Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR
K211633 · Canon Medical Systems Corporation · LNH · Jul 22, 2021 · Radiology
Device Facts
Record ID
K211633
Device Name
Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR
Applicant
Canon Medical Systems Corporation
Product Code
LNH · Radiology
Decision Date
Jul 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
MR Image Noise Reduction
Deep learning-based image reconstruction
—
—
—
—
—
—
Indications for Use
Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA. MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are: ·Proton density (PD) (also called hydrogen density) ·Spin-lattice relaxation time (T1) ·Spin-spin relaxation time (T2) ·Flow dynamics ·Chemical Shift Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
Device Story
Vantage Orian 1.5T is a 1.5 Tesla MRI system; utilizes 1.4m short magnet; incorporates Pianissimo/Pianissimo Zen noise reduction. Inputs: proton density, T1/T2 relaxation, flow dynamics, chemical shift signals. Transforms inputs via pulse sequences and AiCE (Advanced Intelligent Clear-IQ Engine) reconstruction; AiCE uses deep learning to remove noise based on g-factor. Outputs: cross-sectional images (transaxial, coronal, sagittal, oblique). Used in clinical settings by physicians/technicians. Healthcare providers interpret images for diagnosis; output aids clinical decision-making by providing anatomical/pathological visualization. Benefits: high-resolution imaging, reduced scan times via acceleration techniques (Compressed SPEEDER, Exsper, Fast 3D), and improved image homogeneity via RX/TX Correction plus.
Clinical Evidence
Bench testing only. Performance verified via phantom imaging to assess image homogeneity (RX/TX Correction Plus) and image quality metrics (distortion, homogeneity, low contrast, SNR) for Exsper acceleration factors up to six. No clinical patient data presented.
Technological Characteristics
1.5T MRI system; 1.4m short magnet; XGO gradient system (45 mT/m, 200 mT/m/ms slew rate). Connectivity: networked. Software: AiCE (Deep Learning-based reconstruction). Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 62304, NEMA MS 1-5. Sterilization: N/A (non-invasive).
Indications for Use
Indicated for diagnostic imaging of head or body anatomic structures in patients requiring cross-sectional transaxial, coronal, sagittal, or oblique MRI, including non-contrast MRA. No specific age or gender contraindications stated beyond standard MRI safety precautions.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/10 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
Canon Medical Systems Corporation % Ms. Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA, INC. 2441 Michelle Drive TUSTIN CA 92780
## Re: K211633
Trade/Device Name: Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: May 25, 2021 Received: May 27, 2021
Dear Ms. Reyes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
July 22, 2021
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
Device Name
Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR
Indications for Use (Describe)
Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces cross-sectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
·Proton density (PD) (also called hydrogen density) ·Spin-lattice relaxation time (T1)
·Spin-spin relaxation time (T2)
·Flow dynamics
·Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff @fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92
## 1. CLASSIFICATION and DEVICE NAME
| Classification Name: | Magnetic Resonance Diagnostic Device |
|-------------------------|---------------------------------------------------------------------------------------|
| Regulation Number: | 90-LNH (Per 21 CFR § 892.1000) |
| Trade Proprietary Name: | Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction<br>Processing Unit for MR |
| Model Number: | MRT-1550 |
#### 2. SUBMITTER'S NAME
Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
## 3. OFFICIAL CORRESPONDENT
Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance Canon Medical Systems Corporation
## 4. CONTACT PERSON, U.S. AGENT and ADDRESS
#### Contact Person
Janine F. Reyes Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 669-7853 Fax: (714) 730-1310 E-mail: jfreyes@us.medical.canon
## Official Correspondent/U.S. Agent
Paul Biggins Senior Director, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-7808 Fax: (714) 730-1310 E-mail: pbiggins@us.medical.canon
{4}------------------------------------------------
## 5. MANUFACTURING SITE
Canon Medical Systems Corporation 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan
- 6. ESTABLISHMENT REGISTRATION 9614698
- 7. DATE PREPARED May 25, 2021
## 8. DEVICE NAME
Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR
## 9. TRADE NAME
Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR
## 10. CLASSIFICATION NAME
Magnetic Resonance Diagnostic Device (MRDD)
## 11. CLASSIFICATION PANEL
Radiology
## 12. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.1000, Magnetic Resonance Diagnostic Device)
## 13. PRODUCT CODE
90-LNH
## 14. PREDICATE DEVICE
Predicate Device (system): Vantage Orian 1.5T, MRT-1550, V6.0 (K202210) Reference Devices: Vantage Orian 1.5T, MRT-1550, V7.0 with AiCE Reconstruction Processing Unit for MR (K203053) and Vantage Orian 1.5T, MRT-1550, V6.0 with AiCE Reconstruction Processing Unit for MR (K202767).
