← Product Code [FXX](/productcode/FXX) · K211497

# Surgical Mask (K211497)

_Honeywell Safety Products · FXX · Jul 28, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211497

## Device Facts

- **Applicant:** Honeywell Safety Products
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Jul 28, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Honeywell Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

## Device Story

Honeywell Procedure Mask is a single-use, disposable, non-sterile, flat-pleated face mask. It consists of three layers: inner and outer spun-bond polypropylene layers and a middle melt-blown polypropylene layer. It features a polyethylene-coated steel nose wire and round knit polyester/Lycra earloops. The device is used in clinical settings by healthcare personnel and patients to provide a physical barrier against microorganisms, body fluids, and particulate matter. It functions as a passive protective barrier; it does not involve electronic components, software, or active sensing. The mask is intended to reduce exposure to blood and body fluids during infection control practices. It is classified as an ASTM Level 2 surgical mask.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing included ASTM F1862 (fluid resistance), ASTM F2299 (PFE ≥99.45%), ASTM F2101 (BFE ≥98.2%), EN 14683 (differential pressure ≤4.8 mmH2O/cm²), and 16 CFR 1610 (flammability Class 1). Biocompatibility testing per ISO 10993-1, -5, and -10 confirmed the device is non-cytotoxic, non-irritating, and non-sensitizing.

## Technological Characteristics

Three-layer construction: spun-bond polypropylene (inner/outer), melt-blown polypropylene (middle). Features polyethylene-coated steel nose wire and round knit polyester/Lycra earloops. Dimensions: 173mm x 95mm. ASTM Level 2 performance. Non-sterile, single-use. No software, electronics, or energy source.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Nordiwell Medical Face Mask ([K210445](/device/K210445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.

July 28, 2021

Honeywell Safety Products Ivy Grieco Senior Advanced Quality Engineer 10 Thurber Boulevard Smithfield, Rhode Island 02917

Re: K211497

Trade/Device Name: Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: May 11, 2021 Received: May 14, 2021

Dear Ivy Grieco:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K211497

Device Name Honeywell Procedure Mask

Indications for Use (Describe)

The Honeywell Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device, provided nonsterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary (K211497)

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

## I. SPONSOR

Honeywell Safety & Productivity Solutions 10 Thurber Blvd. Smithfield, RI 02917 USA

#### II. CONTACT PERSON

Ivy Grieco Title: Sr. Adv. Quality Engineer Phone: 401-864-0496 Email: Ivy.Grieco@Honeywell.com Date Prepared: July 21, 2021

#### III. DEVICE

Name of Device: Honeywell Procedure Mask Common or Usual Name: Procedure Mask Classification Name: Surgical Apparel (21 CFR §878.4040) Requlatory Class: II Product Code: FXX

#### IV. PREDICATE DEVICE

Manufacturer: Changzhou Combat Protective Equipment Co., Ltd. Device: Nordiwell Medical Face Mask 510(k) Number: K210445 This predicate has not been subject to a design-related recall.

#### V. DEVICE DESCRIPTION

The Honeywell Procedure Mask is composed of three layers that are a flat, pleated style mask with earloops to secure it over the users' mouth and face. The inner and outer layers are manufactured from spun-bond polypropylene. The middle layer is made of melt-blown polypropylene. The mask is a single use, disposable device, provided nonsterile and is not made from natural rubber latex.

