Blue Nitrile Exam Gloves

K211341 · Jiangxi Surefine Medical Co., Ltd. · LZA · Jul 29, 2021 · General Hospital

Device Facts

Record IDK211341
Device NameBlue Nitrile Exam Gloves
ApplicantJiangxi Surefine Medical Co., Ltd.
Product CodeLZA · General Hospital
Decision DateJul 29, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

The Blue Nitrile Exam Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner.

Device Story

Blue Nitrile Exam Gloves are disposable, non-sterile, ambidextrous gloves. Worn by examiners to provide a barrier against contamination between patient and examiner. Manufactured from nitrile rubber. Available in sizes S, M, L, and XL. Used in clinical or medical settings. No electronic components or software.

Clinical Evidence

No clinical data. Bench testing only. Testing included biocompatibility (ISO 10993-10 irritation/sensitization, ISO 10993-5 cytotoxicity), residual powder (ASTM D6124), freedom from holes (ASTM D5151), and physical properties/dimensions (ASTM D6319). All results met acceptance criteria.

Technological Characteristics

Material: Nitrile. Form factor: Ambidextrous, non-sterile, powder-free. Dimensions: S-XL sizes, finger/palm thickness ≥0.05mm. Standards: ASTM D6124 (residual powder), ASTM D5151 (holes), ASTM D6319 (nitrile exam gloves).

