SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
K211401 · Gyrus Acmi, Inc. · GEX · Jun 3, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211401
Device Name
SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
Applicant
Gyrus Acmi, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 3, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Device Story
Thulium laser system (1940 nm) delivers pulsed near-infrared light via optical fibers to target tissues; used for soft tissue surgery and lithotripsy. System comprises laser console, sterile fibers (150-940 microns), footswitch, and accessories. Operated by clinicians in OR/clinic settings. Integrated LED touchscreen controls laser parameters; auxiliary monitor displays data. Software v2.1 provides factory presets for procedures (e.g., ureteral stone). Laser energy absorption by tissue enables incision, ablation, or stone fragmentation. Benefits include precise tissue management and endoscopic stone treatment.
Clinical Evidence
Bench testing only. Software verification testing was conducted to confirm that the updated GUI and new factory presets met all performance requirements and acceptance criteria.
Technological Characteristics
Thulium laser (1940 nm) with 550 nm aiming beam. Max power: 60W (Premium), 35W (Pro). Max frequency: 2400 Hz (Premium), 100 Hz (Pro). Fiber diameters: 150-940 microns. Connectivity: wired/wireless footswitch, auxiliary video output. Sterilization: Ethylene Oxide (10^-6 SAL). Software: v2.1, GUI-based control. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and ISO 10993 standards.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2021
Gyrus ACMI, Inc. Julie Acker Program Manager, Regulatory Affairs 800 West Park Drive Westborough, Massachusetts 01581
Re: K211401
Trade/Device Name: SOLTIVE Laser System (SOLTIVE Pro SuperPulsed Laser, SOLTIVE Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 5, 2021 Received: May 6, 2021
Dear Julie Acker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K211401
#### Device Name
SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories)
#### Indications for Use (Describe)
The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Urology:
· Ablation of Benign Prostatic Hypertrophy) [BPH]• Laser Resection of the Prostrate (LRP)• Laser Enucleation of the Prostate (LEP) · Laser Ablation of the Prostate (LAP) · Transurethral Incision of the Prostate (TUIP) · Condylomas · Urethral strictures · Lesions of external genitalia · Bladder neck incisions (BN) · Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors Endoscopic fragmentation of ureteral, bladder, and renal calculi• Treatment of distal impacted fragments remaining in the ureters following lithotripsy
Lithotripsy and Percutaneous Urinary Lithotripsy Indications:
· Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones • Endoscopic fragmentation of calculi • Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
Gastroenterology:
Open and endoscopic gastroenterology surgery (incision, resection, ablation,
vaporization, coagulation and haemostasis) including:
· Appendectomy · Angiodysplasia · Polyps · Colorectal cancer · Biopsy · Telangiectasias
• Gall Bladder calculi • Telangiectasias of the Osler-Weber-Renu disease • Biliary/Bile duct calculi • Vascular
Malformation . Ulcers . Gastric ulcers . Esophagitis . Duodenal ulcers . Esophageal ulcers . Non Bleeding Ulcers · Varices · Pancreatitis · Colitis · Haemorrhoids · Mallory-Weiss tear · Cholecystectomy · Gastric Erosions • Benign and Malignant Neoplasm
Gynecology:
Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) of soft tissue
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# SOLTIVE™ LASER SYSTEM
# 510(k) PREMARKET NOTIFICATION K211401
# SECTION 5.0 - 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
Gyrus ACMI, Inc. SOUTHBOROUGH, MA
{5}------------------------------------------------
### 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. SOLTIVE™ LASER SYSTEM (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories)
#### General Information
| Contract Manufacturer: | IPG Medical Corporation<br>377 Simarano Drive<br>Marlborough, MA 01752, USA |
|------------------------------------|------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | Gyrus ACMI, Inc.<br>800 West Park Drive<br>Westborough, MA 01581 |
| Establishment Registration Number: | 3003790304 |
| Contact Person: | Julie Acker<br>Program Manager Regulatory Affairs<br>Phone: 484-294-6981<br>Email: julie.acker@olympus.com |
| Date Prepared: | May 5, 2021 |
**Device Description**
| Trade Name: | SOLTIVET™ Laser System<br>(SOLTIVET™ Pro SuperPulsed Laser,<br>SOLTIVE™ Premium SuperPulsed<br>Laser, SOLTIVET™ Laser Fibers, and<br>Accessories) |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Generic/Common Name: | Laser Instrument for Use in General<br>Surgery |
| Classification Name: | Laser Instrument for Use in General<br>Surgery |
| Regulation/CFR Citation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Classification: | Class II |
| Review Panel: | Gastroenterology/Urology |
**Predicate Devices**SOLTIVE Laser Systems
K183647
{6}------------------------------------------------
#### Comparison to Predicate Device:
The subject of this 510(k) is a modification to the SOLTIVE Laser Systems to update the application software to Version 2.1. The device hardware design is identical to the predicate. The subject change to the software updates the graphical user interface and corresponding changes to the Instructions for Use. The predicate Soltive application software offers standard factory presets for procedures commonly performed using the Soltive Laser System. The modified software Version 2.1 adds a factory preset within the Lithotripsy menu for Ureteral Stone procedures, as well as other enhancements.
