The iovera? system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The ioveraº system is not indicated for treatment of central nervous system tissue.
Device Story
Portable cryogenic surgical device; uses nitrous oxide (N2O) to cool a stainless steel Smart Tip via Joule-Thomson effect/latent heat of vaporization. System comprises battery-powered handpiece, charging dock, N2O cartridge, and single-patient-use Smart Tips. Physician inserts Smart Tip into target tissue; initiates cooling cycle via handpiece control. Handpiece provides user feedback via LCD display. Device freezes surrounding tissue to destroy it or create peripheral nerve lesions for pain blocking. Used in clinical settings; provides localized cryotherapy to manage pain or perform surgical tissue destruction. Benefits include non-pharmacological pain relief and precise tissue targeting.
Clinical Evidence
Bench testing only. Design verification and validation testing included functional testing, product performance (handpiece and cartridge), Smart Tip testing, user interface testing, and electrical/EMC safety testing. All tests met predetermined acceptance criteria.
Indicated for patients requiring tissue destruction during surgery, peripheral nerve lesioning for pain management, or relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. Contraindicated for central nervous system tissue.
Regulatory Classification
Identification
A cryogenic surgical device is a device used to destroy nervous tissue or produce lesions in nervous tissue by the application of extreme cold to the selected site.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it. The full name reads "U.S. Food & Drug Administration".
September 10, 2021
Pacira Pharmaceuticals Inc. Kent Jones Senior Director QA/RA 46400 Fremont Blvd Fremont, California 94538
Re: K211334
Trade/Device Name: Iovera System Regulation Number: 21 CFR 882.4250 Regulation Name: Cryogenic Surgical Device Regulatory Class: Class II Product Code: GXH Dated: August 10, 2021 Received: August 11, 2021
Dear Kent Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211334
Device Name Iovera System
### Indications for Use (Describe)
The iovera? system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The ioveraº system is not indicated for treatment of central nervous system tissue.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| <div><span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Pacira Pharmaceuticals, Inc. The logo is in purple and gray. The words "PACIRA" and "PHARMACEUTICALS, INC." are in large, bold letters. Below that, in smaller letters, it says "A Subsidiary of Pacira BioSciences, Inc."
# 510(k) Summary
## a. Date Summary Prepared:
Date: 09/10/2021
| 510(k) Type: | Special |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Name of Device: | iovera° System |
| Device Class: | Class II |
| Product Code: | GXH-Cryogenic Surgical device |
| Regulation: | 21 CFR Part 882.4250 |
| Sponsor/Submitter: | Pacira Pharmaceuticals, Inc.<br>46400 Fremont Blvd<br>Fremont, CA 94538<br>Ph: 510-933-1500<br>Fax: 510-933-1501 |
| Establishment Registration: | 3009131456 |
| Correspondent Contact<br>Information: | Kent Jones,<br>Senior Director QA/RA<br>Pacira Pharmaceuticals Inc.<br>Email: kent.jones@pacira.com<br>Ph: 510-933-1513 |
## b. Predicate Device Information:
The iovera° system is substantially equivalent to the following currently legally marketed device:
| 510(k) Number | Product | Sponsor |
|---------------|----------------|------------------------|
| K161835 | iovera° system | Pacira Pharmaceuticals |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Pacira Pharmaceuticals, Inc. The logo features the company name in a stylized font, with the letters "PACIRA" in a larger, bolder font than the rest of the name. Above the company name is a curved line. Below the company name is the text "A Subsidiary of Pacira BioSciences, Inc."
## ioveraº system
#### Description of Device: ﻥ
The iovera° system is a portable cryogenic surgical device used to destroy tissue and/or produce lesions in nervous tissue through application of extreme cold to the selected site. The device is based on introduction of a Smart Tip internally cooled by the cryogenic fluid (nitrous oxide, N₂O) to a selected area. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization.
### Device Design
The device comprises four main components:
- . A reusable Handpiece
- . A Charging Dock
- An assortment of single-patient-use Smart Tips
- . A Cartridge containing nitrous oxide
The iovera® Handpiece is battery powered and provides feedback to the user during device preparation and use. The Handpiece connects to both the Cartridge and to the Smart Tip. The user activates a treatment cycle through a control on the Handpiece, which starts and stops the treatment. The Handpiece contains an LCD display for providing feedback to the user when the device is ready to use. The Charging Dock stores the Handpiece between uses and provides power for charging the battery.
