K211744 · Health & Life Co., Ltd. · FLL · Nov 24, 2021 · General Hospital
Device Facts
Record ID
K211744
Device Name
Non-Contact Infrared Thermometer, model HL710H
Applicant
Health & Life Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Nov 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Non-Contact Infrared Thermometer, model HL710H is intend for the measurement of human body temperature from the forehead. The device is indicated for use by people of all ages in the home.
Device Story
Non-contact infrared thermometer; measures infrared energy radiated from forehead via thermopile sensor; converts energy to temperature value; displays on LCD. Features Bluetooth for transmitting results to paired devices. Includes 'High Temperature Notice' (flashing result, red LED backlight, beep) for temperatures 37.5–43.0°C. Used in home environment by patients or caregivers. Provides temperature readings to assist in monitoring body temperature.
Clinical Evidence
Bench testing only. Compliance with ISO 80601-2-56:2017 and ASTM E1965-98:2016 standards for clinical accuracy of infrared thermometers. Biocompatibility (ISO 10993-1, 5, 10), EMC/Electrical safety (IEC 60601-1, 60601-1-2, 60601-1-11), and software validation (IEC 62304) were performed.
Technological Characteristics
Infrared thermopile sensor; ABS housing/button; 3.0V DC (2x AAA alkaline batteries); LCD display; Bluetooth connectivity; measurement range 22.0–43.0°C; accuracy ±0.2°C (35.0–42.0°C). Standards: ISO 80601-2-56, ASTM E1965-98, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993, IEC 62304.
Indications for Use
Indicated for measurement of human body temperature from the forehead for people of all ages in the home.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Microlife Digital Infrared Ear Thermometer, Model IR1DN1 (IR210) (K183663)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 24, 2021
Health & Life Co., Ltd Melody Jiang Regulatory Affairs 9F, No. 186, Jian Yi Road Zhonghe District New Taipei, 23553 Taiwan
Re: K211744
Trade/Device Name: Non-Contact Infrared Thermometer, model HL710H Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 28, 2021 Received: November 2, 2021
Dear Melody Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K211744
Device Name
Non-Contact Infrared Thermometer, model HL710H
Indications for Use (Describe)
The Non-Contact Infrared Thermometer, model HL710H is intend for the measurement of human body temperature from the forehead. The device is indicated for use by people of all ages in the home.
Type of Use (Select one or both, as applicable)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# PREMARKET NOTIFICATION
# K211744 - 510(k) SUMMARY
#### (As Required By 21 CFR 807.92)
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
NOV 24 2021 Date:
#### 1. Submitter:
Health & Life Co., Ltd. 9F. No.186, Jian Yi Road, Zhonghe District, New Taipei City, Taiwan, R.O.C TEL: +886-2-8227-1300 FAX: +886-2-8227-1301
Contact person: Peggy Su / RA & QA Div. Manager E-mail: peggy.su@hlmt.com.tw Tel: 886-2-8227-1300 ext. 11205 Fax: 886-2-8227-1301
#### 2. Name of the Device:
Trade Name: Non-Contact Infrared Thermometer, model HL 710H Common Name: Clinical electronic thermometer Regulation Name: Clinical electronic thermometer Classification: Class II Regulation: 21 CFR 880.2910 Classification Panel: 880 General Hospital and Personal Use Devices Product Code: FLL
# 3. Information for the 510(k) Cleared Device (Predicate Device): Microlife Non-Contact Infrared Forehead Thermometer, Model: FR1DG1(NC200) (K191829)
1
{4}------------------------------------------------
### 4. Device Description:
The Non-Contact Infrared Forehead Thermometer, Model HL710H is an electronic thermometer using an infrared sensor which can put out different signal when measuring different object temperature or in different ambient temperature to measure infrared energy radiated from the forehead. This energy is collected through the Infrared sensor and converted to a temperature value, then display it by LCD.
The subject device HL710H features a Bluetooth transmission function, which enables the device transmit measured results to paired Bluetooth-enabled device after measurement.
When Bluetooth transmission connection established, the device would transmit measured temperature results with time to the Bluetooth enabled device. Before attempting to sync the thermometer with your Bluetooth enabled device, make sure Bluetooth function is turned ON in both your Bluetooth enabled device and the thermometer.
Additional, HL710H has some feature as below:
#### High Temperature Notice
HL710H has a built-in "High Temperature Notice" feature.
