K211512 · Smith & Nephew, Inc. · HSX · Sep 16, 2021 · Orthopedic
Device Facts
Record ID
K211512
Device Name
VISIONAIRE UK Patient Matched Cutting Guides
Applicant
Smith & Nephew, Inc.
Product Code
HSX · Orthopedic
Decision Date
Sep 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Indications for Use
Smith & Nephew's VISIONAIRE UK Patient Matched Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use: - JOURNEY II Unicompartmental Knee (JOURNEY II UK) System The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for single use only.
Device Story
Patient-specific surgical instrumentation; assists in positioning knee replacement components and marking bone intra-operatively. Input: patient imaging data (MRI, X-Ray). Process: additive manufacturing (Nylon 12) creates guides based on reconstructed bony/cartilaginous anatomy; guides provide mechanical alignment via intimate fit with proximal tibia or distal femur. Output: physical cutting guides used by orthopedic surgeons in clinical settings. Benefit: facilitates accurate implant positioning during unicompartmental knee replacement procedures.
Clinical Evidence
Bench testing only; cadaveric testing performed to demonstrate performance equivalence to predicate devices. No clinical data required.
Technological Characteristics
Material: Nylon 12. Manufacturing: Additive manufacturing. Design: Patient-matched via MRI/X-ray reconstruction. Form factor: Patient-specific cutting guides/blocks. Connectivity: N/A. Sterilization: Not specified. Software: Non-medical device software used for design/manufacturing; no changes to previously cleared VISIONAIRE software.
Indications for Use
Indicated for patients undergoing knee replacement surgery requiring patient-specific surgical instrumentation for positioning components and marking bone, provided anatomic landmarks are identifiable on imaging scans. Intended for use with the JOURNEY II Unicompartmental Knee (UK) System.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
October 29, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smith & Nephew, Inc. Leah Andre Regulatory Affairs Specialist II 7135 Goodlett Farms Parkway Cordova. Tennessee 38016
Re: K211512
Trade/Device Name: VISIONAIRE UK Patient Matched Cutting Guides Regulation Number: 21 CFR 888.3520 Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSX, OOG, MBH, JWH
Dear Leah Andre:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated September 16, 2021. Specifically, FDA is updating this SE Letter as an administrative correction. The 510(k) summary that was included in the SE package included the VISIONAIRE lateral cutting guides; however, the VISIONAIRE lateral cutting guides have been removed from the submission. The updated 510(k) summary does not include the lateral cutting guides and will be included in the SE Package.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ting Song, Ph.D., R.A.C., OHT6: Office of Orthopedic Devices, (301) 769-7677.
Sincerely,
Ting Song-S
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
September 16, 2021
Image /page/1/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Smith & Nephew, Inc. Leah Hawkins Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova. Tennessee 38016
## Re: K211512
Trade/Device Name: VISIONAIRE UK Patient Matched Cutting Guides Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX, OOG, MBH, JWH Dated: August 19, 2021 Received: August 20, 2021
#### Dear Leah Hawkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
#### K211512
Device Name VISIONAIRE UK Patient Matched Cutting Guides
#### Indications for Use (Describe)
Smith & Nephew's VISIONAIRE UK Patient Matched Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of knee replacement components intra-operatively and in guiding the marking. of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for use with the following existing Smith & Nephew, Inc. Knee Systems and their cleared indications for use:
- JOURNEY II Unicompartmental Knee (JOURNEY II UK) System
The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for single use only.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style=""> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| 510(k) Summary Submitted by: | Smith & Nephew, Inc.<br>Orthopaedic Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Date of Submission: | May 13, 2021 |
| Primary Contact Person: | Leah Hawkins, Regulatory Affairs Specialist II |
| | T (901) 800-3355 |
| | M (901) 463-8447 |
| Secondary Contact Person: | Michelle Huettner, Director, Regulatory Affairs |
| | T (901) 800-3241 |
| | M (765) 426-6070 |
| Name of Device: | VISIONAIRE UK Patient Matched Cutting Guides |
| Common Name: | Knee Prosthesis |
| Device Classification Name and<br>Reference: | 21 CFR 888.3520 - Knee Joint femorotibial<br>metal/polymer non-constrained cemented<br>prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopaedics/87 |
| Product Code: | HSX - Knee joint femorotibial metal/polymer non-<br>constrained cemented prosthesis |
| | OOG - Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained<br>cemented prosthesis. |
| | MBH - Knee joint patellofemorotibial<br>metal/polymer porous-coated uncemented<br>prosthesis. |
| | |
{5}------------------------------------------------
JWH - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
#### Predicate Device:
Primary Predicate: Materialise PKA Guide System - K173970 (S.E. 7/6/2018) HSX, OOG
Secondary Predicate: VISIONAIRE Patient Matched Cutting Blocks - K200826 (S.E. 4/24/2020) JWH, MBH, OOG
