K211471 · Ortho Development Corporation · HSX · Dec 16, 2021 · Orthopedic
Device Facts
Record ID
K211471
Device Name
Balanced Knee System Uni
Applicant
Ortho Development Corporation
Product Code
HSX · Orthopedic
Decision Date
Dec 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1. Non-inflammatory degenerative joint diseases (NIDJD), e.g., osteoarthritis, avascular necrosis 2. Traumatic arthritis 3. Previous tibial condyle or plateau fracture with loss of anatomy or function 4. Varus deformities 5. Revision of the tibial bearing insert of a previously implanted unicompartmental knee system provided that the tibial tray mechanism is not compromised, and femoral and tibial tray components remain well fixed and undamaged. The BKS Uni is intended for unicompartmental knee arthroplasty procedures. The system is single-use and intended for implantation with bone cement.
Device Story
Balanced Knee System (BKS) Uni is a single-compartment knee replacement prosthesis; used for unicompartmental knee arthroplasty. System comprises cobalt chromium femoral component, titanium (Ti-6Al-4V ELI) tibial tray, and crosslinked Vitamin E UHMWPE tibial insert. Device implanted by orthopedic surgeons in clinical/OR settings using bone cement. Functions as a non-constrained cemented prosthesis to restore joint function in patients with degenerative or traumatic joint disease. Output is a mechanical replacement of the femorotibial joint surface; provides structural support and articulation to alleviate pain and improve mobility. Single-use device.
Clinical Evidence
No clinical testing was conducted. Substantial equivalence is supported by bench testing, including range of motion, constraint, component interlock strength, tibial and femoral fatigue, contact area/stress, and wear test analysis.
Technological Characteristics
Materials: Cobalt Chromium (femoral), Ti-6Al-4V ELI (tibial tray), crosslinked Vitamin E UHMWPE (insert). Principle: Non-constrained cemented femorotibial prosthesis. Sterilization: Gamma radiation and ethylene oxide (ISO 11137-1, ISO 11137-2, ISO 11135). Biocompatibility: ISO 10993-1. Packaging: ISO 11607-1/2.
Indications for Use
Indicated for patients with non-inflammatory degenerative joint diseases (osteoarthritis, avascular necrosis), traumatic arthritis, previous tibial condyle/plateau fracture, or varus deformities requiring unicompartmental knee arthroplasty. Also indicated for revision of tibial bearing inserts in previously implanted unicompartmental systems with well-fixed, undamaged components.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
December 16, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Ortho Development Corporation Drew Weaver Director of Regulatory & Clinical Affairs 12187 South Business Park Drive Draper. Utah 84020
# Re: K211471
Trade/Device Name: Balanced Knee® System Uni Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: November 10, 2021 Received: November 12, 2021
### Dear Drew Weaver:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K211471
Device Name Balanced Knee® System Uni
Indications for Use (Describe)
1. Non-inflammatory degenerative joint diseases (NIDJD), e.g., osteoarthritis, avascular necrosis
2. Traumatic arthritis
3. Previous tibial condyle or plateau fracture with loss of anatomy or function
4. Varus deformities
5. Revision of the tibial bearing insert of a previously implanted unicompartmental knee system provided that the tibial tray mechanism is not compromised, and femoral and tibial tray components remain well fixed and undamaged.
The BKS Uni is intended for unicompartmental knee arthroplasty procedures. The system is single-use and intended for implantation with bone cement.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K211471 page 1 of 4
# Section 5
# 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
| Name of the Sponsor: | Ortho Development® Corporation<br>12187 South Business Park Drive<br>Draper, Utah 84020 |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Primary Contact: | Name: Divya Palan<br>Position: Regulatory Affairs Specialist<br>Telephone: (617) 459-9652<br>Email: DPalan@orthodevelopment.com |
| 510(k) Secondary Contact: | Name: Darlene Hull<br>Position: Director of Regulatory and Clinical Affairs.<br>Telephone: (801) 619-3499<br>Email: DHull@orthodevelopment.com |
| Date Prepared: | May 10, 2021 |
| Submission Type: | Traditional 510(k) |
| Proprietary Name: | Balanced Knee® System Uni |
| Common Name: | Knee joint femorotibial metal/polymer non-constrained<br>cemented prosthesis |
| Classification: | 21 CFR 888.3520 |
| Device Class: | Class II |
| Device Product Code: | HSX |
| Primary Predicate Device: | Zimmer Unicompartmental Knee System (K033363) |
| Secondary Predicate Device: | EMPOWR Partial Knee System (K191325) |
{4}------------------------------------------------
# K211471 page 2 of 4
### 5.1 Device Description:
The Balanced Knee® System (BKS®) Uni is a single compartment knee replacement. The BKS Uni is indicated for cemented use only.