To user who have a high temperature between 37.5 ~ 43.0 ℃ (99.5 °F ~ 109.4 °F), "High Temperature Notice" function will inform them with: Flashing result, red LED backlight, and a series of 'beep' sound.
### 5. Indications for Use
The Non-Contact Infrared Thermometer, model HL710H is intend for the measurement of human body temperature from the forehead. The device is indicated for use by people of all ages in the home.
{5}------------------------------------------------
# 6. Substantial Equivalence Comparison:
Product Specification Comparison Table of Subject Device HL710H, and Predicate Device Microlife Non-Contact Infrared Forehead Thermometer FR1DG1(NC200) (K191829)
| Comparison<br>Items | Subject Device Non-<br>Contact Infrared<br>Thermometer HL710H | Predicate Device<br>Microlife Non-Contact<br>Infrared Forehead<br>Thermometer<br>FR1DG1(NC200)<br>K191829 | Results |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| <b>Indications for use</b> | The Non-Contact<br>Infrared Thermometer,<br>model HL710H is intend<br>for the measurement of<br>human body<br>temperature from the<br>forehead. The device is<br>indicated for use by<br>people of all ages in<br>home. | The Microlife Non-<br>Contact Infrared Forehead<br>Thermometer, Model<br>FR1DG1 (NC200) is<br>intended for the<br>intermittent measurement<br>and monitoring of human<br>body temperature. The<br>device is indicated for use<br>by people of all ages in the<br>home. | same |
| <b>Thermometer type</b> | Infrared thermometer<br>Non-Contact | Infrared thermometer<br>Non-Contact | same |
| <b>Device<br/>Measurement<br/>Technology</b> | Infrared | Infrared | same |
| <b>Temperature<br/>Measurement<br/>distance</b> | Appropriate within 5 cm | Appropriate within 5 cm | same |
| <b>Measuring<br/>Location (human)</b> | Forehead | Forehead | same |
| <b>Mode of<br/>operation(Body<br/>temperature)</b> | Adjusted mode | Adjusted mode | same |
| Appearance | | | Different |
| Physical<br>dimension | 147 mm x 37 mm x 34 mm | 156.7 x 43 x 47 mm | Different |
| Weight | 62.1g±10g (2.44 g ±<br>0.35 oz)<br>(Battery Excluded) | 68.5 g (without batteries) | Similar |
| Power supply | 3.0V DC with 2 AAA<br>1.5V alkaline batteries | 3.0V DC with 2 AAA<br>batteries | same |
| Display<br>resolution | 0.1°C or 0.1°F | 0.1°C or 0.1°F | same |
| Measuring<br>range | 22.0 °C ~ 43.0 °C<br>(71.6 °F~ 109.4 °F) | 32.0-43.0 °C (89.6~109.4 °F) | Different |
| Accuracy | ±0.2 °C: 35.0 ~ 42.0 °C;<br>±0.3 °C: 22.0 ~ 33.9 °C,<br>34.0 ~ 34.9 °C, 42.1 ~ 43.0 °C;<br>±0.4 °F: 95.0 ~ 107.6 °F;<br>±0.5 °F: 71.6~93.0 °F,<br>93.2 ~94.8 °F;<br>107.8~109.4 °F | ±0.2 °C: 35.0 ~ 42.0 °C;<br>±0.3 °C: 34.0 ~ 34.9 °C;<br>42.1 ~ 43.0 °C;<br>±0.4 °F: 95.0 ~ 107.6 °F;<br>±0.5 °F: 93.2 ~94.8 °F:<br>107.8~109.4 °F | Different |
| Operating<br>Conditions<br>(Body) | 10 °C~40 °C(50.0<br>°F~104.0 °F), 15%~95<br>% R.H. | 15~40°C (59°F~104°F),<br>15~95 % R.H. | Different |
| Storage<br>conditions | -20.0 °C ~ 70.0 °C<br>(-4.0 °F ~ 131.0 °F)≤95<br>% R.H. | -25 ~ 55 °C(-13°F ~131°F)<br>15-95 % relative<br>maximum humidity | Different |
| Display type | LCD display | LCD display | same |
| Memory | 12 sets memories | 30 sets memories | Different |
| Measurement time | 1-2 seconds | 3 seconds | Different |
| Error | Error Symbol.<br>Image: Error Symbol<br>System error. Please<br>return the device to local<br>distributor. | Display<br>Image: Er0<br>or<br>Image: Er6<br>when system has a<br>malfunction | Different |
| Patient-Contact<br>Button material | ABS | PMMA | Different |
| Housing and<br>battery cover<br>material | ABS | ABS | Same |
| Probe cover | - | - | none |
| Sensor type | Thermopile | Thermopile | Same |
| reference body<br>site | Oral | Oral | Same |
| Backlight | Green and red backlight<br>according to the<br>measured temperature | Green and red backlight<br>according to the measured<br>temperature | Same |
