The Venclose Maven System (digiRF Generator and Maven Catheter) is intended for endovascular coagulation of blood vessels in patients with perforator and tributary vein reflux.
Device Story
The Venclose Maven System consists of a digiRF Generator and a sterile, single-use Maven Catheter. The system is used for endovenous radiofrequency (RF) ablation to treat incompetent perforator and tributary veins. The physician operates the system, which delivers temperature-controlled RF energy to the catheter's heating coil (0.5 cm length). This energy heats the vein wall, causing collagen denaturation, irreversible luminal occlusion, subsequent fibrosis, and vein resorption. The generator features a touchscreen interface that automatically sets non-adjustable treatment parameters. An optional foot pedal is supported. The device is used in a clinical setting to provide a minimally invasive alternative to surgical vein treatment, benefiting patients by effectively occluding refluxing veins.
Clinical Evidence
A prospective, literature-controlled, confirmatory, single-investigator, nonrandomized study enrolled 27 subjects to treat incompetent perforator veins (IPVs). The study confirmed the device can be used as intended without significant adverse outcomes. At 30-day follow-up, the occlusion rate was 80% and the reflux-free rate was 93%.
Technological Characteristics
System comprises a 6F (2.0 mm) catheter with a 0.5 cm heating coil and a multi-voltage RF generator. Operates via resistive RF ablation with temperature control. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, and IEC 60601-1-6/IEC 62366. Biocompatibility per ISO 10993-1. Software verification and validation performed per FDA guidance.
Indications for Use
Indicated for patients with perforator and tributary vein reflux requiring endovascular coagulation of blood vessels.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 3, 2021
Venclose, Inc. Ms. Mai-Ly Wilcox, BS, RAC Sr. Director, Regulatory and Clinical Affairs 2570 N. First Street, Second Floor, #221 San Jose, California 95131
Re: K211806
Trade/Device Name: Venclose Maven System (digiRF Generator and Maven Catheter) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: June 10, 2021 Received: June 11,2021
Dear Ms. Wilcox:
We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
encies. You must comply with all the Act's
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A ) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211806
### Device Name
Venclose Maven System (digiRF Generator and Maven Catheter)
Indications for Use (Describe)
The Venclose Maven System (digiRF Generator and Maven Catheter) is intended for endovascular coagulation of blood vessels in patients with perforator and tributary vein reflux.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for VENCLOSE. The logo features a stylized letter V made of several parallel lines that transition in color from yellow to orange to red to purple. Below the V is the word "VENCLOSE" in a simple, sans-serif font.
### 510(k) SUMMARY
#### SUBMITTER l.
Venclose, Inc. 2570 N. First Street 2nd Floor, #221 San Jose, CA 95131
Phone: 844-834-6292
Contact Person: Ms. Mai-Ly Wilcox, BS
Date Prepared: August 2, 2021
#### ll. DEVICE
| Device Name: | Venclose Maven System (digiRF Generator and<br>Maven Catheter) |
|-----------------------|---------------------------------------------------------------------------------------|
| Common or Usual Name: | Electrosurgical Cutting and Coagulating Instrument |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR § 878.4400) |
| Regulatory Class: | II |
| Product Code | GEI |
#### III. PREDICATE DEVICES
| Primary Predicate Device: | VNUS RFS and RFSFlex Catheters |
|----------------------------------------------------|---------------------------------------------------------------------------------------|
| Primary Predicate Device 510(k) No.: | K052003 |
| Primary Predicate Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR § 878.4400) |
| Primary Predicate Regulatory Class: | II |
| Primary Predicate Product Code: | GEI |
| | |
| Secondary Predicate Device: | Venclose digiRF Generator & EVSRF Catheters |
| Secondary Predicate Device 510(k)<br>No.: | K160754 |
| Secondary Predicate Device<br>Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories (21 CFR § 878.4400) |
| Secondary Predicate Regulatory Class: | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for VENCLOSE. The logo consists of a stylized letter V formed by several parallel lines that transition in color from yellow to orange to pink to purple. Below the V is the word "VENCLOSE" in a simple, sans-serif font, with a trademark symbol to the right of the word.
#### DEVICE DESCRIPTION IV.
Venclose Maven System consists of two (2) main components:
- 1) Maven Catheter with integrated cable connector and,
- digiRF Generator which delivers radiofrequency (RF) energy to the Maven catheter. 2)
The Maven catheter is a sterile single-use disposable medical device for endovenous radiofrequency ablation. The primary components of the catheter include the shaft, handle and integrated connector cable. The catheter shaft is 6 French (6F = 2.0 mm in diameter) in profile, with an insertable length of 40 cm and a 0.5 cm heating coil. The catheter is energized by the digiRF Generator which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter.
