K211797 · GS Medical Co., Ltd. · NDN · Oct 28, 2021 · Orthopedic
Device Facts
Record ID
K211797
Device Name
TRACKER Plus Kyphoplasty System
Applicant
GS Medical Co., Ltd.
Product Code
NDN · Orthopedic
Decision Date
Oct 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TRACKER Plus Kyphoplasty System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, and calcaneus. This includes percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
Device Story
The TRACKER Plus Kyphoplasty System is an inflatable bone tamp used for percutaneous vertebral augmentation and fracture reduction in cancellous bone. The system comprises a balloon catheter, cement dispenser kit, and cement mixing system. During a procedure, the balloon catheter is inserted into the target bone; the balloon is inflated to create a void or reduce a fracture; the balloon is then deflated and removed. Cleared PMMA bone cement is subsequently injected into the created void using the dispenser and mixing components. The device is operated by clinicians in a surgical or clinical setting. The output is a physical void in the bone and subsequent stabilization via cement, which aids in fracture reduction and pain management. The device is a mechanical tool; it does not involve electronic processing or software-based decision support.
Clinical Evidence
Bench testing only. Performance testing conducted per ISO 10555-4 to evaluate balloon deflation, burst pressure, fatigue strength, unconstrained burst strength, inflated dimension, tensile bond strength, and insertion/withdrawal force. All results met performance specifications.
Indicated for reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, and calcaneus, including percutaneous vertebral augmentation, using cleared spinal PMMA bone cements.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
October 28, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
GS Medical Co. Ltd. % Barry E. Sands Founder and President RQMIS, Inc. 110 Haverhill Road, Suite 524 Amesbury, Massachusetts 01913
Re: K211797
Trade/Device Name: TRACKER Plus Kyphoplasty System Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: September 23, 2021 Received: September 24, 2021
Dear Barry Sands:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
Device Name
TRACKER Plus Kyphoplasty System
#### Indications for Use (Describe)
The TRACKER Plus Kyphoplasty System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, and calcaneus. This includes percutaneous vertebral augmentation. The system is to be used with cleared spinal polymethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!"
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Pishlishing Services (301) 443-6740 EF
{3}------------------------------------------------
## 510(k) SUMMARY
GS Medical 's TRACKER Plus Kyphoplasty System 510k Submission
## I. SUBMITTER
GS Medical Co. Ltd. 90, Osongsaengmyeong 4-ro Osong-eup, Heungdeok-gu, Cheongju-si Chungcheongbuk-do 28161 Korea Phone: 352.327.4641 Fax: 352.378.2617
#### Contact Person:
Barry Sands RQMIS, Inc. 110 Haverhill Road, Suite 526 Amesbury, MA 01913 Phone: +1 (978) 358-7307 Fax: +1 (978) 477-0206
Date Prepared: September 23, 2021
## II. DEVICE
Trade/Device Name: TRACKER Plus Kyphoplasty System Common or Usual Name: Inflatable Bone Tamp Classification Name: Orthopedics Regulation Number: 21 CFR 888.3027 Regulatory Class: Class II Product codes NDN, HRX
{4}------------------------------------------------
## III. PREDICATE DEVICES
| Device | Predicate | Regulation | Product<br>Code | Manufacturer | 510(k)<br>Number |
|----------------------------------------------|-----------|--------------------|-----------------|------------------------|------------------|
| TRACKER<br>Kyphoplasty<br>System | Primary | 21 CFR<br>888.3027 | NDN, HRX | GS Medical Co.<br>Ltd. | K192335 |
| 11G InterV<br>Kyphoplasty<br>Catheter (Flex) | Secondary | 21 CFR<br>888.3027 | NDN, HRX | Pan Medical Ltd. | K162453 |
# IV. DEVICE DESCRIPTION
The TRACKER Plus Kyphoplasty System is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine, tibia, radius, and calcaneus. This includes percutaneous vertebral augmentation. This system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation, such as kyphoplasty.
The TRACKER Plus Kyphoplasty System consists of the balloon catheter, cement dispenser kit, and cement mixing system. The cement dispenser kit and the cement mixing system are intended to be used with the balloon catheter, but are sold separately.
# V. SUBSTANTIAL EQUIVALENCE
# Technological Comparison
The TRACKER Plus Kyphoplasty System is substantially equivalent to other legally marketed inflatable bone tamps. Specifically, the TRACKER Plus Kyphoplasty System is substantially equivalent to TRACKER Kyphoplasty System. The TRACKER Plus Kyphoplasty System has the same general intended use and indications for use and has only minor differences in technological characteristics and principles of operation as the previously cleared primary predicate TRACKER Kyphoplasty System (K192335). The TRACKER Plus Kyphoplasty System also has the same general intended use, technological characteristics, and principles of operation as the previously cleared secondary predicate 11G InterV Kyphoplasty Catheter (Flex) (K162453).
