Ez3D-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist. Ez3D-i is intended for use as software to load, view and save DICOM images from CT, panorama, cephalometric and intraoral imaging equipment and to provide 3D visualization, 2D analysis, in various MPR (Multi-Planar Reconstruction) functions.
Device Story
Ez3D-i is dental imaging software for viewing, processing, and analyzing DICOM-compliant images (CT, panorama, cephalometric, intraoral). It provides 3D visualization, 2D analysis, and Multi-Planar Reconstruction (MPR). Features include implant simulation, canal drawing, and bone density assessment. Used in dental clinics by radiologists and dentists; software runs on standard IBM-compatible PCs or networks. Clinicians use the output for diagnostic interpretation and treatment planning. The device facilitates effective communication between doctors and patients and supports precise clinical decision-making for dental procedures.
Clinical Evidence
No clinical data. Performance established via software verification/validation and measurement accuracy testing. Device passed all tests against pre-determined pass/fail criteria.
Technological Characteristics
Software-only device; no patient contact. Operates on IBM-compatible PCs/networks (Windows 8, 10). Supports DICOM image format. Features include MPR, 3D reformation, implant simulation, and measurement tools (length, angle, volume). Connectivity via standard PC network. No specific hardware requirements or sterilization needed.
Indications for Use
Indicated for maxillofacial radiographic imaging in patients requiring dental radiology analysis. Intended for use by trained medical professionals, including radiologists and dentists, to view and interpret DICOM-compliant images from CT, panoramic, cephalometric, and intraoral imaging equipment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
August 20, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Ewoosoft Co., Ltd. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 18881 Von Karman Ave., STE 160 IRVINE CA 92612
Re: K211791
Trade/Device Name: Ez3D-i/E3 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: June 4, 2021 Received: June 10, 2021
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K211791
Device Name Ez3D-i /E3
#### Indications for Use (Describe)
Ez3D-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist.
Ez3D-i is intended for use as software to load, view and save DICOM images from CT, panorama, cephalometric and intraoral imaging equipment and to provide 3D visualization, 2D analysis, in various MPR (Multi-Planar Reconstruction) functions.
| Type of Use (Select one or both, as applicable) | <input checked="true" type="checkbox"/> <input type="checkbox"/> |
|-------------------------------------------------|------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv
(K211791)
This summary of 510(k) information is being submitted in accordance with requirements of 21 CFR Part 807.92.
#### 1. Date: 8/9/2021
## 2. Applicant / Submitter
Ewoosoft Co., Ltd. 801-ho, Vatechnetworks Bldg., 13, Samsung 1-ro 2-gil, Hwaseong-si, Gyeonggido , Republic of Korea Tel: +82 31 8015 6172 Fax: +82 31 8015 6196 Contact person: Young Seok Kim Email: eddie.kim@ewoosoft.com
## 3. U.S. Designated Agent
Priscilla Chung LK Consulting Group USA, Inc. 18881 Von Karman Ave. STE 160 Irvine, CA 92612 Fax: 714.409.3357 Tel: 714.202.5789 Email: juhee.c@LKconsultingGroup.com
#### 4. Device Information:
- Trade/Device Name: Ez3D-i /E3
- Regulation Number: 21 CFR 892.2050
- Regulation Name: Medical Image Management and Processing System ●
- Regulatory Class: Class II
- Product Code: LLZ
## 5. Predicate Device:
- Manufacturer: Ewoosoft Co., Ltd. ●
- Trade/Device name: Ez3D-i /E3
- 510(k) number: K200178
- Regulation number 21 CFR 892.2050 ●
- . Regulation name: Picture Archiving and Communications System
- Regulatory Class: Class II ●
- . Classification Product Code: LLZ
#### Page 1 of 4
{4}------------------------------------------------
#### 6. Device Description:
Ez3D-i is 3D viewing software for dental CT images in DICOM format with a host of useful functions including MPR, 2-dimensional analysis and 3-dimensional image reformation. It provides advanced simulation functions such as Implant Simulation, Drawing Canal, and Implant Environ Bone Density, etc. for the benefit of effective doctor and patient communication and precise treatment planning.
## 7. Indication for use:
Ez3D-i is dental imaging software that is intended to provide diagnostic tools for maxillofacial radiographic imaging. These tools are available to view and interpret a series of DICOM compliant dental radiology images and are meant to be used by trained medical professionals such as radiologist and dentist.
Ez3D-i is intended for use as software to load, view and save DICOM images from CT, panorama, cephalometric and intraoral imaging equipment and to provide 3D visualization, 2D analysis, in various MPR (Multi-Planar Reconstruction) functions.
