The CenterVue EIDON FA is a confocal scanning ophthalmoscope indicated for color, infrared and autofluorescence imaging and fluorescein angiography of a human retina with or without the use of a mydriatic agent.
Device Story
Confocal scanning ophthalmoscope; uses white, infrared, and blue LEDs to capture retinal images. Subject device adds optional Ultra Widefield Lens (UWFL) to increase field of view from 60° to 80°. Input: retinal light reflections; Output: digital images (color, infrared, autofluorescence, fluorescein angiography). Used in clinical settings by eye care professionals. Features automated alignment, focus, exposure, and capture. Output displayed on 10.1" multi-touch tablet. Assists clinicians in visualizing retinal structures for diagnostic assessment. Benefits include expanded retinal view in single exposure, aiding comprehensive posterior eye evaluation.
Clinical Evidence
Bench-based clinical comparison study conducted at four sites in Italy. Compared images from EIDON FA (60°) and EIDON FA with UWFL (80°) across color, infrared, autofluorescence, and fluorescein angiography modalities. Sample included eyes with and without pathology. Qualitative side-by-side comparison of the central 60° field demonstrated that images from the subject device are similar to the predicate device. No clinical performance metrics (sensitivity/specificity) reported; study focused on image comparability.
Technological Characteristics
Confocal scanning ophthalmoscope; LED-based (IR 825-870nm, white 440-650nm, blue 440-475nm). 14 Mpixel sensor. Field of view: 60° (standard) or 80° (with UWFL). Connectivity: 10.1" multi-touch tablet. Standards: ANSI Z80-36 (light hazard), ISO 15004-1, ISO 10940, IEC 62304. Class I, Type B. IPX0. Dimensions: 620x590x360mm; 25kg.
Indications for Use
Indicated for color, infrared, autofluorescence imaging, and fluorescein angiography of the human retina. Applicable for patients requiring retinal imaging, with or without mydriatic agents.
Regulatory Classification
Identification
An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.
Special Controls
*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
August 19, 2021
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized human figure. The U.S. Food & Drug Administration logo is on the right and features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
CenterVue SpA Luca Scienza Quality and Regulatory Affairs Manager Via S. Marco 9H Padova, 35129 Italv
Re: K211328
Trade/Device Name: Eidon Fa, Eidon, Eidon Af, Eidon Uwfl Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: Class II Product Code: MYC Dated: July 12, 2021 Received: July 14, 2021
Dear Luca Scienza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number: K211328
Device Name
EIDON FA
Indications for Use (Describe)
The CenterVue EIDON FA is a confocal scanning ophthalmoscope indicated for color, infrared and autofluorescence imaging and fluorescein angiography of a human retina with or without the use of a mydriatic agent.
Type of Use (Select one or both, as applicable)
▽ Prescription Use (Part 21 CFR 801 Subpart D)
] Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
> PSC Publishing Services (301) 443-6740 El
{3}------------------------------------------------
# centervue
## 510(k) Summary
| Proprietary Name(s): | EIDON FA, EIDON, EIDON AF, EIDON UWFL |
|-------------------------|------------------------------------------------------------------------|
| Type of submission: | Special |
| Date of preparation: | August 10th, 2021 |
| Manufacturer: | CENTERVUE S.p.A. |
| | Via San Marco 9H |
| | 35129 Padova - ITALY |
| Submitter and contact | Mr. Luca Scienza |
| | Centervue S.p.A. |
| | Manager of Quality and Regulatory Affairs |
| | Via San Marco 9H, 35129 Padova - ITALY |
| | Phone: +39 049 501 8399 |
| | Fax: +39 049 501 8398 |
| | Email: luca.scienza@icare-world.com |
| Product Code: | MYC |
| Regulation Number: | 886.1570 |
| Common Name: | Ophthalmoscope |
| Panel: | Ophthalmic |
| Class: | Class II |
| Indications for Use: | The CenterVue EIDON FA is a confocal scanning ophthalmoscope |
| | indicated for color, infrared and auto-fluorescence imaging and |
| | fluorescein angiography of a human retina with or without the use of a |
| | mydriatic agent. |
| Legally marketed device | EIDON FA (K180526) |
(Predicate)
{4}------------------------------------------------
#### Device description
The CenterVue EIDON FA with Software version 4.0 and Ultra Widefield Lens is a modification of the CenterVue EIDON FA (K180526).
The CenterVue EIDON FA is a confocal scanning ophthalmoscope indicated for color, infrared and autofluorescence imaging and fluorescein angiography of a human retina with or without the use of a mydriatic agent.