#### Product Description
The SOLTIVE Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is a thulium laser, producing a pulsed beam of coherent near-infrared light (1940 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LED touch screen gives the user control over the necessary laser system parameters. The SOLTIVE Laser System is equipped with a 550 nm aiming beam.
The SOLTIVE Laser System is produced in two models, the SOLTIVE™ Premium SuperPulsed Laser and SOLTIVE™ Pro SuperPulsed Laser. The Premium has a maximum power output of 60 Watts and a maximum Frequency output of 2400 Hz. A secondary, foldable screen is provided. The Pro can operate at a maximum power of 35 Watts and Frequency limited to 100Hz. Both systems are operated with a wireless footswitch or wired footswitch, and both systems can utilize an auxiliary video monitor to display operating parameters. The Premium and Pro Laser must be used with a SOLTIVE Laser Fiber, offered in diameters from 150 - 940 microns.
The system includes:
- Laser console ●
- Laser fibers sterile single use and reusable ●
- Foot pedal, wireless or wired ●
- Accessories power cord, HDMI cable, safety goggles/glasses, fiber cutter, .
- fiber cleaver, fiber gripper, sterilization trav, cart .
The devices do not incorporate medicinal substances, tissues, or blood products. The laser fibers are single-use and reusable, and both single-use and reusable are initially sterilized by Ethylene Oxide. Fibers are compatible with the Olympus endoscope models as listed on the IFU.
{7}------------------------------------------------
#### Technological Characteristics
The proposed SOLTIVE Laser System with application software Version 2.1 has the identical intended use, design, and scientific technology as the predicate SOLTIVE Laser System (K183647). The subject modification is solely an update to the application software to add a Ureteral Stone preset to the menu of factory presets, and other enhancements. There are no new issues of safety or effectiveness with the proposed device.
#### Material
No material changes were made to the SOLTIVE Laser System cleared under K183647.
#### Indications for Uses
The indications for use of the modified device, as described in its labeling, have not changed as a result of the modification.
The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
### Urology
- · Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- · Laser Resection of the Prostrate (LRP)
- · Laser Enucleation of the Prostate (LEP)
- · Laser Ablation of the Prostate (LAP)
- · Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral strictures
- · Lesions of external genitalia
- · Bladder neck incisions (BNI)
- · Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- · Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
· Treatment of distal impacted fragments remaining in the ureters following lithotripsy
#### Lithotripsy and Percutaneous Urinary Lithotripsy Indications
• Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- · Endoscopic fragmentation of calculi
{8}------------------------------------------------
· Treatment of distal impacted fragments of steinstrasse when guide wire cannot be Passed
#### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including:
• Vascular Malformation
- · Appendectomy
- Polyps · Biopsy
- · Angiodysplasia • Colorectal cancer
- Telangiectasias
• Gastritis
• Colitis
• Esophagitis
- Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
• Gall Bladder calculi
• Non Bleeding Ulcers
- · Ulcers
- Gastric ulcers
- Duodenal ulcers
- · Varices
- · Pancreatitis
- Mallory-Weiss tear
• Esophageal ulcers
- Haemorrhoids • Cholecystectomy
- · Gastric Erosions
- · Benign and Malignant Neoplasm
#### Gynecology
Open and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
### Compliance to Voluntary Standards
The design of the SOLTIVE Laser System complies with the following standards:
| Standard | Title |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN ISO 14971:2019 | Medical devices -- Application of risk management to medical<br>devices |
| ISO 15223-1:2016 | Medical devices -- Symbols to be used with medical device<br>labels, labeling and information to be supplied -- Part 1:<br>General requirements |
| IEC 60601-<br>1:2005+A1:2012 | Medical electrical equipment - Part 1: General requirements<br>for basic safety and essential performance |
| IEC 60601-1-2:2014 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility -<br>Requirements and tests |
| IEC 60825-1:2014 | Safety of laser products - Part 1: Equipment classification<br>and requirements |
| IEC 62366-1:2015 | Medical devices - Part 1: Application of usability<br>engineering to medical devices |
| IEC 60601-2- | Medical electrical equipment - Part 2-22: Particular |
| Standard | Title |
| 22:2007+A1:2012 | requirements for basic safety and essential performance of<br>surgical, cosmetic, therapeutic and diagnostic laser<br>equipment |
| IEC 60601-1-<br>6:2010+A1:2013 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Usability |
| IEC<br>62304:2006+A1:2015 | Medical device software - Software life cycle processes |
| ISO 11607-1:2019 | Packaging for terminally sterilized medical devices -- Part 1:<br>Requirements for materials, sterile barrier systems and<br>packaging systems |
