Agfa's DX-D Imaging Package with XD Detectors is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of the human anatomy. The DX-D Imaging Package with XD Detectors may be used wherever conventional screen-film systems may be used. Agfa's DX-D Imaging Package with XD Detectors s not indicated for use in mammography.
Device Story
Direct radiography (DR) system capturing general projection radiographic images; utilizes flat-panel detectors (scintillator-photodetector type: Cesium Iodide or Gadolinium Oxysulfide) and NX workstation with MUSICA image processing. Operates by converting X-ray input into digital images for display; includes offline workflow capability storing up to 200 images on the detector. Used in clinical settings by radiographers/technicians. Output viewed by clinicians to support diagnostic decision-making. Benefits include digital image capture replacing conventional screen-film systems, improved workflow efficiency, and diagnostic quality imaging for general patient populations.
Clinical Evidence
No clinical trials or animal studies performed. Bench testing only. Image quality evaluated using anthropomorphic adult and pediatric phantoms by internal experts. Performance/functionality and usability testing confirmed device meets acceptance criteria. Software verification and validation completed for NX 23.
Technological Characteristics
Solid-state flat-panel detectors (CsI or GOS scintillator); NX workstation; MUSICA image processing. Connectivity: Wireless/Wired. Standards: IEC 60601-1, IEC 60601-1-2, ISO 13485, ISO 14971, DICOM (PS3.1-3.20), IEC 62366-1, IEC 62304. Software level of concern: Moderate.
Indications for Use
Indicated for general projection radiographic applications to capture diagnostic quality images of human anatomy in adult and pediatric patients. Not indicated for mammography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
July 30, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Agfa N.V. % Mrs. ShaeAnn Cavanagh Regulatory Affairs Manager, North America Agfa US Corp. 10 South Academy Street GREENVILLE SC 29601
Re: K211790
Trade/Device Name: DX-D Imaging Package with XD Detectors Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: June 9, 2021 Received: June 10, 2021
Dear Mrs. Cavanagh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211790
Device Name DX-D Imaging Package with XD Detectors
### Indications for Use (Describe)
Agfa's DX-D Imaging Package with XD Detectors is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of the human anatomy. The DX-D Imaging Package with XD Detectors may be used wherever conventional screen-film systems may be used.
Agfa's DX-D Imaging Package with XD Detectors s not indicated for use in mammography.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
### Agfa N.V. DX-D Imaging Package with XD Detectors K211790
#### I. SUBMITTER
Agfa N.V. Septestraat 27 B-2640 Mortsel Belgium Contact: Wim Govaerts, Prepared: June 9, 2021 Telephone: + 32 3444 6246
#### II. DEVICE
Name of Device: DX-D Imaging Package with XD Detectors
Common Name: Solid-State X-Ray Imager (Flat Panel/Digital Imager) Classification Name: Stationary X-ray System Regulatory Classification: Class II, 21 CFR 892.1680 Product Code: MQB
#### III. PREDICATE DEVICE(S)
This is a 510(k) for Agfa's DX-D Imaging Package with XD Detectors which is a solid state x-ray system, a direct radiography (DR) system. It is substantially equivalent to Vieworks Co. Ltd's predicate device, Vivix-S VW (K200418) and Agfa's reference device, DX-D Imaging Package -DX-D 40 (K142184).
Predicate Device: Vivix-S VW Common Name: Solid-State X-Ray Imager (Flat Panel/Digital Imager) Classification Name: Stationary X-ray System Regulatory Classification: Class II, 21 CFR 892.1680 Product Code: MQB
Reference Device: DX-D Imaging Package - DX-D 40 Common Name: Solid-State X-Ray Imager (Flat Panel/Digital Imager) Classification Name: Stationary X-ray System Regulatory Classification: Class II, 21 CFR 892.1680 Product Code: MQB
Neither the Vivix-S VW (K200418) predicate device nor the DX-D Imaging Package - DX-D 40 (K142184) reference device have not been subject to a design-related recall. However, several Agfa X-Ray systems that use the DX-D Imaging Package have identified "unintended movement" adverse events. Agfa has taken these reported defects into consideration and is mindful of these events while developing the subject device, DX-D Imaging Package with XD Detectors.
