K211822 · Elastic Care Incorporated · DRG · Feb 3, 2022 · Cardiovascular
Device Facts
Record ID
K211822
Device Name
LifePath Remote Patient Monitoring Platform
Applicant
Elastic Care Incorporated
Product Code
DRG · Cardiovascular
Decision Date
Feb 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The LifePath Remote Patient Monitoring Platform connects with compatible physiological measurement devices (blood pressure meters, pulse oximeters, thermometers, glucometers) to collect vital sign using a Patient App and securely transmit them, unaltered, to LifePath servers. The LifePath Platform is intended to be used at home or in healthcarerelated environments such as long-term care facilities or clinics, nursing homes and hospitals. The LifePath Remote Patient Monitoring Platform serves as Software as a Medical Device and can be used only with specific FDA cleared third-party measurement devices. The LifePath Remote Patient Monitoring Platform securely stores the collected information on LifePath servers and make it available for viewing by remotely located clinician App installed on their smartphones. The clinician can analyze the data, prescribe medications or contact the patient if needed. Clinicians can set thresholds individually for each patient; notifications are sent to the clinician for threshold breaches. The LifePath Platform does not interpret, make diagnoses or serve as a substitute for medical care. It is not intended to provide real time data. The LifePath Platform is contraindicated for use in high-acuity environments, such as the ICU or operating rooms, and is not intended for use on acutely or critically ill patients or those requiring time-critical or emergency intervention. It is intended for patients who are willing and capable of managing its use or where its operations can be performed by a caregiver. Clinical judgment and experience of a Clinician are required to check and interpret the information delivered.
Device Story
LifePath RPM Platform is a software-based telemedicine system; collects vital signs from FDA-cleared third-party Bluetooth-enabled devices (pulse oximeters, thermometers, glucometers, BP monitors). Patient App on mobile device receives data via Bluetooth LE; transmits encrypted, unmodified data to central LifePath server. Clinician App on mobile device allows remote viewing of stored data, threshold setting, and communication (chat/video). Clinicians use output to monitor patient status, prescribe medications, or contact patients. System provides notifications for threshold breaches. Used in home or clinical settings by patients/caregivers and clinicians. Does not provide real-time data, diagnosis, or interpretation. Benefits include remote patient management and improved clinician-patient communication.
Clinical Evidence
No clinical data. Safety and performance established via software verification and validation testing per FDA guidance.
Technological Characteristics
Software as a Medical Device (SaMD). Connectivity: Bluetooth LE (2.4 GHz) for device-to-app; Wi-Fi/cellular for app-to-server. Platform: Android (v8+), mobile smartphones. Security: HIPAA-compliant encryption. Data processing: Central server-based storage and threshold analysis. No hardware components; utilizes COTS third-party medical devices.
Indications for Use
Indicated for adult patients with non-critical conditions requiring remote monitoring of vital signs (pulse rate, SpO2, skin temperature, blood pressure, blood glucose) at home or in healthcare facilities. Contraindicated for high-acuity environments (ICU, OR), acutely/critically ill patients, or those requiring time-critical/emergency intervention.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 3, 2022
Elastic Care Incorporated % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K211822
Trade/Device Name: LifePath Remote Patient Monitoring Platform Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency physiological signal transmitter and receiver Regulatory Class: Class II Product Code: DRG Dated: January 4, 2022 Received: January 5, 2022
#### Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology. Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K211822
Device Name LifePath Remote Patient Monitoring Platform
#### Indications for Use (Describe)
The LifePath Remote Patient Monitoring Platform connects with compatible physiological measurement devices (blood pressure meters, pulse oximeters, thermometers, glucometers) to collect vital sign using a Patient App and securely transmit them, unaltered, to LifePath servers. The LifePath Platform is intended to be used at home or in healthcarerelated environments such as long-term care facilities or clinics, nursing homes and hospitals. The LifePath Remote Patient Monitoring Platform serves as Software as a Medical Device and can be used only with specific FDA cleared third-party measurement devices.