| ISO 11135:2014 | Sterilization of health care products -- Ethylene oxide:<br>Requirements for development, validation and routine<br>control of a sterilization process for medical devices |
| ISO 10993-1:2018 | Biological evaluation of medical devices -- Part 1:<br>Evaluation and testing |
| ISO 10993-5:2009 | Biological evaluation of medical devices -- Part 5: Tests for<br>in vitro cytotoxicity |
| ISO 10993-<br>7:2008+COR1:2009 | Biological evaluation of medical devices -- Part 7: Ethylene<br>oxide sterilization residuals |
| ISO 10993-10:2010 | Biological evaluation of medical devices -- Part 10: Tests<br>for irritation and delayed-type hypersensitivity |
| ISO 10993-11:2017 | Biological evaluation of medical devices -- Part 11: Tests<br>for systemic toxicity |
| ASTM F 1980-16 | Standard Guide for Accelerated Aging of Sterile Barrier<br>Systems for Medical Devices |
| ASTM<br>F1886/F1886M:2016 | Standard Test Method for Determining Integrity of Seals for<br>Flexible Packaging by Visual Inspection |
| ASTM F88-<br>15/F88M-15 | Standard Test Method for Seal Strength of Flexible Barrier<br>Materials |
| ISO 17665-<br>1:2006/R2013 | Sterilization of health care products -- Moist heat -- Part 1:<br>Requirements for the development, validation and routine<br>control of a sterilization process for medical devices |
| ISO<br>14937:2009/R2013 | Sterilization of health care products -- General requirements<br>for characterization of a sterilizing agent and the<br>development, validation and routine control of a sterilization<br>process for medical devices |
| ISTA Procedure 3A<br>2018 | Packaged Products for Parcel Delivery System Shipment 70kg<br>(150 lb) or less |
| ASTM D4169-16 | Performance Testing of Shipping Containers and Systems |
| ASTM F2096-11 | Standard Test Method for Detecting Gross Leaks in Porous<br>Medical Packaging by Internal Pressurization (Bubble Leak) |
{9}------------------------------------------------
{10}------------------------------------------------
#### Summary of Sterilization and Shelf-Life Discussion
SOLTIVE SuperPulsed Laser Fibers Single-use and SOLTIVE SuperPulsed Laser Fibers Reusable are provided sterile. The devices are sterilized by Ethylene Oxide to a provide sterility assurance level of 10-6.
The Shelf-Life period for the Single-use fibers was determined via testing and through an analysis of the shelf-life stability of the materials used in the design of the device, as well as an analysis of the packaging materials and processes used with other Gyrus ACMI devices. Shelf-life studies are on file to support the labeled shelf life.
#### Summary of All Performance Testing (bench, clinical, non-clinical)
Software verification tests were conducted in support of the subject software modification. Testing demonstrated that all performance requirements met the prescribed acceptance criteria.
#### Substantial Equivalence
Substantial equivalence is demonstrated by acknowledged verification/validation activities. The subject devices have identical technology, performance, dimensions, and materials. The differences to the predicate device SOLTIVE Laser Systems are:
- Revised software to update the graphical user interface to add a Ureteral Stone . preset to the menu of Lithotripsy presets and other enhancements.
- . Updated Instructions for Use to align with updated graphical user interface.
{11}------------------------------------------------
| Gyrus ACMI SOLTIVE Laser System | | | |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Design<br>Feature | Proposed<br>(Software Version 2.1) | Predicate K183647<br>(Software Version 2.0) | Comparison |
| Intended Use | The SOLTIVE Laser<br>System is intended for use<br>in surgical procedures such<br>as open laparoscopic and<br>endoscopic, involving<br>endoscopic ablation,<br>vaporization, excision,<br>incision, resection,<br>coagulation and hemostasis<br>of soft tissue in medical<br>specialties including:<br>Urology, lithotripsy,<br>gastroenterology,<br>arthroscopy, discectomy,<br>gynecology, ENT and<br>general surgery. | The SOLTIVE Laser<br>System is intended for use<br>in surgical procedures such<br>as open laparoscopic and<br>endoscopic, involving<br>endoscopic ablation,<br>vaporization, excision,<br>incision, resection,<br>coagulation and hemostasis<br>of soft tissue in medical<br>specialties including:<br>Urology, lithotripsy,<br>gastroenterology,<br>arthroscopy, discectomy,<br>gynecology, ENT and<br>general surgery. | Identical |
| Function of<br>Device | Thulium Laser<br>generator with fiber<br>optic delivery | Thulium Laser<br>generator with fiber<br>optic delivery | Identical |
| Mechanics of<br>Action | Light/energy absorption<br>by target tissues | Light/energy absorption<br>by target tissues | Identical |
| Laser Source | Thulium | Thulium | Identical |
| Accessories | Surgical fibers (reusable<br>and single use) available in<br>following nominal core<br>diameters (150, 200, 365,<br>550, 940 μm) | Surgical fibers (reusable and<br>single use) available in<br>following nominal core<br>diameters (150, 200, 365,<br>550, 940 μm) | Identical |
| Biocompatible | Yes | Yes | Identical |
| Sterilization<br>(Single-use<br>fibers) | Ethylene Oxide 10-6 | Ethylene Oxide 10-6 | Identical |
| Single Use &<br>Reusable<br>Fibers | Yes | Yes | Identical |
### Equivalence Comparison Table:
#### Conclusion:
In summary, the Gyrus ACMI SOLTIVE Laser System is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.