← Product Code [FGB](/productcode/FGB) · K211686

# Flexible Video-Choledo-Cysto-Ureteroscope System (K211686)

_Shanghai SeeGen Photoelectric Technology Co., Ltd. · FGB · Feb 22, 2022 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211686

## Device Facts

- **Applicant:** Shanghai SeeGen Photoelectric Technology Co., Ltd.
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Feb 22, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

## Device Story

Flexible Video-Choledo-Cysto-Ureteroscope System consists of a sterile flexible endoscope and a non-sterile imaging processor with LED light source. Endoscope features a distal CMOS sensor, light-guiding section, and mechanical control section for tip deflection. Input: optical images from human cavities (bladder, ureter, biliary/pancreatic ducts). Transformation: CMOS sensor converts light to electrical signals; imaging processor board processes signals for display. Output: real-time video on monitor for physician observation. Used in clinical settings by physicians to guide diagnostic/therapeutic procedures. Benefits: enables minimally invasive visualization and access to internal anatomy for diagnosis and treatment.

## Clinical Evidence

No clinical data included. Substantial equivalence supported by bench testing, including functional performance per ISO 8600-1/3/4, electrical safety, electromagnetic compatibility, biocompatibility, and sterilization validation.

## Technological Characteristics

Flexible endoscope with CMOS distal sensor and LED illumination. Features mechanical wire-rope deflection control. Imaging processor includes image processing board and light source. Connectivity via video interface (NTSC/PAL/DVI). Sterilization: ETO. Compliance: ISO 8600-1/3/4, IEC 60601-1, IEC 60601-2-18, IEC 60601-1-2, ISO 10993.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Flexible Video-Uretero-Choledochoscope System ([K142556](/device/K142556.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym with the full name of the agency on the right. The FDA part is in blue, with the acronym in a solid blue square and the full name written in a smaller font next to it.

February 22, 2022

Shanghai SeeGen Photoelectric Technology Co., Ltd % Nick Wang RA Specialist Shanghai Vanhe Consulting Co., Ltd 2F, Building No.1, 3938 Huqingping Road Shanghai, Shanghai 201713 China

Re: K211686

Trade/Device Name: Flexible Video-Choledo-Cysto-Ureteroscope System Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FGB, FBN, FET, FAJ, FGA Dated: January 17, 2022 Received: January 21, 2022

Dear Nick Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Mark J. Antonino, M.S. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K211686

Device Name

Flexible Video-Choledo-Cysto-Ureteroscope System

Indications for Use (Describe)

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts surgically, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

|  | <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> | <span>                 ☐ Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |
|--|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|--|-----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for SEEGEN. The logo is composed of the word "SEEGEN" in a bold, red font, followed by Chinese characters. Below the Chinese characters is the phrase "vision borderless" in a smaller, light gray font. The logo is simple and modern, and the use of red and gray gives it a professional look.

K211686
Page 1 of 6

# Section 5 510(k) Summary

{4}------------------------------------------------

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

#### 5.1 Submitter

| Submitted by:   | Shanghai SeeGen Photoelectric Technology Co., Ltd         |  |
|-----------------|-----------------------------------------------------------|--|
|                 | Address: 3 Floor, Building No.1, 4299 JinDu Road, Minhang |  |
|                 | District, Shanghai, China                                 |  |
| Contact Person: | Nick Wang                                                 |  |
|                 | RA Specialist                                             |  |
|                 | Shanghai Vanhe Consulting Co., Ltd                        |  |
|                 | Address: 2F, Building No.1, 3938 Huqingping Road, Qingpu  |  |
|                 | District, Shanghai, China.                                |  |
|                 | Phone: 0086-13585860297                                   |  |
|                 | Email: vanheconsulting@126.com                            |  |
| Date Prepared:  | May 27, 2021                                              |  |

# 5.2 Device

| Device Name:       | Flexible Video-Choledo-Cysto-Ureteroscope System                                                                                                                                                                                                |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:       | Ureteroscope And Accessories, Flexible/Rigid;<br>Choledochoscope And Accessories, Flexible/Rigid;<br>Endoscopic Video Imaging System/Component,<br>Gastroenterology-Urology;<br>Cystoscope and Accessories, Flexible/Rigid;<br>Kit, Nephroscope |
| Regulatory Class:  | II                                                                                                                                                                                                                                              |
| Regulation Number: | 21 CFR 876.1500                                                                                                                                                                                                                                 |
| Regulation Name:   | Endoscope and Accessories                                                                                                                                                                                                                       |
| Product Code:      | FGB, FBN, FET, FAJ, FGA                                                                                                                                                                                                                         |

