OGM 1 Polymethylmethacrylate (PMMA) bone cement, OGM 1A Polymethylmethacrylate (PMMA) bone cement
K211869 · Ormed Grup Medikal Turizm Saglik Hizmetleri Sanayi VE · LOD · May 19, 2022 · Orthopedic
Device Facts
Record ID
K211869
Device Name
OGM 1 Polymethylmethacrylate (PMMA) bone cement, OGM 1A Polymethylmethacrylate (PMMA) bone cement
Applicant
Ormed Grup Medikal Turizm Saglik Hizmetleri Sanayi VE
Product Code
LOD · Orthopedic
Decision Date
May 19, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
OGM® 1 is a PMMA bone cement intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone. OGM® 1A is indicated for use in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
Device Story
OGM® 1 and OGM® 1A are PMMA-based bone cements for orthopedic fixation. OGM® 1A includes the antibiotic Gentamicin for infection management in revision surgeries. Both devices consist of a two-component system: a monomer liquid and a polymer powder, mixed by the clinician during surgery. The cement is applied to fix prosthetic implants to bone. The device provides mechanical stability and, in the case of OGM® 1A, local antibiotic release. Performance is verified through mechanical testing (compressive strength, bending modulus) and biocompatibility assessments. The device is intended for use by surgeons in clinical settings.
Clinical Evidence
No clinical data provided. Evidence is based on bench testing, including mechanical characterization (ISO 5833, ASTM F2118, ASTM D732), biocompatibility (ISO 10993), endotoxin testing (USP), and antibacterial efficiency testing (AATC) against E. coli. Gentamicin release profiles were compared to predicate devices.
Technological Characteristics
PMMA-based bone cement; two-component system (monomer liquid, polymer powder). Materials: poly(methyl acrylate, methyl methacrylate), barium sulfate (radiopaque), benzoyl peroxide (initiator), N,N-dimethyl-p-toluidine (accelerator). OGM® 1A includes 1g Gentamicin sulfate. Sterilization: Ethylene oxide and sterile filtration per ISO 11135-1. High-viscosity formulation. Mechanical properties characterized per ISO 5833 and ASTM F2118.
Indications for Use
Indicated for patients undergoing arthroplastic procedures of the hip, knee, or other joints requiring fixation of polymer or metallic prosthetic implants to living bone (OGM® 1) or patients requiring second-stage revision for total joint arthroplasty following clearance of initial infection (OGM® 1A).
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized image of an eagle, while the FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue as well. The logo is simple and professional, and it is easily recognizable.
May 19, 2022
Ormed Grup Medikal Turizm Saglik Hizmetleri Sanayi Ve % Mehmet Ormeci Consultant Medcer Uluslararasi Medikal Belgelendirme Anonim Sirketi Taspinar Mahallesi 2800. Caddesi A-2 Apt. No:6 B/49 Ankara, 06830 Turkey
Re: K211869 Trade/Device Name: OGM® 1 OGM® 1A Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: LOD, MBB Dated: March 28, 2022 Received: April 8, 2022
Dear Mehmet Ormeci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For:
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211869
Device Name OGM® 1
Indications for Use (Describe)
OGM® 1 is a PMMA bone cement intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211869
Device Name OGM® 1A
Indications for Use (Describe)
OGM® 1A is indicated for use in the second stage revision for total joint arthroplasty after the initial infection has been cleared.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for OGM. The logo is blue and black, and it features the letters "OGM" in a stylized font. Below the letters are the words "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI" in a smaller font. The logo is simple and modern, and it is likely used to represent a company or organization.