# VI. INDICATIONS FOR USE

The Honeywell Procedure Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids, and particulate material. These face masks are intended for use in infection control practices to reduce potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

{4}------------------------------------------------

# VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The Honeywell Procedure Mask is compared with the predicate device Nordiwell Face Mask (K210445)). The results are shown below in the Technological Characteristics Comparison Table:

| Item                                          | Subject Device<br>Honeywell Procedure Mask                                                                                                                                                                                                                                                                                                                                                                 | Predicate Device<br>Nordiwell Medical Face Mask                                                                                                                                                                                                                                                                                                                                                              | Comparison |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) Number                                 | K211497                                                                                                                                                                                                                                                                                                                                                                                                    | K210445                                                                                                                                                                                                                                                                                                                                                                                                      |            |
| Intended Use<br>Statement                     | The Honeywell Procedure Mask is<br>intended to be worn to protect both the<br>patient and healthcare personnel from<br>transfer of microorganisms, body fluids,<br>and particulate material.<br><br>These face masks are intended for use<br>in infection control practices to reduce<br>potential exposure to blood and body<br>fluids. This is a single use, disposable<br>device, provided non-sterile. | The Nordiwell Medical Face Mask is<br>intended to be worn to protect both<br>patient and healthcare personnel from<br>transfer of microorganisms, body<br>fluids, and particulate matters.<br><br>These face masks are intended for use<br>in infection control practices to reduce<br>potential exposure to blood and body<br>fluids. This is a single use, disposable<br>device(s) and provided nonsterile | Same       |
| Materials                                     |                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Outer Layer                                   | Polypropylene Spun-bond                                                                                                                                                                                                                                                                                                                                                                                    | Polypropylene Spun-bond                                                                                                                                                                                                                                                                                                                                                                                      | same       |
| Middle Layer                                  | Polypropylene Melt-blown                                                                                                                                                                                                                                                                                                                                                                                   | Polypropylene Melt-blown                                                                                                                                                                                                                                                                                                                                                                                     | same       |
| Inner Layer                                   | Polypropylene Spun-bond                                                                                                                                                                                                                                                                                                                                                                                    | Polypropylene Spun-bond                                                                                                                                                                                                                                                                                                                                                                                      | same       |
| Earloop                                       | Round Knit Polyester and Lycra                                                                                                                                                                                                                                                                                                                                                                             | Spandex and Nylon                                                                                                                                                                                                                                                                                                                                                                                            | different  |
| Nose Wire                                     | Polyethylene Coated Steel Wire                                                                                                                                                                                                                                                                                                                                                                             | Polyethylene Coated Steel Wire                                                                                                                                                                                                                                                                                                                                                                               | same       |
| Design Features                               |                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Color                                         | White                                                                                                                                                                                                                                                                                                                                                                                                      | White                                                                                                                                                                                                                                                                                                                                                                                                        | same       |
| Style                                         | Flat, Pleated                                                                                                                                                                                                                                                                                                                                                                                              | Flat, Pleated                                                                                                                                                                                                                                                                                                                                                                                                | same       |
| Layers                                        | 3                                                                                                                                                                                                                                                                                                                                                                                                          | 3                                                                                                                                                                                                                                                                                                                                                                                                            | same       |
| Physical State                                | Single Use Only                                                                                                                                                                                                                                                                                                                                                                                            | Single Use Only                                                                                                                                                                                                                                                                                                                                                                                              | same       |
| Sterility                                     | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                | Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                  | same       |
| Dimensions                                    |                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Length                                        | 173mm ± 3mm                                                                                                                                                                                                                                                                                                                                                                                                | 160mm ± 5mm                                                                                                                                                                                                                                                                                                                                                                                                  | different  |
| Width                                         | 95mm ± 3mm                                                                                                                                                                                                                                                                                                                                                                                                 | 106mm ± 5mm                                                                                                                                                                                                                                                                                                                                                                                                  | different  |
| Performance Specifications & Testing          |                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| ASTM Level                                    | 2                                                                                                                                                                                                                                                                                                                                                                                                          | 2                                                                                                                                                                                                                                                                                                                                                                                                            | same       |
| Fluid Resistance<br>(ASTM F1862)              | Passed at 120 mmHg                                                                                                                                                                                                                                                                                                                                                                                         | Passed at 120 mmHg                                                                                                                                                                                                                                                                                                                                                                                           | same       |
| Particulate<br>Filtration<br>Efficiency (PFE) | Passed at ≥98% @ 0.1 micron                                                                                                                                                                                                                                                                                                                                                                                | Passed at ≥98% @ 0.1 micron                                                                                                                                                                                                                                                                                                                                                                                  | same       |
| Bacterial Filtration<br>Efficiency (BFE)      | Passed at ≥98%                                                                                                                                                                                                                                                                                                                                                                                             | Passed at ≥98%                                                                                                                                                                                                                                                                                                                                                                                               | same       |
| Differential<br>Pressure                      | Passed at <5.0 mmH $2$ O/cm $2$                                                                                                                                                                                                                                                                                                                                                                            | Passed at <5.0 mmH $2$ O/cm $2$                                                                                                                                                                                                                                                                                                                                                                              | same       |
| Flammability<br>(16 CFR 1610)                 | Class 1                                                                                                                                                                                                                                                                                                                                                                                                    | Class 1                                                                                                                                                                                                                                                                                                                                                                                                      | same       |
| Biocompatibility                              |                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| Cytotoxicity                                  | Non-Cytotoxic                                                                                                                                                                                                                                                                                                                                                                                              | Non-Cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                | same       |
| Skin Irritation                               | Non-Irritating                                                                                                                                                                                                                                                                                                                                                                                             | Non-Irritating                                                                                                                                                                                                                                                                                                                                                                                               | same       |
| Dermal<br>Sensitization                       | Non-Sensitizing                                                                                                                                                                                                                                                                                                                                                                                            | Non-Sensitizing                                                                                                                                                                                                                                                                                                                                                                                              | same       |