# 5.3 Predicate Device

| Device Name:       | Flexible Video-Uretero-Choledochoscope System, K142556                                            |
|--------------------|---------------------------------------------------------------------------------------------------|
| Common Name:       | Ureteroscope And Accessories, Flexible/Rigid;<br>Choledochoscope And Accessories, Flexible/Rigid; |
| Regulatory Class:  | Class II                                                                                          |
| Regulation Number: | 21 CFR 876.1500                                                                                   |
| Regulation Name:   | Endoscope and Accessories                                                                         |
| Product Code:      | FGB, FBN                                                                                          |

{5}------------------------------------------------

# 5.4 Device Description

Flexible Video-Choledo-Cysto-Ureteroscope System is composed of sterile Flexible Video-Choledo-Cysto-Ureteroscope and an non-sterile Imaging Processor System (Including Light Source).The Flexible Video-Choledo-Cysto-Ureteroscope is composed of Control Section, Light-guiding Section, Connector Section, Insertion Tube and Distal End. The Control Section is pulled by the wire rope to control the bending direction of the Distal End.

The Light-guide section transmits the illumination light from the image processor to the Distal End. The Insertion Tube is used to guide it into other instruments or cavity of the body. The Distal End contains a camera system and a lighting system for illumination and observation.

Imaging Processor System (Including Light Source) is composed of lighting system, image processing board. The lighting system provides the light source for the endoscope probe at the back end. The image processing board receives electronic signals from the front-end camera module and processes them, and finally transmits them to the display through the video interface.

Flexible Video-Choledo-Cysto-Ureteroscope is a kind of medical electronic optical instrument, also known as optical camera, which can enter into the human bladder, ureter, biliary and pancreatic duct for observation and diagnosis. The operator delivers the optical camera system to the site of diagnosis and treatment by means of a mechanical part with a flexible insertion tube and a system of bends.

The product is equipped with tiny size digital imaging parts --photoelectric sensors "CMOS", on which the objects in human cavity will be transferred though lens optical system, and converts light signals into electrical signals. The electrical signal will be transferred to Imaging Processor System (Including Light Source) and display images on it's monitor output for doctor observation and diagnosis.

#### 5.5 Indication for Use:

The Flexible Video-Choledo-Cysto-Ureteroscope System is indicated for endoscopic examination in the urinary tract and can be used percutaneously to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures. The Flexible Video-Choledo-Cysto-Ureteroscope System is also indicated for the examination of bile ducts, and using additional accessories, to perform various diagnostic and therapeutic procedures during cholecystectomy.