{4}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The DX-D Imaging Package, previously cleared under K142184, is a solid state x-ray system, a direct radiography (DR) system (product code MQB) intended to capture general radiographic images of the human body. It is a combination of Agfa's NX workstation with MUSICA TM image processing and one or more flat-panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS). It is capable of replacing other direct radiography, including computed radiography systems with conventional or phosphorous film cassettes.
This submission is to add the XD Detectors (XD 10/10+, XD 14/14+ and XD 17/17+) Flat Panel Detectors to Agfa's DX-D Imaging Package portfolio. Agfa's XD Detectors are currently marketed by Vieworks Co. Ltd. as FXRD-4343VAW/VAW Plus. FXRD-3643VAW/VAW Plus. and FXRD-2530VAW/VAW Plus which is the predicate for this submission (K200418).
The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are identical to those previously cleared in the DX-D Imaging Package - DX-D 40 (K142184reference) device and other devices in Agfa's radiography portfolio today. The addition of the offline workflow is identical to the Vivix-S VW (K200418) predicate device.
Principles of operation and technological characteristics of the new, predicate and reference devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically similar to the predicate device (K200418) which includes the addition of an offline workflow capable of storing up to 200 images on the flat-panel detector for later viewing. It uses the same NX workstation with MUSICA™ image processing as the reference device (K142184) and the same flat panel detectors of the scintillator-photodetector type (Cesium Iodide - Csl or Gadolinium Oxysulfide - GOS) to capture and digitize the images as the predicate device (K200418). Laboratory data and image quality evaluations conducted with internal specialists confirm that performance is equivalent to the predicate. Differences in devices do not alter the intended diagnostic effect nor do they impact the safety and effectiveness of the device.
Performance characteristics between the subject device and reference device are summarized on the next page in tabular format.
{5}------------------------------------------------
| Performance | DX-D 40 Flat-<br>Panel<br>Detector<br>(K142184) | DX-D 45 Flat-<br>Panel Detector<br>(K142184) | DX-D 60 Flat-<br>Panel Detector<br>(K142184) | XD 10/10+<br>Wireless Detector<br>(new device) | XD 14/14+<br>Wireless Detector<br>(new device) | XD 17/17+<br>Wireless Detector<br>(new device) |
|-----------------------|-------------------------------------------------|----------------------------------------------|----------------------------------------------|------------------------------------------------|------------------------------------------------|------------------------------------------------|
| Characteristics | | | | | | |
| Model | 6007/100 &<br>6007/200 | 6007/101 &<br>6007/201 | 6007/110 &<br>6007/111 | FXRD-2530VAW<br>(PLUS) | FXRD-3643VAW<br>(PLUS) | FXRD-4343VAW<br>(PLUS) |