{5}------------------------------------------------
## 8. Substantial Equivalence:
| | Subject Device | Predicate Device |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Ez3D-i v5.3(E3) | Ez3D-i v5.2(E3) |
| 510K number | K211791 | K200178 |
| Manufacturer | Ewoosoft | Ewoosoft |
| Indications for use | Ez3D-i is dental imaging software that is<br>intended to provide diagnostic tools for<br>maxillofacial radiographic imaging. These<br>tools are available to view and interpret a<br>series of DICOM compliant dental radiology<br>images and are meant to be used by trained<br>medical professionals such as radiologist<br>and dentist.<br>Ez3D-i is intended for use as software to<br>load, view and save DICOM images from<br>CT, panorama, cephalometric and intraoral<br>imaging equipment and to provide 3D<br>visualization, 2D analysis, in various MPR<br>(Multi-Planar Reconstruction.) functions | Ez3D-i is dental imaging software that is<br>intended to provide diagnostic tools for<br>maxillofacial radiographic imaging. These<br>tools are available to view and interpret a<br>series of DICOM compliant dental<br>radiology images and are meant to be used<br>by trained medical professionals such as<br>radiologist and dentist.<br>Ez3D-i is intended for use as software to<br>load, view and save DICOM images from<br>CT, panorama, cephalometric and intraoral<br>imaging equipment and to provide 3D<br>visualization, 2D analysis, in various MPR<br>(Multi-Planar Reconstruction.) functions |
| Platform | IBM-compatible PC or PC network | IBM-compatible PC or PC network |
| Operating System | Microsoft Window 8, 10 | Microsoft Window 7, 8, 10 |
| User Interface | Mouse, Keyboard | Mouse, Keyboard |
| Image Input Sources | Images can be scanned, loaded from digital<br>cameras or card readers, or imported from a<br>radiographic imaging device | Images can be scanned, loaded from digital<br>cameras or card readers, or imported from a<br>radiographic imaging device |
| 32 bit / 64 bit | 32 / 64 bit | 32 / 64 bit |
| Image format | DICOM | DICOM |
| Image Measurement<br>Tools | Length, Multi Length, Angle, Multi Angle,<br>Circle, ROI/Area, Volume, Profile | Length, Multi Length, Angle, ROI/Area,<br>Volume, Profile |
| Image viewing | Full, side by side, gallery, thumbnail | Full, side by side, gallery, thumbnail |
| Image manipulation | Grayscale, invert, emboss, brightness,<br>contrast, gamma, sharpen, median,<br>despeckle, hue, saturation, equalize, flip,<br>mirror, masking, rotate, magnify, annotation,<br>cephalometric tracing, ceph growth<br>projections, implant simulations | Grayscale, invert, emboss, brightness,<br>contrast, gamma, sharpen, median,<br>despeckle, hue, saturation, equalize, flip,<br>mirror, masking, rotate, magnify,<br>annotation, cephalometric tracing, ceph<br>growth projections, implant simulations |
| 3D imaging<br>capability | Ez3D-i can view, transfer and process 3D<br>radiographs. Furthermore, it supports Smart<br>Click, Smart Clipping, Implant Simulation<br>and Canal Draw. | Ez3D-i can view, transfer and process 3D<br>radiographs. Furthermore, it supports Smart<br>Click, Smart Clipping, Implant Simulation<br>and Canal Draw. |
| Image annotation | Text, paint, ellipse, pointer, select, draw,<br>magnify, line, rectangle, polygon, ruler,<br>protractor, smile library, smudge, brush,<br>redeye reduction, select region, copy / paste | Text, paint, ellipse, pointer, select, draw,<br>magnify, line, rectangle, polygon, ruler,<br>protractor, smile library, smudge, brush,<br>redeye reduction select region copy/paste |
Ez3D-i described in this 510(k) has the same intended use and the same technical characteristics as the unmodified device (K200178, Ez3D-i /E3).
The subject device and the predicate devices are substantially equivalent, having the same indications for use and functionalities like operation software, computer platform, picture
{6}------------------------------------------------
archiving and communication format, image format, image processing features, windowing, 3D image construction, image edit, measurements and manipulation.
The differences are that the subject device has additional features such as changes in PC system requirement information, login/logout function, the location of the User Account Management to Server Web Console EzServer, license plan options setting, export icon function, show/hide patient information icon function, adding and deleting VR coloring graph function, MPR interval function, and export to DICOM function. These differences are not significant since they do not raise any new or potential safety risks to the user or patient and questions of safety or effectiveness. Based on the test results submitted in this 510K, we conclude that the subject device is substantially equivalent to the predicate devices.
## 9. Technological Characteristics:
Ez3D-i is a software device that does not contact the patient, nor does it control any life sustaining devices. Results produced by the software's diagnostic, treatment planning and simulation tools are dependent on the interpretation of trained and licensed radiologists. clinicians and referring physicians as an adjunctive to standard radiology practices for diagnosis.
#### 10. Performance Data:
SW verification/validation and the measurement accuracy test were conducted to establish the performance, functionality and reliability characteristics of the modified devices. The device passed all of the tests based on pre-determined Pass/Fail criteria.
#### 11. Conclusion:
The subject device is substantially equivalent in the areas of technical characteristics, general function, application, and indications for use. The new device does not introduce a fundamentally new scientific technology, and the device has been validated through system level test. Therefore, we conclude that the subject device described in this submission is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.