← Product Code [LZA](/productcode/LZA) · K211781

# Disposable Nitrile Gloves (K211781)

_Jiangsu Jinke Protective Equipment Co., Ltd. · LZA · Sep 29, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211781

## Device Facts

- **Applicant:** Jiangsu Jinke Protective Equipment Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Sep 29, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The Disposable nitrile gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable nitrile examination gloves; non-sterile; ambidextrous; powder-free. Used by healthcare professionals during patient examinations to provide a barrier against contamination. Device acts as a physical barrier; material properties (nitrile) provide tensile strength and elongation. Manufactured to meet ASTM D6319-19 specifications for dimensions, physical properties, and freedom from holes. Used in clinical settings; operator is the clinician or examiner. Benefits include protection of both patient and examiner from cross-contamination.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D6124 (residual powder), ASTM D5151 (freedom from holes), and ASTM D6319 (physical properties/dimensions). Biocompatibility assessed via ISO 10993-10 (irritation/sensitization), ISO 10993-5 (cytotoxicity), and ISO 10993-11 (systemic toxicity).

## Technological Characteristics

Nitrile rubber material; powder-free; ambidextrous; non-sterile. Dimensions and physical properties comply with ASTM D6319-19. Biocompatibility per ISO 10993-5, 10, and 11. Residual powder per ASTM D6124-06. Freedom from holes per ASTM D5151-19.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color ([K171422](/device/K171422.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

September 29, 2021

Jiangsu Jinke Protective Equipment Co., Ltd % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room608. No.738. Shangcheng Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K211781

Trade/Device Name: Disposable Nitrile Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA Dated: September 7, 2021 Received: September 23, 2021

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Clarence W. Murray III -S

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211781

Device Name Disposable nitrile gloves

Indications for Use (Describe)

The Disposable nitrile gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary K211781

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 submitter's information

Name: Jiangsu Jinke Protective Equipment Co., Ltd Address: West Shuangyang Road(South Side), Sihong Economic Development Zone, Suqian, Jiangsu, 223900, China Phone Number: +86-18058186617 Contact: Mia Lai Date of Preparation: 2021.05.27

#### Designated Submission Correspondent

Mr. Boyle Wanq Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device information

Trade name: Disposable nitrile gloves

Common name: Patient Examination Gloves

Classification name: Non-powdered patient examination glove

Model(s): S, M, L, XL 3.0 Classification

LZA Production code: Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate device information

Manufacturer: Ever Global (Vietnam) Enterprise Corp

{4}------------------------------------------------

Disposable Powder Free Nitrile Examination Glove, White/ Device:

{5}------------------------------------------------

Blue/ Black/ Pink Color 510(k) number: K171422

#### 5.0 Intended use

The Disposable nitrile gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device description

The proposed device is Powder Free Disposable nitrile gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124, ASTM D5151, and ASTM D6319. The proposed device is non-sterile.

#### 7.0 Summarv of Technological Charateristic Comparison

| Item                     | Proposed device                                                                                                                                                                                     | Predicated device                                                                                                                                                                                                                                                | Comparison |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) number            | K211781                                                                                                                                                                                             | K171422                                                                                                                                                                                                                                                          | Same       |
| Product Code             | LZA                                                                                                                                                                                                 | LZA                                                                                                                                                                                                                                                              | Same       |
| Regulation No.           | 21CFR880.6250                                                                                                                                                                                       | 21CFR880.6250                                                                                                                                                                                                                                                    | Same       |
| Class                    | I                                                                                                                                                                                                   | I                                                                                                                                                                                                                                                                | Same       |
| Intended Use             | The Disposable nitrile<br>gloves is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | The Disposable Powder<br>Free Nitrile Examination<br>Glove, White/ Blue/ Black/<br>Pink Color is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | Same       |
| Powdered or Powered free | Powdered free                                                                                                                                                                                       | Powdered free                                                                                                                                                                                                                                                    | Same       |
| Design Feature           | ambidextrous                                                                                                                                                                                        | ambidextrous                                                                                                                                                                                                                                                     | Same       |
| Labeling Information     | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Disposable<br>nitrile gloves, Non-Sterile                                                          | Single-use indication,<br>powder free, device color,<br>device name, glove size<br>and quantity, Disposable<br>Powder Free Nitrile<br>Examination Glove,<br>Non-Sterile                                                                                          | Same       |