| Manufacturer | Vieworks Co.<br>Ltd. | Vieworks Co.<br>Ltd. | Vieworks Co.<br>Ltd. | Vieworks Co. Ltd. | Vieworks Co. Ltd. | Vieworks Co. Ltd. |
| Spatial<br>Resolution | 3.5 lp/mm | 4.0 lp/mm | 3.5 lp/mm | 4.0 lp/mm | 3.5 lp/mm | 3.5 lp/mm |
| DQE | 1 lp/mm –<br>CsI: 0.494<br>GOS: 0.259 | 1 lp/mm –<br>CsI: 0.524<br>GOS: 0.282 | 1 lp/mm –<br>CsI: 0.520<br>GOS: 0.294 | 1 lp/mm –<br>XD: 0.500<br>XD+: 0.587 | 1 lp/mm –<br>XD: 0.425<br>XD+: 0.587 | 1 lp/mm –<br>XD: 0.412<br>XD+: 0.587 |
| | 2 lp/mm –<br>CsI: 0.379<br>GOS: 0.157 | 2 lp/mm –<br>CsI: 0.450<br>GOS: 0.203 | 2 lp/mm –<br>CsI: 0.443<br>GOS: 0.212 | 2 lp/mm –<br>XD: 0.401<br>XD+: 0.445 | 2 lp/mm –<br>XD: 0.321<br>XD+: 0.399 | 2 lp/mm –<br>XD: 0.345<br>XD+: 0.407 |
| | 3 lp/mm –<br>CsI: 0.215<br>GOS: 0.061 | 3 lp/mm –<br>CsI: 0.314<br>GOS: 0.118 | 3 lp/mm –<br>CsI: 0.301<br>GOS: 0.105 | 3 lp/mm –<br>XD: 0.288<br>XD+: 0.316 | 3 lp/mm –<br>XD: 0.206<br>XD+: 0.257 | 3 lp/mm –<br>XD: 0.220<br>XD+: 0.280 |
| MTF | 1 lp/mm –<br>CsI: 0.685<br>GOS: 0.589 | 1 lp/mm –<br>CsI: 0.685<br>GOS: 0.594 | 1 lp/mm –<br>CsI: 0.698<br>GOS: 0.588 | 1 lp/mm –<br>XD: 0.729<br>XD+: 0.650 | 1 lp/mm –<br>XD: 0.751<br>XD+: 0.635 | 1 lp/mm –<br>XD: 0.726<br>XD+: 0.656 |
| | 2 lp/mm –<br>CsI: 0.386<br>GOS: 0.266 | 2 lp/mm –<br>CsI: 0.383<br>GOS: 0.277 | 2 lp/mm –<br>CsI: 0.405<br>GOS: 0.272 | 2 lp/mm –<br>XD: 0.424<br>XD+: 0.315 | 2 lp/mm –<br>XD: 0.446<br>XD+: 0.302 | 2 lp/mm –<br>XD: 0.428<br>XD+: 0.311 |
| | 3 lp/mm –<br>CsI: 0.209<br>GOS: 0.115 | 3 lp/mm –<br>CsI: 0.208<br>GOS: 0.129 | 3 lp/mm –<br>CsI: 0.224<br>GOS: 0.117 | 3 lp/mm –<br>XD: 0.236<br>XD+: 0.157 | 3 lp/mm –<br>XD: 0.247<br>XD+: 0.152 | 3 lp/mm –<br>XD: 0.231<br>XD+: 0.161 |
Agfa N.V. Premarket Notification: DX-D Imaging Package - XD Detectors
Table 1: Detector Performance Characteristics
Configuration information for the flat-panel detectors can be found in the XD 10/10+, XD 14/14+, and XD 17/17+ (K200418) and DX-D 40/45/60 (K142184) User Manuals. The XD 10/10+, XD 14/14+, and XD 17/17+ detectors can be integrated in an X-ray system that communicates to a workstation. The Service Manual details the possible configurations and integrations with the NX workstation and X-ray generator. All of Agfa's DR X-ray systems (i.e. DX-D 100-K103597, DX-D 300-K103050, DX-D 600-K112670, DR 400-K141192, DR 600-K152639, DR 800-K183275) will integrate with the detectors.
#### V. INDICATIONS FOR USE
Agfa's DX-D Imaging Package with XD Detectors is indicated for use in general projection radiographic applications to capture for display diagnostic quality radiographic images of the human anatomy. The DX-D Imaging Package with XD Detectors may be used wherever conventional screen-film systems may be used.
Agfa's DX-D Imaging Package with XD Detectors is not indicated for use in mammography.
NOTE: The mammography applications embedded in the MUSICA software are for previously cleared CR imaging applications (K081963) and not intended for direct radiography (DR) imaging.
{6}------------------------------------------------
Furthermore, the additional mammography software is only available through additional license keys that must be purchased. These license keys are only available outside of the USA.