Indications for Use

Indicated for use as a disposable medical glove worn on the examiner's hands to prevent cross-contamination between patient and examiner. Suitable for use by healthcare professionals and for over-the-counter use.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA seal. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". July 29, 2021 Jiangxi Surefine Medical Co., Ltd. % Boyle Wang General Manager Shanghai Truthful Information Technology Co., Ltd. Room608, No.738, Shangcheng Rd., Pudong Shanghai, 200120 China Re: K211341 Trade/Device Name: Blue Nitrile Exam Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA Dated: July 6, 2021 Received: July 12, 2021 Dear Boyle Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Clarence W. Murray, III, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K211341 Device Name Blue Nitrile Exam Gloves Indications for Use (Describe) The Blue Nitrile Exam Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary This summary of 510(k) is being submitted in accordance with 21 CFR 807.92. #### 1.0 submitter's information Name: Jiangxi Surefine Medical Co., Ltd. Address: Jinshan Rd., Yugan Hi-tech Industrial Park, Shangrao,Jiangxi Province, 335100, China Phone Number: +86-13805860705 Contact: Steve Zhu Date of Preparation: 2021.04.21 ## Designated Submission Correspondent Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Tel: +86-21-50313932 Email: Info@truthful.com.cn # 2.0 Device information Trade name: Blue Nitrile Exam Gloves Common name: Patient Examination Gloves Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL #### 3.0 Classification Production code: LZA Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital #### 4.0 Predicate device information Ever Global (Vietnam) Enterprise Corp Manufacturer: Device: Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color 510(k) number: K171422 {4}------------------------------------------------ #### 5.0 Intended use The Blue Nitrile Exam Gloves are disposable devices intended for medical purposes that are worn on the examiner's hands to prevent contamination between patient and examiner. #### 6.0 Device description The proposed device is Powder Free Blue Nitrile Exam Gloves. The proposed device is blue. The design of proposed device is addressing the standards as ASTM D6124,ASTM D5151, and ASTM D6319. The proposed device is non-sterile. #### 7.0 Technological Characteristics Comparison | Item | Proposed device | Predicated device | Comparison | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | 510(k) number | K211341 | K171422 | | | Product Code | LZA | LZA | Same | | Regulation No. | 21CFR880.6250 | 21CFR880.6250 | Same | | Class | I | I | Same | | Intended Use | The Blue Nitrile Exam Gloves is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | The Disposable Powder Free Nitrile Examination Glove, White/ Blue/ Black/ Pink Color is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner. | Same | | Powdered or Powered free | Powdered free | Powdered free | Same | | Design Feature | ambidextrous | ambidextrous | Same | | Labeling Information | Single-use indication, powder free, device color, device name, glove size and quantity, Blue Nitrile Exam Gloves, Non-Sterile | Single-use indication, powder free, device color, device name, glove size and quantity, Disposable Powder Free Nitrile Examination Glove, Non-Sterile | Same | #### Table1-General Comparison ## Table2 Device Dimensions Comparison | Predicate | Designation | Size | | | | | Tolerance | |-----------------|-------------|------|---|---|---|----|-----------| | Device(K171422) | | XS | S | M | L | XL | | {5}------------------------------------------------ | | Length, mm | 230 | 230 | 230 | 230 | 230 | min | | |-----------------|----------------|------|-----|-----|-----|-----|-----------|--| | | Width, mm | 75 | 85 | 95 | 105 | 115 | $\pm 5$ | | | | Thickness, mm: | | | | | | | | | | Finger | 0.05 | | | | | min | | | | Palm | 0.05 | | | | | min | | | Proposed Device | Designation | Size | | | | | Tolerance | | | | | S | M | L | XL | | | | | | Length, mm | 220 | 230 | 230 | 230 | | min | | | | Width, mm | 80 | 95 | 110 | 120 | | $\pm 10$ | | | | Thickness, mm: | | | | | | | | | | Finger | 0.05 | | | | | min | | | | Palm | 0.05 | | | | | min | | | Remark | Analysis 1 | | | | | | | | ## Table3 Performance Comparison | Item | Proposed device | Predicated device | Remark | | | |------------------------|---------------------------------------------------------------------------------|---------------------------------------------------------------------------|----------------------------------------|------------|------| | Colorant | blue | White/ Blue/ Black/ Pink | Analysis2 | | | | Physical<br>Properties | Before<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME | | | | Ultimate<br>Elongation | 500%min | 500%min | SAME | | | After<br>Aging | Tensile<br>Strength | 14MPa, min | 14MPa, min | SAME | | | | Ultimate<br>Elongation | 400%min | 400%min | SAME | | | | Comply with ASTM D6319 | Comply with ASTM D6319 | SAME | | | Freedom from Holes | Be free from holes<br>when tested in<br>accordance with<br>ASTMD5151<br>AQL=2.5 | Be free from holes when<br>tested in accordance with<br>ASTMD5151 AQL=2.5 | SAME | | | | | Powder Content | 0.03 | Meet the requirements of<br>ASTM D6124 | SIMILAR | | Analysis1: The proposed device has different sizes to the predicate device, but all