Medline Digital Thermometer Non-Lubricated Probe Sheath
K211931 · Medline Industries, Inc. · FLL · Sep 20, 2021 · General Hospital
Device Facts
Record ID
K211931
Device Name
Medline Digital Thermometer Non-Lubricated Probe Sheath
Applicant
Medline Industries, Inc.
Product Code
FLL · General Hospital
Decision Date
Sep 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
These devices are indicated for use as a barrier accessory to digital thermometers while taking rectal, oral or axillary temperature measurements. These sheaths are non-sterile and intended for single use only. This accessory is contraindicated for use with broken skin.
Device Story
Flexible plastic sheath; acts as barrier accessory for digital stick thermometers; prevents contamination of thermometer transducer during temperature measurement. Used in clinical or home settings; operated by clinicians or patients. Sheath covers thermometer probe; provides physical barrier between device and patient tissue (oral, axillary, rectal). Single-use; disposable; non-lubricated; non-sterile. Benefits patient by reducing cross-contamination risk during temperature monitoring.
Clinical Evidence
No clinical data. Bench testing only. Performance testing included compatibility and leakage testing per ASTM E1104-98(2016). Biocompatibility testing included cytotoxicity (ISO 10993-5), skin sensitization (ISO 10993-10), and skin irritation (ISO 10993-10).
Technological Characteristics
Flexible plastic probe sheath; single-use; non-sterile; non-lubricated. Conforms to ASTM E1104-98. Biocompatibility evaluated per ISO 10993-1, ISO 10993-5, and ISO 10993-10.
Indications for Use
Indicated for use as a barrier accessory to digital thermometers for rectal, oral, or axillary temperature measurements. Contraindicated for use on broken skin.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Medline Digital Rectal Thermometer Sheath (K183431)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2021
Medline Industries, Inc. Kelsey Closen Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60093
Re: K211931
Trade/Device Name: Medline Digital Thermometer Non-Lubricated Probe Sheath Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: June 15, 2021 Received: June 22, 2021
Dear Kelsey Closen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211931
Device Name
Medline Digital Thermometer Non-Lubricated Probe Sheath
#### Indications for Use (Describe)
These devices are indicated for use as a barrier accessory to digital thermometers while taking rectal, oral or axillary temperature measurements. These sheaths are non-sterile and intended for single use only. This accessory is contraindicated for use with broken skin.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image features the logo of Medline, a healthcare company. The logo consists of the word "MEDLINE" in a bold, sans-serif font, positioned to the left of a stylized white cross. The cross is composed of four triangular shapes converging at a central point. The logo is set against a blue square background.
edline Industries. Inc Three Lakes Drive Northfield. IL 60093
# 510(k) SUMMARY K211931
# [AS REQUIRED BY 21CFR807.92(c)]
#### Submitter / 510(k) Sponsor
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
Registration Number: 1417592
#### Contact Person
Contact Person: Kelsey Closen, Regulatory Affairs Specialist Phone: 847-949-2283 Email: KClosen(@medline.com
Summary Preparation Date September 20, 2021
Type of 510(k) Submission
Traditional
## Device Name / Classification
Trade Name: Medline Digital Thermometer Non-Lubricated Probe Sheath Common Name: Clinical Electronic Thermometer Regulation Name: Clinical electronic thermometer Product Code: FLL Classification Panel: General Hospital Regulatory Class: II Regulation Number: 21 CFR 880.2910
Predicate Device Medline Digital Rectal Thermometer Sheath, K183431
## Device Description
The Medline Digital Thermometer Non-Lubricated Probe Sheath is an accessory to a clinical electronic thermometer, which is intended to measure body temperature. The Medline Digital Thermometer Non-Lubricated Probe Sheath is a flexible plastic sheath that is used to cover the transducer of clinical electronic thermometers while taking body temperature measurements from the oral, axillary, or rectal measurement sites and acts as a barrier to prevent possible contamination of the device during
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Medline logo. The logo consists of the word "MEDLINE" in a sans-serif font, stacked above a stylized starburst symbol. The logo is set against a blue square background. There is a light gray bar on the right side of the square.
dline Industries. Three Lakes Drive Northfield. IL 60093
temperature measurement. The sheath is an over-the-counter (OTC), disposable, single use, nonlubricated, and non-sterile device that is compatible with most digital stick thermometers.
#### Indications for Use
These devices are indicated for use as a barrier accessory to digital thermometers while taking rectal, oral or axillary temperature measurements. These sheaths are non-sterile and intended for single use only. This accessory is contraindicated for use with broken skin.