| System | Subject Device | Predicate Device | Reference Device | Reference Device |
|-------------------|---------------------------------------------------------------------------------------------|---------------------------------------|---------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| | Vantage Orian 1.5T,<br>MRT-1550, V7.0 with<br>AiCE Reconstruction<br>Processing Unit for MR | Vantage Orian 1.5T,<br>MRT-1550, V6.0 | Vantage Orian 1.5T,<br>MRT-1550, V7.0 with<br>AiCE Reconstruction<br>Processing Unit for MR | Vantage Orian 1.5T,<br>MRT-1550, V6.0 with<br>AiCE Reconstruction<br>Processing Unit for MR |
| Marketed<br>By | Canon Medical Systems<br>USA, Inc. | Canon Medical Systems<br>USA, Inc. | Canon Medical Systems<br>USA, Inc. | Canon Medical Systems<br>USA, Inc. |
| 510(k)<br>Number | This Submission | K202210 | K203053 | K202767 |
| Clearance<br>Date | | September 22, 2020 | December 2, 2020 | January 15, 2021 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Canon Medical Systems USA, INC. The word "Canon" is in red, and the words "CANON MEDICAL SYSTEMS USA, INC." are in black. The logo is simple and clean, and it is likely used on the company's website and marketing materials. The logo is likely used to promote the company's products and services.
### 15. REASON FOR SUBMISSION
Modification of a cleared device
#### 16. SUBMISSION TYPE
Traditional 510(k) Premarket Notification
#### 17. DEVICE DESCRIPTION
The Vantage Orian (Model MRT-1550) is a 1.5 Tesla Magnetic Resonance Imaging (MRI) System. The Vantage Orian uses 1.4 m short and 4.1 tons light weight magnet. It includes the Canon Pianissimo™ and Pianissimo Zen technology (scan noise reduction technology). The design of the gradient coil and the whole body coil of the Vantage Orian provides the maximum field of view of 55 x 55 x 50 cm. The Model MRT-1550/ UC, UD, UG, UH, UK, UL, UO, UP includes the XGO gradient system.
This system is based upon the technology and materials of previously marketed Canon Medical Systems MRI systems and is intended to acquire and display cross-sectional transaxial, coronal, sagittal, and oblique images of anatomic structures of the head or body. The Vantage Orian MRI System is comparable to the current 1.5T Vantage Orian MRI System (K202210), cleared September 22, 2020 with the following modifications.
#### 18. SUMMARY OF CHANGE(S)
This submission is to report the following changes:
#### Summary of Hardware Changes: (subject modification)
- Gradient amplifier for XGO is applied (K203053). Maximum gradient amplitude (per axis) [mT/m], maximum slew rate [mT/m/ms], and rise time [ms] changes:
- O Increased Gradient Strength: increased gradient strength from 34 [mT/m] to 45 [mT/m].
- O Maximum Slew Rate Change: maximum slew rate of gradient field was changed from 155 [mT/m/ms] to 200 [mT/m/ms].
- O Rise Time Change: rise time changed from 0.220 ms to 0.225 ms.
- Gradient coil minor change: minor change to Zch wire position.
