Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses
Applicant
Pegavision Corporation
Product Code
LPL · Ophthalmic
Decision Date
Dec 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with nondiseased eyes in powers from +6.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myoropia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters. Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition from +0.25D to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear.
Device Story
Ionic hydrogel contact lenses (Etafilcon A) for daily wear; available in spherical, aspherical, toric, and multifocal designs. Manufactured via cast-molding; photo-polymerized HEMA/MAA copolymer. Contains UV-absorbing monomer (benzotriazole) and visibility tint (Reactive Blue 19). Supplied sterile in blister packs with borate buffered saline solution containing Tween 80, Sodium Hyaluronate, Polyethylene Glycol, and Cyanocobalamin. Prescribed by eye care professionals for ametropia correction. Benefits include vision correction and UV protection. Daily disposable or frequent/planned replacement options available.
Clinical Evidence
No clinical data provided. Safety and effectiveness established via non-clinical bench testing, including physicochemical properties, biocompatibility (ISO 10993-5, -10, -11), extractables, and stability/sterility testing.
Technological Characteristics
Material: Etafilcon A (HEMA/MAA copolymer); Group IV high water ionic hydrogel. Water content: 58%. Refractive index: 1.402. Oxygen permeability: 19.73 x 10^-11 (cm²/sec)(mlO2/ml-mmHg). UV blocking: <5% UVB, <50% UVA. Manufacturing: Cast-molded. Packaging: Borate buffered saline with Tween 80, Sodium Hyaluronate, Polyethylene Glycol, and Cyanocobalamin.
Indications for Use
Indicated for daily wear correction of ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic/non-aphakic persons with non-diseased eyes. Power ranges +6.00 to -12.25D; astigmatism correction up to -3.50D; reading add +0.25D to +3.00D.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses (K200296)
Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 6, 2021
Pegavision Corporation Estela Lin Regulatory Affair Senior Engineer 2F-1, No.5, Shing Yeh St. Taoyuan, Guishan Dist. 33341 Taiwan
Re: K211448
Trade/Device Name Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses, Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN. Dated: October 28, 2021 Received: November 2, 2021
Dear Estela Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation titled "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-devicereporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211448
#### Device Name
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
Indications for Use (Describe)
Spherical and Aspherical
Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with nondiseased eyes in powers from +6.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Toric
Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myoropia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters.
#### Multifocal
Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition from +0.25D to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear.
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
| Type of Use (Select one or both, as applicable) | | | | |
#### This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211448
#### Device Name
Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses
Indications for Use (Describe)
Spherical and Aspherical
Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Toric
Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25diopters and astigmatic corrections from -0.25 to -3.50 diopters.
#### Multifocal
Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition from +0.25D to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eve care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
The following 510(K) Summary is being submitted as required by 21CFR 807.92(a).
| Company: | PEGAVISION CORPORATION<br>2F-1 No.5, Shing Yeh St., Guishan Dist., Taoyuan City 333,<br>Taiwan |
|-----------------|------------------------------------------------------------------------------------------------|
| Contact Person: | Mr. TS Yang, President |
| Phone: | 886-3-329-8808 |
| Fax: | 886-3-329-8897 |
| E-Mail: | TSYang@pegavision.com |
| Date Prepared: | December 01, 2021 |
## Submitter Information
## Identification of the Predicate Device
| Trade Name: | Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic)<br>Contact Lenses<br>Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Soft (hydrophilic) Contact Lenses (daily wear) |
| Classification<br>Name: | Lenses, Soft Contact, Daily Wear 21CFR. 886.5925, Product<br>Code LPL<br>Lens. Soft Contact (Disposable). 21CFR. 886.5925, Product<br>Code MVN |
| FDA Classification: | Class II |
| Predicate Device<br>Name: | K200296 Pegavision Products<br>Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact<br>Lenses<br>Pegavision (Etafilcon A) Color Daily Disposable Soft<br>(Hydrophilic) Contact Lenses |
## Description of Device
The Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses
{5}------------------------------------------------
and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are clear, visibility-tinted, UV absorber-containing and are available in spherical, aspherical, toric and multifocal designs. It is an ionic hydrogel lens derived from Etafilcon A material which is a co-polymer of 2-Hydroxyethyl Methacrylate (HEMA) and Methacrylic Acid (MAA), cross-linked with Ethylene Glycol Dimethacrylate (EGDMA) and 1,1,1-Trimethylolpropane Trimethacrylate (TMPTMA) and made by photo-polymerization. The lens consists of 42% Etafilcon A and 58% water by weight when immersed in buffered saline solution. The lens polymer further contains benzotriazole, a UV absorbing monomer, thus is able to block the UV radiation to the ocular. . The average transmittance characteristics of theses lenses are less than 5% in the UVB range of 280-315 nm and less than 50% in the UVA range of 315-380mm. The lens is visibly tined with "Reactive Blue19" color additive, 21 CFR section 73.3121. Lenses are supplied sterile in the sealed blister package containing sterile isotonic borate buffered saline solution with Tween 80, Sodium Hyaluronate, Polyethylene Glycol, and Cyanocobalamin as a color indicator to induce the package solution as pink color.
## Indications for use
## 1. Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses Spherical and Aspherical
Aquamax (Etafilcon A) SPHERE and ASPHERE Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
#### Toric
Aquamax (Etafilcon A) Toric Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters.