NOTE: DSA is not part of this submission; therefore, the functionality mentioned in the labeling documentation is out of scope for this clearance. DSA software is controlled using license keys and will not be accessible in this release of NX 23. Agfa intends to submit a 510(k) premarket notification for DSA on the DR 800 later this year.
## PEDIATRIC USE SUMMARY
The DX-D Imaging Package with XD Detectors is intended for general populations, including adult and pediatric patients of all ages. There are no specific pediatric and neonatal design features; however, the DX-D Imaging Package with XD Detectors provides the following specific design features and instructions that enable safer use of the device with pediatric and neonatal patients:
| Pediatric Imaging Design Features | Standard or Optional |
|-----------------------------------|--------------------------------------------------------------------|
| Protocols or exposure indices | Standard - make own exam tree<br>optional - make use of age groups |
| Post-processing application | Standard |
| Reconstruction algorithm | Standard |
## Table 2: Pediatric Design Features
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES
The new device. predicate device (K200418) and the reference device (K142184) are solid state imaging devices, Product Code MQB. Agfa's DX-D Imaging Package with XD Detectors is substantially equivalent to both the predicate device (K200418) and the reference device (K142184) in that it uses the same basic technology to capture and transmit images.
Principles of operation and technological characteristics of the new, predicate and reference devices are the same. There are no changes to the intended use/indications of the device. The new device is physically and electronically similar to the predicate device (K200418) which includes the addition of an offline workflow capable of storing up to 200 images on the flat-panel detector for later viewing. It uses the same NX workstation with MUSICA™ image processing as the reference device (K142184) and the same flat panel detectors of the scintillator-photodetector type (Cesium Iodide - CsI or Gadolinium Oxysulfide - GOS) to capture and digitize the images as the predicate device (K200418). Differences in devices do not alter the intended diagnostic effect nor do they impact the safety and effectiveness of the device.
The optional image processing allows users to conveniently select image processing settings for different patient sizes and examinations. The image processing algorithms in the new device are identical to those previously cleared in the DX-D Imaging Package - DX-D 40 (K142184reference) device and other devices in Agfa's radiography portfolio today. The addition of the offline workflow is identical to the Vivix-S VW (K200418) predicate device.
{7}------------------------------------------------
### Agfa N.V. Premarket Notification: DX-D Imaging Package - XD Detectors
The DX-D Imaging Package with XD Detectors indications for use statement is substantially equivalent to both the predicate device (K200418) and the reference device (K142184). The new device and both the predicate device (K200418) and the reference device (K142184) include the statement that the device is not indicated for mammography The new device and both the predicate device (K200418) and the reference device (K142184) have similar indications for use statements in that both devices are intended to display general radiographic images of the human body and can be used in place of conventional film/screen systems. Intended uses are the same. The devices have the same technological characteristics.
The only difference between the new device and predicate device (K200418) Indication for Use statement is that the new device's Indication For Use statement includes specific references to the human body; however, standard practice dictates the predicate device is used for diagnostic procedures that include human anatomy. The differences between the subject device, DX-D Imaging Package with XD Detectors, predicate device K200418 and reference device K142184 do not seem to impact safety or effectiveness.
Performance data, image quality clinical evaluations, and usability/functionality data are adequate to ensure equivalence.
Similarities and differences between the subject device and predicate are summarized on the next page in tabular format.