CenterVue EIDON FA (K180526) is a scanning ophthalmoscope white light to obtain color images of the retina, infrared light to obtain infrared-reflectance images of the retina, and blue light to obtain autofluorescence and fluorescence images.
The CenterVue EIDON FA is part of a family of devices, which includes three models: EIDON FA, EIDON AF, and EIDON. EIDON is the base model, which features the following imaging modalities: infrared reflectance, color and red-free. EIDON AF adds autofluorescence imaging to the base model.
EIDON FA is the fully featured device, which adds fluorescein angiography to the capabilities of the EIDON AF and encompasses the features and functionality of the other models.
With respect to the CenterVue EIDON FA, EIDON FA with software version 4.0 provides additional features to acquire retinal images with the Ultra Widefield Lens. The optional Ultra Widefield Lens increases the field of view of the devices from 60° to 80°. All the features related to the acquisition without the Ultra Widefield Lens are unchanged.
The modified devices use the same base technology and maintains the same Intended Use and Indications for Use of the predicate device.
The fundamental scientific technology of the subject device is unchanged from the predicate and remains confocal, line scanning, LED-based, imaging.
The functional differences between EIDON FA and EIDON FA with software version 4.0 are as follows:
- . EIDON FA with software version 4.0 allows to capture retinal picture with an increased field of view, from 60° to 80° to allow the user to view a greater proportion of the posterior eye in one view.
- . EIDON FA with software version 4.0 allows to capture Ultra Widefield images with the same modalities of the unmodified device, namely: color images, infrared-reflectance images, autofluorescence images and fluorescence images.
No technological differences between EIDON FA and EIDON FA with SW version 4.0 exist.
- The modified device can be used with or without the Ultra Widefield Lens. The technological principle of retinal acquisitions remains the same in the modified device.
- . The device software has been modified to implement the acquisition and management of images taken with the Ultra Widefield Lens.
{5}------------------------------------------------
## centervue
The EIDON UWFL – Ultra Widefield Lens accessory is composed of a lens doublet, assembled in a custom lens holder, which can be mounted on the standard objective of the parent device by means of an incorporated threaded ring to increase the field of view of the parent device from 60° (standard objective) to 80° (with UWFL).
The CenterVue EIDON UWFL – Ultra Widefield Lens is an optional accessory to extend the field of view of the CenterVue EIDON, EIDON AF and EIDON FA from 60° to 80°. It is indicated for color, infrared and autofluorescence imaging and fluorescein angiography of a human retina with or without the use of a mydriatic agent.
The principle of operation of the EIDON UWFL is an increase of the standard objective to achieve a greater field of view.
The mechanism of action and the conditions for use are the same of the predicate device.
Other than the above, no differences exist between the EIDON FA with Ultra Widefield Lens, in particular the principle of operation, mechanism of action and interaction with the patient are unchanged.
#### TECHNICAL SPECIFICATIONS
#### Class and type of applied part
- Class I, Type B (according to IEC 60601-1). ●
#### IP classification:
- . IPX0 (according to the degree of protection provided by the enclosure with respect to harmful penetration of particulate matter or water).
#### Image acquisition:
| • Minimum pupil size: | 2.5 mm (non-mydriatic use) |
|-------------------------|-----------------------------------------------------------------------|
| • Field of view: | 60° (H) x 55° (V) captured in a single exposure |
| | 80° (H) x 75° (V) captured in a single exposure, with Ultra Widefield |
| | Lens |
| • Sensor size: | 14 Mpixel (4608x3288) |
| • Light source: | infrared LED (825-870 nm), white LED (440-650 nm), |
| | blue LED (440-475 nm) |
| • Imaging modalities: | color, red-free, IR reflectance, autofluorescence (AF), fluorescein- |
| | angiography (FA) |
| • Working distance: | 28 mm |
| | 16 mm, with Ultra Widefield Lens |
| • Resolution: | 60 pixels/deg (centre of the field) |
| | 45 pixels/deg (centre of the field), with Ultra Widefield Lens |
| • Resolution on retina: | 15 microns |
| | 17 microns with Ultra Widefield Lens |
{6}------------------------------------------------
- Pixel pitch: 4.9 microns
6.34 microns, with Ultra Widefield Lens
- . 1840x1622 pixels FA video resolution:
- 5 fps . FA video acquisition rate:
### Other features:
•
| • Automatic operation: | auto-alignment, auto-focus, auto-exposure, auto-capture |
|-----------------------------|---------------------------------------------------------|
| • Focus adjustment range: | -12 D to +15 D |
| | -8 D to +8 D with Ultra Widefield Lens |
| • Internal fixation target: | dynamic, programmable |
| • Display: | 10.1" multi-touch, color tablet |
#### Dimensions:
| • Weight: | 25 Kg (55 lbs) |
|-----------|--------------------------------------------|
| • Size: | 620 X 590 X 360 mm (14.2" x 23.2" x 24.4") |
#### Power supply:
| · Voltage: | 12 V DC |
|----------------------|---------|
| · Power consumption: | 60 W |
{7}------------------------------------------------
#### Comparison with predicate device
The predicate device selected for comparison with the CenterVue EIDON FA with software version 4.0 and Ultra Widefield Lens is identified as follows:
| Proprietary Name: | EIDON FA |
|----------------------|---------------------------------|
| Manufacturer: | CenterVue S.p.A. |
| 510(k) Number: | K180526 |
| Clearance Date: | 12 July 2018 |
| FDA Product Code: | MYC |
| Classification Name: | Ophthalmoscope, Laser, Scanning |
| Regulation Number: | 886.1570 |
The Subject Device and the Predicate Device share many identical or similar properties and features. In particular, the fundamental scientific technology of the subject device is unchanged from the predicate and remains confocal, line scanning, LED-based, ophthalmoscopy. Also, the principle of operation, mechanism of action and interaction with the patient are unchanged.
