BioTeke Sterile Disposable Virus Sampling Kit

K211707 · Bioteke Corporation (Wuxi) Co., Ltd. · JSM · May 4, 2022 · Microbiology

Device Facts

Record IDK211707
Device NameBioTeke Sterile Disposable Virus Sampling Kit
ApplicantBioteke Corporation (Wuxi) Co., Ltd.
Product CodeJSM · Microbiology
Decision DateMay 4, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.2390
Device ClassClass 1

Indications for Use

The BioTeke Sterile Disposable Virus Sampling Kit is intended for the collection and transport of clinical specimens to the laboratory for standard diagnostic / identification techniques. The BioTeke Sterile Disposable Virus Sampling Kit consists of a sterile swab and culture-based media that can be used for respiratory viral diagnostic assays including, Influenza, Respiratory Syncytial Virus (RSV) A and B, Parainfluenza, Adenovirus, and Rhinovirus.

Device Story

Specimen collection and transport kit; includes sterile swab and 10mL polymer tube with 3mL Virus Transport Media (VTM). Swab collects nasal or throat samples; transferred into VTM tube for preservation during transport to laboratory. Swab features notched breakpoint for tube closure. Used by healthcare providers in clinical settings. Output is preserved clinical specimen for subsequent laboratory viral culture or diagnostic assays. Benefits patient by maintaining viral viability during transit, ensuring accurate diagnostic results.

Clinical Evidence

Bench testing only. Virus recovery and viability testing performed per CLSI-M40-A2. Study assessed viability of Influenza, RSV A/B, Parainfluenza, Adenovirus, and Rhinovirus at room temperature (23-25°C) over 120 hours. Results showed log reduction ≤ 0.22 PFU/mL at 48 hours, demonstrating maintenance of viral viability.

Technological Characteristics

Specimen collection kit; flocked polyester swab with ABS shaft (150mm length, 80mm breakpoint). Transport media: 3mL modified Hank's buffered salt solution containing NaCl, KCl, CaCl2, KH2PO4, Glucose, Gentamicin, Polymyxin B, MgSO4, MgCl2, Na2HPO4, Sodium bicarbonate, and Phenol Red. Single-use, sterile, non-propagating transport medium. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for collection and transport of clinical specimens (throat or nasal) for laboratory diagnostic/identification of respiratory viruses including Influenza, RSV A and B, Parainfluenza, Adenovirus, and Rhinovirus.