{8}------------------------------------------------
| PRODUCT COMPARISON TABLE | | | | | | | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|
| | DX-D Imaging Package with<br>XD Detectors<br>(New Device) | Vieworks - Vivix-S VW<br>(PREDICATE –<br>K200418) | Explanation of Differences | | | | | |
| Communications | Same as Predicate | Wireless/Wired | N/A | | | | | |
| Flat Panel | Same as Predicate | Flat Panel Detector | N/A | | | | | |
| Detector Material | Same as Predicate | Cesium Iodide (CsI)<br>scintillator | N/A | | | | | |
| Detector Sizes | Same as Predicate | 25x30cm, 35x43cm/<br>14x17in, 43x43cm/<br>17x17in | N/A | | | | | |
| Interface to<br>Generator | Same as Predicate | AED & Synchronized | N/A | | | | | |
| Pixel size | Same a Predicate | XD 10/10+ - 124 μm<br>XD 14/14+ - 140 μm<br>XD 17/17+ - 140 μm | N/A | | | | | |
| Dynamic Range | Same as Predicate | 16 bit | N/A | | | | | |
| Power Supply | Same as Predicate | Battery: replaceable &<br>rechargeable | N/A | | | | | |
| Operator<br>Workstation | NX Workstation | N/A, workstation is not<br>included in the system | DX-D Imaging Package includes<br>the NX workstation for image<br>acquisition. The predicate device<br>only includes the flat panel<br>detector. | | | | | |
| Image processing | MUSICA Imaging Processing | 3rd Party Software | Agfa's proprietary image<br>processing software | | | | | |
| Indications for Use | Agfa's DX-D Imaging Package<br>with XD Detectors is indicated<br>for use in general projection<br>radiographic applications to<br>capture for display diagnostic<br>quality radiographic images of<br>human anatomy. The DX-D<br>Imaging Package with XD<br>Detectors may be used<br>wherever conventional screen-<br>film systems may be used.<br>Agfa's DX-D Imaging Package<br>with XD Detectors is not<br>indicated for use in<br>mammography. | The Vivix-S VW series is<br>used for the general-<br>purpose diagnostic<br>procedures, as well as<br>intended to replace<br>radiographic film/ screen<br>systems.<br>The Vivix-S VW is not<br>intended for<br>mammography<br>applications. | New device specifically<br>references human anatomy,<br>whereas the predicate device<br>states general-purpose diagnostic<br>procedures. However, standard<br>practice dictates diagnostic<br>procedures would include human<br>anatomy. | | | | | |
### Table 3: Device Comparison Table
#### PERFORMANCE DATA VII.
Laboratory testing and software testing (for a moderate level of concern device) using equivalent test protocols were evaluated by qualified individuals to demonstrate that adequate design controls (according to 21 CFR 820.30) were in place.
{9}------------------------------------------------
Verification and validation testing confirmed the device meets performance, safety, usability and security requirements. Pediatric indications were also taken into account. Results were verified and validated.
No clinical trials were performed in the device. No animal or clinical studies were performed in the development of the new device. No patient treatment was provided or withheld.
# Bench Testing
Image quality evaluations for adults and pediatric patients, performance/functionality and usability data has been provided.
- o Technical and acceptance testing was completed on the DR Retrofit and XD Detectors in order to confirm the medical device functions and performs as intended. All deviations or variances are documented in a defect database and addressed in the CRD documentation and verified. All mitigations have been tested and passed. All design input requirements have been tested and passed. All planned verification activities have been successfully completed.
- Performance functionality evaluations were conducted with four qualified experts. The results of these tests fell within the acceptance criteria for the DX-D Imaging Package with XD Detectors; therefore, the DX-D Imaging Package (DR Retrofit) supports XD Detectors and an offline workflow.
- Clinical image validation was conducted using anthropomorphic adult and pediatric phantoms and evaluated by qualified internal experts. The radiographers evaluated the DX-D Imaging Package with XD Detectors with DX-D 10/20 and DX-D 40/45/60 flat-panel detectors by comparing overall image quality. The image quality testing also validated MUSICA software of the imaging processing for general radiography (GenRad) applications for the DX-D Imaging Package with XD Detectors.
Clinical image quality evaluation is not essential in establishing substantial equivalence for the DX-D Imaging Package with XD Detectors. Adequate Bench Testing results should be sufficient to determine device safety and effectiveness.