# 510k Summary OGM® 1 and OGM® 1A K211869
## 510(k) Summary
| 510(k) Submitter Name | ORMED GRUP SAGLIK MEDIKAL TURIZM SAGLIK HIZMETLERI SANAYI VE TICARET LTD. STI. |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter Adress | Macun Mahallesi 177 Sokak No:19 H/7b Yenimahalle, Ankara, TURKEY |
| 510(k) Submitter Telephone No | +90 312 211 01 92 |
| Primary Correspondent Contact Information | Mehmet Fatih Ormeci (Consultant)<br>+90-532-397-3043<br>mformeci@gmail.com<br>Fax Number: + 90 312 211 01 93 |
| Summary Preparation Date | 03/28/2022 |
| Trade Or Proprietary<br>Name | OGM® 1, OGM® 1A |
|------------------------------|-------------------------------------------|
| Common Name | PMMA Bone Cement |
| Classification Name | Bone Cement |
| Regulation Name | Polymethylmethacrylate (PMMA) bone cement |
| Regulation Number | 21 CFR 888.3027 |
| Product Code | LOD, MBB |
| Regulatory Class | Class II |
| Subject Device 510k No | Primary Predicate Device<br>510k No | Primary Predicate Device Manufacturer |
|------------------------|-------------------------------------|--------------------------------------------------------------------------|
| OGM® 1<br>K211869 | PALACOS® R<br>K202475 | Heraeus Medical GmbH<br>Philipp-Reis-Str. 8/13<br>61273 Wehrheim Germany |
| OGM® 1A<br>K211869 | PALACOS® R+G<br>K202475 | Heraeus Medical GmbH<br>Philipp-Reis-Str. 8/13<br>61273 Wehrheim Germany |
| Subject Device 510k No | Seconder Predicate<br>Device 510k No | Primary Predicate Device Manufacturer |
|------------------------|----------------------------------------------|-------------------------------------------------------|
| OGM® 1A<br>K211869 | DePuy 1 Gentamicin Bone<br>Cement<br>K023103 | DePuy, Inc.<br>700 Orthopaedic Drive Warsaw, IN 46581 |
{5}------------------------------------------------
## Device Description:
A bundled Traditional 510(k) submission is being supplied to the U.S. FDA to gain clearance for modifications to OGM® 1 and OGM® 1A®. This submission encompasses multiple devices that have similar intended use and indications for use as well as rely on similar data.
## OGM® bone cements without Gentamicin
OGM® 1 is polymethylmethacrylate (PMMA) bone cement:
- OGM® 1: is a standard-setting, high-viscosity, PMMA-based bone cement for orthopaedic surgery,
The OGM® 1 consists of two components, a monomer liquid and a polymer powder. The liquid
component contains the monomer, accelerator, and a stabilizer. The powder contains the polymer,
X-Ray-opacifier, and initiator. They are intended for single-use and are provided sterile (ethylene
oxide and sterile filtration).
## OGM® bone cements with Gentamicin
OGM® 1A is polymethylmethacrylate (PMMA) bone cement, containing the antibiotic Gentamicin: - OGM® 1A is a standard-setting, high-viscosity, PMMA-based bone cement for orthopaedic surgery, The OGM® 1A consists of two components, a monomer liquid and a polymer powder. The liquid component contains the monomer, accelerator, and a stabilizer. The powder contains the polymer, X-Ray-opacifier, initiator and the antibiotic Gentamicin. They are intended for single-use and are provided sterile (ethylene oxide and sterile filtration).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for OGM, which is a medical company. The logo is blue and black, and it features the company's name in a stylized font. Below the logo, the text "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI" is written in a smaller font. The logo is simple and modern, and it conveys a sense of professionalism and expertise.
### Indications For Use:
OGM® 1 is a PMMA bone cement intended for use in arthroplastic procedures of the hip, knee and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
OGM® 1A is indicated for use in the second stage of a two-stage revision for total joint arthroplasty after the initial infection has been cleared.
## Comparison of technological characteristics:
## OGM® 1
Device comparison demonstrated that the OGM® 1 is substantially equivalent to the previously cleared PALACOS® R (K202475) regarding intended use, technological characteristics (device design, material and performance) as well as operating principle. At a high level, the subject device and predicate device are based on the following same or similar technological elements:
- PMMA bone cement (same),
- Chemical composition (similar),
- Sterilized with an established method (ethylene oxide) as per DIN EN ISO 11135-1 and DIN EN ISO 10993-7 (same),
- Side by side mechanical testing was performed according to the ISO 5833 and ASTM D732. The mechanical characteristics are similar when it is compared with predicate device.
