The navigated instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as a skull, a long bone, or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. When used with a Medtronic StealthStation™ Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery.
Device Story
StealthStation Spinous Process Clamps are reusable orthopedic stereotaxic instruments designed for use with the Medtronic StealthStation Navigation System. The device provides rigid attachment between a patient and a patient reference frame during surgery. It functions by clamping onto the spinous process of a vertebra, allowing the navigation system to track the patient's anatomy relative to pre-operative or intra-operative imaging (CT, MR, or fluoroscopy). The clamps are operated by surgeons in an OR setting. By maintaining a stable reference frame, the device enables the navigation system to provide real-time guidance, potentially increasing surgical accuracy and reducing risks associated with spinal procedures. The device is constructed from titanium and stainless steel and is steam-sterilizable.
Clinical Evidence
Bench testing only. Evidence includes functional verification, useful life testing, navigation accuracy testing, and packaging verification (ASTM D4169, ISTA 2A). Biological safety testing conducted per ISO 10993-1:2018 confirms the device is non-cytotoxic, non-sensitizing, non-irritating, non-toxic, and non-pyrogenic.
Technological Characteristics
Materials: Titanium, 17-4 Stainless Steel, 300 Series Stainless Steel. Principle: Mechanical clamping for rigid fixation of navigation reference frames. Form factor: Reusable clamps with varying jaw shapes (concave/non-concave) and sizes (Short, Medium, Tall). Sterilization: Steam sterilization. Connectivity: Passive component used with StealthStation navigation system.
Indications for Use
Indicated for patients requiring stereotactic surgery where rigid anatomical structures (skull, long bone, vertebra) can be referenced against CT/MR models, fluoroscopy, or digitized landmarks. Used for rigid attachment of patient reference frames during surgery.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
July 8, 2021
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Medtronic Navigation Victoria Baldock Associate Regulatory Affairs Specialist 826 Coal Creek Circle Louisville, Colorado 80027
## Re: K211442
Trade/Device Name: StealthStation Spinous Process Clamps Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: May 7, 2021 Received: May 10, 2021
## Dear Victoria Baldock:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211442
Device Name StealthStation™ Spinous Process Clamps
### Indications for Use (Describe)
The navigated instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as a skull, a long bone, or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
When used with a Medtronic StealthStation™ Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------|
| Prescription Use (Ref 21 CFR 201 Subpart D) | <span style="text-decoration: overline;">X</span> |
| Over-The-Counter Use (21 CFR 201 Subpart C) | <span style="text-decoration: overline;"> </span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K211442
Page 1 of 4
## 510(k) Summary
## May 07, 2021
### l. Company: Medtronic Navigation 826 Coal Creek Circle Louisville, CO 80027 Telephone Number: (720) 890-3500
Contact: Victoria Baldock Associate Regulatory Affairs Specialist Telephone Number: (423) 863-5907 Email: tori.a.baldock@medtronic.com
> Rishi Sinha (Alternate) Regulatory Affairs Director Telephone Number: (720) 890-2485 Email: rishi.k.sinha@medtronic.com
- Proprietary Trade Name: StealthStation™ Spinous Process Clamps II.
- lll. Common Name: Orthopedic Stereotaxic Instrument
- IV. Classification Name: Stereotaxic Instrument (21 CFR 882.4560)
- V. Classification: Class II
- VI. Product Code: OLO (Stereotaxic Instrument)
### VII. Predicate Devices:
The legally marketed predicate devices are identified below:
| Predicate | 510(k) Clearance |
|--------------------------------------------------------------------------------------|------------------|
| Navigated CAPSTONE® Trials, CLYDESDALE® Trials, and CAPSTONE® & CLYDESDALE® Inserter | K131425 |
#### VIII. Product Description:
The Spinous Process Clamps are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery. The subject devices are designed for use with the StealthStation™ System and are intended to be reusable.
### IX. Indications for Use:
The navigated instruments are specifically designed for use with the StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure such as a skull, a long bone, or vertebra can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
{4}------------------------------------------------
When used with a Medtronic StealthStation™ Navigation System, the Spine Referencing fixation devices are intended to provide rigid attachment between patient and patient reference frame for the duration of the surgery.