The LifePath Remote Patient Monitoring Platform securely stores the collected information on LifePath servers and make it available for viewing by remotely located clinician App installed on their smartphones. The clinician can analyze the data, prescribe medications or contact the patient if needed. Clinicians can set thresholds individually for each patient; notifications are sent to the clinician for threshold breaches.
The LifePath Platform does not interpret, make diagnoses or serve as a substitute for medical care. It is not intended to provide real time data.
The LifePath Platform is contraindicated for use in high-acuity environments, such as the ICU or operating rooms, and is not intended for use on acutely or critically ill patients or those requiring time-critical or emergency intervention. It is intended for patients who are willing and capable of managing its use or where its operations can be performed by a caregiver. Clinical judgment and experience of a Clinician are required to check and interpret the information delivered.
| Type of Use (Select one or both, as applicable) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Elastic Care. The logo consists of a stylized letter "E" in black, with a red line extending from the "E" to a partial circle on the right, resembling an electrocardiogram (ECG) waveform. Below the logo are the words "ELASTIC" and "CARE" stacked vertically in a simple, sans-serif font.
# 510(k) Summary
[As per the requirements of 21 CFR Part 807.92]
# 1 SUBMITTER IDENTIFICATION
| Company name: | ELASTIC CARE INCORPORATED |
|------------------|---------------------------|
| Company Address: | 95 Apple Creek Blvd., |
| | Markham, Ontario L3R 1C7 |
| | Canada |
| Phone: | (416)-704-5141 |
| Fax: | (416)-214-6238 |
| Contact Person: | Ashok Kalle |
| Contact Email: | ashok.kalle@pathcom.com |
| Date Prepared: | 17 May 2021 |
## 2 SUBMITTER CORRESPONDENT
| Contact Person: | Prithul Bom |
|-----------------|------------------------------------------------------------------------------------------|
| Address: | Regulatory Technology Services, 1000 Westgate Drive,<br>Suite #510k Saint Paul, MN 55114 |
| Phone: | 612-963-0379 |
| Contact Email: | prithul.bom@rts3pro.com |
## 3 DEVICE IDENTIFICATION
| Trade Name: | LifePath Remote Patient Monitoring Platform |
|-----------------------------------|--------------------------------------------------------------|
| Common Name: | LifePath RPM Platform |
| Device Class: | II |
| Classification Regulation Number: | 870.2910 |
| Classification Name: | Radiofrequency physiological signal transmitter and receiver |
| Product Code: | DRG |
| Classification Panel: | Cardiovascular |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Elastic Care. The logo features the letters 'EC' in a stylized design. The 'E' is formed by a black line with a red horizontal line running through it. The 'C' is a black circle with a red line that resembles an electrocardiogram (ECG) tracing. Below the symbol are the words 'ELASTIC CARE' stacked vertically in a simple, sans-serif font.
## 4 LEGALLY MARKETED PREDICATE DEVICE(S)
Predicate Device: Comarch e-Care Platform, K181248
# 5 DEVICE DESCRIPTION
The LifePath Remote Patient Monitoring Platform (the "LifePath RPM Platform") is a software application. It is used in conjunction only with FDA cleared Commercial, Off-The-Shelf (COTS) Third-Party Medical Devices (TPDs). These TPDs are with Bluetooth functionality to collect, aggregate, and present, without modification, vital signs and symptoms required by clinicians to assess a patient's health. The device is intended for patients who have non-critical conditions and are located either in their own homes or in professional healthcare facilities such as hospitals, skilled nursing facilities, nursing homes, or long-term care facilities. The LifePath RPM Platform constitutes a telemedicine system for remote monitoring of patients with the connected Clinician. Both the Patient App and Clinician App are accessed via authentication using a one-time security code by the respective users.
Image /page/4/Figure/6 description: The image shows a diagram of the LifePath Processing Software system. The system includes third-party devices such as a blood pressure meter, pulse oximeter, glucometer, and thermometer. These devices connect to the LifePath Patient App, which then connects to the LifePath Clinician App. The LifePath Processing Software includes features such as authentication/authorization, vital data processing and storage, appointments booking, video call and chat functions, alerts, prescriptions creation/storage, notifications handling, and patient data storage.
