K211395 · Einstein Works, LLC · FMI · Jul 28, 2021 · General Hospital
Device Facts
Record ID
K211395
Device Name
NIO-I
Applicant
Einstein Works, LLC
Product Code
FMI · General Hospital
Decision Date
Jul 28, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The NIO-I is an intraosseous device intended to provide vascular access in the proximal tibia of infants as an alternative to emergency IV access. It is indicated for use in infants between gestational age of 36 weeks and weight of at least 5 pounds (2.3 Kg) and up to 3 years of age.
Device Story
Manual intraosseous infusion device; provides vascular access in proximal tibia for fluid/drug delivery. Device consists of handle, needle gripper, needle housing, trocar needle (trocar + cannula), and needle cover. Operation: manual insertion through bone cortex until haptic feedback from stepped-needle design indicates penetration of medullary cavity; handle disconnected; trocar removed; cannula remains in situ. Used in hospital, clinic, or emergency care settings by healthcare providers. Provides alternative to emergency IV access; facilitates rapid infusion.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Manual intraosseous device. Materials: Stainless steel 302 (trocar), 316L (cannula); Polycarbonate Makrolon Rx2530 and ABS PA757 (handle). 15mm needle length, 18G penetration/14G cannula. Luer Lock hub. Sterilization: Gamma irradiation. Complies with ISO 10993-1, ISO 594-1/2, ISO 7864, ISO 9626, ISO 11137, ISO 11607, IEC 62366-1, ASTM F1980.
Indications for Use
Indicated for infants (gestational age ≥36 weeks, weight ≥2.3 kg, up to 3 years old) requiring vascular access in the proximal tibia as an alternative to emergency IV access.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG ADMINISTRATION" in blue as well. The FDA is a federal agency responsible for regulating and supervising the safety of food, drugs, and other products.
July 28, 2021 Einstein Works LLC Roy Bachrach Managing Director 5312 Elm St Houston, Texas 77081
Re: K211395
Trade/Device Name: NIO-I Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI, MHC Dated: April 28, 2021 Received: May 5, 2021
Dear Roy Bachrach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Alan Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211395
Device Name NIO-I
Indications for Use (Describe)
The NIO-I is an intraosseous device intended to provide vascular access in the proximal tibia of infants as an alternative to emergency IV access. It is indicated for use in infants between gestational age of 36 weeks and weight of at least 5 pounds (2.3 Kg) and up to 3 years of age.
Type of Use (Select one or both, as applicable)
| <div> <span style="text-decoration: underline;">X</span> Prescription Use (Part 21 CER 801 Subpart D) </div> |
|-----------------------------------------------------------------------------------------------------------------|
| <div> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
Einstein Works, LLC. info@einsteinworksllc.com www.einsteinworksllc.com
Image /page/3/Picture/1 description: The image shows the logo for Einstein Works. On the left is a red graphic of a stylized letter 'e'. To the right of the graphic is the text 'EINSTEIN' on the top line and 'WORKS' on the bottom line. The text is in a dark blue font.
#### 510(K) SUMMARY FOR EINSTEIN WORKS' NIO-I K211395
#### 1. Submitter Information
Einstein Works LLC. 5312 Elm St. Houston, TX 77081 Phone: 001-713-7236000 E mail: roy@ps-med.com
Contact person name: Mr. Roy Bachrach - Managing Director Phone: 001-713-7236000 Email: roy@ps-med.com
#### 2. Date of Preparation July 28, 2021
#### 3. Identification of Predicate
Waismed LTD's NIO-I, Intraosseous infusion device cleared under 510(k) number K190538 on July 10, 2019.
## 4. Identification of Subject Device
a. Regulatory Information Regulation Name: Needle, Interosseous Regulation Number: 880.5570 Product Code: Primary: FMI Secondary: MHC Common Name: Intraosseous infusion device Proprietary/Trade name: NIO-I
..............................................................................................................................................................................
..............................................................................................................................................................................