## Summary of Software Changes:
- . FASE 3D/FFE3D:
- o Compressed SPEEDER (Compressed Sensing 3D): This application allows acceleration factors for shorter scan times or higher resolution in FASE3D and FFE3D in addition to FSE2D, imaging based upon the principle of compressed sending in combination with parallel imaging (K203053).
- FE 3D: ●
- o Fat Fraction Quantification: Data is acquired with several different TE and provides PDFF image, R2* image, water image, fat image, in phase image and out of phase image (total 6 kind of images). Proton Density Fat Fraction data and R2* data is supporting fat content ratio (K203053).
{6}------------------------------------------------
- EPI: ●
- ୍ Exsper (Expanded SPEEDER): This application allows reducing scan time for diffusion imaging, Exsper technique scans center of k-space data but surrounding area data is undersampled. It finds coefficient from the data of center k-space and synthesizes the undersampled data by using surrounding data and the coefficient (K203053).
- ୍ Exsper (Expanded SPEEDER): The number of maximum acceleration factor changes to 6 (subject modification).
- FFE 3D:
- o Fast 3D for SSFP: This application allows Fast 3D mode in SSFP 3D sequences, which helps to reduce the scan time by about 50% (K203053).
- 3D Time-of-Flight (TOF): ●
- o Fast 3D for TOF: This application allows reducing the scan time while maintaining image quality by up to half for TOF images by adjusting data acquisition ratio (K203053).
- Short T2* map (multi echo-UTE): UTE imaging is available for acquisition of different TE data for T2* mapping of tissues with short T2* (K203053).
- Advanced DWI-Diffusion time: Various diffusion times are available (K203053).
- AiCE (Advanced Intelligent Clear-IQ Engine):
- o AiCE anatomical region expansion: Head (brain), MSK (shoulder, elbow, wrist/hand, hip, knee, foot/ankle), Spine (cervical, thoracic, lumbar), Pelvis (soft tissue pelvis, female, prostate), Abdomen (liver, renal, pancreas), Breast, Cardiac (K202767).
- o AiCE noise estimation improvement: By taking into consideration g-factor, noise can be removed from the part where g-factor was considered (K203053).
- RX/TX Correction plus: The intensity nonuniformity due to the sensitivity distribution of the coil during transmission and reception can be corrected. The RX/TX Correction plus function can be used for the chest, extremities, and joints, for which the RX/TX Correction function cannot be used (subject modification).
- Double Coverage Interleave: The planned slices are divided into coverages at intervals of N-1 slices, where N is the number of coverages, and then the coverages are excited sequentially. In addition, Interleave acquisition is applied in each coverage (subject modification).
## Summary of Accessory Changes:
- RF Coils: The following RF coils, previously cleared by the OEM (Quality Electrodynamics, LLC) via 510(k) premarket notification K200477, are added:
- Shape Coil, MJAB-207A (subject modification) o
- Shape Coil W, MJAB-217A (subject modification) O
- Patient Pads: surface material change.
- Respiratory Gating System Belt: material change.
{7}------------------------------------------------
#### Summary of labeling Change:
- M-Power: "M-Power" marketing name is removed (subject modification). ●
#### 19. SAFETY PARAMETERS
| Item | Subject Device:<br>Vantage Orian 1.5T,<br>MRT-1550, V7.0 | Predicate Device:<br>Vantage Orian 1.5T,<br>MRT-1550, V6.0 | Notes |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|-------|
| Static field strength | 1.5T | 1.5T | Same |
| Operational Modes | Normal and 1st Operating Mode | Normal and 1st Operating Mode | Same |
| i. Safety parameter display | SAR, dB/dt | SAR, dB/dt | Same |
| ii. Operating mode access<br>requirements | Allows screen access to 1st level<br>operating mode | Allows screen access to 1st level<br>operating mode | Same |
| Maximum SAR | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015) | 4W/kg for whole body (1st<br>operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015) | Same |
| Maximum dB/dt | 1st operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015 | 1st operating mode specified in IEC<br>60601-2-33:<br>2010+A1:2013+A2:2015 | Same |
| Potential emergency<br>condition and means<br>provided for shutdown | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Shutdown by Emergency Ramp<br>Down Unit for collision hazard for<br>ferromagnetic objects | Same |
#### 20. IMAGING PERFORMANCE PARAMETERS
No change from the previous predicate submission, K202210.