The LifePath Patient App connects through a short-range radio system using Bluetooth to FDA 510(k) cleared devices (listed in Table 1 below) to measure physiological parameters of adult patients and securely transmit unmodified data to the LifePath Processing Software residing on the LifePath Server (or central server) for storage, formatting, remote display, and download. The information is available for secure viewing by remotely-located healthcare professionals (eg.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo for a company called "ELASTIC CARE". The logo features the letters "EC" in a stylized font, with the "E" in black and the "C" formed by a black outline and a red heartbeat line. The words "ELASTIC CARE" are written in a simple, sans-serif font below the letters.
Clinicians). The data may be retrieved and viewed via the Clinician App and Patient App on smartphones through internet access using local Wi-Fi 802.11 or a 3G/4G/LTE cellular network. By using the Clinician App a Clinician can view the vital signs of the connected patient, analyze the data, and prescribe medications. Clinicians will also be able to set thresholds for each patient such that notifications will be sent to the Clinician in the event of any threshold breach. The LifePath RPM Platform will be prescribed by the Clinician to the patient as per the need to remotely monitor the patient. The patient, using the Patient App, can book appointments with the Clinician. The Patient App also has the provision for the patient to respond to the Clinician's questions on their prior health conditions, medical conditions, any allergic conditions, etc., that will enable Clinicians to get a clear medical status of the patient. The patient and the Clinician can interact through chats and video calls using their respective apps.
The LifePath RPM Platform is not intended for use in high-acuity environments, such as the ICU or operating rooms, and is not intended for use on acutely or critically ill patients or those requiring time-critical or emergency intervention.
## 5.1 Physiological Parameters from Third-Party Medical Devices
The LifePath RPM Platform gathers physiological parameter readings transmitted by Third-Party Medical Devices ("TPDs") using Bluetooth LE. These third-party bedside or mobile devices shall each have individually received FDA 510k clearance and the LifePath RPM Platform will have been configured to receive physiological parameter readings from them by integrating the device-specific software(s). Physiological vital signs and parameters measured include the following medical devices:
- Pulse rate from a Pulse Oximeter ●
- Peripheral oxygen saturation (SpO2) from a Pulse Oximeter ●
- Skin temperature from a digital thermometer ●
- Systolic and diastolic blood pressure from a non-invasive BP Monitor ●
- Blood sugar (glucose) from a Glucometer
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains a logo for a company called "ELASTIC CARE". The logo features the letters "EC" in a stylized design. The "E" is a square shape, and the "C" is a curved shape. A red line runs through the middle of the "E" and continues into the "C", forming a heartbeat symbol. Below the logo, the words "ELASTIC CARE" are written in a simple, sans-serif font.
#### The following table lists the FDA 510k cleared third-party medical devices from which the LifePath RPM Platform receives physiological parameters:
| Physiological<br>Parameters | Device | Manufacturer | Model | 510 (k)<br>number | Date cleared<br>(MM/DD/YYYY) |
|-------------------------------------------------------|---------------------------------------------------------|-------------------------------------------------------------|-----------|-------------------|------------------------------|
| Pulse rate,<br>Peripheral oxygen<br>saturation (SpO2) | Pulse Oximeter | Shanghai Berry<br>Electronic Tech<br>Co., Ltd | BMC1000C | K172141 | 04/04/2018 |
| Skin temperature | Clinical Electronic<br>Thermometer | Shenzhen Jumper<br>Medical Equipment<br>Co., Ltd. | JPD-FR400 | K172795 | 03/14/2018 |
| Blood sugar<br>(glucose) | Blood Glucose<br>Monitoring<br>System | Tyson Bio | HT100B | K170079 | 10/06/2017 |
| | | All Medicus Co.,<br>Ltd. | ABM-513S | K170241 | 10/13/2017 |
| Systolic and<br>diastolic blood<br>pressure | Non-invasive<br>Blood Pressure<br>Measurement<br>System | Shenzhen Belter<br>Health<br>Measurement And<br>Analysis TE | ePA-46B4 | K134029 | 07/25/2014 |
Table 1
## 5.2 LifePath RPM Platform - MAJOR COMPONENTS
The LifePath RPM Platform consists of the following:
- a) The LifePath Patient Application ("Patient App"). This is downloaded and installed on a mobile phone which is in the proximity of the person whose vital signs are being monitored. The patient may seek assistance from a caregiver in operating the App and collecting vital signs.
