← Product Code [FXX](/productcode/FXX) · K211832

# Fluid Resistant Procedure Mask, Fluid Resistant Surgical Mask (K211832)

_Zhejiang Lanhine Medical Products, Ltd. · FXX · Sep 8, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211832

## Device Facts

- **Applicant:** Zhejiang Lanhine Medical Products, Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Sep 8, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Fluid Resistant Procedure/Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

## Device Story

Fluid Resistant Procedure/Surgical Masks (models 15603F, 15703F) are single-use, non-sterile, flat-pleated, three-layer face masks. The device consists of inner and outer non-woven polypropylene layers and a middle melt-blown polypropylene filter layer. Model 15603F utilizes elastic polyurethane ear loops; model 15703F utilizes non-woven polypropylene tie-on bands. Both models include a polypropylene-coated iron nose clip for facial fit. The masks are worn over the mouth and nose to provide a physical barrier against microorganisms, body fluids, and particulates. Used in clinical or general settings to support infection control practices. The device is disposable and intended for single use.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862 (synthetic blood penetration), ASTM F2299 (particulate filtration efficiency >99%), ASTM F2101 (bacterial filtration efficiency >99%), EN 14683 (differential pressure <6.0 mmH2O/cm2), and 16 CFR 1610 (flammability Class 1). Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

Three-layer construction: inner/outer non-woven polypropylene, middle melt-blown polypropylene filter. Nose clip: polypropylene-coated iron. Ear loops: polyurethane; Tie-on bands: non-woven polypropylene. Dimensions: 17.5cm x 9.5cm. Non-sterile, single-use. ASTM F2100 Level 2 compliant.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Surgical Face Masks ([K153496](/device/K153496.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services - USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top of the word "ADMINISTRATION".

September 8, 2021

Zhejiang Lanhine Medical Products LTD % Ivy Wang Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China

Re: K211832

Trade/Device Name: Fluid Resistant Procedure Mask/Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX, Dated: June 14, 2021 Received: June 14, 2021

Dear Ivy Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211832

Device Name Fluid Resistant Procedure/Surgical Mask

#### Indications for Use (Describe)

The Fluid Resistant Procedure/Surgical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

## K211832

Document prepared date: 9/8/2021

#### A. Applicant:

Name: Zhejiang Lanhine Medical Products LTD. Address: 1989 Cidong Road, Cidongbinhai District, 315300 Cixi City, Zhejiang Province, China Contact Person: Jerry Gu Tel: +86-015255006789 Mail: mark@kanghuimedical.com

Submission Correspondent: Primary contact: Ivy Wang Shanghai SUNGO Management Consulting Co., Ltd. 14th floor 1500# Century Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo 14th floor 1500# Century Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@gmail.com

#### B. Device:

Trade Name: Fluid Resistant Procedure/Surgical Mask Common Name: Surgical Face Mask Model(s): 15603F, 15703F

Regulatory Information Classification Name: Surgical Face Mask Classification: Class II Product code: FXX Regulation Number: 878.4040 Review Panel: Surgical Apparel

## C. Predicate device:

| 510K    | Device name                       | ASTM F2100<br>level | Manufacturer                                 |
|---------|-----------------------------------|---------------------|----------------------------------------------|
| K153496 | Disposable Surgical Face<br>Masks | Level2              | Xiantao Rayxin Medical Products Co.,<br>Ltd. |

#### D. Indications for use of the device:

The Fluid Resistant Procedure/Surgical Masks (model: 15603F, 15703F) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and 

{4}------------------------------------------------

particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This a single use, disposable device(s), provided non-sterile.

# E. Device Description:

The proposed device (model: 15603F) is blue color, and Flat Pleated type mask, utilizing Ear Loops way for wearing, and it has Nose clips design for fitting the face mask around the nose.

The proposed device (model: 15703F) is blue color, and Flat Pleated type mask, utilizing tie-on way for wearing, and it has Nose clips design for fitting the face mask around the nose.

The proposed devices are manufactured with three layers, the inner and outer layers are made of Non-woven Fabric(polypropylene), and the middle layer is made of Melton brown Fabric (Polypropylene). The 15603F model of proposed device, ear loops, is held in place over the users' mouth and nose by two elastic ear loops welded to the face mask. The elastic ear loops are made of polyurethane. The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Polypropylene coating iron. The 15703F model of proposed device, tie-on, is held in place over the users' mouth and nose by two tie-on bands welded to the face mask. The tie-on bands are made of non-woven Fabric (Polypropylene). The nose piece contained in the proposed device(s) is in the layers of face mask to allow the user to fit the face mask around their nose, which is made of Polypropylene coating iron. The proposed devices are sold non-sterile and are intended to be single use, disposable device.