- Mixing and application properties (same), and
- Clinical use of the devices including the anatomical location of exposure (same) and intended use population (same).
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI. The logo is blue and black. The logo has the letters OGM in it.
Regarding chemical composition comparison, while the predicate device contains zirconium dioxide, the subject device contains barium sulfate in the cement powder as an X ray contrast medium. Both materials are used in commercialized bone cements as an X ray contrast medium. Chemical formulas between predicate device and subject device have slight differences. The predicate device contains chlorophyll additive serves as marking of the bone cement at the site of the operation. The subject device do not contain chlorophyll, but it is still visible as off white instead of green which chlorophyll provides for predicate device.
#### OGM® 1A
Device comparison demonstrated that the OGM® 1A is substantially equivalent to the previously cleared PALACOS® G (K202475) and DePuy 1 Gentamicin Bone Cement (K023103) regarding intended use, technological characteristics (device design, material and performance) as well as operating principle. At a high level, the subject device and predicate device are based on the following same or similar technological elements:
- PMMA bone cement (same),
- Chemical composition (similar),
- Sterilized with an established method (ethylene oxide) as per DIN EN ISO 11135-1 and DIN EN ISO 10993-7 (same),
- Side by side mechanical testing was performed according to the ASTM F2118, ISO 5833 and ASTM D732. The mechanical characteristics are similar when it is compared with predicate device (K023103).
- Mixing and application properties (same) and
- Clinical use of the devices including the anatomical location of exposure (same), and intended use population (same).
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for OGM, which is a medical company. The logo is blue and black, and it features the company's name in a stylized font. Below the logo, the text "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI" is written in a smaller font. The logo is simple and professional, and it conveys the company's focus on medical services.
- 1g Gentamicin is utilized to powder (similar for primary predicate and same for secondary predicate).
- Gentamicin release profiles are compared with the secondary predicate device (K023103) and no significant difference was observed.
#### Discussion of Non-Clinical Tests:
For the OGM® 1 and OGM® 1A bone cements the stability of liquid component, maximum temperature, setting time, intrusion, compressive strength, bending modulus and bending strength of was characterized per ISO 5833.
Mechanical tests were also performed according to ASTM F2118 standards. ASTM F451 requirements are met.
ATCC antibacterial efficiency testing was performed to OGM 1A.
Endotoxin testing was performed to OGM® 1 and OGM® 1A according to USP Endotoxin Reference Standard. The results meet the limit value (20 EU/device).
EtO sterilization was validated per ISO 11135.
Biocompatibility testing, including cytotoxicity, irritation, acute systemic toxicity, subacute systemic toxicity, implantation and genotoxicity was performed per ISO 10993. Because OGM® 1A can be regarded as worst case product, the results are also valid for OGM® 1.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for "OGM", which is in blue and black. Below the logo is the text "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI", which is in a smaller font size. The logo and text are centered in the image.