The Patient App receives the vital signs data from the third-party medical devices over Bluetooth LE and transmits them, encrypted and unmodified, over the Internet to the "LifePath Processing Software" installed on a secure, remote central server. The Patient App must be turned on and be available within Bluetooth range of the third-party devices at the location of the individual(s) whose vital signs and parameters are being monitored.
- b) The LifePath Processing Software ("Processing Software") receives, stores, formats, and makes the vital signs information accessible for viewing by Clinicians. Notifications will be sent to the Clinician for any threshold breaches
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Elastic Care. The logo consists of a stylized letter "E" in black, with a red line extending from the right side of the "E" that forms a heart rate waveform. The waveform ends inside a black circle that is connected to the "E". Below the symbol are the words "ELASTIC" and "CARE" stacked on top of each other in a sans-serif font.
set by the Clinician. The Processing Software resides in the central server (LifePath Server).
- c) A LifePath Clinician Application ("Clinician App"). This is downloaded and installed on a mobile phone used by the Clinician who is monitoring the patient. The Clinician App is an interface to observe physiological parameters, to view and respond to alerts, and to perform other administrative functions that do not involve vital signs.
Elastic Care onboards healthcare professionals such as Clinicians bv background verification of their credentials. They are required to have skills to interpret physiological parameters and vital signs.
### 5.3 LifePath RPM Platform Feature Summary
The LifePath RPM Platform includes the following features and functions:
- a. Creation of a master patient record on the central server which contains patient information.
- b. Setting of individual thresholds by a Clinician for each vital sign or parameter, and for sending alerts to the Clinician when these thresholds are crossed.
- c. The ability for the reviewing clinician to add annotations and notes to the patients' record.
- d. Reminders and alerts for patients and Clinicians related to medical and nonmedical activities.
- e. Scheduling of appointments by the patient with the Clinician using the Patient App.
- f. Issue of secure electronic prescriptions by the supervising Clinician from the Clinician App in conformance to legal requirements.
- g. Communications between the Clinician and the patient via video and chat.
- h. The ability for the clinicians and patients to manually enter patients' vital signs and symptoms into the patient records using the Patient App.
- i. Storage and display of historical patient vital signs.
- i. Maintenance of records of consultation for Clinician billing and audit.
## 5.4 Mode of Operation
The LifePath RPM Platform is a software application constituting a telemedicine system for remote monitoring of patients with the connected Clinician. It is used in
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for a company called "Elastic Care". The logo features the letters "EC" in a stylized design. The "E" is a black square with an open side, and the "C" is a black arc. A red line runs horizontally through the "E" and continues into the "C", where it becomes a red heartbeat symbol. Below the letters, the words "ELASTIC" and "CARE" are stacked on top of each other in a simple, sans-serif font.
combination with FDA cleared TPDs with Bluetooth functionality to collect, aggregate, and present, without modification, vital signs and symptoms required by clinicians to assess a patient's health. The Patient's vital sign data can be viewed by the Clinician using the Clinician App over cellular or Wi-Fi networks to analyze and prescribe medications or provide medical advice. Clinicians can also set thresholds for individual patients and receive notifications for any breach. The LifePath RPM Platform also provides the capability for the patient to book consultations with the Clinician and interact using either chat or video call facility. All patients shall have non-critical conditions and shall not require in-person emergency medical care.
