The MIS Precision Chevron Bunion System is indicated for fixing and stabilizing the elective osteotomies of the mid-foot bones and the metatarsal and phalanges of the foot only.
Device Story
Kit for minimally invasive chevron bunion osteotomy and fixation; includes cannulated headless compression screws (titanium Ti-6AL-4V ELI) and single-use instruments (polycarbonate, PEEK, stainless steel). Surgeon uses osteotomy guide, targeting guide, screw guide, drill, and K-wires to perform osteotomy and place screw. Used in clinical/surgical settings by foot and ankle physicians. Output is stabilized bone fixation. Benefits include precise osteotomy placement and secure fixation via minimally invasive approach. Procedure verified via fluoroscopy.
Clinical Evidence
No clinical data. Performance evaluated via 25 simulated surgical procedures on cadaveric tissue (14 male, 11 female feet). Endpoints included successful procedure completion, proper guide positioning, and correct implant placement confirmed via fluoroscopy. Bench testing included biocompatibility (ISO 10993), sterility (ISO 11137), and mechanical testing (ASTM F543).
Technological Characteristics
Cannulated headless compression screws (3.5mm diameter, 24/27mm length) made of Ti-6AL-4V ELI titanium alloy. Single-use instruments made of injection-molded polycarbonate, PEEK, and stainless steel. Mechanical performance tested per ASTM F543. Sterilization via radiation (ISO 11137). Packaging integrity verified per ASTM F1886, F2096, and F88. Accelerated aging per ASTM F1980.
Indications for Use
Indicated for fixation and stabilization of elective osteotomies of mid-foot bones, metatarsals, and phalanges in adult patients requiring bunion correction.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 31, 2022
RELJA Innovations, LLC % Allison Komiyama Principal AcKnowledge Regulatory Strategies, Inc. 2251 San Diego Ave, Suite B-257 San Diego, California 92110
Re: K211628
Trade/Device Name: MIS Precision Chevron Bunion System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: December 30, 2021 Received: December 30, 2021
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211628
Device Name MIS Precision Chevron Bunion System
Indications for Use (Describe)
The MIS Precision Chevron Bunion System is indicated for fixing and stabilizing the elective osteotomies of the mid-foot bones and the metatarsal and phalanges of the foot only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K211628
| 1. Submitted By: | RELJA Innovations LLC<br>17145 W. Bluemound Rd.<br>Suite #J-265<br>Brookfield, WI 53005 |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Date: | January 27, 2022 |
| Contact Person: | David Pancratz<br>Chief Operating Officer<br>Phone: 210-383-3907<br>Email: dave@reljainnovations.com |
| Regulatory Consultant: | Allison Komiyama, Ph.D., RAC<br>Lucie Dalet, Ph.D.<br>RQM+<br>Phone: 412-816-8253<br>Email: akomiyama@rqmplus.com<br>ldalet@rqmplus.com |
| 2. Proprietary Name: | MIS Precision Chevron Bunion System |
| Common Name: | Screw, Fixation, Bone |
| Classification Name and Reference: | 21 CFR 888.3040 – Class II<br>Smooth or threaded metallic bone fixation fastener |
| Device Product Code, Device Panel: | HWC, Orthopedic |
| 3. Primary Predicate: | Stryker Fixos Screw System, K070039 |
### 4. Device Description:
The MIS Precision Chevron Bunion System™ is a kit designed to allow surgeons to perform a chevron bunion osteotomy, and subsequent fixation of the osteotomy, with a minimally invasive surgical technique. The MIS Precision Chevron Bunion System™ consists of a
{4}------------------------------------------------
single, sterile-packaged SKU that contains both implants and instruments needed for the procedure.
The implants in the kit are cannulated headless compression screws, made from titanium (Ti-6AL-4V ELI). There are two implants in the kit, however only one is used in a surgical procedure. Two implants are provided to allow the physician to select the proper size for the patient at the time of surgery. The implants are 3.5 mm diameter, with lengths of 24 mm and 27 mm.
There are also single-use instruments, made of injection molded polycarbonate, PEEK, and stainless steel included in the MIS Precision Chevron Bunion System™. Instruments include an osteotomy guide (one each for left and right feet), a targeting guide, a screw guide, screwdriver, drill bit, and several K-wires and a guidewire.
Additional general surgical instrumentation is utilized, but not part of the RELJA system, such as sagittal saw, drivers, and scalpel.
- 5. Indications for Use
The MIS Precision Chevron Bunion System is indicated for fixing and stabilizing the elective osteotomies of the mid-foot bones and the metatarsal and phalanges of the foot only.
- 6. Technological Characteristics Comparison
The MIS Precision Chevron Bunion System™ utilizes a 3.5 mm diameter cannulated compression screw made of titanium alloy for fixation of a chevron osteotomy of the hallux. Similarly, the Memometal Fixos Screw System recommends a 3.5 mm diameter cannulated compression screw made of titanium alloy for fixation of a chevron osteotomy of the hallux. The Memometal Fixos Screw System also includes numerous other sizes of screws recommended for other surgical procedures.
The MIS Precision Chevron Bunion System™ also includes single-use sterile instrumentation to assist in making the chevron osteotomy and placing the implant. The Memometal Fixos system utilizes metal and/or reusable instruments for performing the same procedure.
- 7. Substantial Equivalence Non-Clinical Evidence
- Biocompatibility: The subject device was subjected to biocompatibility testing in compliance to ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and ISO 10993-10 Biological evaluation of medical devices — Part 10: Tests for irritation and skin sensitization.
{5}------------------------------------------------
- Sterility and shelf life: The final, finished subject device was subjected to sterility testing in compliance with ISO 11137-1 Sterilization of health care products - Radiation - Part 1 : Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose. Bacterial endotoxin testing in accordance with recommendations in the FDA Guidence Document, Pyrogen and Endotoxin Testing: Ouestions and Answers and shelf-life testing confirming five-year shelf-life was also completed on the final finished device. Accelerated aging according to ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems and Medical Devices was completed. Packaging integrity on the aged devices was verified according to ASTM F1886 Standard Test Method for Determining Integrity of Seals for Medical Packaging by Visual Inspection, ASTM F2096 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test), and ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials.
- Performance evaluation: Testing included Static Torsion and Driving Torque per ASTM -F543 Standard Specification and Test Methods for Metallic Medical Bone Screws, and predicted Axial Pullout Strength in accordance with FDA Guidance Document. The results of the Static Torsion testing show that the RELJA screw is stronger than the predicate in torsion. The results of Driving Torque and Axial Pullout Strength analysis show that the RELJA screw meets FDA requirements for non-spinal metallic screws.
- Performance evaluation: Performance and usability testing of the subject device by trained foot and ankle physicians were conducted in twenty-five simulated surgical procedures using cadaveric tissue, on five separate dates. The procedure was successfully completed on male (14 total) and female (11 total) feet ranging in size from small to large. Results show that the precision osteotomy guide is able to be properly positioned on any size foot by following the surgical technique. Fixation and fluoroscopy images showed proper location of the osteotomy and placement of the implant after each simulated procedure.
The safety and effectiveness of the MIS Precision Chevron Bunion System™ is adequately supported by testing, substantial equivalence information, materials information and comparison of design characteristics provided within this premarket submission.
- 8. Substantial Equivalence Clinical Evidence Not applicable.
- 9. Substantial Equivalence Conclusions
The design characteristics of the subject devices do not raise any new types of questions regarding safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate systems and are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.