Cobalt Chromium Femoral Component
The femoral components are cobalt chromium (Co-Cr-Mo). The femoral components are right and left specific and are intended for cemented use only.
Titanium Tibial Trays The titanium tibial trays (Ti-6Al-4V ELI) are left and right specific and are for cemented use only.
Polyethylene Tibial Insert The E-Vitalize tibial insert is manufactured from crosslinked Vitamin E Ultra High Molecular Weight Polyethylene. The inserts match the respective size of the tibial tray used.
### 5.2 Indication for Use:
- 1. Non-inflammatory degenerative joint diseases (NIDJD), e.g., osteoarthritis, avascular necrosis
- 2. Traumatic arthritis
- 3. Previous tibial condyle or plateau fracture with loss of anatomy or function
- 4. Varus deformities
- 5. Revision of the tibial bearing insert of a previously implanted unicompartmental knee system provided that the tibial tray mechanism is not compromised, and femoral and tibial tray components remain well fixed and undamaged.
The BKS Uni is intended for unicompartmental knee arthroplasty procedures. The system is single-use and intended for implantation with bone cement.
#### 5.3 Comparison of Technological Characteristic:
The Balanced Knee® System Uni is technologically similar to the already cleared predicate device Zimmer Unicompartmental Knee System (K033363) and EMPOWR Partial Knee System (K191325) in terms of indication for use/intended use, technological characteristics, basic design, device material, and principle of operation.
{5}------------------------------------------------
# K211471 page 3 of 4
### 5.4 Performance Data:
### Sterilization
The Balanced Knee® System Uni is gamma radiation and ethylene oxide sterilized and was validated to a sterility assurance level of 10° in accordance with the following standards:
- ISO 11137-1:2006, Am1:2013, Sterilization of health care products Requirements . for validation and routine control - Radiation sterilization; and
- . ISO 11137-2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose.
- ISO 11135:2014/Amd.1:2018(E), Sterilization of health-care products -. Requirements for the development, validation and routine control of sterilization process for medical devices.
Validation results indicate that the Balanced Knee® System Uni complies with the standards.
#### Shelf Life
The packaging of Balanced Knee® System Uni was validated in accordance with the following standards:
- ISO 11607-1:2006 Packaging for terminally sterilized medical devices Part 1: . requirements for materials, sterile barrier systems and packaging systems; and
- ISO 11607-2:2006 Packaging for terminally sterilized medical devices Part 2: . Validation requirements for forming, sealing and assembly processes.
Validation results indicate that the packaging for the Balanced Knee® System Uni complies with the standards.
#### Biocompatibility
The Balanced Knee® System Uni contact materials were verified in accordance with the following standards:
- ISO 10993-1:2009, Biological evaluation of medical device Part 1: Evaluation and . testing within a risk management process.
{6}------------------------------------------------
# K211471 page 4 of 4
Validation results indicate that the biological evaluation complies with the standard.
# Mechanical Testing
The following non-clinical mechanical tests and analyses were conducted on the subject device.
- Range of Motion Test .
- Constraint Test .
- Component Interlock Strength Test .
- Tibial Component Fatigue Test .
- . Contact Area/Stress Test
- Femoral Fatigue Test ●
- Wear Test Analysis .
# Clinical Testing
No clinical testing of the Balanced Knee® System Uni has been conducted.
# 5.5 Substantial Equivalence Conclusion:
Verification and Validation activities were conducted to establish the performance of Balanced Knee® System Uni. The results of these activities demonstrate that Balanced Knee® System Uni is as safe, as effective, and performs as well as the legally marketed predicates.
Based on similarities in indication for use/intended use, technological characteristic, basic design, device material, and principle of operation, Balanced Knee® System Uni is considered substantially equivalent to the previously cleared predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.