# F. Comparison with predicate device/

Table 1 General Comparison

| Device              | Proposed Device                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                          | Comparison                           |                            |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|----------------------------|
| Manufacturer        | Zhejiang Lanhine Medical Products LTD.                                                                                                                                                                                  | Xiantao Rayxin Medical Products Co., Ltd.                                                                                                                                                                                                 | -                                    |                            |
| 510K number         | K211832                                                                                                                                                                                                                 | K153496                                                                                                                                                                                                                                   | -                                    |                            |
| Classification      | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                   | Class II Device, FXX (21 CFR878.4040)                                                                                                                                                                                                     | Same                                 |                            |
| Indications for use | The Fluid Resistant Procedure/Surgical Mask (model: 15603F, 15703F) are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. | The Disposable Surgical Face Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection | Similar                              |                            |
|                     |                                                                                                                                                                                                                         |                                                                                                                                                                                                                                           |                                      |                            |
| Styles              |                                                                                                                                                                                                                         | Ear Loops, Tie-On,<br>Flat Pleated, 3 layers                                                                                                                                                                                              | Ear Loops, Flat Pleated,<br>3 layers | Different                  |
|                     |                                                                                                                                                                                                                         | Outer layer                                                                                                                                                                                                                               | Non-woven Fabric<br>(Polypropylene)  | Spun-bond<br>polypropylene |
| Mat<br>erial        | Middle<br>layer                                                                                                                                                                                                         | Melton brown Fabric<br>(Polypropylene)                                                                                                                                                                                                    | Melt blown<br>polypropylene filter   | Same                       |
|                     | Inner layer                                                                                                                                                                                                             | Non-woven Fabric<br>(Polypropylene)                                                                                                                                                                                                       | Spun-bond<br>polypropylene           | Same                       |
|                     | Nose clip                                                                                                                                                                                                               | Polypropylene<br>coating iron                                                                                                                                                                                                             | Malleable aluminum<br>wire           | Different                  |
|                     | Tie-on<br>bands                                                                                                                                                                                                         | Non-woven Fabric<br>(Polypropylene)                                                                                                                                                                                                       | /                                    | --                         |
|                     | Ear loops                                                                                                                                                                                                               | Polyurethane                                                                                                                                                                                                                              | Polyester                            | Different                  |
| Color               |                                                                                                                                                                                                                         | Blue                                                                                                                                                                                                                                      | Blue                                 | Same                       |
|                     | Dimension(Length)                                                                                                                                                                                                       | $17.5 cm +/- 0.5 cm$                                                                                                                                                                                                                      | $17.5 cm +/- 0.1 cm$                 | Different                  |
|                     | Dimension (Width)                                                                                                                                                                                                       | $9.5 cm +/- 0.5 cm$                                                                                                                                                                                                                       | $9.5 cm +/- 0.1 cm$                  | Different                  |
| OTC use             |                                                                                                                                                                                                                         | Yes                                                                                                                                                                                                                                       | Yes                                  | Same                       |
| Sterility           |                                                                                                                                                                                                                         | Non-Sterile                                                                                                                                                                                                                               | Non-Sterile                          | Same                       |
| Use                 |                                                                                                                                                                                                                         | Single Use                                                                                                                                                                                                                                | Single Use, Disposable               | Same                       |
| ASTM F2100 level    |                                                                                                                                                                                                                         | Level 2                                                                                                                                                                                                                                   | Level 2                              | Same                       |
|                     | Biocompatibility                                                                                                                                                                                                        | ISO10993                                                                                                                                                                                                                                  | ISO10993                             | Same                       |

{5}------------------------------------------------

Difference analysis: The proposed device has different nose clip & ear loops material as well as dimension to the predicate device, but the performance and biocompatibility of the device has been tested.