The proposed conditions for use of the LifePath RPM Platform are listed below.
#### 5.4.1 Hardware hosts for Patient and Clinician Applications:
- a. Android (version 8 and above).
- b. Smartphones with screen sizes equal to, or greater than, 5.7 inches diagonally.
- c. Processor: 2 core 1.2 GHz or faster processor.
- d. RAM: Minimum 2 GB or more.
- e. ROM: Minimum 2 GB or more.
- f Network Bandwidth: More than 1 MBPS (Internet access using local Wi-Fi 802.11 or 3G/4G/LTE cellular.
- g. Bluetooth: version 2.0 and higher and operating frequency of 2.4 GHz.
The LifePath RPM Platform ensures the security of transmitted data via encryption in compliance with HIPAA requirements.
## 6 INDICATION FOR USE
The LifePath Remote Patient Monitoring Platform connects with compatible physiological measurement devices (blood pressure meters, pulse oximeters, thermometers, glucometers) to collect vital sign using a Patient App and securely transmit them, unaltered, to LifePath servers. The LifePath Platform is intended to be used at home or in healthcare-related environments such as long-term care facilities or clinics, nursing homes and hospitals. The LifePath Remote Patient Monitoring Platform serves as Software as a Medical Device and can be used only with specific FDA cleared third-party measurement devices.
The LifePath Remote Patient Monitoring Platform securely stores the collected information on LifePath servers and make it available for viewing by remotely located clinicians using a Clinician App installed on their smartphones. The clinician can analyze the data, prescribe medications or contact the patient if needed. Clinicians can set thresholds individually for
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains a logo for Elastic Care. The logo features the letters 'EC' in a stylized design, with the 'E' in black and the 'C' formed by a black arc. A red line runs horizontally through the letters, forming a heartbeat waveform in the middle. Below the letters, the words 'ELASTIC CARE' are stacked vertically in a simple, sans-serif font.
each patient; notifications are sent to the clinician for threshold breaches.
The LifePath Platform does not interpret, make diagnoses or serve as a substitute for medical care. It is not intended to provide real time data.
The LifePath Platform is contraindicated for use in high-acuity environments, such as the ICU or operating rooms, and is not intended for use on acutely or critically ill patients or those requiring time-critical or emergency intervention.
It is intended for patients who are willing and capable of managing its use or where its operations can be performed by a caregiver. Clinical judgment and experience of a Clinician are required to check and interpret the information delivered.
# 7 COMPARISON OF TECHNOLOGICAL
## 7.1 Characteristics with Legally Marketed Predicate Device
Table 2: The table below describes the similarities and differences with respect to the intended use, principle of operation, and technological characteristics between the subject device and the predicate device.
| SI<br># | Feature | Predicate Device<br>Comarch E-Care Platform<br>(K181248) | Subject Device<br>Lifepath Remote Patient<br>Monitoring Platform (Pending) |
|---------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Device<br>Classification | II | II |
| | Product Code | DRG | DRG |
| | Regulation<br>Number | 870.2910 | 870.2910 |
| 1. | Indications for<br>Use | The Comarch e-Care Platform is<br>intended to connect with<br>physiological measurement devices<br>(weight scales, blood pressure<br>meters, pulse oximeters, peak flow<br>meters, thermometers, spirometers,<br>glucometers) intended to use at<br>home and send the measurement<br>results to central server. Comarch | The LifePath<br>Remote Patient<br>Monitoring Platform connects with<br>compatible<br>physiological<br>measurement devices to collect<br>vital sign measurements intended<br>to be used at home or<br>healthcare-related environments and<br>securely transmit them unaltered to<br>the LifePath server. LifePath<br>Remote Patient Monitoring Platform |
| 2. | Intended use | | |
