K211527 · Exo Imaging, Inc. · IYN · Aug 20, 2021 · Radiology
Device Facts
Record ID
K211527
Device Name
Exo Iris
Applicant
Exo Imaging, Inc.
Product Code
IYN · Radiology
Decision Date
Aug 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Exo Iris is indicated for use by qualified and trained healthcare professionals in environments where healthcare is provided to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric patients for the following clinical applications: Peripheral Vessel (including carotid, deep vein thrombosis and arterial studies), Small Organ (including thyroid, scrotum and breast), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional), Musculoskeletal (superficial). Modes of operation include: B-mode, B-mode + Color Doppler.
Device Story
Exo Iris is a handheld, battery-operated diagnostic ultrasound system. It utilizes a single 2D phased array transducer to capture ultrasound signals, which are processed and displayed on a commercial off-the-shelf (COTS) mobile device (iPhone) via a proprietary mobile application. The system supports B-mode and B-mode + Color Doppler imaging. Operators use the app's touchscreen interface to control imaging parameters and view anatomical structures. The device is intended for use by trained healthcare professionals in various clinical settings to assist in diagnostic imaging and measurements. The output provides real-time visual feedback to the clinician, facilitating clinical decision-making for various anatomical assessments. The device is non-sterile and relies on standard ultrasound physics for image generation.
Clinical Evidence
No clinical data were required for this device. Substantial equivalence is supported by bench testing, including electrical safety, electromagnetic compatibility, mechanical safety, and ultrasound acoustic output measurements in compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993, and NEMA UD-2 standards. Software verification and validation, including cybersecurity, were performed.
Technological Characteristics
Handheld 2D phased array ultrasound transducer; battery-operated; COTS mobile device (iPhone) display. Compliant with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasound safety), ISO 10993 (biocompatibility), and NEMA UD-2 (acoustic output). Non-sterile. Software-based control via proprietary mobile application.
Indications for Use
Indicated for adult and pediatric patients for diagnostic ultrasound imaging and measurement of anatomical structures and fluids in peripheral vessel, small organ, cardiac, abdominal, urological, fetal/obstetric, gynecological, and musculoskeletal applications. For use by qualified, trained healthcare professionals in clinical environments.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
August 20, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Exo Imaging, Inc. % Antoanela Gomard Senior Director of Quality and Regulatory Affairs 3600 Bridge Parkway, Suite 102 REDWOOD CITY CA 94065
#### Re: K211527
Trade/Device Name: Exo Iris Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 22, 2021 Received: July 27, 2021
Dear Antoanela Gomard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known)
## K211527
Device Name Exo Iris
#### Indications for Use (Describe)
Expiration Date: 06/30/2023 See PRA Statement below.
Form Approved: OMB No. 0910-0120
Exo Iris is indicated for use by qualified and trained healthcare professionals in environments where healthcare is provided to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric patients for the following clinical applications: Peripheral Vessel (including carotid, deep vein thrombosis and arterial studies), Small Organ (including thyroid, scrotum and breast), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional), Musculoskeletal (superficial). Modes of operation include: B-mode, B-mode + Color Doppler.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Prepared May 13th, 2021
# K211527
| Sponsor: | Exo Imaging Inc.<br>3600 Bridge Parkway<br>Redwood City, CA, 94065 |
|-----------------------|-----------------------------------------------------------------------|
| Contact Person: | Antoanela Gomard<br>Senior Director of Quality and Regulatory Affairs |
| Telephone: | 650 283 0458 |
| Email: | antoanela@exo.inc |
| Submission Date: | May 13th, 2021 |
| Device Name: | Exo Iris |
| Common Name: | Diagnostic Ultrasound System |
| Trade Name: | Exo Iris |
| Classification: | |
| Regulatory Class: | II |
| Review Category: | 21CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 |
| Classification Panel: | Radiology |
Classification Name and Regulation Number Product Code:
| | Regulation Number | Product Code |
|------------------------------------------|-------------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90 - IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90 - IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90 - ITX |
{4}------------------------------------------------
#### A. Legally Marketed Predicate Devices
The predicate device is the Butterfly iQ Ultrasound System, manufactured by Butterfly Network, Inc. (K202406).