#### G. Non-Clinical Test Conclusion

Non-clinical tests were conducted to verify that the proposed device met all design specifications as was similar to the predicate device. The test results demonstrated that the proposed device complies 

{6}------------------------------------------------

1989 Cidong Road, Cidongbinhai District, 315300 Cixi City, Zhejiang Province, China

with the following standards and the requirements stated in the Guidance for Industry and FDA Staff: Surgical Masks - Premarket Notification [510(k)] Submission issued on March 5, 2004:

- ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro A Cytotoxicity
- > ISO 10993-10: 2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- > ASTM F2100, Standard Specification for Performance of Materials Used In Medical Face Masks
- > ASTM F1862, Standard Test Method for Resistance of Medical Face Masks To Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume At A Known Velocity);
- > EN 14683, Medical Face Masks-Requirements and Test Methods;
- > ASTM F2101, Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol of Staphylococcus Aureus;
- > ASTM F2299, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres;
- A 16 CFR 1610, Standard for the Flammability of clothing textiles;

| Item                                                  | Purpose                                                                                                           | Proposed device<br>(model: 15603F)                                                                                  | Proposed device<br>(model: 15703F)                                                                                  | Acceptance<br>Criteria                                          | Result |
|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|--------|
| Fluid<br>Resistance<br>Performance<br>ASTM F1862      |                                                                                                                   | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>32 out of 32 pass at 120<br>mmHg             | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>32 out of 32 pass at 120<br>mmHg             | 29 out<br>of 32<br>pass<br>at 120<br>mmH<br>g for<br>level<br>2 | PASS   |
| Particulate<br>Filtration<br>Efficiency<br>ASTM F2299 | The purpose<br>of the<br>performance<br>testing is to<br>demonstrate<br>the<br>functionality<br>of the<br>subject | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 99.15%<br>Lot2: 99.22%<br>Lot3: 99.22% | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 99.03%<br>Lot2: 99.10%<br>Lot3: 99.30% | ≥ 98%                                                           | PASS   |
| Bacterial<br>Filtration<br>Efficiency<br>ASTM F2101   |                                                                                                                   | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 99.40%<br>Lot2: 99.50%                 | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 99.20%<br>Lot2: 99.30%                 | ≥ 98%                                                           | PASS   |

Table 2 - Performance Testing

{7}------------------------------------------------

Zhejiang Lanhine Medical Products LTD 1989 Cidong Road, Cidongbinhai District, 315300 Cixi City, Zhejiang Province, China

|                                                           | device. | Lot3: 99.60%                                                                                                                                | Lot3: 99.30%                                                                                                                                |                        |      |
|-----------------------------------------------------------|---------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|------------------------|------|
| Differential<br>Pressure (Delta<br>P) EN 14683<br>Annex C |         | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 3.55 mmH2O/cm2<br>Lot2: 3.59 mmH2O/cm2<br>Lot3: 3.60 mmH2O/cm2 | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Lot1: 3.59 mmH2O/cm2<br>Lot2: 3.58 mmH2O/cm2<br>Lot3: 3.56 mmH2O/cm2 | < 6.0m<br>mH2O<br>/cm2 | PASS |
| Flammability<br>16 CFR 1610                               |         | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Class 1                                                              | 3 non-consecutive lots<br>tested, using a sample<br>size of 32/lot.<br>Class 1                                                              | Class 1                | PASS |

Table 3 Biocompatibility Comparison

| Item          | Proposed device                                                        | Predicate device                                                          | Acceptance Criteria | Result |
|---------------|------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------|--------|
| Cytotoxicity  | Under the conditions of<br>the study, the device is<br>non-cytotoxic.  | Under the conditions<br>of the study, the<br>device is<br>non-cytotoxic.  | Non-Cytotoxic       | PASS   |
| Irritation    | Under the conditions of<br>the study, the device is<br>non-irritating. | Under the conditions<br>of the study, the<br>device is<br>non-irritating. | Non-Irritating      | PASS   |
| Sensitization | Under the conditions of<br>the study, the device is<br>non-sensitizing | Under the conditions<br>of the study, the<br>device is<br>non-sensitizing | Non-Sensitizing     | PASS   |

## H. Clinical Test Conclusion

No clinical study is included in this submission.

## I. Conclusion

Based on the nonclinical tests performed, the subject devices (Model:15603F & 15703F) are as safe, as effective, and performs as well as the legally marketed predicate device, K153496, BYD Precision Manufacturer Co. Ltd, Single-use Surgical Masks.

---

**Source:** [https://fda-staging.innolitics.com/device/K211832](https://fda-staging.innolitics.com/device/K211832)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com