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Einstein Works. The logo consists of a stylized letter 'e' in red, followed by the words 'EINSTEIN' and 'WORKS' stacked on top of each other in blue. The 'e' is formed by two curved lines that create an oval shape, and there is a thin red line underneath the entire logo.
#### 5. Device Description
The NIO-I is a manual intraosseous device. It includes a lightweight ergonomic handle for easy and fast manual insertion of the needle into the medullary cavity. The device is designed specifically for infants between 36 weeks and weight of at least 5 pounds (2.3 Kg) and up to 3 years of age. The NIO-I provides haptic feedback via the steppedneedle design that helps preventing over-penetration into the distal cortex. After insertion of the needle, the user disconnects the handle and pulls out the trocar leaving the cannula inside the medullary cavity. Through the cannula, the caregiver can infuse fluids and drugs that will reach the vascular system.
#### 6. Indications for Use Statement
| Characteristics | Subject Device | Predicate Device |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Einstein Works LLC NIO-I | Waismed Ltd NIO-I<br>K190538 |
| Indications for Use | The NIO-I is an intraosseous device intended to provide vascular access in the proximal tibia of infants as an alternative to emergency IV access. It is indicated for use in infants between gestational age of 36 weeks and weight of at least 5 pounds (2.3 Kg) and up to 3 years of age. | The NIO-I is a intraosseous device intended to provide vascular access in the proximal tibia of infants as an alternative to emergency IV access. It is indicated for use in infants between gestational age of 36 weeks and weight of at least 5 pounds (2.3 Kg) and up to 3 years of age. |
| Prescription Only<br>or Over the<br>counter | Prescription | Prescription |
#### Discussion of Substantial Equivalence of in Indications for Use statement:
Both devices have the exact same intended use, indications and clinical purpose.
Therefore, in terms of indications for use statement, the NIO-I is substantially equivalent to the selected predicate device.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Einstein Works. The logo consists of a stylized red oval shape on the left, followed by the words "EINSTEIN" and "WORKS" stacked on top of each other in blue. A thin red line is present at the bottom of the logo.
# 7. Summary of Technological Characteristics
| Item | Subject Device | Predicate Device | Comment |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Einstein Works LLC NIO-<br>I | Waismed Ltd NIO-I<br>K190538 | |
| Target<br>Population | Emergency Care for infant<br>patients (birth to 29 days<br>with minimal limit of<br>gestational age of 36<br>weeks and weight of 5lb<br>[2.3Kg]). | Emergency Care for infant<br>patients (birth to 29 days<br>with minimal limit of<br>gestational age of 36<br>weeks and weight of 5lb<br>[2.3Kg]). | Equivalent |
| Anatomical Sites | Proximal Tibia. | Proximal Tibia. | Equivalent |
| Environment<br>Used | Hospital, Clinic,<br>Emergency Care. | Hospital, Clinic,<br>Emergency Care. | Equivalent |
| Design | Consists of: handle, needle<br>gripper, needle housing,<br>trocar needle (trocar +<br>cannula) and needle cover,<br>allowing intraosseous<br>access for the delivery of<br>drugs and fluids. | Consists of: handle, needle<br>gripper, needle housing,<br>trocar needle (trocar +<br>cannula) and needle cover,<br>allowing intraosseous<br>access for the delivery of<br>drugs and fluids. | Equivalent |
| Mechanism of<br>Action | Manually in three stages:<br>(1) The NIO-I needle is<br>manually inserted through<br>the cortex of the bone until<br>change is resistance is felt,<br>indicating the needle<br>penetrated the bone<br>marrow cavity.<br>(2) The insertion handle is<br>then disconnected from the<br>needle and, (3) the trocar<br>is removed by pulling it<br>upwards. | Manually in three stages:<br>(1) The NIO-I needle is<br>manually inserted through<br>the cortex of the bone until<br>change is resistance is felt,<br>indicating the needle<br>penetrated the bone<br>marrow cavity.<br>(2) The insertion handle is<br>then disconnected from the<br>needle and, (3) the trocar<br>is removed by pulling it<br>upwards. | Equivalent |
