ZEN-2090 Turbo, C-Arm mobile is used for providing fluoroscopic and radiographic images of patient anatomy in a hospital or medical clinics.
Device Story
Mobile C-arm X-ray system; provides real-time fluoroscopic and radiographic imaging of patient anatomy. Components include X-ray tube, X-ray tube assembly, X-ray controller, and XTV camera. Operated by clinicians in hospitals or clinics during diagnostic, surgical, and interventional procedures. System captures X-ray signals, processes them via controller, and displays real-time images on monitors. Enables visualization of anatomy without film development; assists physicians in surgical guidance and interventions. Benefits include improved image quality compared to predicate, facilitating precise procedural guidance.
Clinical Evidence
Bench testing only. Compliance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-43, IEC 60601-2-54, IEC 62366, and 21 CFR 1020.30/31/32. Software verification and validation testing performed per FDA guidance.
Technological Characteristics
Mobile C-arm fluoroscopic X-ray system. High-frequency inverter generator; stationary X-ray tube (0.6mm focal spot). Image intensifier (E5830SD-P4A) with 9" active area. Max output 2.0kW. Connectivity includes XTV camera and monitor output. Compliant with IEC 60601 series and 21 CFR 1020 performance standards.
Indications for Use
Indicated for providing fluoroscopic and radiographic imaging of patient anatomy in hospital or medical clinic settings for diagnostic, surgical, and interventional procedures.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
March 9, 2022
GENORA Y Co., Ltd. % Kaitlynn Min Business Development Manager GENORAY America Inc. 147 E. Bristol Lane ORANGE CA 92780
Re: K211780
Trade/Device Name: ZEN-2090 Turbo Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB. JAA. OXO Dated: February 17, 2022 Received: February 18, 2022
Dear Ms. Kaitlynn Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K211780
Device Name ZEN-2090 Turbo
Indications for Use (Describe)
ZEN-2090 Turbo, C-Arm mobile is used for providing fluoroscopic images of patient anatomy in a hospital or medical clinics.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Exhibit 5 510(k) Summary
Date of Summary Preparation: June. 04, 2021
- 1. Submitter and US Official Correspondent
| Submitter: | GENORAY Co., Ltd. |
|----------------|--------------------------------------|
| Address: | 512, 560, Dunchon-daero, Jungwon-gu, |
| | Seongnam-si, Gyeonggi-Do, Korea |
| Telephone No.: | +82-31-5178-5500 |
| Fax: | +82-31-5178-5599 |
Official Correspondent (U.S): Kaitlynn Min - Business Manager
| Correspondent: | GENORAY America Inc. |
|----------------|-------------------------------------------|
| Address: | 147 E. Bristol Lane, Orange, CA 92865 USA |
| Telephone No.: | +1-855-436-6729 |
| Fax: | +1-714-786-8919 |
| Email: | kaitlynn@genorayamerica.com |
# 2. Establishment Registration Number 3005843418
#### Device Information 3.
| Trade/Device Name: | ZEN-2090 Turbo |
|---------------------------|-------------------------------------------------------------------------------------------------------------------|
| Regulation Name: | Fluoroscopic X-ray System |
| Classification Name: | Interventional Fluoroscopic X-Ray System |
| Product Code: | OWB / Interventional Fluoroscopic X-Ray System |
| Subsequence product code: | JAA / System, X-Ray, Fluoroscopic, Image-Intensified<br>OXO / Image-intensified fluoroscopic x-ray system, mobile |
| Device Class: | Class II per regulation 21 CFR 892.1650 |
{4}------------------------------------------------
- Predicate Device (Equivalent Legally Marketed Device) 4.
| Manufacturer: | GENORAY Co., Ltd |
|-------------------------|-------------------------------------------------------------------------------------------------------------------|
| Device Name: | ZEN-2090 Pro |
| 510(k) Number: | K091918 (Decision Date - October 07, 2009) |
| Classification Name: | Interventional Fluoroscopic X-Ray System |
| Primary Product Code: | OWB / Interventional Fluoroscopic X-ray System |
| Secondary product code: | JAA / system, x-ray, fluoroscopic, image-intensified<br>OXO / image-intensified fluoroscopic x-ray system, mobile |
| Device Class: | Class II per regulation 21 CFR 892.1650 |
#### Description of the Device 5.
ZEN-2090 Turbo, C-arm mobile is used for providing fluoroscopic image of patient anatomy, especially during diagnostic, surgical and interventional procedures. The fluoroscopic mode of operation is very useful to the attending physician to see the images on real time without the need to develop individual films.
ZEN-2090 Turbo is consisting of the X-ray tube, X-ray tube assembly, X-ray controller, XTV Camera and some accessories.
### 6. Indications for use
ZEN-2090 Turbo, C-Arm mobile is used for providing fluoroscopic and radiographic images of patient anatomy in a hospital or medical clinics.
