← Product Code [PBX](/productcode/PBX) · K211461

# Family of Venus RF Systems - Venus Freedom (K211461)

_Venus Concept USA, Inc. · PBX · Oct 14, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211461

## Device Facts

- **Applicant:** Venus Concept USA, Inc.
- **Product Code:** [PBX](/productcode/PBX.md)
- **Decision Date:** Oct 14, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Venus Freedom device is intended for the treatment of the following medical conditions; using the Pearl, Diamond and Slim applicators for delivery of non-thermal RF combined with massage and magnetic field pulses: - Relief of minor muscle aches and pain, relief of muscle spasm. - Temporary improvement of local blood circulation - Temporary reduction in the appearance of cellulite.

## Device Story

Non-invasive, non-ablative RF system; delivers radiofrequency (RF) energy and pulsed electro-magnetic field (PEMF) combined with mechanical massage to skin/subcutaneous tissue. Generates heat via electrical impedance in epidermis, dermis, and muscular tissue. Used in professional healthcare facilities by clinicians. Console includes power supply, RF/PEMF generators, CPU, and display; utilizes Pearl, Diamond, or Slim applicators with single-use disposable tips. Clinician operates device via footswitch and interface to trigger tissue changes; output is thermal/non-thermal energy delivery. Benefits include pain/spasm relief, improved circulation, and cellulite reduction.

## Clinical Evidence

Bench testing only. Verification testing confirmed max RF power output and PEMF power density. Bench testing demonstrated ability to maintain safe skin surface temperatures using protocols consistent with the predicate device. Biocompatibility testing (ISO 10993-1, -5, -10) passed for disposable tips.

## Technological Characteristics

RF frequency: 1MHz; PEMF: 15 Gauss (15Hz). Applicators: Pearl (6 electrodes, 1 PEMF coil), Diamond (4 electrodes, 1 PEMF coil), Slim (6 electrodes, 1 PEMF coil). Biocompatible materials per ISO 10993-1. Standards: IEC 60601-1, IEC 60601-1-6, IEC 60601-2-2, IEC 60601-1-2, IEC 62304. Console-based system with footswitch.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Venus Heal device ([K182094](/device/K182094.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left is the Department of Health & Human Services logo, and on the right is the FDA logo. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

October 14, 2021

Venus Concept USA Inc. Yoni Iger VP Regulatory, Science & Technology 1880 N Commerce Pkwv. Suite 2 Weston, Florida 33326

Re: K211461

Trade/Device Name: Family of Venus RF Systems - Venus Freedom Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: PBX Dated: August 10, 2021 Received: August 17, 2021

Dear Yoni Iger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020

See PRA Statement below.

510(k) Number (if known) K211461

Device Name

Family of RF Systems - Venus Freedom

Indications for Use (Describe)

The Venus Freedom device is intended for the treatment of the following medical conditions; using the Pearl, Diamond and Slim applicators for delivery of non-thermal RF combined with massage and magnetic field pulses:

- Relief of minor muscle aches and pain, relief of muscle spasm.
- Temporary improvement of local blood circulation
- Temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unlessit displays a currently valid OMB number. "

Page 1 of 1 PSC Publishing Services (301) 443-6740 EF FORM FDA 3881 (7/17)

{3}------------------------------------------------

#### VI. 510(k) SUMMARY

## VENUS FREEDOM DEVICE - K211461

- Applicant Name: Venus Concept USA Inc. 1880 N Commerce Pkwy, Ste 2 Weston, FL33326, USA Tel: +1 888 907-0115
- Contact Person: Dr. Yoni Iger VP Regulatory, Science & Technology Venus Concept USA Inc. Yoni@venusconcept.com Tel: +1 888 907-0115

Date Prepared: October 13, 2021

Trade Name: Venus Freedom Device

Classification Name: 21 CFR 878.4400 Electrosurgical cutting and coagulation device and accessories

Product Code: PBX

Classification: Class II Medical Device

Classification Panel: General & Plastic Surgery Panel

Predicate Device: Venus Heal device (K182094)

## Intended Use/Indication for Use:

The Venus Freedom device is intended for the treatment of the following medical conditions; using the Pearl, Diamond and Slim applicators for delivery of non-thermal RF combined with massage and magnetic field pulses:

·Relief of minor muscle aches and pain, relief of muscle spasm

·Temporary improvement of local blood circulation

· Temporary reduction in the appearance of cellulite

{4}------------------------------------------------

# K211461

## Device Description:

The Venus Freedom device, part of the Family of Venus RF devices, is a noninvasive, non-ablative device that delivers Radiofrequency (RF) energy, along with the Pulsed Electro-Magnetic Field (PEMF), into the skin to generate heat through electrical impedance in the epidermis, dermis, subcutaneous layers, and muscular tissue.