#### 21. INDICATIONS FOR USE
Vantage Orian 1.5T systems are indicated for use as a diagnostic imaging modality that produces crosssectional transaxial, coronal, sagittal, and oblique images that display anatomic structures of the head or body. Additionally, this system is capable of non-contrast enhanced imaging, such as MRA.
MRI (magnetic resonance imaging) images correspond to the spatial distribution of protons (hydrogen nuclei) that exhibit nuclear magnetic resonance (NMR). The NMR properties of body tissues and fluids are:
- Proton density (PD) (also called hydrogen density)
- Spin-lattice relaxation time (T1) ●
- Spin-spin relaxation time (T2)
- Flow dynamics
- Chemical Shift
Depending on the region of interest, contrast agents may be used. When interpreted by a trained physician, these images yield information that can be useful in diagnosis.
{8}------------------------------------------------
*Made For life*
#### 22. SUMMARY OF DESIGN CONTROL ACTIVITIES
Risk Management activities for new software functionalities and pulse sequences are included in this submission. The test methods used are the same as those submitted in the previously cleared submission of the predicate device, Vantage Orian 1.5T, MRT-1550, V6.0 (K202210). A declaration of conformity with design controls is included in this submission.
#### 23. SAFETY
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
This device is based upon the same technologies, materials and software as the predicate device. Risk activities were conducted in concurrence with established medical device development standards and guidance. Additionally, testing was done in accordance with applicable recognized consensus standards published by the International Electrotechnical Commission (IEC) for medical devices and the National Electrical Manufacturers Association (NEMA):
#### LIST OF APPLICABLE STANDARDS
- ANSI AAMI ES60601-1:2005/(R)2012 ● and A1:2012
- IEC60601-1-2 (2014)
- IEC60601-1-6 (2010), Amd.1 (2013)
- IEC60601-2-33 (2010), Amd.1 (2013), Amd.2 (2015)
- . IEC60825-1 (2007)
- IEC62304 (2006), Amd.1 (2015) ●
- IEC62366 (2007), Amd.1 (2014)
- NEMA MS 1 (2008)
- NEMA MS 2 (2008)
- NEMA MS 3 (2008)
- NEMA MS 4 (2010)
- NEMA MS 5 (2018)
#### 24. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrate that the system requirements have been met. Additionally, image quality testing was completed which demonstrated that the subject device meets predetermined acceptance criteria.
MR image quality metrics were performed, utilizing phantom images, to assess Rx/Tx Correction Plus with regards to image homogeneity. It was concluded that Rx/Tx Correction Plus increases the homogeneity of the image compared to the image without intensity correction.
MR image quality metrics were performed, utilizing phantom images, to assess Exper (expanded SPEEDER) maximum acceleration factor of up to six with regards to image distortion, homogeneity, low contrast and SNR. It was confirmed that the distortion due to magnetic field inhomogeneity was reduced by increasing the Exsper acceleration factor.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
{9}------------------------------------------------
#### 25. SUBSTANTIAL EQUIVALENCE
Canon Medical Systems Corporation believes that the Vantage Orian 1.5T, MRT-1550, V7.0 Magnetic Resonance Imaging (MRI) System is substantially equivalent to the previously cleared predicate device, Vantage Orian 1.5T, MRT-1550, V6.0, referenced in this submission.
Canon Medical Systems Corporation believes that the changes incorporated into the Vantage Orian 1.5T, MRT-1550, V7.0 are substantially equivalent to the previously cleared predicate device.
#### 26. CONCLUSION
The modifications incorporated into the Vantage Orian 1.5T, MRT-1550, V7.0 do not change the indications for use or the intended use of the device. Based upon bench testing, volunteer clinical imaging, successful completion of software validation and application of risk management and design controls, it is concluded that the subject device is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.