## Software Verification and Validation Testing
Verification and validation plans comprise of test protocols. The complete device has been certified and validated. During the final risk analysis meeting, the risk management team concluded that the medical risk is no greater than with conventional x-ray film previously released to the field.
Software verification testing for NX 23 was completed in scope of the DX-D Imaging Package with XD Detectors. Two software iterations were tested, including the MUSICA image processing and VDI with the three XD Detectors (XD 10/10+, XD 14/14+, XD 17/17+). HERDE defects were identified; however, they were solved between test execution and the completion of the final report and will be part of the next maintenance software release.
{10}------------------------------------------------
For the NX 23 (NX Orion) software there are a total of 535 risks in the broadly acceptable region and 37 risks in the ALARP region with only four of these risks identified. Zero risks were identified in the Not Acceptable Region. Therefore, the device is assumed to be safe, the benefits of the device are assumed to outweigh the residual risk. The software risk assessment is assessed on solution level for the DX-D Imaging Package with XD Detectors (DR Retrofit).
The term "Level of Concern" means the level of risk that the software device is determined to be if the software were to fail. The Level of Concern for the DX-D Imaging Package with XD Detectors and NX 23 has been determined to be moderate.
NOTE: DSA is not part of this submission; therefore, the functionality mentioned in the labeling documentation is out of scope for this clearance. DSA software is controlled using license keys and will not be accessible in this release of NX 23. Agfa intends to submit a 510(k) premarket notification for DSA on the DR 800 later this year.
## Electrical Safety and Electromagnetic Compatibility (EMC) Testing:
- IEC 60601-1: 2005 Medical Electrical Equipment: General Requirements for Safety and Essential Performance.
- IEC 60601-1-2: 2014 Medical Electrical Equipment Part 1-2: General Requirements for ● Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements and Tests.
The DX-D Imaging Package with XD Detectors do not fall under the scope of FDA Subchapter J mandated performance standards 21 CFR 1020.30 and 1020.31 since there is no risk of radiation leakage or change in radiation during the use of this device.
Agfa's in-house standard operating procedures were also used for the development of the device and software; these procedures conform to the following standards:
- ISO 13485:2015 Medical Devices Quality Management Systems ●
- ISO 14971:2012 Application of Risk Management to Medical Devices
- ACR/NEMA PS3.1-3.20: 2011 Digital Imaging and Communications in Medicine (DICOM) ●
- IEC 62366-1:2015 Medical Devices Part 1: Application of Usability Engineering to Medical ● Devices
- IEC 62304:2006 Medical Device Software Software Lifecycle Processes [Including ● Amendment 1 (2016)]
### Guidance Documents
Agfa utilized the following guidance documents in the development of the DX-D Imaging Package with XD Detectors:
- Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices (September ● 2016).
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005)
- Cybersecurity for Networked Medical Devices Containing Off-The-Shelf (OTS) Software (Jan 2005)
- Off-the-Shelf Software Use in Medical Devices (September 2019) ●
{11}------------------------------------------------
- . Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (Oct 2018)
- Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications (Nov ● 2017)
## Summary
Based on the performance data as documented in the above testing, the DX-D Imaging Package with XD Detectors is found to have a safety and effectiveness profile that is similar to the predicate device.
#### CONCLUSIONS VIII.
Agfa's DX-D Imaging Package with XD Detectors has indications for use that is consistent with that of the legally marketed predicate device (K200418) and reference device (K142184). Intended uses are the same. Laboratory tests conclude that the device is substantially equivalent to the predicate. Differences in devices do not alter the intended diagnostic effect nor do they impact the safety and effectiveness of the device.
The new device and the Vivix-S VW predicate device (K200418) and reference device, DX-D Imaging Package - DX-D 40 (K142184) are solid-state X-Ray Imagers, Product Code MQB. Agfa's DX-D Imaging Package with XD Detectors device is substantially equivalent to both the predicate device (K200418) and the reference device (K142184) in that it uses the same basic technology to capture and transmit images.
This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.