#### Multifocal
Aquamax (Etafilcon A) Multifocal Daily Disposable Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with
{6}------------------------------------------------
non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition from +0.25D to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses are intended for single-use disposable wear.
## 2. Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses Spherical and Aspherical
Aquamax (Etafilcon A) SPHERE and ASPHERE Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia and hyperopia) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.00 diopters. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
### Toric
Aquamax (Etafilcon A) Toric Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of ametropia (myopia or hyperopia with astigmatism) in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and astigmatic corrections from -0.25 to -3.50 diopters.
#### Multifocal
Aquamax (Etafilcon A) Multifocal Soft (Hydrophilic) Contact Lenses are indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) and presbyopia in aphakic and/or non-aphakic persons with non-diseased eyes in powers from +6.00 to -12.25 diopters and with non-diseased eyes who may require a reading addition from +0.25D to +3.00D. The lenses may be worn by persons who exhibit astigmatism of 2.00 diopters or less that does not interfere with visual acuity.
The lenses are intended for frequent/planned replacement wear with cleaning, rinsing, disinfection and scheduled replacement as prescribed by the eye care professional. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical (not heat) lens care system only.
{7}------------------------------------------------
## Summary of Clinical Study
Etafilcon A lenses have been used widely. Their safety and effectiveness have been well documented. Their safety and effectiveness can be further exemplified by the lenses cleared by FDA.
Clinical studies for Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses of the present device are not required for the premarket notification, as the USAN name and process are the same as the above-mentioned predicate devices.
## Summary of Non-clinical Study
#### a) Non-Clinical Testing performed
The following tests were conducted as recommended by the Premarket Notification 510(k) Guidance Document for Daily Wear Contact Lenses, revised May 1994. Includes:
- -Physicochemical Properties Test
- Biocompatibility Test -
| Contact Lenses (Etafilcon A) |
|------------------------------|
|------------------------------|
- Cytotoxicity Test (according to ISO 10993-5)
- Maximization Sensitization Test (according to ISO 10993-10)
- . Ocular Irritation Test (according to ISO 10993-10)
- . Acute Systemic Toxicity Test (according to ISO 10993-11)
Packaging Solution
- Cytotoxicity Test (according to ISO 10993-5) .
- . Ocular Irritation Test (according to ISO 10993-10)
Plastic Container (Foil& PP Blister)
- Cytotoxicity Test (according to ISO 10993-5) .
- . Ocular Irritation Test (according to ISO 10993-10)
- Acute Systemic toxicity test (according to ISO 10993-11) .
Extractable Test -
- -Shelf Life Test and Sterility Test
The non-clinical performance tests had been performed to demonstrate the safety and effectiveness of Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses and establish substantial equivalence to predicate lenses Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses and Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296). The evidence of substantial equivalence to the predicate lenses is described below.
{8}------------------------------------------------
#### Technological characteristics studies b)
The technological characteristics of Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are illustrated in the following Table.
| | Proposed Device | K200296 Predicate |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Production Method | Cast-Molded | Cast-Molded |
| USAN Name | Etafilcon A | Etafilcon A |
| Material Classification | Group IV<br>high water ionic | Group IV<br>high water ionic |
| Water Content (%) | 58% | 58% |
| Refractive Index | 1.402 | 1.402 |
| Oxygen Permeability (edge<br>corrected) @ 35°C | 19.73 x 10-11<br>(cm²/sec)(mlO2/ml-mmHg) | 19.73 x 10-11<br>(cm²/sec)(mlO2/ml-mmHg) |
| Transmittance<br>Visible light @ 380~780nm<br>UVA @ 380-315nm<br>UVB @ 315-280nm | > 95%<br>< 50%<br>< 5% | > 95%<br>< 50%<br>< 5% |
| Lens design | Spherical and Aspherical<br>Toric<br>Multifocal | Spherical and Aspherical<br>Toric<br>Multifocal |
| Packaging solution | Borate buffered saline (with Tween<br>80, Sodium hyaluronate,<br>Polyethylene Glycol, and<br>Cyanocobalamin) | Borate buffered saline<br>(with Tween 80, Sodium<br>hyaluronate, Polyethylene Glycol) |
#### Substantial Equivalence Statement
Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses for daily wear submitted in this 510(k) file is substantially equivalent in intended use, main materials, design, safety and performance claims to the predicate devices - Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses, Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296). Successful results from chemical/physical, stability, biocompatibility tests, extractable test confirm the lenses are within established finished product specification, remain stable, and are non-toxic and biocompatible with the ocular environment. Differences between the devices are the package solution, which cited in this section do not raise any new issues of substantial equivalence.
{9}------------------------------------------------
## Conclusion
A series of pre-clinical tests were performed to demonstrate the safety and effectiveness of the Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses and to establish substantial equivalence to the predicate devices. Information submitted in the 510(k) also establishes that the Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses do not raise questions of safety and effectiveness. Therefore, Aquamax (Etafilcon A) Daily Disposable Soft (Hydrophilic) Contact Lenses and Aquamax (Etafilcon A) Soft (Hydrophilic) Contact Lenses are substantially equivalent to the predicate devices - Pegavision (Etafilcon A) Color Soft (Hydrophilic) Contact Lenses, Pegavision (Etafilcon A) Color Daily Disposable Soft (Hydrophilic) Contact Lenses (K200296).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.