| | | e-Care Platform serves as Software as a Medical Device and can be used only with FDA cleared measurement devices.<br><br>Comarch e-Care Platform displays the collected measurements on the Web application and securely stores them in a database server, where the caregiver can view the results, analyze them, leave comments and contact patient if necessary. Caregivers are able to set thresholds individually for each patient. Measurement results sent to e-Care Platform from connected devices are analyzed and if result is beyond the threshold, caregiver gets the notification.<br><br>The Comarch e-Care Platform is not interpretive, nor is it intended for diagnosis or as a substitute for medical care, and it is not intended to provide real-time data. It is made available to patients when time-critical care is not required.<br><br>The Comarch e-Care Platform is contraindicated for patients requiring direct medical supervision or emergency intervention. It is intended for patients who are willing and capable of managing its use. Clinical judgment and experience by a caregiver are required to check and interpret the information delivered. | serves as Software as a Medical Device and can be used only with specific FDA cleared third-party measurement devices.<br><br>The LifePath server securely stores the collected information and makes it available for viewing to remotely located clinicians using the Clinician App installed on their smartphones. The clinician can analyze the data, prescribe medications or contact the patient if needed. Clinicians can set thresholds individually for each patient and for any threshold breach, notification is sent to the clinician.<br><br>The LifePath RPM Platform does not interpret, make diagnoses nor serve as a substitute for medical care.<br><br>The LifePath RPM Platform is contraindicated for use in high-acuity environments, such as the ICU or operating rooms, and is not intended for use on acutely or critically ill patients or those requiring time-critical or emergency intervention.<br><br>It is intended for patients who are willing and capable of managing its use or can be assisted by a caregiver. Clinical judgment and experience by a Clinician are required to check and interpret the information delivered. |
| | | Telemedicine system | Telemedicine system |
| 3. | Intended<br>users | Home users and healthcare<br>providers. | Home users and healthcare providers. |
| 4. | Site of Use | Healthcare related environment or<br>home | Healthcare-related environment or<br>home |
| 5. | Patient<br>population | The medical device is intended for<br>adult users. | The medical device is intended for<br>adult users. |
| 6. | OTC and/or Rx | Rx | Rx |
| 7. | Contraindicati<br>on | The Comarch e-Care Platform is<br>contraindicated for patients<br>requiring direct - medical<br>supervision or emergency<br>intervention. | The LifePath RPM Platform is not<br>intended for use in high-acuity<br>environments, such as the ICU or<br>operating rooms, and is not intended<br>for use on acutely or critically ill<br>patients or those requiring time-critical<br>or emergency intervention. |
| 8. | Data<br>collection<br>software | Comarch e-Care software<br>application, Comarch SMA software<br>application | LifePath proprietary server software,<br>LifePath Patient Software Application |
| 9. | Data<br>collection<br>software<br>functionality | Transmit data from measuring<br>devices and store data in a central<br>database. | Collect data from measuring devices<br>and transmit them to a central server<br>(LifePath Server). |
| 10. | Communicati<br>on method of<br>hub with<br>central server | Any Internet access method with<br>Ethernet or Wi-Fi 802.11 b/g/n<br>interface or 3G/4G network. | Internet access using local Wi-Fi<br>802.11 or 3G/4G/LTE cellular network. |
| 11. | Types of<br>measuring<br>devices which<br>can be<br>interfaced to<br>receiver hub | Medical devices designed for home<br>use: glucose meters weight scales,<br>blood pressure meters, pulse<br>oximeters, peak flow meters,<br>thermometers, spirometers. | Medical Devices designed for home<br>use: FDA 510K cleared glucometers,<br>pulse oximeters, blood pressure<br>meters, and thermometers |
| 12. | Measuring<br>device<br>software | Measuring device software<br>unchanged | Measuring device software unchanged |
| 13. | Communicati<br>on with | Short-range radio system using<br>Bluetooth | Short-range radio system using<br>Bluetooth |
| | measuring<br>devices | | |
| 14. | Method<br>implemented<br>for collecting data<br>from<br>measuring<br>devices | Bluetooth v2.0 and Bluetooth v4.0 | Bluetooth v2.0 and higher |
| 15. | Measuring<br>devices<br>communicatio<br>n frequency | Bluetooth 2.4 GHz…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.