| Biocompatibility | Cytotoxicity: ISO<br>10993-5:2009<br>Sensitization: ISO<br>10993-10: 2010<br>Irritation or<br>intracutaneous<br>reactivity: ISO 10993-<br>10: 2010 | Change in Material same<br>as K183663 (Microlife<br>Digital Infrared Ear<br>Thermometer, Model IR1<br>DN1 (IR210) ) | Same |
{6}------------------------------------------------
{7}------------------------------------------------
For those differences, we did the relevant evaluations to demonstrate that those differences don't affect the safety and performance of subject device HL 710H. The detail discussions are as follows,
{8}------------------------------------------------
# 1) Materials and Physical Composition
Both the predicate device FR1DG1 (NC200) (K191829) and the subject device HL710H's sensor, microprocessor, LCD are using same materials. The body-contacting part of the subject device HL710H passed biocompatibility evaluations and shows that the subject device complies with standards ISO 10993-1, ISO 10993-5, and ISO 10993-10. Thus, this difference does not raise any new safety or performance questions.
#### 2) Energy Source
Both the predicate device FR1DG1 (NC200) and the subject device HL710H obtain its energy from 2 AAA (1.5V, LR03) Alkaline Batteries. Besides, the subject device HL710H passed EMC and Electrical Safety validations as well as predicate device FR1DG1 (NC200) (K191829). Therefore, there are no issues associated with energy source relevant to this analysis. This has been tested and confirmed according to IEC 60601-1-2: 2014; IEC 60601-1, AAMI/ANSI ES60601-1: 2005/(R2012).
# 3) Appearance and Physical Dimension
The physical dimension of the subject device Non-Contact Infrared Thermometer HL710H is 147 x 37 x 34mm, while predicate device FR1DG1 (NC200) is 156.7 x 43 x 47 mm. The difference is caused because of their different appearance, but these difference does not raise any new safety and effectiveness questions. This has been tested and confirmed according to IEC 60601-1-2: 2014; IEC 60601-1, AAMI/ANSI ES60601-1: 2005/(R2012) and ISO 80601-2-56 Second edition 2017-03.
# 4) Weight
The weight of the subject device Non-Contact Infrared Thermometer HL710H is 62.1g±10g, while predicate device FR1DG1 (NC200) is 68.5 g. The weight is similar around 62-69 g, this is because of their different appearance, people feel same when holding the product. The difference does not raise any new safety and effectiveness questions.
# 5) Measurement range
The subject device HL710H body mode measurement range is 22.0 ℃ ~ 43.0 °C(71.6 °F ~ 109.4 °F), whereas the predicate device FR1DG1 (NC200) measurement range is 32.0-43.0 ℃ (89.6~109.4 °F). Despite minor differences in the
{9}------------------------------------------------
measurement ranges between the subject and predicate device, these are within the ISO 80601-2-56 Second edition 2017-03 and ASTM E1965-98(Reapproved 2016) requirements, and therefore, meet this performance standard. This does not introduce any new risk to the device.
#### 6) Accuracy
The subject device HL710H and the predicate device FR1DG1 (NC200) can both measure human body's temperature, it does not have significant difference of performance. The subject device HL710H extended the temperature as ±0.3 ℃: 22.0 ~ 33.9 ℃; ±0.5 °F: 71.6~93.0 °F, wider than the predicate device. Moreover, HL710H's accuracy is inclusive of the predicate's accuracy; this difference has been verified in compliance with the standard. The difference does not affect the performance and accuracy which was evaluated in the performance testing.
### 7) Operating conditions
The operating temperature range of subject device HL710H body mode is 10 ℃~40 °C(50 °F~104.0 °F), 15%~95 % R.H, whereas he temperature range of predicate device FR1DG1 (NC200) is 15~40℃ (59°F~104°F), 15~95 % R.H. these are within the IEC 60601-1-11: 2015, ISO 80601-2-56 Second edition 2017-03 and ASTM E1965-98 (Reapproved 2016) requirements, and therefore, meet these performance standards. The difference does not affect the performance and accuracy which was evaluated in the performance testing.