## Table1-General Comparison

{6}------------------------------------------------

| Predicate       | Designation    | Size |     |     |     | Tolerance |     |
|-----------------|----------------|------|-----|-----|-----|-----------|-----|
| Device(K171422) |                | XS   | S   | M   | L   | XL        |     |
|                 | Length, mm     | 230  | 230 | 230 | 230 | 230       | min |
|                 | Width, mm      | 75   | 85  | 95  | 105 | 115       | ±5  |
|                 | Thickness, mm: |      |     |     |     |           |     |
|                 | Finger         | 0.05 |     |     |     |           | min |
|                 | Palm           | 0.05 |     |     |     |           | min |
| Proposed Device | Designation    | Size |     |     |     | Tolerance |     |
|                 |                | S    | M   | L   | XL  |           |     |
|                 | Length, mm     | 220  | 230 | 230 | 230 | min       |     |
|                 | Width, mm      | 80   | 95  | 110 | 120 | ±10       |     |
|                 | Thickness, mm: |      |     |     |     |           |     |
|                 | Finger         | 0.05 |     |     |     |           | min |
|                 | Palm           | 0.05 |     |     |     |           | min |
| Remark          | Analysis 1     |      |     |     |     |           |     |

#### Table2 Device Dimensions Comparison

Analysis1: The sizes and tolerances of proposed device are different with those of the predicate, but they all meet the requirements of ASTM D6319-19.

| Item                   |                 |                        | Proposed device                                                                 | Predicated device                                                         | Comparison |
|------------------------|-----------------|------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------|
| Colorant               |                 |                        | blue                                                                            | White/ Blue/ Black/ Pink                                                  | Analysis2  |
| Physical<br>Properties | Before<br>Aging | Tensile<br>Strength    | 14MPa, min                                                                      | 14MPa, min                                                                | SAME       |
|                        |                 | Ultimate<br>Elongation | 500%min                                                                         | 500%min                                                                   | SAME       |
|                        | After<br>Aging  | Tensile<br>Strength    | 14MPa, min                                                                      | 14MPa, min                                                                | SAME       |
|                        |                 | Ultimate<br>Elongation | 400%min                                                                         | 400%min                                                                   | SAME       |
| Comply with ASTM D6319 |                 |                        |                                                                                 | Comply with ASTM D6319                                                    | SAME       |
| Freedom from Holes     |                 |                        | Be free from holes<br>when tested in<br>accordance with<br>ASTMD5151<br>AQL=2.5 | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5 | SAME       |
| Powder Content         |                 |                        | 0.06 mg/glove                                                                   | Meet the requirements of<br>ASTM D6124                                    | SIMILAR    |

#### Table3 Performance Comparison

Analysis 2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility test.

{7}------------------------------------------------

| Item               |                   | Proposed device                                                                                                 | Predicated device                                | Comparison |
|--------------------|-------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------|------------|
| Material           |                   | Nitrile                                                                                                         | Nitrile                                          | SAME       |
| Biocompatibility   | Irritation        | Under the conditions of the study, not an irritant                                                              | Comply with ISO10993-10                          | SAME       |
|                    | Sensitization     | Under conditions of the study, not a sensitizer.                                                                |                                                  |            |
|                    | Cytotoxicity      | Under the conditions of the study, the device is potentially cytotoxic                                          | Comply with ISO10993-5                           | Analysis3  |
|                    | Systemic toxicity | Under the conditions of the study, the device does not elicit a systemic toxicity response in the model animal. | Complies with ISO 10993-11 Third edition 2017-09 |            |
| Label and Labeling |                   | Meet FDA's Requirement                                                                                          | Meet FDA's Requirement                           | SAME       |

#### Table 4 Safety Comparison

Analysis3: The proposed device is potentially cytotoxic, but all proposed devices are conducted the systemic toxicity test, the test results show that the proposed device is safe.

#### 8.0 Discussion of non-clinical testing

In summary, the performance testing and biocompaitbility of the subject device was conducted to demonstrate the subject device met the specification and the acceptance criteria in the standards listed below.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D6319-19, Standard Specification For Disposable nitrile gloves For Medical Application.

ASTM D7160-16, Standard Practice for Determination of Expiration Dating for Medical Gloves

ISO 10993-10:2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ISO 10993-5:2009 Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity

ISO 10993-11 Third edition 2017-09, Biological evaluation of medical devices -Part 11: Tests for systemic toxicity.