Reference Device: VISIONAIRE Adaptive Guides - K170282 (S.E. 5/22/2017) JWH, MBH, OOG
{6}------------------------------------------------
#### Device Description
The purpose of this Traditional 510(k) is to notify the FDA of our intent to market the Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides with the VISIONAIRE Patient Matched Technology. The VISIONAIRE Unicompartmental Knee (UK) Patient Matched Cutting Guides are to be used as patient-specific surgical instrumentation to assist in the positioning of partial knee replacement implant components intra-operatively and in guiding the marking of bone before cutting. The subject device is designed and manufactured from patient imaging data (i.e. MRI, X-Ray) using additive manufacturing of Nylon 12 material to create the patientmatched guides. The blocks achieve mechanical alignment via an intimate fit with the patient's proximal tibia or distal femur. This fit is achieved by reconstructing the patient's bony and cartilaginous anatomy from the MRI scans of the patient's knee. The VISIONAIRE UK Patient Matched Cutting Guides are intended to be used for medial femoral and tibia resections in conjunction with previously cleared Smith & Nephew JOURNEY II Unicompartmental Knee (JOURNEY II UK) System (K190085). Similar unicompartmental knee patient-specific surgical instrumentation used with JOURNEY II UK System has been previously cleared as Materialise PKA Guide System (Primary Predicate - K173970).
The VISIONAIRE UK Patient Matched Cutting Guides utilizes VISIONAIRE Patient Matched Technology to design and manufacture the patient-matched guides. As a result of this submission, the subject device will use the VISIONAIRE Patient Matched Technology process in previously cleared VISIONAIRE Patient Matched Cutting Blocks (Secondary Predicate -K200826). The only modification to the VISIONAIRE Patient Matched Technology process is the addition of VISIONAIRE UK implant options.
The VISIONAIRE UK Patient Matched Cutting Guides are a line extension to the currently marketed VISIONAIRE Adaptive Guides (Reference Device – K170282) and will utilize similar manufacturing and technological processes.
{7}------------------------------------------------
## Indications for Use
Smith & Nephew's VISIONAIRE UK Patient Matched Cutting Guides are intended to be used as patient-specific surgical instrumentation to assist in the positioning of knee replacement components intra-operatively and in guiding the marking of bone before cutting provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.
The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for use with the following existing Smith & Nephew. Inc. Knee Systems and their cleared indications for use:
- JOURNEY II Unicompartmental Knee (JOURNEY II UK) System (K190085)
The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are intended for single use only.
#### Technological Characteristics
Smith & Nephew conducted cadaveric testing of the VISIONAIRE UK Patient Matched Cutting Guides. The results of this testing demonstrated that the patient matched cutting blocks/guides designed using the case processing applications perform equivalent to the predicate devices listed in the Table 5.1 below.
The associated software in order to manufacture the subject device demonstrate that there are no new issues related to the safety and effectiveness of the predicate device and the software will perform as intended. All changes are made to the non-medical device software as part of the subject device submission. There are no changes required to the previously cleared VISIONAIRE Patient Matched Cutting Blocks software (Secondary Predicate - K200826) that are involved in creating the VISIONAIRE UK Patient Matched Cutting Guides. Additional details related to the software are provided in Section 16.
{8}------------------------------------------------
Based on the testing within this premarket notification, there are no new issues related to the safety and effectiveness of the subject device. Clinical data was not needed to support the safety and effectiveness of the subject device.
## Substantial Equivalence Information
The Smith & Nephew VISIONAIRE UK Patient Matched Cutting Guides are substantially equivalent in function, equivalent design features, intended use, indications for use, operational principles, manufacturing processes, and materials as the commercially available predicate devices listed below in Table 5.1.
#### Table 5.1: Predicate Devices
| Predicate Type | Manufacturer | Description | Submission<br>Number | Clearance<br>Date |
|---------------------|----------------------|-------------------------------------------|----------------------|-------------------|
| Primary Predicate | Materialise N.V. | Materialise PKA Guide System | K173970 | 07/06/2018 |
| Secondary Predicate | Smith & Nephew, Inc. | VISIONAIRE Patient Matched Cutting Blocks | K200826 | 4/24/2020 |
| Reference Device | Smith & Nephew, Inc. | VISIONAIRE Adaptive Guides | K170282 | 05/22/2017 |
## Conclusion
As previously noted. this 510(k) Premarket Notification is being submitted to request clearance for the VISIONAIRE UK Patient Matched Cutting Guides. Based on the similarities to the predicate cutting blocks and guide system, cadaveric testing, and a review of the software validation testing performed, the subject device is substantially equivalent to the commercially available predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.