| Over-<br>Penetration | Stepped-needle Exists to<br>provide sensory feedback | Stepped-needle Exists to<br>provide sensory feedback | Equivalent |
| Item | Subject Device | Predicate Device | Comment |
| | Einstein Works LLC NIO-I | Waismed Ltd NIO-I K190538 | |
| Safety Mechanisms | upon cortex penetration to<br>prevent over-penetration. | upon cortex penetration to<br>prevent over-penetration. | |
| Insertion Site | In proximal tibia | In proximal tibia | Equivalent |
| Insertion Site Identification | Extending the patient's leg<br>at the tibia site, the<br>insertion site is located<br>approximately 1cm medial<br>to the tibial tuberosity, or<br>just below the patella<br>(approximately 1cm or one<br>finger width) and slightly<br>medial (approximately<br>1cm or one finger width),<br>along the flat aspect of the<br>tibia. Then insertion takes<br>place. | Extending the patient's leg<br>at the tibia site, the<br>insertion site is located<br>approximately 1cm medial<br>to the tibial tuberosity, or<br>just below the patella<br>(approximately 1cm or one<br>finger width) and slightly<br>medial (approximately<br>1cm or one finger width),<br>along the flat aspect of the<br>tibia. Then insertion takes<br>place. | Equivalent |
| Materials | Needle:<br>Stainless steel 302<br>(Trocar), stainless steel<br>316L (Cannula)<br><br>Plastic components:<br>Polycarbonate Makrolon<br>Rx2530 and ABS PA757<br>(Handle only, no body<br>contact with the patient). | Needle:<br>Stainless steel 302<br>(Trocar), stainless steel<br>316L (Cannula)<br><br>Plastic components:<br>Polycarbonate Makrolon<br>Rx2530 and ABS PA757<br>(Handle only, no body<br>contact with the patient). | Equivalent |
| Biocompatibility | Biocompatible. Complies<br>with ISO 10993-1 | Biocompatible. Complies<br>with ISO 10993-1 | Equivalent |
| Needle dwelling time | <24 Hours | <24 Hours | Equivalent |
| Hub Interface | The cannula hub is a<br>standard hub Luer Lock | The cannula hub is a<br>standard hub Luer Lock | Equivalent |
| Item | Subject Device | Predicate Device | Comment |
| | Einstein Works LLC NIO-<br>I | Waismed Ltd NIO-I<br>K190538 | |
| | appropriate for connecting<br>to any standard infusion<br>system. | appropriate for connecting<br>to any standard infusion<br>system. | |
| Needle length | 15 mm (0.6") with stepped<br>needle for pediatric use. | 15 mm (0.6") with stepped<br>needle for pediatric use. | Equivalent |
| Needle gauge | 18G penetration<br>(and 14G above the needle<br>step) | 18G penetration<br>(and 14G above the needle<br>step) | Equivalent |
| Single use | Single use | Single use | Equivalent |
| Sterilization | Sterile | Sterile | Equivalent |
| Sterilization<br>method | Gamma irradiation | Gamma irradiation | Equivalent |
| Sterile<br>Packaging | Sterile barrier | Sterile barrier | Equivalent |
# Table 1: Technological Comparison with the Predicates
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Einstein Works. The logo consists of a red elliptical shape on the left and the words "EINSTEIN WORKS" in blue on the right. A thin red line is located underneath the logo.
Einstein Works, LLC. info@einsteinworksllc.com www.einsteinworksllc.com
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Einstein Works. The logo consists of a stylized letter "e" in red on the left, followed by the words "EINSTEIN" and "WORKS" stacked on top of each other in blue. A thin red line is present below the text.
Einstein Works, LLC. info@einsteinworksllc.com www.einsteinworksllc.com
Discussion of Substantial Equivalence of in Technological Characteristics: Both devices have the exact same technological characteristics and principle of operation. Therefore, there are no differences raising any questions of safety and effectiveness.