An assortment of Smart Tips is available for the iovera° system. All Smart Tip needles are made of stainless steel and have a closed-end design that fully contains the cryogen so that it does not enter the target tissue. The Smart Tip is the only patient contacting component of the iovera° system. The user removes the Smart Tip from its sterile packaging and attaches it to the Handpiece immediately before use.
The iovera° system uses a commercially available nitrous oxide cylinder. The Cartridge is filled with pure N₂O.
## Device Functionality/Scientific Concepts:
The device functionality is based on the user introducing the Smart Tip into the selected treatment area or the target nervous tissue. The user then initiates the flow of cryogen by pressing the on/off button. Liquid cryogen flows from the Handpiece into the closed-end Smart Tip. The Smart Tip is cooled by the Joule-Thomson Effect and/or latent heat of vaporization; as the liquid cryogen expands into a gas, the temperature drops around the external surface of the Smart Tip, causing the surrounding tissue to freeze. The treatment is completed after a pre-programmed amount of time, at which time the user can safely remove the Smart Tip from the treatment site.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Pacira Pharmaceuticals, Inc. The logo is in purple and features the company name in a bold, sans-serif font. Above the company name is a curved line. Below the company name is the text "A Subsidiary of Pacira BioSciences, Inc."
## d. Indications for Use:
The iovera system is used to destroy tissue during surgical procedures by applying freezing cold. It can also be used to produce lesions in peripheral nervous tissue by the application of cold to the selected site for the blocking of pain. It is also indicated for the relief of pain and symptoms associated with osteoarthritis of the knee for up to 90 days. The iovera° system is not indicated for treatment of central nervous system tissue.
## e. Technological Characteristics Comparison to Predicate Device:
The iovera° system is substantially equivalent in intended use, technology, design, and materials to the above listed legally marketed predicate device.
| Technological<br>Characteristics | |
|-----------------------------------------------------|----------------|
| Predicate Device | Subject Device |
| Cryogenic device | Same |
| Nitrous oxide coolant, pressurized cylinder | Same |
| Reusable handpiece, battery powered | Same |
| Single use tip for subdermal cooling, EO sterilized | Same |
| Charging dock | Same |
| Sensors monitor nitrous oxide delivery | Same |
#### Performance Data: f.
Design Verification testing was performed for the iovera° system to demonstrate that the device meets product specifications. Using a risk-based approach for the modifications, design verification testing was performed according to recognized standards, and is consistent with the predicate device and test methods described in previous 510(k) submissions for the iovera° system. The verification methods used and applied are appropriate for the changes. The verification and validation testing assures the modified subject device meet design input requirements, product specifications and relevant standards.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Pacira Pharmaceuticals, Inc. The logo is in purple and gray. The word "PACIRA" is in large, purple letters, and the words "PHARMACEUTICALS, INC." are in smaller, gray letters below. There is a gray arch above the word "PACIRA". The text "A Subsidiary of Pacira BioSciences, Inc." is below the company name.
## ioveraº system
Verification and validation performed on the subject device supports the substantial equivalence of the modified iovera9 system to the predicate device:
- Functional Testing ●
- Product Performance (Handpiece and Cartridge)
- Smart Tip Testing
- User Interface Testing
- . Electrical /EMC safety Testing
The subject ioveraº system device met the design verification and validation inputs, passing all predetermined acceptance criteria. No new issues of safety or effectiveness were identified during design verification and validation testing.
## g. Conclusion:
The subject ioveraº system, with the same intended use, indications for use, design, materials, technological features, and principles of operation as the cleared iovera° system (K161835), is substantially equivalent to the predicate device. The modification proposed in this 510(k) does not affect the safety and effectiveness of the device as the intended therapeutic use for the creation of lesions with the application of cold in peripheral nerves to block pain has not changed.
The modifications described, and the data presented, in this special premarket notification demonstrates that iovera® system is safe and effective, does not introduce or raise new questions of safety and effectiveness, and is substantially equivalent in intended use, technology, design, and materials to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.