{5}------------------------------------------------

#### VIII. SUMMARY OF NON-CLINICAL TESTING

#### Performance Testing

Mask performance was tested to ASTM's F2100 standard. Each type of performance testing used 32 samples in each of 3 different, non-consecutive lots. Testing consisted of the following:

| Test Methodology | Purpose                                         | Acceptance Criteria                        | Results                 |
|------------------|-------------------------------------------------|--------------------------------------------|-------------------------|
| ASTM F1862       | Fluid Resistance/Synthetic Blood<br>Penetration | 120 mmHg                                   | 120 mmHg                |
| ASTM F2299       | Particulate Filtration Efficiency               | ≥98%                                       | ≥99.45% @ 0.1<br>micron |
| ASTM F2101       | Bacterial Filtration Efficiency                 | ≥98%                                       | ≥98.2% @ 3.0<br>micron  |
| EN 14683 Annex C | Differential Pressure<br>"Breathability"        | <6.0 mmH2O/cm²                             | ≤4.8 mmH2O/cm²          |
| 16 CFR Part 1610 | Flammability                                    | IBE or ≥3<br>seconds burn<br>time, Class 1 | IBE, Class 1            |

# Biocompatibility testing

The biocompatibility evaluation for the Honeywell Procedure Mask device was conducted in accordance with the FDA's Guidance "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" September 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:

| Test Methodology | Purpose              | Acceptance Criteria | Results                                                              |
|------------------|----------------------|---------------------|----------------------------------------------------------------------|
| ISO 10993-5      | Cytotoxicity         | Non-Cytotoxic       | This device is non-<br>cytotoxic                                     |
| ISO 10993-10     | Dermal Sensitization | Negligible          | This device is not<br>considered a contact<br>sensitizer             |
| ISO 10993-10     | Skin Irritation      | Negligible          | This device is classified<br>as negligibly irritating to<br>the skin |

{6}------------------------------------------------

# IX. SUMMARY OF CLINICAL TESTING

Clinical testing is not needed for this device.

# X. CONCLUSION

The proposed device has the same indication for use and similar technological characteristic as the predicate device. Non-clinical testing demonstrates that the proposed device performs as safe and effective as the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K211497](https://fda-staging.innolitics.com/device/K211497)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).