#### 8. Non-Clinical Performance Testing
Body contact materials were evaluated for biocompatibility in accordance with FDA's Guidance for Use of ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process", dated September 04, 2020 and ISO 10993 international standards series as detailed below. Biocompatibility tests included, sensitization, hemolysis, pyrogenicity, cytotoxicity, acute system toxicity and intracutaneous reactivity.
The biocompatibility tests were conducted to verify that the proposed device is not adverse to human tissue based on the following standards:
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Einstein Works. The logo consists of a stylized letter "e" in red, followed by the words "EINSTEIN" and "WORKS" stacked on top of each other in blue. The logo is simple and modern, and the colors are eye-catching.
- ISO 10993-1:2018 Biological evaluation of medical devices Part 1: A Evaluation and testing within a risk management process.
- > ISO 10993-4:2017 Biological evaluation of medical devices Part 4: Selection of tests for interactions with blood.
- > ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization.
- > ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for systemic toxicity.
Other non-clinical tests were conducted to verify that the proposed device met all design specifications and is Substantially Equivalent (SE) to the predicate device. Tests identified under subsections 'a', 'c', 'd', 'e', 'f', 'h' below are based on the company's internal V&V tests' plan, as following described.
Tests identified under subsections 'b', 'g', 'i', 'j', 'k', 'l', 'm' are in compliance with applicable standards as follows:
- a) Force required for needle cover removal;
- b) Visual inspection for corrosion, per ISO 9626;
- c) Force required for needle insertion;
- d) Minimal force for insertion of the needle;
- e) Force for detachment of the needle housing from the needle gripper;
- f) Maximum force for trocar removal after insertion to the platform;
- g) Cannula effective length, per ISO 9626 and ISO 7864;
- h) Maximum force for cannula removal after insertion;
- Usability study, per IEC 62366-1; i)
- Shelf life, per ISO 11607-1 and ASTM F1980-16; i)
- k) Sterilization validation, per ISO 11137-1, ISO 11137-2, ISO 13004, ANSI AAMI ST72:
- Limits for acidity or alkalinity, per ISO 7864; I)
- m) Conical fitting, per ISO 594-1 and ISO 594-2.
The test results demonstrated that the proposed device complies with the following standards:
- > ISO 594-1:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General requirements.
- > ISO 594-2:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock fittings.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Einstein Works. The logo consists of a stylized red orbit shape on the left and the words "EINSTEIN WORKS" in blue on the right. The word "EINSTEIN" is above the word "WORKS".
- ISO 11137-1:2006 Sterilization of health care products Radiation Part 1: A Requirements for development, validation and routine control of a sterilization process for medical devices.
- A ISO 11137-2:2013 - Sterilization of Health Care Products - Radiation - Part 2: Establishing the Sterilization Dose.
- ISO 13004:2013 Sterilization of health care products Radiation -A Substantiation of selected sterilization dose: Method VDmaxSD.
- > ANSI AAMI ST72:2011/(R)2016 Bacterial endotoxins -Test methods, routine monitoring, and alternatives to batch testing.
- > IEC 62366-1:2015 Medical devices Part 1: Application of usability engineering to medical devices.
- A ASTM F1980-16 - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
- ISO 11607-1:2019 Packaging for terminally sterilized medical devices Part 1: A Requirements for materials, sterile barrier systems and packaging systems.
- > ISO 7153-1:1999 Surgical instruments Metallic materials Part 1: Stainless steel.
- > ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods.
- > ISO 7864:2016 Sterile hypodermic needles for single use Requirements and test methods.
All tests' results support Einstein Works LLC labeling claims in order to establish substantial equivalency.
#### 9. Clinical Test Conclusion
No clinical Study is included in this submission.
## 10. Substantially Equivalent (SE) Conclusion
..............................................................................................................................................................................
The evaluation of the subject device performance demonstrates that it is as safe and as effective as the legally marketed predicate device.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.