## B. Device Description:
Exo Iris is a hand-held, general purpose diagnostic imaging system used to enable visualization of anatomical structures and fluid of adult and pediatric patients. The system is intended to be used by trained healthcare professionals.
The system generates 2D images using a single ultrasound transducer with broad imaging capabilities. The images are displayed on a commercial off-the-shelf mobile device (iPhone) by means of a proprietary mobile application (Exo Iris app) provided by Exo Imaging. Images can be displayed in the following modes: B-Mode, B-Mode + Color Doppler.
The mobile application's user interface includes touchscreen menus, buttons, controls, indicators, and navigation icons that allow the operator to control the system and to view ultrasound images.
#### C. Intended Use / Indications for Use
Exo Iris is indicated for use by qualified and trained healthcare professionals in environments where healthcare is provided to enable diagnostic ultrasound imaging and measurement of anatomical structures and fluids of adult and pediatric for the following clinical applications: Peripheral Vessel (including carotid, deep vein thrombosis and arterial studies), Small Organ (including thyroid, scrotum and breast), Cardiac, Abdominal, Urology, Fetal/Obstetric, Gynecological, Musculoskeletal (conventional), Musculoskeletal (superficial).
Modes of operation include: B-mode, B-mode + Color Doppler.
#### D. Substantial Equivalence
The Exo Iris Ultrasound System is substantially equivalent to the Butterfly iQ Ultrasound System, manufactured by Butterfly Network, Inc. (K202406). They are both portable ultrasound systems for diagnostic use. The indications for use and the core technology of the devices are similar as illustrated in the tables below.
{5}------------------------------------------------
| Device | Subject Device | Predicate Device |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Exo Iris Ultrasound System<br>(this submission) | Butterfly iQ Ultrasound System<br>510(k): K202406 |
| Indications<br>for Use | Exo Iris is indicated for use by qualified<br>and trained healthcare professionals in<br>environments where healthcare is provided<br>to enable diagnostic ultrasound imaging<br>and measurement of anatomical structures<br>and fluids of adult and pediatric patients for<br>the following clinical applications:<br>Peripheral Vessel (including carotid, deep<br>vein thrombosis and arterial studies), Small<br>Organ (including thyroid, scrotum and<br>breast), Cardiac, Abdominal, Urology,<br>Fetal/Obstetric, Gynecological,<br>Musculoskeletal (conventional),<br>Musculoskeletal (superficial).<br>Modes of operation include: B-mode, B-<br>mode + Color Doppler. | The Butterfly iQ Ultrasound System is indicated<br>for use by trained healthcare professionals in<br>environments where healthcare is provided to<br>enable diagnostic ultrasound imaging and<br>measurement of anatomical structures and<br>fluids of adult and pediatric patients for the<br>following clinical applications: Peripheral<br>Vessel (including carotid, deep vein thrombosis<br>and arterial studies), Procedural Guidance,<br>Small Organs (including thyroid, scrotum and<br>breast), Cardiac, Abdominal, Urology,<br>Fetal/Obstetric, Gynecological,<br>Musculoskeletal (conventional),<br>Musculoskeletal (superficial) and Ophthalmic.<br>Modes of operation include B-mode, B-mode +<br>M-mode, B-mode + Color Doppler, B-mode +<br>Power Doppler. |
| Manufacturer | Exo Imaging Inc. | Butterfly Network, Inc. |
| 510(k)<br>number | TBD | K202406 |
| Regulation<br>medical<br>specialty | Radiology | Radiology |
| Product code | IYN, IYO. ITX | IYN, IYO. ITX |
| Regulation<br>number | 21CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 | 21CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 |
| Regulation<br>description | Diagnostic Ultrasound System | Diagnostic Ultrasound System |
| Classification | II | II |
| Intended<br>Users | Trained healthcare professionals | Trained healthcare professionals |
| 510(k) Track | Track 3 | Track 3 |
| Imaging<br>Modes | B-mode (Anatomy) / B-mode +Color<br>Doppler (Flow) | B-mode, B-mode + M-mode, B-mode + Color<br>Doppler, B-mode + Power Doppler. |