The device consists of a console (main unit) and three applicators (Pearl, Diamond and Slim).

The Venus Freedom RF energy, PEMF and tissue manipulation (massage), are utilized to trigger changes in the tissue which result in local blood circulation improvement, muscle spasm relief along with pain relief. Temporary reduction in the appearance of cellulite is similarly achieved by the combination of these three main mechanisms: tissue manipulation (massage), RF delivery and PEMF delivery.

Consistent with the previous clearance, the device is intended to be used in professional healthcare facilities (prescription use) just as the predicate.

## Comparison of Technological Characteristics:

The following table compares the subject Venus Freedom device to its predicate device with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.

| Parameter                                                   | Proposed Device:<br>Venus Freedom<br>Venus Concept Inc.                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device:<br>Venus Heal<br>Venus Concept Inc.<br>(K182094)                                                                                                                                                                                                                                                                                                                                                                | Substantial<br>Equivalence                                                                                                           |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Product Class, Code                                         | Class II, PBX                                                                                                                                                                                                                                                                                                                                                                                                                                   | Class II, PBX                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                 |
| Indications for Use                                         | The Venus Freedom device is<br>intended for the treatment of<br>the following medical<br>conditions; using the Pearl,<br>Diamond and Slim<br>applicators for delivery of<br>non-thermal RF combined<br>with massage and magnetic<br>field pulses:<br>• Relief of minor muscle aches<br>and pain, relief of muscle<br>spasm<br>• Temporary improvement of<br>local blood circulation<br>• Temporary reduction in the<br>appearance of cellulite. | The Venus Heal device<br>is intended for the treatment<br>of the following medical<br>conditions; using the Large<br>and Small applicators for<br>delivery of non-thermal RF<br>combined with massage and<br>magnetic field pulses:<br>• Relief of minor muscle aches<br>and pain, relief of muscle<br>spasm<br>• Temporary improvement of<br>local blood circulation<br>• Temporary reduction in the<br>appearance of cellulite. | Same                                                                                                                                 |
| Parameter                                                   | Proposed Device:<br>Venus Freedom<br>Venus Concept Inc.                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device:<br>Venus Heal<br>Venus Concept Inc.<br>(K182094)                                                                                                                                                                                                                                                                                                                                                                | Substantial<br>Equivalence                                                                                                           |
| Energy Used/<br>Delivered                                   | RF Energy<br>Pulsed Electro-Magnetic Field (PEMF)                                                                                                                                                                                                                                                                                                                                                                                               | RF Energy<br>Pulsed Electro-Magnetic Field (PEMF)                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                 |
| Tissue Manipulation<br>Mechanism                            | Massage mechanism                                                                                                                                                                                                                                                                                                                                                                                                                               | Massage mechanism                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                 |
| Design                                                      | Use of RF and PEMF combined<br>energies delivered through applicator<br>and massage of the skin.                                                                                                                                                                                                                                                                                                                                                | Use of RF and PEMF combined<br>energies delivered through applicator<br>and massage of the skin.                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                 |
| Components                                                  | Console, including:<br>- Power supply<br>- RF generator<br>- Main CPU<br>- Display panel<br>- PEMF generator<br>Applicators:<br>- Pearl<br>- Diamond<br>- Slim<br>Footswitch                                                                                                                                                                                                                                                                    | Console, including:<br>- Power supply<br>- RF generator<br>- Main CPU<br>- Display panel<br>- PEMF generator<br>Applicators:<br>- Large<br>- Small                                                                                                                                                                                                                                                                                | Same<br>New<br>Applicator<br>Addition of<br>Footswitch                                                                               |
| Applicators                                                 | Pearl Applicator:<br>6 electrodes<br>1 PEMF coil<br>Diamond Applicator:<br>4 electrodes<br>1 PEMF coil<br>Slim Applicator:<br>6 electrodes<br>1 PEMF coil                                                                                                                                                                                                                                                                                       | Large Applicator:<br>8 electrodes<br>8 PEMF coils<br>Small Applicator:<br>4 electrodes<br>4 PEMF coils                                                                                                                                                                                                                                                                                                                            | Same<br>delivery of<br>Bipolar RF<br>and PEMF,<br>although<br>different<br>number of<br>electrodes<br>and coils<br>New<br>Applicator |
| Applicator's<br>disposable tip                              | Single use disposable per applicator:<br>- Pearl disposable tip<br>- Diamond disposable tip<br>- Slim disposable tip                                                                                                                                                                                                                                                                                                                            | None, all applicators are reusable                                                                                                                                                                                                                                                                                                                                                                                                | Different                                                                                                                            |
| Parameter                                                   | Proposed Device:<br>Venus Freedom<br>Venus Concept Inc.                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device:<br>Venus Heal<br>Venus Concept Inc.<br>(K182094)                                                                                                                                                                                                                                                                                                                                                                | Substantial<br>Equivalence                                                                                                           |
| Device Dimensions                                           | 16.5x16.5x10.4 in (W × D × H)<br>42x42x26.5 cm (W × D × H)                                                                                                                                                                                                                                                                                                                                                                                      | 16.5x16.5x10.4 in (W × D × H)<br>42x42x26.5 cm (W × D × H)                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                 |
| Performance                                                 | Frequency: 1MHz<br>Max. RF output power:<br>80W (Pearl, Diamond and Slim<br>applicators)                                                                                                                                                                                                                                                                                                                                                        | Frequency: 1MHz<br>Max. RF output power:<br>100W (Large and Small applicators)                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                 |
|                                                             | PEMF Power: 15 Gauss (15Hz)                                                                                                                                                                                                                                                                                                                                                                                                                     | PEMF Power: 15 Gauss (15Hz)                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                 |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different                                                                                                                            |
| Power density per<br>effective treatment<br>area (Watt/cm²) | Pearl Applicator:<br>3.75W/cm²<br>Diamond Applicator:<br>24.24W/cm²<br>Slim Applicator:<br>4.5W/cm² (1 pair of electrodes)                                                                                                                                                                                                                                                                                                                      | Large Applicator:<br>2.6W/cm²<br>Small Applicator:<br>20.4 W/cm²                                                                                                                                                                                                                                                                                                                                                                  | Different                                                                                                                            |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different                                                                                                                            |
| Materials                                                   | Biocompatible materials per ISO<br>10993-1                                                                                                                                                                                                                                                                                                                                                                                                      | Biocompatible materials per ISO<br>10993-1                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                 |