{5}------------------------------------------------
## 7. Substantial equivalence chart
| | Proposed device | Predicate device |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | ZEN-2090 Turbo | ZEN-2090 Pro |
| Manufacturer | GENORAY Co., Ltd. | GENORAY Co., Ltd. |
| 510(k) No. | - | K091918 |
| Indications<br>for use | ZEN-2090 Turbo, C-Arm mobile is used for providing<br>fluoroscopic and radiographic images of patient anatomy<br>in a hospital or medical clinics. | ZEN-2090 Pro is a mobile digital C-arm designed to provide<br>fluoroscopic and radiographic images of the patient during<br>diagnostic, surgical and interventional procedures. Examples of<br>clinical application may include cholangiography, endoscopy,<br>orthopedic, neurologic, stone localization, critical care and<br>emergency room procedures i.e. surgical interventions needing<br>X-ray imaging and/or guidance and interventions inside and<br>outside the operating room. |
| Generator | High Frequency Inverter | High Frequency Inverter |
| Max. output<br>power | 2.0kW | 2.2kW |
| X-ray tube | Stationary tube | Stationary tube |
| | 0.6 mm | Large : 1.5 mm / Small : 0.5 mm |
{6}------------------------------------------------
K211780
| Name | Proposed device | Predicate device |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| | ZEN-2090 Turbo | ZEN-2090 Pro |
| Detector<br>(Image<br>Intensifier) | E5830SD-P4A | E5830SD-P4A |
| | • Active image area : 9" (9"/6"/4.5") | • Active image area : 9" (9"/6"/4.5") |
| | • Min. Central Resolution at the monitor:<br>- 9" (23cm): 2.2 lp/mm<br>- 6" (16cm): 2.8 lp/mm<br>- 4.5" (12cm): 3.0 lp/mm | • Min. Central Resolution at the monitor:<br>- 9" (23cm): 2.2 lp/mm<br>- 6" (16cm): 2.8 lp/mm<br>- 4.5" (12cm): 3.0 lp/mm |
| | • DQE: 65% | • DQE: 65% |
| | | |
| Fluoroscopy | 40-110 kV / 0.2-12.0 mA | 40-110 kV/ 0.2-6.0 mA |
| Pulsed<br>Fluoroscopy | 1.0 - 24.0 mA | 0.2-6.0 mA |
| Dimensions | SID : 960 mm | SID : 960 mm |
| | Panning Rotation: ±12.5° | Panning Rotation: ±12.5° |
| | Orbital Rotation: 135° | Orbital Rotation: 120° |
| | Vert. Travel: 500 mm | Vert. Travel: 400mm |
| | Horiz. Travel: 200 mm | Horiz. Travel: 200mm |
{7}------------------------------------------------
### K211780
The proposed ZEN-2090 Turbo is based on the predicate device, ZEN-2090 Pro (K091918).
ZEN-2090 Pro supports both Radiography mode and Fluoroscopy mode, while ZEN-2090 Turbo supports only Fluoroscopy mode.
For Fluoroscopy mode, ZEN-2090 Pro consists of single dose mode, and the maximum X-ray output power is 440W at 110xV, 4mA and the focal size is 0.5 mm with small focus. This is the same for both Continuous Fluoroscopy mode.
ZEN-2090 Turbo can select 3 dose modes (low, normal, high mode), and the focal size is 0.6 mm with single focus. Normal mode of ZEN-2090 Turbo in Continuous Fluoroscopy mode is the same as ZEN-2090 Pro, and high mode is limited to 880W at 110kV, 8mA, it is the twice than normal mode. ZEN-2090 Turbo in Pulsed Fluoroscopy mode is set higher than ZEN-2090 Pro in both normal and high mode.
Among them, high mode is limited to 2000W at 110kV, 18.1mA.
According to the above, ZEN-2090 Turbo is 2 times more than ZEN-2090 Pro in Continuous Fluoroscopy mode, and about 5 times difference in Pulsed Fluoroscopy mode.
Therefore, we have justified that ZEN-2090 Turbo has better image quality than predicate device, ZEN-2090 Pro (K091918).
{8}------------------------------------------------
#### Safety, EMC and Performance data comparison to Predicate 8.
ZEN-2090 Turbo has successfully completed verification and validation testing per GENORAY quality system as well engineering bench testing in support of this submission. The system has been tested and is compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-43, IEC 60601-2-54 and IEC 62366. Also all applicable 21 CFR performance standards (21 CFR 1020.30, 1020.31, 1020.32) are met.
And Software verification testing of the functional requirements as well as performance and safety has been performed to verify that all the requirements of System Requirements Specification as well as the safety risk control measures from the Detailed Risk Management Matrix and the Privacy and Security requirements and mitigations have been implemented. FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical devices", FDA Guidance "Guidance for the content of premarket submissions for management of cyber security" to ensure substantial equivalence. Results demonstrated that all executed verification tests were passed.
Non-clinical validation testing has been performed to validate that ZEN-2090 Turbo conforms to the intended use, claims, user needs, effectiveness of safety measures and instructions for use.
As a results, all test results were satisfactory and the result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate device.
#### 9. Conclusion
In reference to the comparison information provided in substantial equivalence chart, and the most of functions and electronic features are similar with predicate device. We believe that the ZEN-2090 Turbo is safe and effective as predicate device, and has no new indication for use. Therefore, ZEN-2090 Turbo is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.