## 8) Storage conditions
The storage temperature range of subject device HL710H is -20.0 ℃ ~ 55.0 °C (-4.0 °F~ 131.0 °F) with ≤95 % R.H., whereas storage temperature range of the predicate device FR1DG1 (NC200) is -25 ~ 55 °C (-13°F ~131°F) with 15-95 % relative maximum humidity. Despite minor differences in the storage environment between the subject and predicate device, these are within the IEC 60601-1-11: 2015, ISO 80601-2-56 Second edition 2017-03 and ASTM E1965-98(Reapproved 2016) requirements, and therefore, meet these performance standards. The difference does not affect the performance and accuracy which was evaluated in the performance testing.
#### 9) Memory
{10}------------------------------------------------
The memory of subject device HL710H is 12 sets memories, whereas the memory of predicate device FR1DG1 (NC200) is 30 sets memories.
The "Memory" and "Operating condition" of subject device is similar with predicate device, the software verification and validation test met the requirements. The performance testing shows that the subject device complies with performance standard. Thus, the differences between the predicate device and subject device will not affect the safety and effectiveness of the subject device.
#### 10) Measurement time
The measurement time of subject device HL710H is 1-2 seconds, whereas the measurement time of predicate device FR1DG1 (NC200) is 3 seconds. Both of HL710H and FR1DG1 (NC200) measure temperature in a short time, no more than 3 seconds.
#### 11) Error
Both subject device HL710H and the predicate device FR1DG1 (NC200)(K191829) have a self-test function. The principle of the self-test is the same. Devices can perform a self-test every time when it is switched on to always guarantee the specified accuracy of any measurement, when it has error, it will display signal. HL710H device will display icon "ErC", FR1DG1(NC200) will display icon "ErU" or "Er6". Although the icons are different, but the purpose is the same. It does not affect the device's performance, safety or effectiveness.
#### 12) Patient-Contact Button material
The Patient-Contact Button material of HL710H is ABS (PA757), the Patient-Contact Button material of FR1DG1 (NC200) is PMMA.
The body-contacting part of the subject device HL710H passed biocompatibility evaluations with reports, shows that the subject device complies with standards ISO 10993-1 Fifth edition (2018-08), ISO 10993-5 Third edition (2009-06-01), and ISO 10993-10 Third Edition (2010-08-01). Thus, this difference does not raise any new safety or performance questions.
#### 7. Discussion of Clinical Tests Performed:
The subject device HL710H is compliant to the standard of ISO 80601-2-56:2017 (Particular requirements for basic safety and essential performance of clinical
{11}------------------------------------------------
thermometers for body temperature measurement) and ASTM E1965-98:2016 (Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature) to validate the clinical accuracy of IR thermometer. All the relevant activities were performed by designate individual(s) and the results demonstrated that the predetermined acceptance criteria were fully met.
{12}------------------------------------------------
# 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence:
The subject device was tested to evaluate its safety and effectiveness, including the following:
# EMC/Electrical Safety:
- ANSI/AAMI ES60601-1:2005/(R2012) and A1:2012, C1:2009/(R)2012 and • A2:2010/(R)2012: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012: Amendment 1 -Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-11:2015: Medical Electrical Equipment – Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-1-2: 2014: Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
## Performance:
- ASTM E1965-98 (Reapproved 2016) Standard Specification for Infrared • Thermometers for Intermittent Determination of Patient Temperature.
- . ISO 80601-2-56 Second edition 2017-03 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement [Including: Amendment 1 (2018)]
#### Risk Management:
- ISO 14971: 2007, Medical devices -- Application of risk management to medical devices
#### Software Design and Development:
- . IEC 62304: 2006/A1:2016 Medical device software - Software life cycle
{13}------------------------------------------------
processes [Including Amendment 1 (2016)]
# Biocompatibility:
- ISO 10993-1 Fifth edition (2018-08), Part 1: Evaluation and testing within a risk . management process
- . ISO 10993-5 Third edition (2009-06-01) Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- . ISO 10993-10 Third Edition (2010-08-01) Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
### FCC Test:
- FCC 47 CFR Part 15, Subpart B, C •
### 9. Conclusions:
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The subject device was tested and fulfilled the requirements of the standards mentioned above. The Non-Contact Infrared Thermometer, model HL710H is substantially equivalent to the Microlife Non-Contact Infrared Forehead Thermometer FR1DG1(NC200) cleared under K191829 with respect to the indications for use and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.