{5}------------------------------------------------

{6}------------------------------------------------

## Performance Data:

Venus Concept conducted several performance tests to demonstrate that the subject Venus Freedom device complies with performance standards and that it functions as intended per its design.

- Verification test demonstrating that the Venus Freedom device meets the system's technical specification for the max RF power output and max PEMF power density.

- A Bench test which demonstrated the ability of the subject device to maintain a safe treatment temperature on the surface of the human skin with the same protocol that was used for the predicate device, Venus Heal (K182094).

{7}------------------------------------------------

# K211461

Electrical Safety and Electromagnetic Compatibility: The Venus Freedom system was tested per the applicable electrical safety and electromagnetic compatibility standards listed below, and all results were passing.

- IEC 60601-1: (2005) and A1:2012, Medical Electrical Equipment Part 1-1: General requirements – for basic safety and essential performance
- IEC 60601-1-6: (2010/AMD2013), Medical Electrical Equipment Part 1-6, General requirements for basic safety and essential performance
- IEC 60601-2-2:2017, Medical Electrical Equipment Part 2-2: Particular requirements for basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-1-2: 2014, Medical Electrical Equipment, Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 62304: (2006/AMD2015) Medical device software Software life cycle processes -

# Software Testing:

Software verification and validation testing has been performed; the device functioned as intended and the results observed were as expected.

# Biocompatibility:

Biocompatibility testing was performed on the Venus Freedom's disposable tip's materials according to the FDA Guidance and ISO 10993-1 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. Testing included cytotoxicity, sensitization and irritation according to standards set forth in ISO 10993- 5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity and ISO 10993-10 Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization. All testing passed.

# Conclusions:

The Venus Freedom device is as safe and effective as its predicate device, Venus Heal, cleared under K182094 for the same requested intended use. The Venus Freedom device has the same intended use and indications for use, similar technological characteristics, and the same principle of operation as its predicate device. Performance data demonstrated that the technological differences in the Venus Freedom do not raise any issues of safety or effectiveness in comparison to the predicate device. Thus, the Venus Freedom device is substantially equivalent to its predicate, Venus Heal (K182094) for the requested intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K211461](https://fda-staging.innolitics.com/device/K211461)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