# Table 1 Comparison of Indications for Use
{6}------------------------------------------------
| Parameters | Subject Device | Predicate Device (K202406) |
|-----------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|
| Trade Name | Exo Iris | Butterfly iQ Ultrasound System |
| Manufacturer | Exo Imaging, Inc. | Butterfly Network, Inc. |
| General Device Description | | |
| Device type | Handheld portable diagnostic<br>ultrasound system | Handheld portable diagnostic<br>ultrasound system |
| Transducer Characteristics | | |
| Array Type | Single probe 2D phased array | Single probe 2D phased array |
| Other Relevant Similarities | | |
| Source of Energy | Battery-operated | Battery-operated |
| Electrical Safety | Yes, compliant with applicable<br>electrical safety standards | Yes, compliant with applicable electrical<br>safety standards |
| Mechanical Safety | Meets mechanical safety standards for<br>a class II medical device | Meets mechanical safety standards for<br>a class II medical device |
| Biocompatibility | Yes, compliant with ISO 10993 | Yes, compliant with ISO 10993 |
| Sterility | Non-sterile | Non-sterile |
| Display | COTS Device Display (iPhone) | COTS device display |
#### Table 2. Substantial Equivalence Comparison for Technological Characteristics
Based on the comparison of indications for use and technological characteristics, the subject device is substantially equivalent to the predicate device.
#### E. Performance Data
All specifications for Exo Iris have been verified and validated as required by the risk analysis. All design verification and validation activities were performed by the designated individual(s) according to the company's Design Control Process and the results demonstrated that the predetermined acceptance criteria were met.
The verification and validation testing included testing to the following applicable standards:
- IEC 60601-1: Edition 3.0 2005 (3.0 2005 + CORR. 1:2006 + CORR. 2:2007 + -
A1:2012). Medical Electrical Equipment - Part 1: General Requirements for Safety.
- IEC 60601-1-2: Edition 4.0 2014, Medical Electrical Equipment Part 1-2: -General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
{7}------------------------------------------------
- IEC 60601-2-37: Edition 2.0 Am 1 2015, Medical Electrical Equipment Part 2--37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment.
- -ISO 10993:2009 Biological Evaluation of Medical Devices. Part 1
- -NEMA UD-2: 2004 Rev 3, Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment.
Verification and validation testing were completed in accordance with the company's Design Control process in compliance with 21 CFR Part 820.30. Exo Imaging certifies that all verification and validation activities provided in this submission were performed by designated individuals and results demonstrate that predetermined acceptance criteria were met. Successful results for the following tests were included in the submission as performance data supporting substantial equivalence:
- 1. Bench testing for electrical and mechanical safety in compliance with the standards cited above
- 2. Bench testing for ultrasound in compliance with the standards cited above and applicable Guidance published by FDA.
- 3. Software testing, consisted of verification and validation testing including test cases related to off the shelf software, as well as cybersecurity features.
Clinical data were not required for this type of device.
# F. Conclusion
Potential risks were identified according to the ISO 14971. The risks were analyzed with regard to risk/benefit category and mitigations were implemented and tested as part of the performance testing described above. All risk mitigations were satisfactorily verified and validated. Where there were technological differences from the predicate, these were shown not to result in any new issues of safety or efficacy according to the performance data submitted.
Therefore, the Exo Iris Ultrasound System is substantially equivalent to the predicate device with regards to intended use and technological characteristics. Results of performance testing demonstrated that the device met the design requirements.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.