenerative disc disease is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should have had at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. ORIO-Ti cervical intervertebral body fusion implants are used to facilitate intervertebral body fusion in the cervical spine and are placed via an anterior approach at the C-3 to C-7 disc levels using autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft. ORIO-Ti cervical intervertebral body fusion implant is to be used with supplemental fixation that has been cleared for use in the cervical spine.
The ORIO-Ti lumbar intervertebral body fusion cage is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients may have had a previous non-fusion spinal surgery at the involved spinal level(s). These patients should have had at least six (6) months of non-operative treatment. ORIO-Ti lumbar intervertebral body fusion implants are to be used with autogenous and/or allogeneic bone graft comprised of cancellous and/or corticocancellous bone graft and implanted via a transforaminal, open posterior, anterior/anterolateral, or lateral approach. The ORIO-Ti lumbar intervertebral body fusion implants are to be used with supplemental fixation that have been cleared for use in the lumbosacral spine.
### Non-clinical Test Summary: 7.
The following tests were conducted:
- . Static and dynamic axial compression (ASTM F2077-18)
- . Static and dynamic shear-compression (ASTM F2077-18)
- Static and dynamic Torsion (ASTM F2077-18) ●
- . Subsidence (ASTM F2267-04 Reapproved 2018)
- Expulsion (MED-SPN-EXP 2nd Edition, Rev. 0) ●
The results of this testing were compared to predicate systems, with the results being equal to or higher than the predicate systems.
### 8. Clinical Test Summary
No clinical studies were performed
### Conclusions Nonclinical and Clinical 9.
Based on the supporting evidence provided, we believe that the subject ORIO-Ti Cage System is substantially equivalent to the predicate devices in terms of indications for use, design, material, performance and function.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.