## 510k Summary OGM® 1 and OGM® 1A K211869
## Substantially Equivalence Discussion
| | Subject Device<br>K211869 | Predicate Device<br>K202475 | Comparison |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer | Image: COGM | Image: Heraeus | |
| Device Name | OGM® 1 | Palacos® R | |
| Intended Use | OGM® 1 is a PMMA bone cement intended<br>for use in arthroplastic procedures of the<br>hip, knee and other joints for the fixation<br>of polymer or metallic prosthetic implants<br>to living bone. | Palacos® R is indicated for use in<br>arthroplastic procedures of the hip, knee,<br>and other joints for the fixation of polymer or<br>metallic prosthetic implants to living bone. | same |
| Components | 40g powder (polymer) and 20ml liquid<br>(monomer) is individually packed. | 40g powder (polymer) and 20ml liquid<br>(monomer) is individually packed. | same |
| Sterilization | ethylene oxide and sterile filtration | ethylene oxide and sterile filtration | same |
| Contact Duration | Permanent / implantable | Permanent / implantable | same |
| Powder Mass | 40g | 40g | same |
| Liquid Volume | 20ml | 20ml | same |
| Powder Chemical<br>Composition | poly(methyl acrylate, methyl methacrylate)<br>35.6 g | poly(methyl acrylate, methyl methacrylate)<br>33.8 g | similar |
| | Barium sulfate<br>4 g | zirconium dioxide<br>5.9 g | similar |
| | hydrous benzoyl peroxide<br>0.4 g | hydrous benzoyl peroxide<br>0.3 g | similar |
| | | chlorophyll VIII trace amount | different |
| Liquid Chemical<br>Composition | methyl methacrylate<br>19.6 g | methyl methacrylate<br>18.4 g | similar |
| | N,N-dimethyl-p-toluidine<br>0.4 g | N,N-dimethyl-p-toluidine<br>0.4 g | Same |
| | | liquid: chlorophyll VIII in an oily solution<br>trace amount | different |
| | Hydroquinone trace amount | Hydroquinone trace amount | same |
Regarding chemical composition comparison, while the predicate device contains zirconium dioxide, the subject device contains barium sulfate in the cement powder as an X ray contrast medium. Both materials are used in commercialized bone cements as an X ray contrast medium. Chemical formulas between predicate device and subject device have slight differences. The predicate device contains chlorophyll additive serves as marking of the bone cement at the site of the operation. The subject device do not contain chlorophyll, but it is still visible as off white instead of green which chlorophyll
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for OGM, which is a medical group. The logo is in blue and black, with the letters "OGM" in a stylized font. Below the logo is the text "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI" in a smaller font. The logo is simple and professional, and it is likely used to represent the company's brand.
provides for predicate device. Subject device has been carried out the ISO 5833 performance tests and ISO 10993 biocompatibility tests. ISO 5833 requirements are fulfilled. ISO 10993 test results demonstrates biological safety. Endotoxin testing was performed to OGM® 1 according to USP Endotoxin Reference Standard. The results meet the limit value (20 EU/device).
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for a company called "OCM ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI". The logo consists of the letters "OCM" in a stylized font, with a graphic to the left of the letters. The letters are in blue, with a black bar above the "M". Below the letters is the company name in smaller letters.
## Substantially Equivalence Discussion
| | Subject Device<br>K211869 | Primary Predicate Device<br>K202475 | Secondary Predicate Device<br>K023103 | Primary<br>Comparison | Secondary<br>Comparison |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|-------------------------|
| Manufacturer | | Heraeus | | - | - |
| Device Name | OGM® 1A | Palacos® R+G | DePuy 1 Gentamicin Bone<br>Cement | - | - |
| Intended Use | OGM® 1A is indicated for use in<br>the second stage of a two-stage<br>revision for total joint<br>arthroplasty after the initial<br>infection has been cleared. | Palacos® R+G is indicated for use in<br>the second stage of a two-stage<br>revision for total joint arthroplasty<br>after the initial infection has been<br>cleared. | DePuy 1 Gentamicin is indicated for<br>use in the second stage of a two<br>stage revision for total joint<br>arthroplasty after the initial<br>infection has been cleared. | same | same |
| Components | 40g powder (polymer) and 20ml<br>liquid (monomer) is individually<br>packed. | 40g powder (polymer) and 20ml<br>liquid (monomer) is individually<br>packed. | 40g powder (polymer) and 20ml<br>liquid (monomer) is individually<br>packed. | same | same |
| Sterilization | ethylene oxide and sterile<br>filtration | ethylene oxide and sterile filtration | ethylene oxide and sterile filtration | same | same |
| Contact Duration | Permanent / implantable | Permanent / implantable | Permanent / implantable | same | same |
| Powder Mass | 40g | 40g | 40g | same | same |