#### Summary of Technological Characteristics
# TABLE 1: COMPARISON OF PROPOSED AND PREDICATE DEVICES
| Device Characteristic | Proposed Device | Predicate Device | Comparison<br>Analysis |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product Name | Medline Digital<br>Thermometer Non-<br>Lubricated Probe Sheath | Medline Digital Rectal<br>Thermometer Sheath | N/A |
| 510(k) Reference | TBD | K183431 | N/A |
| Product Owner | Medline Industries, Inc. | Medline Industries, Inc. | Same |
| Indications for Use | These devices are indicated<br>for use as a barrier accessory<br>to digital thermometers while<br>taking rectal, oral or axillary<br>temperature measurements.<br>These sheaths are non-sterile<br>and intended for single use<br>only. This accessory is<br>contraindicated for use with<br>broken skin. | These devices are indicated for<br>use as a barrier accessory to<br>digital thermometers while<br>taking rectal temperature<br>measurements. These sheaths<br>are nonsterile and intended for<br>single use only. This accessory<br>is contraindicated for use with<br>broken skin. | Different |
| Design Standard | Conforms to Standard ASTM<br>E1104-98 | Conforms to Standard ASTM<br>E1104-98 | Same |
| Design Features &<br>Configurations | One size, Clear, Flexible,<br>peel- away package, marking<br>for device insertion | One size, Clear, Flexible, peel-<br>away package, marking for<br>device insertion | Same |
| Prescription vs. OTC | OTC | OTC | Same |
| Application Site | Oral, axillary, rectal | Rectal | Different |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Medline. The logo is set against a blue square background. The word "MEDLINE" is written in white, block letters on the left side of the square. To the right of the word is a white symbol that resembles a stylized cross or plus sign, with the vertical line extending downwards beyond the horizontal line.
ledline Industries, Inc. Three Lakes Drive Northfield. IL 60093
| Sterile vs. Non-Sterile | Non-Sterile | Non-Sterile | Same |
|-----------------------------------|-------------|-------------|------|
| Disposable vs. Non-<br>Disposable | Disposable | Disposable | Same |
| Single Use vs. Reusable | Single Use | Single Use | Same |
#### Discussion
The application site is different between the proposed device and the predicate. Biocompatibility was conducted per ISO 10993, "Biological Evaluation of the Medical Devices – Part 1: Evaluation of Testing within a Risk Management Process." to address this difference.
#### Summary of Non-Clinical Testing
Testing was conducted to demonstrate safety and effectiveness of the Medline Digital Thermometer Non-Lubricated Probe Sheath. A summary of testing is presented below with more information provided in the applicable sections.
## Biocompatibility Testing
The biological evaluation for the Medline Digital Thermometer Non-Lubricated Probe Sheath was conducted in accordance with FDA guidance document, "Use of International Standard ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and ISO 10993-1 "Biological Evaluation of the Medical Devices – Part 1: Evaluation of Testing within a Risk Management Process."
The following Biocompatibility tests were performed:
- ISO 10993-5: Cytotoxicity ●
- ISO 10993-10: 2010: Skin Sensitization ●
- ISO 10993-10: 2010: Skin Irritation ●
# Performance Testing (Bench)
Performance testing on the Medline Digital Thermometer Non-Lubricated Probe Sheath was conducted for compatibility and leakage per the ASTM E1104-98 (2016): Standard Specification for Clinical Thermometer Probe Covers and Sheaths and tensile strength.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Medline. The logo is set against a blue square background. The word "MEDLINE" is written in white, slanted, sans-serif font. A white starburst-like symbol is behind the text, with the text centered on the starburst. There is a gray bar at the top of the blue square.
Medline Industries, Inc. Three Lakes Drive Northfield, IL 60093
#### TABLE 2: Performance Testing
| Test | Testing Standards | Acceptance Criteria |
|---------------|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Compatibility | ASTM E1104-98(2016) 5.4 | Probe cover shall not degrade the measurement time or<br>accuracy of the<br>temperature-taking device such that the probe and temperature-<br>taking unit fails to<br>meet the requirements of ASTM E1112 – § 4.2 |
| Leakage | ASTM E1104-98 (2016) 5.3 | There shall be no continuous bubble stream observed from the<br>probe cover within 5<br>seconds of applying an internal pressure of 8.4 kPa. |
## Summary of Clinical Testing
Not applicable.
#### Conclusion
The differences between the predicate and the subject device do not raise and new or different questions of safety or effectiveness. Medline Industries, Inc. concludes that the Medline Digital Thermometer Non-Lubricated Probe Sheath are substantially equivalent to the predicate device Medline Digital Rectal Thermometer Sheath cleared under K183431.
{7}------------------------------------------------
5-004
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.