proposed devices are conducted the properties test, the test results shown that the sizes comply with the requirements of standard ASTM D6319-19, Standard Specification For Blue Nitrile Exam Gloves For Medical Application. Analysis2: The proposed device has different color to the predicate device, but all proposed devices are conducted the biocompatibility and performance tests were comparable. {6}------------------------------------------------ | Item | | Proposed device | Predicated device | Remark | |--------------------|---------------|--------------------------------------------------------------------------------|-------------------------|---------| | Material | | Nitrile | Nitrile | SAME | | Biocompatibility | Irritation | Under the conditions of the study, not an irritant | Comply with ISO10993-10 | SAME | | | Sensitization | Under conditions of the study, not a sensitizer. | | | | | Cytotoxicity | Under conditions of the study, did not show potential toxicity to L-929 cells. | Comply with ISO10993-5 | SIMILAR | | Label and Labeling | | Meet FDA's Requirement | Meet FDA's Requirement | SAME | # Table4 Biocompatibility Testing Comparison # 8.0 Summary of Non-Clinical Performance Testing The following performance data has been provided to demonstrate that the subject device meets the acceptance criteria in the standard. # Table 5 Summary of Non-Clinical Performance Testing | No. | Name of the Test<br>Methodology / Standard | Purpose | Acceptance Criteria | Results | |-----|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | ISO 10993-10:2010<br>Biological Evaluation Of<br>Medical Devices - Part<br>10: Tests For Irritation<br>And Skin Sensitization. | This part of ISO<br>10993 assesses<br>possible contact<br>hazards from<br>chemicals released from<br>medical devices, | Skin Sensitization<br>Test:<br>provided grades less than 1,<br>otherwise<br>sensitization. | All grades are 0.<br><br>All animals survived and no<br>abnormal signs were observed<br>during the study. | | 2 | | which may<br>produce skin and<br>mucosal irritation,<br>eye irritation or<br>skin sensitization. | Skin Irritation Test:<br>If the primary irritation<br>index is 0-0,4, the<br>response category is<br>Negligible.<br>0,5-1,9 means slight<br>2-4,9 means moderate<br>5-8 means severe | The primary irritation index is 0.<br>The response of the proposed<br>device was categorized as<br>negligible under the test<br>condition | | | | | | | | 3 | ISO 10993-5:2009<br>Biological Evaluation Of<br>Medical Devices - Part<br>5: Tests For In Vitro<br>Cytotoxicity | This part of ISO<br>10993 describes<br>test methods to<br>assess the in vitro<br>cytotoxicity of<br>medical devices. | The viab.% of the<br>100% extract of the<br>test article is the final<br>result, and if viability is<br>reduced to <70% of<br>the blank, it has<br>cytotoxic potential. | Viab.% of 100% test article<br>extract is 86.5%<br>It means the proposed device<br>have no potential toxicity to L-<br>929 in the MTT method | | 4 | ASTM D6124-06<br>(Reapproved 2017),<br>Standard Test Method<br>for Residual Powder on<br>Medical Gloves | This standard is<br>designed to<br>determine the<br>amount ofresidual<br>powder (or<br>filter-retained<br>mass) found<br>on medical gloves | powder residue limit of<br>2.0 mg | 0.03mg | | 5 | ASTM<br>D5151-06(Reapproved2<br>015), Standard Test<br>Method for Detection of<br>Holes in Medical Gloves. | This test method<br>covers the<br>detection of holes<br>in<br>medical gloves. | Samples number: 125<br>gloves<br>AQL: 2.5 (ISO 2859)<br>Criterion ≤ 7 gloves<br>for water leakage | no glove water leakage found | | 6 | ASTM<br>D6319-10(Reapproved<br>2015),Standard<br>Specification For Nitrile<br>Examination Gloves For<br>Medical Application. | This specification<br>covers certain<br>requirements for<br>nitrile rubber<br>gloves used in<br>conducting<br>medical<br>examinations and<br>diagnostic and<br>therapeutic<br>procedures. | Sterility: no need<br>Freedom from holes:<br>pl. Refer to No. 5 in<br>table 5<br>Dimensions:<br>S: width 80±10mm<br>Length ≥220 mm<br>M: width 95±10mm<br>Length ≥230 mm<br>L: width 110± 10mm<br>Length ≥230 mm<br>XL: width 120±10mm<br>Length ≥230 mm<br>Thickness:<br>Finger ≥0.05 mm<br>Palm ≥0.05 mm<br>Physical properties:<br>Before aging<br>Tensile strength | N.A.<br>Please refer to No. 5 in table 5<br>Dimensions:<br>S: width: 83-86 mm<br>Length 253-266 mm<br>M: width 96-98 mm<br>Length 243-262 mm<br>L: width 106-110 mm<br>Length 247-254 mm<br>XL: width 112-118 mm<br>Length 245-252 mm<br>Thickness:<br>Finger 0.09-0.11 mm<br>Palm 0.08-0.11 mm<br>Physical properties:<br>Before aging<br>Tensile strength 15.7-17.7 MPa<br>Ultimate Elongation 532.284% -<br>552.072% | {7}------------------------------------------------ {8}------------------------------------------------ | | Ultimate Elongation ≥<br>500%<br>After Accelerated<br>Aging<br>Tensile strength ≥<br>14MPa<br>Ultimate Elongation ≥<br>400%<br><br>Powder-free Residue:<br>pl. Refer to No. 4 in<br>table 5 | Tensile strength 15.2-17.8 MPa<br>Ultimate Elongation 525.947% -<br>548.352%<br><br>Powder-free Residue:<br>pl. Refer to No. 4 in table 5 | |--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| |--|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| # 9. Summary of Clinical Performance Test No clinical study is included in this submission. # 10.0 Conclusion The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...