### X. Comparison of the Technological Characteristics:
The subject Spinous Process Clamps (Generation 4) utilize the same fundamental scientific technology and have the same mode of operation and functionality as the predicate Spinous Process Clamps (Generation 2) (K131425). The subject devices have equivalent materials and sterilization/reprocessing methods as the predicate (K131425). The changes in design features between the subject and the predate do not raise any new risks or any concerns about the safety and effectiveness. The subject devices also meet the established navigational accuracy requirements.
| Feature | Subject Devices<br>StealthStation™ Spinous<br>Process Clamps | Predicate Devices<br>Navigated CAPSTONE®<br>Trials, CLYDESDALE®<br>Trials, and CAPSTONE®<br>& CLYDESDALE®<br>Inserter (K131425) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | When used with a<br>Medtronic StealthStation™<br>Navigation System, the<br>Spine Referencing fixation<br>devices are intended to<br>provide rigid attachment<br>between patient and patient<br>reference frame for the<br>duration of the surgery. The<br>devices are intended to be<br>reusable. | When used with a<br>Medtronic StealthStation™<br>Navigation System, the<br>Spine Referencing fixation<br>devices are intended to<br>provide rigid attachment<br>between patient and<br>patient reference frame for<br>the duration of the surgery.<br>The devices are intended<br>to be reusable. |
| Indications for Use | The navigated instruments<br>are specifically designed for<br>use with the<br>StealthStation™ System,<br>which is indicated for any<br>medical condition in which<br>the use of stereotactic<br>surgery may be<br>appropriate, and where<br>reference to a rigid<br>anatomical structure such<br>as a skull, a long bone, or<br>vertebra can be identified<br>relative to a CT or MR<br>based model, fluoroscopy<br>images, or digitized<br>landmarks of the anatomy. | The navigated instruments<br>are specifically designed<br>for use with the<br>StealthStation™ System,<br>which is indicated for any<br>medical condition in which<br>the use of stereotactic<br>surgery may be<br>appropriate, and where<br>reference to a rigid<br>anatomical structure such<br>as a skull, a long bone, or<br>vertebra can be identified<br>relative to a CT or MR<br>based model, fluoroscopy<br>images, or digitized<br>landmarks of the anatomy. |
## Table 1. Comparison of Technological Characteristics
{5}------------------------------------------------
| Sterilization Method | Non-sterile, Steam sterilization, Reusable | Non-sterile, Steam sterilization, Reusable |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Material | Titanium, 17-4 Stainless Steel, 300 Series Stainless Steel | Titanium, 17-4 Stainless Steel, 300 Series Stainless Steel |
| Thread Relief | Screw engaged threads with clamping mechanism at all time, due to full threaded screw design. Removal of two washer components due to screw changes. | Unengaged screw when clamps are open, due to thread relief in screw design. |
| Jaw Shape | Single Clamps: Concave shape of jaws to align to the spinous process bulbous top<br>Double Clamps: Non-concave | Single Clamps: Non-concave<br>Double Clamps: Non-concave |
| Teeth | Single Clamps: Small teeth on the upper (concave) part of jaws, same size teeth as predicate on lower part of jaws<br>Double Clamps: Teeth all same size | Single Clamps: Teeth all same size<br>Double Clamps: Teeth all same size |
| Distal End Shape | Single Clamps: Pointed jaw at distal end<br>Double Clamps: Blunt jaw at distal end | Single Clamps: Blunt jaw at distal end<br>Double Clamps: Blunt jaw at distal end |
| Clamping Range | Single Clamps: 11 mm<br>Double Clamps: 9 mm | Single Clamps: 9 mm<br>Double Clamps: 9 mm |
| Size Offerings | Short, Medium, and Tall | Short and Tall |
### XI. Discussion of the Performance Testing:
Testing conducted to demonstrate equivalency of the subject device to the predicate is summarized as follows:
| Performance Testing<br>Activity | Description |
|---------------------------------|------------------------------------------------------------------------------------------------|
| Functional Verification | Confirms that the Spinous Process Clamp design<br>satisfies functional requirements. |
| Useful Life Testing | Confirms that the Spinous Process Clamps can<br>operate normally throughout their useful life. |
{6}------------------------------------------------
| Navigation Accuracy Testing | Verifies the robustness and navigational accuracy of<br>the Spinous Process Clamps. |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Packaging Verification | Confirms that the Spinous Process Clamps when<br>packaged in their packaging can withstand ship<br>testing per ASTM D4169 and ISTA 2A. |
Additionally, biological endpoint testing, conducted per recommendations from ISO 10993-1:2018, indicates that the subject devices are non-cytotoxic, non-sensitizing, nonirritating, non-toxic, and non-pyrogenic and pose a negligible risk of adverse biological effects to patients when used as intended.
### XII. Conclusion:
The subject devices have shown through comparison to be substantially equivalent to the identified predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.