Regulatory Classification

Identification

A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ## I Background Information: A 510(k) Number K211707 B Applicant BioTeke Corporation (WuXi) Co., Ltd C Proprietary and Established Names BioTeke Sterile Disposable Virus Sampling Kit D Regulatory Information | Product Code(s) | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology | | LIO | Class I, reserved | 21 CFR 866.2900 - Microbiological specimen collection and transport device | MI - Microbiology | ## II Submission/Device Overview: ### A Purpose for Submission: To make a substantial equivalence determination for the BioTeke Sterile Disposable Virus Sampling Kit for collection and transport of clinical respiratory specimens (Influenza, Respiratory Syncytial Virus (RSV) A and B, Parainfluenza, Adenovirus, and Rhinovirus) to the laboratory for standard diagnostic/identification techniques. ### B Type of Test: Transport Culture Medium Device Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993-0002 www.fda.gov {1} K211707 - Page 2 of 6 ## III Intended Use/Indications for Use: ### A Intended Use(s): The BioTeke Sterile Disposable Virus Sampling Kit is intended for collection and transport of clinical specimens to the laboratory for standard diagnostic/identification techniques. The BioTeke Sterile Disposable Virus Sampling Kit consists of a sterile swab and culture-based media that can be used for respiratory viral diagnostic assays including, Influenza, Respiratory Syncytial Virus (RSV) A and B, Parainfluenza, Adenovirus, and Rhinovirus. ### B Indication(s) for Use: NA ### C Special Conditions for Use Statement(s): $\mathrm{R_X}$ - For Prescription Use Only ### D Special Instrument Requirements: None ## IV Device/System Characteristics: ### A Device Description: The BioTeke Sterile Disposable Virus Sampling Swab Kit is a specimen collection and transport kit designed to collect a throat or nasal sample from the patient (utilizing the provided Specimen Collection Swab) and subsequently transferring the swab containing the patient’s microbial sample into the provide 10mL polymer tube containing 3mL of VTM (Virus Transport Media) for purposes of transferring the swab and associated collected microorganisms into a transport container designed to preserve the sample during transport to the laboratory prior to laboratory analysis. The Specimen Collection Swab is designed with a break point (notched area on the shaft) to allow for the top of the VTM tube to close when the inoculated Specimen Collection Swab is placed into the BioTeke VTM tube. The BioTeke Sterile Disposable Virus Sampling Swab Kit consists of the following: - One (1) Specimen Collection Swab (sterile and individually packaged) - One (1) Virus Transport Media (10mL polymer tube containing 3mL of sterile VTM) The Specimen Collection Swab measures 150mm with a breakpoint location (from the tip) at 80mm, a head length at 20mm and a head width of 3mm. The width of the shaft for the Specimen Collection Swab measures 2.5mm. The 10mL polymer tubes are filled with 3mL of Virus Transport Media (VTM) and secured with a leak-proof polymer cap. ### B Principle of Operation: The swab is used to collect nasal or throat samples. ### V Substantial Equivalence Information: {2} A Predicate Device Name(s): Transport Culture Medium Device B Predicate 510(k) Number(s): K113249 C Comparison with Predicate(s): | Table 2. Similarities and Differences Between Bioteke Sterile Disposable Virus Sampling Swab Kit and Predicate Device | | | | --- | --- | --- | | Device & Predicate Device(s): | Device: K211707 | Predicate: K113249 | | | Bioteke Sterile Disposable Virus Sampling Swab Kit | Puritan UTM-RT Collection and Transport System | | General Device Characteristic Similarities | | | | Intended Use | The BioTeke Sterile Disposable Virus Sampling Kit is intended for collection and transport of clinical specimens to the laboratory for standard diagnostic/identification techniques. The BioTeke Sterile Disposable Virus Sampling Kit consists of a sterile swab and culture-based media that can be used for respiratory viral diagnostic assays including, Influenza, Respiratory Syncytial Virus (RSV) A and B, Parainfluenza, Adenovirus, and Rhinovirus. | Puritan UTM -. RT Collection and Transport System is intended for the collection and transport of clinical samples containing viruses, chlamydiae, mycoplasmas and ureaplasmas from the collection site to the testing laboratory. The specimen transported in the Puritan UTM - RT can be used in the laboratory to perform viral, chiamydial, mycoplasmal and ureaplasmal culture. | | Assay Type | Culture Media, Non-propagating Transport | Same. | | Storage Temp (until used) | 5 to 25°C | 2 to 25°C | | General Device Characteristic Differences | | | | Contents of Viral Transport Media | Hank's Balanced Salt Solution, Gentamicin, Polymyxin B, Phenol Red, Sodium Salt | Hank's Balanced Salt Solution L-Glutamic Acid, Bovine Serum Albumin, Phenol Red, Gelatin, Colistin, Sucrose, Amphotericin | K211707 - Page 3 of 6 {3} K211707 - Page 4 of 6 VI Standards/Guidance Documents Referenced: CLSI-M40-A2 Quality Control of Microbiological Transport Systems ISO 10993-1 Biological Evaluation of Medical Devices ISO 11137-32017 Sterilization of health care products Radiation-Part 3. Guidance on dosimetric aspects ISO 11137-1:2018 Sterilization of health care products - Microbiological methods-Part 1: Determination of a population of microorganisms on products ISO 11137-2:2009 Sterilization of health care products Microbiological methods-Part 2 VII Performance Characteristics (if/when applicable): A Analytical Performance: 1. Precision/Reproducibility: Not applicable 2. Linearity: Not applicable 3. Analytical Specificity/Interference: Not applicable 4. Assay Reportable Range: Not applicable 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods): The ability of BioTeke Sterile Disposable Virus Sampling Kit to maintain viability of respiratory viruses was assessed by virus culture after storage at various conditions (time and time). Viruses were diluted in matrix at approximately 3 X the Limit of Detection (LoD) prior to transfer into the VTM. The results demonstrate that the viruses, influenza, RSV A, RSV B, parainfluenza, adenovirus and rhinovirus stored at 23-25°C were maintained within 0.25 log (PFU/mL) of the input concentration of virus for 48 hours as described in the table below. | Virus Viability over Time | | | | | | --- | --- | --- | --- | --- | | Organism | pfu/ml T=0 | pfu/ml T=24 | pfu/ml T=48 | T=48 hours Log reduction | | Influenza A | 85.7 | 64.0 | 54.0 | -0.20 | | RSV A | 105.0 | 85.3 | 63.3 | -0.22 | | RSV B | 106.7 | 84.7 | 64.3 | -0.22 | {4} | Parainfluenza | 167.3 | 141.7 | 104.0 | - 0.21 | | --- | --- | --- | --- | --- | | Adenovirus | 194.7 | 163.0 | 117.0 | - 0.22 | | Rhinovirus | 137.5 | 108.5 | 89 | - 0.21 | RSV=respiratory syncytial virus Additional studies demonstrated that specimens containing the viruses noted in the table above maybe stored under the following conditions: up to 7 days at 2-8°C, or up to 14 days at -70°C. The expiration date of the single-use sterile sampling swab is 12 months when stored at 5-25°C, in the dark.; 6. Detection Limit: Not applicable 7. Assay Cut-Off: Not applicable 8. Accuracy (Instrument): Not applicable 9. Carry-Over: Not applicable ## B Comparison Studies: 1. Method Comparison with Predicate Device: Method comparison is not applicable for a bacterial transport medium. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were done to determine the ability of the Bioteke Sterile Disposable Virus Sampling Swab Kit to maintain viability of viruses commonly found as the etiological agents of respiratory infections. The analytical studies showed acceptable maintenance of respiratory viruses for subsequent viral culture studies. 2. Matrix Comparison: Not applicable ## C Clinical Studies: 1. Clinical Sensitivity: Not applicable 2. Clinical Specificity: K211707 - Page 5 of 6 {5} Not applicable 3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable): Not applicable D. Clinical Cut-Off: Not applicable E. Expected Values/Reference Range: Not applicable F. Other Supportive Performance Characteristics Data: Not applicable VIII. Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. IX. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. K211707 - Page 6 of 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...