{8}------------------------------------------------

| Test<br>Methodology /<br>Standard                                                                                                           | Purpose                                                                                                                                                                                                                   | Acceptance Criteria                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Results                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-<br>10:2010<br>Biological<br>Evaluation Of<br>Medical Devices<br>- Part 10: Tests<br>For Irritation And<br>Skin<br>Sensitization. | This part of ISO<br>10993 assesses<br>possible contact<br>hazards from<br>chemicals released<br>from medical<br>devices,<br>which may produce<br>skin and mucosal<br>irritation, eye irritation<br>or skin sensitization. | Skin Sensitization Test:<br>provided<br>grades less than 1,<br>otherwise sensitization.<br><br>Skin Irritation Test:<br>If the primary irritation<br>index is 0-0,4, the<br>response category is<br>Negligible.<br>0,5-1,9 means slight<br>2-4,9 means moderate<br>5-8 means severe                                                                                                                                                                                                                                                                                 | All grades are 0.<br>All animals were survived<br>and no abnormal signs were<br>observed during the study.<br><br>The primary irritation index<br>is 0.<br>The response of the<br>proposed device was<br>categorized as negligible<br>under the test condition                                                                                                                                                                                                                                                                                                        |
| ISO 10993-<br>5:2009 Biological<br>Evaluation Of<br>Medical Devices<br>- Part 5: Tests<br>For In Vitro<br>Cytotoxicity                      | This part of ISO<br>10993 describes test<br>methods to assess<br>the in vitro<br>cytotoxicity of<br>medical devices.                                                                                                      | The viab.% of the<br>100% extract of the<br>test article is the final<br>result, and if viability is<br>reduced to <70% of<br>the blank, it has<br>cytotoxic potential.                                                                                                                                                                                                                                                                                                                                                                                             | Viab.% of 100% test article<br>extract is 13.7%<br>It means the proposed<br>device have potential<br>toxicity to L-929 in the MTT<br>method                                                                                                                                                                                                                                                                                                                                                                                                                           |
| ISO 10993-11:<br>2017 Biological<br>evaluation of<br>medical devices<br>- Part 11: Tests<br>for systemic<br>toxicity                        | To evaluate the<br>potential for medical<br>device materials to<br>cause adverse<br>systemic reactions.                                                                                                                   | Within the monitoring<br>period (72 h), if the<br>toxicosis response of<br>testing group is not<br>greater than that of<br>control group, the<br>testing sample is<br>regarded as<br>acceptable.                                                                                                                                                                                                                                                                                                                                                                    | There was no evidence of<br>systemic toxicity from the<br>extract.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| ASTM D6124-06<br>(Reapproved<br>2017), Standard<br>Test Method for<br>Residual Powder<br>on Medical<br>Gloves                               | This standard is<br>designed to<br>determine the<br>amount ofresidual<br>powder (or filter-<br>retained mass) found<br>on medical gloves                                                                                  | powder residue limit of<br>2.0 mg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 0.06 mg /glove                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| ASTM D5151-19<br>Standard Test<br>Method for<br>Detection of<br>Holes in Medical<br>Gloves                                                  | This test method<br>covers the detection<br>of holes in<br>medical gloves.                                                                                                                                                | Samples number: 125<br>gloves<br>AQL: 2.5 (ISO 2859)<br>Criterion ≤7 gloves for<br>water leakage                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | no glove water leakage<br>found                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| ASTM D6319-19<br>Standard<br>Specification for<br>Nitrile<br>Examination<br>Gloves for<br>Medical<br>Application                            | This specification<br>covers certain<br>requirements for<br>nitrile rubber gloves<br>used in conducting<br>medical<br>examinations and<br>diagnostic and<br>therapeutic<br>procedures.                                    | Sterility: no need<br>Freedom from holes:<br>pl. Refer to table 3<br><br>Dimensions:<br>S: width 80±10mm<br>Length ≥220 mm<br>M: width 95±10mm<br>Length ≥230 mm<br>L: width 110±10mm<br>Length ≥230 mm<br>XL: width 120±10mm<br>Length ≥230 mm<br><br>Thickness:<br>Finger ≥0.05 mm<br>Palm ≥0.05 mm<br><br>Physical properties:<br>Before aging<br>Tensile strength ≥<br>14MPa<br>Ultimate Elongation ≥<br>500%<br>After Accelerated<br>Aging<br>Tensile strength ≥<br>14MPa<br>Ultimate Elongation ≥<br>400%<br><br>Powder-free Residue:<br>pl. Refer to table 3 | N.A.<br><br>Dimensions:<br>S: width: 84-88 mm<br>Length 251-261 mm<br>M: width 91-95 mm<br>Length 252-256 mm<br>L: width 103-106 mm<br>Length 250-256 mm<br>XL: width 118-121mm<br>Length 253-261 mm<br><br>Thickness:<br>Finger 0.14-0.15 mm<br>Palm 0.08-0.09 mm<br><br>Physical properties:<br>Before aging<br>Tensile strength 14.67-<br>26.48 MPa<br>Ultimate Elongation<br>585.48% - 846.81%<br>After Accelerated Aging<br>Tensile strength 14.40-<br>22.23 MPa<br>Ultimate Elongation<br>527.73% - 775.00%<br><br>Powder-free Residue:<br>pl. Refer to table 3 |

{9}------------------------------------------------

## 9.0 Summary of Clinical Testing:

#### Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.

---

**Source:** [https://fda-staging.innolitics.com/device/K211781](https://fda-staging.innolitics.com/device/K211781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com