{5}------------------------------------------------
# Performance Comparison
In all instances the device functioned as intended and all results were satisfactory and met all performance specifications. The tests performed were:
- Balloon Deflation
- Burst Pressure
- Fatigue Strength
- Unconstrained Burst Strength
- Inflated Dimension
- •Tensile bond strength test
- •Insertion and withdrawal force test
# Comparison to the Predicate Device
| | Subject Device | Primary Predicate<br>Device | Secondary Predicate<br>Device |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | TRACKER Plus<br>Kyphoplasty System | TRACKER Kyphoplasty<br>System | 11G InterV Kyphoplasty<br>Catheter (Flex) |
| 510k Number | - | K192335 | K162453 |
| Manufacturer | GS MEDICAL Co., Ltd. | GS MEDICAL Co., Ltd. | Pan Medical Ltd. |
| Product Code | HRX, NDN | HRX, NDN | HRX, NDN |
| Common Name | Inflatable Bone Tamp | Inflatable Bone Tamp | Inflatable Bone Tamp |
| Indication for<br>use | The TRACKER Plus<br>Kyphoplasty System is<br>intended to be used for<br>the reduction of<br>fractures and/or<br>creation of a void in<br>cancellous bone in the<br>spine, tibia, radius, and<br>calcaneus. This includes<br>percutaneous vertebral<br>augmentation. The<br>system is to be used<br>with cleared spinal<br> | The TRACKER<br>Kyphoplasty System is<br>intended to be used for<br>the reduction of<br>fractures and/or<br>creation of a void in<br>cancellous bone in the<br>spine, tibia, radius, and<br>calcaneus. This includes<br>percutaneous vertebral<br>augmentation. The<br>system is to be used<br>with cleared spinal<br>polymethylmethacrylat<br>e (PMMA) bone<br>cements indicated for<br>use during<br>percutaneous vertebral<br>augmentation, such as<br>kyphoplasty. | 8G InterV Kyphoplasty<br>Catheter, 11G InterV<br>Kyphoplasty Catheter<br>(Mini), 13G InterV<br>Kyphoplasty Catheter<br>(Micro) and 11G InterV<br>Kyphoplasty Catheter<br>(Flex) are intended to<br>be used for reduction<br>and fixation of<br>fractures and/or<br>creation of a void in<br>cancellous bone in the<br>spine during balloon<br>kyphoplasty (for use<br>with cleared spinal<br>polymethylmethacrylat<br>e (PMMA) bone<br>cements). |
| Components | - Bone Catheter<br>- Expander Syringe<br>- Kit<br>(Needle Pipe, Needle<br>Pin, Expander, Cannula,<br>Spacer, Guide Wire,<br>Wire Pin, Cement<br>Pusher, Cement Filler,<br>and Guide wire) | - Bone Catheter<br>- Expander Syringe<br>- Kit<br>(Needle Pipe, Needle<br>Pin, Expander, Cannula,<br>Spacer, Guide Wire,<br>Wire Pin, Cement<br>Pusher, Cement Filler,<br>and Guide wire) | - Balloon Catheter |
| Balloon Size | 10mm, 15mm, 20mm | 10mm, 15mm, 20mm | 20mm |
| Bone Tamp<br>Max. Inflation<br>Pressure | 350 PSI | 350 PSI | 400 PSI |
{6}------------------------------------------------
{7}------------------------------------------------
| Composition of<br>Material | Thermoplastic<br>Polyurethane<br>Platinum<br>Polycarbonate & ABS<br>Stainless Steel & ABS | Thermoplastic<br>Polyurethane<br>Platinum<br>Polycarbonate & ABS<br>Stainless Steel & ABS | Balloon Material:<br>Polyurethan |
|----------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|----------------------------------|
| Guide Wire<br>Material | Stainless Steel, Nitinol | Stainless Steel | Nitinol |
| Packaging | Pouch, Tyvek Blister<br>Tray,<br>Cardboard Box | Pouch, Tyvek Blister<br>Tray,<br>Cardboard Box | Pouch,<br>Cardboard Box |
| Sterilization | EO Sterilization | EO Sterilization | EO Sterilization |
| Biocompatibility | Meets ISO 10993 | Meets ISO 10993 | Meets ISO 10993 |
# VI. NON-CLINICAL TESTING
The performance test was conducted using Tracker Plus Balloon Catheter to evaluate balloon deflation, burst pressure, fatigue strength, unconstrained burst strength, and inflated dimension as per "ISO 10555-4: Intravascular catheters – Sterile and Single use catheters – Part 4: Balloon dilation catheters."
# VII. CONCLUSION
The TRACKER Plus Kyphoplasty System is as safe and effective as the TRACKER Kyphoplasty System (Primary Predicate, K192335) and 11G InterV Kyphoplasty Catheter (Flex) (K162453). The TRACKER Kyphoplasty System has the same intended uses and indications for use and has only minor differences in technological characteristics and principles of operation as the previously cleared primary predicate. The TRACKER Plus Kyphoplasty System also has the same general intended use, technological characteristics, and principles of operation as the previously cleared secondary predicate. The technological differences between the subject device and the predicates do not raise new questions of safety and effectiveness. Performance data demonstrate that the TRACKER Plus Kyphoplasty System is as safe and effective as the primary and secondary predicate. Thus, the TRACKER Kyphoplasty System is substantially equivalent to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.