K211932 · Oak Tree Engineering, LLC · KWQ · Jul 27, 2021 · Orthopedic
Device Facts
Record ID
K211932
Device Name
Fixxsure Anterior Lumbar Plate and Screw System
Applicant
Oak Tree Engineering, LLC
Product Code
KWQ · Orthopedic
Decision Date
Jul 27, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FIXXSURE™ Anterior Lumbar Plate and Screw System is indicated for use via the anterior, lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc conformed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.
Device Story
The FIXXSURE™ Anterior Lumbar Plate and Screw System is a spinal fixation orthosis used by surgeons to stabilize the lumbar and lumbosacral spine. The system consists of titanium alloy plates and screws of varying sizes. Plates feature integrated spikes for stability during screw insertion and a curvature designed for anatomic fit. The device is implanted during anterior, lateral, or anterolateral surgical procedures to provide structural support as an adjunct to spinal fusion. By securing the vertebral bodies, the system helps maintain alignment and stability, potentially reducing pain and facilitating bone fusion in patients with spinal instability or deformity.
Clinical Evidence
Bench testing only. Mechanical performance evaluated via static compression, static tension, static torsion, and dynamic compression testing per ASTM F1717-18. Results showed mechanical values similar to or higher than the predicate device.
Technological Characteristics
Constructed from Ti-6Al-4V ELI titanium alloy (ASTM F136). System includes plates with integrated spikes and screws (5.5-6.0 mm diameter, 20-55 mm length). Mechanical testing performed per ASTM F1717-18. No software or electronic components.
Indications for Use
Indicated for patients requiring spinal fixation as an adjunct to fusion in the lumbar and lumbosacral (L1-S1) spine. Conditions include fracture, tumor, degenerative disc disease, pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities, spinal stenosis, or failed previous surgery.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
Oak Tree Engineering, LLC. % Douglas W. Neary Consultant 419 Main Street, #242 Huntington Beach, CA 92648
Re: K211932
Trade/Device Name: FIXXSURE™ Anterior Lumbar Plate and Screw System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: KWQ Dated: April 5, 2021 Received: June 22, 2021
Dear Douglas W. Neary:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR
July 27, 2021
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211932
### Device Name
FIXXSURE™ Anterior Lumbar Plate and Screw System
### Indications for Use (Describe)
The FIXXSURE™ Anterior Lumbar Plate and Screw System is indicated for use via the anterior, lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc conformed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spinal stenosis, or a failed previous spine surgery.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Oak Tree Engineering. The logo features a white oak tree on the left side of the image. To the right of the tree, the words "OAK TREE ENGINEERING" are written in white, with the word "ENGINEERING" in a smaller font size. Below that, the words "Project & Process Engineering" are written in an even smaller font size. The background of the logo is a dark blue color.
# 510(k) SUMMARY
K211932 Pg. 1 of 3
The following 510(k) summary is being submitted as required by 21 CFR 807.92:
- 1 Manufacturer
| Submitter: | Oak Tree Engineering, LLC. |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| | 419 Main Street, #242, Huntington Beach, California 92648.<br>Phone: 805-797-3003<br>Website: https://www.oaktreellc.com |
| Contact Person: | Douglas W. Neary<br>douglaswneary@aol.com |
| Date | April 5, 2021 |
#### 2. Device Identification
| Trade Name: | FIXXSURE™ Anterior Lumbar Plate and Screw System |
|-------------------------|--------------------------------------------------|
| Common Name: | Spinal Fixation Appliances |
| Product Code: | KWQ |
| Classification: | Class II |
| Classification<br>Name: | Spinal intervertebral body fixation orthosis |
| Regulation No. | 21 CFR 888.3060 |
| Classification Pane | Orthopedic |
#### 3. Predicate or legally marketed devices which are substantially equivalent.
The FIXXSURE™ Anterior Lumbar Plate and Screw System has undergone a comprehensive battery of non-clinical testing, including chemical, The testing supports a determination of substantial equivalence. The FIXXSURE™ Anterior Lumbar Plate and Screw System met all pre-defined acceptance criteria and, in tests where it was compared to either AccelFix Anterior Lumbar Plate and Screw System, the predicate or reference device, was found not to represent a new worst case. Overall, the results of the performance bench tests support the substantial equivalence of the Subject device.
The design feature and indications for use for the subject device 'FIXXSURE™ Anterior Lumbar Plate and Screw System' is substantially equivalent to the following predicate:
- 1) Primary Predicate Device: AccelFix Lumbar Plate System (K192481)
#### 4 Materials
FIXXSURE™ Anterior Lumbar Plate and Screw System is manufactured from Ti-6Al-4V ELI titanium alloy (ASTM F136). This this is the same material used in the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Oak Tree Engineering. The logo is set on a blue background. On the left side of the logo is a white tree. To the right of the tree is the text "OAK TREE ENGINEERING" in white, with the words "Project & Process Engineering" in a smaller font below.
#### న. Description of the Device
The FIXXSURE™ Anterior Lumbar Plate and Screw System's implants are a lumbar Plate System intended for use as an aid in spinal fixation. They are made of Titanium 6AL-4V Alloy (ASTM F136).
The FIXXSURE™ Anterior Lumbar Plate and Screw System consists of a variety of shapes and sizes of plates and screws.
The plate has been designed to include spikes for added stability and alignment during screw insertion. The plates feature a curvature for anatomic fit. The diameter of the screw is available from 5.5 to 6.0 mm and the length from 20 to 55 mm.
#### 6. Indications for Use
The FIXXSURE™ Anterior Lumbar Plate and Screw System is indicated for use via the anterior, lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc conformed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, scoliosis, spondylolisthesis, lordotic deformities of the spine, spinal stenosis, or a failed previous spine surgery.
#### 7. Performance -Bench testing
The FIXXSURE™ Anterior Lumbar Plate and Screw System was tested according to ASTM F 1717-18.
- 1) Static Compression Bending Test ASTM F1717-18
- Static Tension Bending Test ASTM F1717-18 2)
- 3) Static Torsional Test - ASTM F1717-18
- Dynamic Compression Bending Test ASTM F1717-18 4)
Static compression, Static tensile, Static torsion and Dynamic compression testing was performed as described in ASTM F 1717-18 to demonstrate the substantial equivalence of the FIXXSURE™ Anterior Lumbar Plate and Screw System to the predicate devices.
Bench testing to evaluate the mechanical properties of the FIXXSURE™ Anterior Lumbar Plate and Screw System showed similar or higher mechanical value than the predicate marketed devices.
#### 8. Summary of Technology Characteristics
FIXXSURE™ Anterior Lumbar Plate and Screw System is substantially equivalent to the predicate systems in terms of design, materials, indications for use and sizing.
#### 9. Substantial Equivalence:
FIXXSURE™ Anterior Lumbar Plate and Screw System was shown to be substantially equivalent to the predicate devices in indications for use, design, function, materials used and mechanical performance.
#### 10. Conclusion
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Oak Tree Engineering, which includes a white tree on the left and the company name and tagline on the right. The text "K211932 Pg. 3 of 3" is located at the bottom of the image. The logo is set against a blue background. The tagline reads "Project & Process Engineering".
The information presented demonstrates the substantial equivalency of the FIXXSURE™ Anterior Lumbar Plate and Screw System to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.