| Item                     | Flexible Video-Choledo-Cysto-<br>Ureteroscope System(Proposed<br>Device<br>)                                                                                                                                                                                                                                                                                                                                                                                                    | Flexible Video-Uretero-<br>Choledochoscope System,<br>K142556                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use    | The Flexible Video-Choledo-Cysto-<br>Ureteroscope System is indicated for                                                                                                                                                                                                                                                                                                                                                                                                       | The KARL STORZ Flexible<br>Video-Uretero-                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Item                     | Flexible Video-Choledo-Cysto-<br>Ureteroscope System(Proposed<br>Device<br>)                                                                                                                                                                                                                                                                                                                                                                                                    | Flexible Video-Uretero-<br>Choledochoscope System,<br>K142556                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                          | endoscopic examination in the<br>urinary tract and can be used<br>percutaneously to examine the<br>interior of the kidney, and using<br>additional accessories, to perform<br>various diagnostic and therapeutic<br>procedures. The Flexible Video-<br>Choledo-Cysto-Ureteroscope<br>System is also indicated for the<br>examination of bile ducts, and using<br>additional accessories, to perform<br>various diagnostic and therapeutic<br>procedures during cholecystectomy. | Choledochoscope System is<br>indicated for endoscopic<br>examination in the<br>urinary tract and can be used<br>percutaneously to examine the<br>interior of the kidney, and<br>using additional accessories,<br>to<br>perform various diagnostic<br>and therapeutic procedures.<br>The Flexible Video-Uretero-<br>Choledochoscope System is<br>also<br>indicated for the examination<br>of bile ducts, and using<br>additional accessories, to<br>perform various diagnostic<br>and therapeutic<br>procedures during<br>cholecystectomy |
| Imaging<br>Technology    | CMOS chips at distal end                                                                                                                                                                                                                                                                                                                                                                                                                                                        | CMOS chips at distal end                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Illumination<br>Source   | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Distal<br>End(mm)        | UV-US110-H (OD:2.5mm)<br>UV-US100-H (OD:3.0mm)                                                                                                                                                                                                                                                                                                                                                                                                                                  | 2.9                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Outer<br>Diameter        | UV-100-E (OD:4.7mm)<br>UV-100T-E (OD:5.3mm)<br>UV-110-E (OD:4.7mm)<br>UV-110T-E (OD:5.3mm)<br>CHV-US100-H (OD:3.0mm)<br>CHV-100-E(OD:4.7mm)<br>CHV-100T-E(OD:5.3mm)                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Working<br>Length(mm)    | 380, 600, 750                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | 700                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Deflection(°)<br>of view | UV-US110-H<br>(UP/DOWN:275°/275°)<br>UV-US100-H<br>(UP/DOWN:275°/275°)                                                                                                                                                                                                                                                                                                                                                                                                          | Up:270<br>Down:270                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Item                     | Flexible Video-Choledo-Cysto-<br>Ureteroscope System(Proposed<br>Device<br>)                                                                                                                                                                                                                                                                                                                                                                                                    | Flexible Video-Uretero-<br>Choledochoscope System,<br>K142556                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                          | UV-100T-E<br>(UP/DOWN:210°/130°)<br>UV-110-E (UP/DOWN:210°/130° )<br>UV-110T-E<br>(UP/DOWN:210°/130°)<br>CHV-US100-<br>H(UP/DOWN:120°/120°)<br>CHV-100-E(UP/DOWN:160°/130°)<br>CHV-100T-<br>E(UP/DOWN:160°/130°)                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Field of View            | 110°                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 90°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Sterilization<br>Method  | ETO Sterilization                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Reusable                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

#### 5.6 Substantial Equivalence and Technological Characteristics

{6}------------------------------------------------

{7}------------------------------------------------

K211686

Page 5 of 6

| Device Name                                  | Imaging Processor System (Including Light Source) | Image1 HD CCU(Predicate Device)K142556 | HUB |
|----------------------------------------------|---------------------------------------------------|----------------------------------------|-----|
| Brightness Control                           | Yes                                               | Yes                                    |     |
| Enhancement Control(Contrast and Definition) | Yes                                               | Yes                                    |     |
| Shutter Speed                                | 1/1141-1/13860 sec                                | 1/60-1/17000 sec                       |     |
| White Balance                                | Yes                                               | Yes                                    |     |
| Zoom                                         | Yes                                               | Yes                                    |     |
| Output Formats                               | NTSC/PAL/DVI                                      | NTSC/PAL/VGA/DVI/SDI                   |     |
| Image/Video Capture                          | Yes                                               | Yes                                    |     |
| Video Format Inputs(Cameras)                 | SD and HD                                         | SD and HD                              |     |
| Video Format Outputs                         | SD and HD                                         | SD and HD                              |     |

#### 5.7 Substantial Equivalence

Flexible Video-Uretero-Choledochoscope System, K142556 is used as predicate device compared to proposed device Flexible Video-Choledo-Cysto-Ureteroscope System manufactured by Shanghai SeeGen Photoelectric Technology Co., Ltd.

## 5.8 Non-clinical Performance Data

The Flexible Video-Choledo-Cysto-Ureteroscope System has been successfully tested

{8}------------------------------------------------

for its functions and performance per ISO 8600-1/3/4 and mechanical characteristics. Safety testing war performed including electrical safety IEC 60601-1 and IEC 60601-2-18, electromagnetic compatibility per IEC 60601-1-2 and biocompatibility of the patient contact materials per ISO 10993. Additional validations were conducted for the sterilization process and EO residual.

## 5.8 Clinical Test Data

No Clinical Study is included in this submission.

#### 5.10 Conclusion

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, Based on the information provided in this premarket notification, Shanghai SeeGen Photoelectric Technology Co., Ltd has demonstrated that proposed device Flexible Video-Choledo-Cysto-Ureteroscope System is substantially equivalent to Flexible Video-Uretero-Choledochoscope System, K142556.

---

**Source:** [https://fda-staging.innolitics.com/device/K211686](https://fda-staging.innolitics.com/device/K211686)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).