| Liquid Volume | 20ml | 20ml | 20ml | same | same |
| Powder Chemical<br>Composition | poly(methyl acrylate, methyl<br>methacrylate)<br>35.6 g | poly(methyl acrylate, methyl<br>methacrylate)<br>33.6 g | poly(methyl acrylate, methyl<br>methacrylate)<br>84.73 (%w/w) | similar | similar |
| | Barium sulfate<br>4 g | zirconium dioxide<br>6.1 g | Barium sulfate<br>9.10 (%w/w) | similar | similar |
| | hydrous benzoyl peroxide<br>0.4 g | hydrous benzoyl peroxide<br>0.3 g | hydrous benzoyl peroxide<br>1.95 (%w/w) | similar | similar |
| | gentamicin base (as sulphate)<br>1 g | gentamicin base (as sulphate)<br>0.5 g | gentamicin base (as sulphate)<br>1 g | different | same |
| | | chlorophyll VIII trace amount | | different | same |
| Liquid Chemical<br>Composition | methyl methacrylate<br>19.6 g | methyl methacrylate<br>18.4 g | methyl methacrylate<br>98.50 (%w/w) | similar | similar |
| | N,N-dimethyl-p-toluidine<br>0.4 g | N,N-dimethyl-p-toluidine<br>0.4 g | N,N-dimethyl-p-toluidine<br>< 1.50 (%w/w) | same | similar |
| | | liquid: chlorophyll VIII in an oily<br>solution trace amount | | different | same |
| | Hydroquinone trace amount | Hydroquinone trace amount | Hydroquinone trace amount | same | same |
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains a logo for OGM, which is in blue and black. Below the logo is the text "ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI". The text is in a smaller font size than the logo. The logo is simple and modern.
Regarding chemical composition comparison, while the primary predicate device contains zirconium dioxide, the subject device contains barium sulfate in the cement powder as an X ray contrast medium. On the other hand, secondary predicate device contains barium sulfate as the subject device. Both materials are used in commercialized bone cements as an X ray contrast medium.
The subject device and secondary predicate device contain 1g gentamicin sulfate, on the contrary primary predicate device contains 0.5g gentamicin sulfate. Gentamicin sulfate is bactericidal and is active against many strains of Gram-positive and Gram-negative pathogens. As per gentamicin sulfate ratio is higher than the primary predicate device antibacterial efficacy can be considered as substantially equivalent to primary predicate device. Antibacterial efficacy tests are performed to subject device. On the other hand, release ratio of Gentamicin is important both in dosage and total release duration. OGM® 1A is subjected to Gentamicin release test comparatively with the predicate device. Lastly, gentamicin has a negative effect on product mechanical and physical properties. ISO 5833 tests are applied to OGM® 1A and demonstrated compliance.
Chemical formulas between predicate devices and subject device have slight differences. The primary predicate device contains chlorophyll additive serves as marking of the bone cement at the site of the operation. The subject device and secondary predicate device do not contain chlorophyll, but it is still visible as off white instead of green which chlorophyll provides for primary predicate device. Subject device has been carried out the ISO 5833 performance tests and ISO 10993 biocompatibility tests. ISO 5833 requirements are fulfilled. ISO 10993 test results demonstrates biological safety.
Endotoxin testing was performed to OGM® 1A according to USP Endotoxin Reference Standard. The results meet the limit value (20 EU/device).
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for ORMED GRUP MEDIKAL TUR.SAG.HIZ.SAN.VE.TIC.LTD.STI. The logo features the letters "OCM" in a stylized font, with the "O" and "C" in blue and the "M" in black. Below the letters is the company name in smaller, all-caps letters. The logo is simple and professional, and it is likely used on the company's website, marketing materials, and other branding materials.
## Conclusion
The non-clinical tests comply with the requirement of ISO 5833 and ASTM F451. Biological safety has been demonstrated according to the ISO 10993-1. Side-by-side mechanical tests were also performed according to ASTM D732, ASTM F2118 standards and the results are similar when compared to predicate devices. Endotoxin testing was performed to OGM® 1 and OGM® 1A according to USP Endotoxin Reference Standard. The results meet the limit value (20 EU/device). Antibacterial efficiency testing was performed according to the AATC and it is showed that the device is efficient to the E.Coli. Gentamicin release profile is also tested side-by-side with predicate device and the results are similar when compared to predicate devices (K023103). The differences between the predicate devices and the subject device do not raise any new or different questions of safety or effectiveness. The subject devices are substantially equivalent to the predicate devices with respect to the indications for use, target populations, treatment method, and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.