The Venclose Maven System (digiRF Generator and Maven Catheter) uses resistive radiofrequency ablation (via energy delivered to heat the wall of an incompetent vein) with temperature-controlled RF energy, and an already widely accepted procedure to cause irreversible luminal occlusion. This is followed by fibrosis and ultimately resorption of the vein.
The Venclose Maven System (digiRF Generator and Maven Catheter) supports an optional foot pedal.
#### V. INDICATIONS FOR USE:
The Venclose Maven System (digiRF Generator and Maven Catheter) is intended for endovascular coagulation of blood vessels in patients with perforator and tributary vein reflux.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The technology and design features of the subject device are equivalent to the primary predicate device. Specifically, the underlying principle for both devices is to utilize RF based thermal energy to denature collagen and shrink/occlude venous vasculature safely and effectively.
At a high level, the subject and the primary predicate devices are based on the following same technological characteristics:
- . Catheter Diameter
- . Insertable Length
- Heating Element Length ●
- Catheter Structure ●
- Handle, Cable & Connector
- Visualization ●
The following technological differences exist between the subject and the primary predicate devices:
- . Heating Set Temperature Range
- Heating Method
- Treatment Time ●
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for VENCLOSE. The logo consists of a stylized letter V made up of several parallel lines that transition in color from yellow to orange to pink to purple. Below the V is the word "VENCLOSE" in gray sans-serif font.
The subject device is only a slight modification to the secondary predicate device (predecessor), EVSRF system. The subject device uses the same device components as the EVSRF system. The only differences in technological characteristics between the subject and the secondary predicate devices are the following:
- . Insertable Lengths
- Heating Element Lengths .
- Heating Set Temperature Range
#### VII. PERFORMANCE DATA
The following performance data were provided in support of substantial equivalence determination.
## Biocompatibility
The Venclose Maven Catheter is manufactured using identical materials and is processed by similar manufacturing methods as the predecessor device. EVSRF Catheter. Therefore, the biocompatibility data for the EVSRF Catheter was adopted for the Maven Catheter. The biocompatibility testing was conducted in accordance with FDA guidance document "Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" issued on June 16, 2016, and the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a risk management process" as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity
- Sensitization
- Irritation ●
- Systemic Toxicity ●
- Hemocompatibility
- . Pyrogen Testing
The subject device is considered to be an external communicating device in circulating blood with limited contact duration (≤ 24 hours).
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted for the Venclose Maven System (digiRF Generator and Maven Catheter). The system complies with the following standards:
- . IEC 60601-1:2012, Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2:2014, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances - Requirements and tests
- . IEC 60601-2-2:2017, Medical electrical equipment – Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-1-6:2013, Medical electrical equipment Part 1-6: General requirements for . safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for VENCLOSE. The logo consists of the word "VENCLOSE" in a sans-serif font, with a stylized "V" above it. The "V" is made up of several parallel lines that are colored in a gradient from yellow to orange to pink.
## Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA guidance document dated May 11, 2005: "Guidance for the Content of Premarket Submissions for Software Contained Medical Devices".
### Non-Clinical
Device testing was conducted to evaluate conformance to product specifications. Bench testing included but are not limited to the following tests: Fluid Inaress Test, Battery Test, Ablation Cycle Test, Temperature Accuracy Test, Tensile Test.
### Clinical
An IDE prospective, literature controlled, confirmatory, single-investigator, nonrandomized study of the Venclose Maven System (digiRF Generator and Maven Catheter) to treat incompetent perforator veins (IPVs) was performed. The primary objective of the study was to confirm that the Maven device can be used as intended and can treat incompetent perforator veins without significant adverse outcomes related to the use of the device. The study enrolled 27 subjects. Based on the data, the Venclose Maven System (digiRF Generator and Maven Catheter) is safe and effective for treatment of incompetent perforator veins. In addition, the occlusion rate achieved was 80% and the reflux free rate achieved was 93% at the 30-day follow-up.
#### CONCLUSION VIII.
The conclusions drawn from the non-clinical performance tests and the clinical study demonstrate that the Venclose Maven System is as safe, as effective and performs as well as the Predicate Devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.