With respect to EIDON FA, the EIDON FA with software 4.0 includes the possibility to equip the devices of the family with an optional Ultra Widefield Lens, to increase the field of view of the devices from 60° to 80°, using the same base technology and maintaining the same Indications for Use.
The functional differences between EIDON FA and EIDON FA with software version 4.0 are as follows:
- . EIDON FA with software version 4.0 allows to capture retinal picture with an increased field of view, from 60° to 80° to allow the user to view a greater proportion of the posterior eye in one view.
- . EIDON FA with software version 4.0 allows to capture Ultra Widefield images with the same modalities of the unmodified device, namely: color images, infrared-reflectance images, autofluorescence images and fluorescence images.
No technological differences between EIDON FA and EIDON FA with SW version 4.0 exist.
- . The modified device can be used with or without the Ultra Widefield Lens. The technological principle of retinal acquisitions remains the same in the modified device.
- . The device software has been modified to implement the acquisition and management of images taken with the Ultra Widefield Lens.
Other than the above, no differences exist between the EIDON FA with software 4.0 and Ultra Widefield Lens, in particular the principle of operation, mechanism of action and interaction with the patient are unchanged.
None of these differences have any significant effect on safety or effectiveness of the Subject Device.
² Neither the CenterVue EIDON FA nor the EIDON FA with software 4.0 contain lasers but rather use LEDs for confocal imaging
{8}------------------------------------------------
Notably, with respect to the impact of these design modifications on a key part of the product's risk analysis (i.e. optical radiation safety), Centervue has confirmed conformance with the applicable recognized standard, ANSI Z80-36:2016, making these modifications eligible for a Special 510(k) notification.
#### Performance data
Continued conformance with the following standards has been confirmed in support of the substantial equivalence determination:
#### ANSI Z80-36 (Light Hazard Protection)
EIDON FA with software version 4.0 and Ultra Widefield Lens fulfills the requirements for a group 2 determination according to ANSI Z80-36, i.e. ophthalmic instruments with potential light hazard.
#### ISO 15004-1
The subject device complies with the ISO 15004-1:2006 standard for ophthalmic instruments.
#### IEC 62304
The subject device software complies with IEC 62304.
#### ISO 10940
EIDON FA with software version 4.0 and Ultra Widefield Lens fully complies with the applicable requirements of ISO 10940 - Ophthalmic instruments — Fundus Cameras when used without the optional lens, and partially complies with applicable requirements of ISO 10940 when used with the optional lens mounted.
#### Clinical summary
The imaging capabilities of EIDON FA with UWFL were assessed by comparing the following Subject and Predicate devices in the listed Imaging Modalities:
| SUBJECT DEVICE | PREDICATE DEVICE | IMAGING MODALITIES |
|-------------------------------------------------------------------|------------------|------------------------------------------------------------------|
| EIDON FA with<br>Software version 4.0 and<br>Ultra Widefield Lens | EIDON FA | Infrared<br>Color<br>Autofluorescence<br>Fluorescein Angiography |
The data were collected at four different sites located in Italy. Each eye was tested with Eidon FA with and without the lens in random order. Patients were tested according to booked appointments for the day: no patients pre-selection was performed based on gender, age, ethnicity, or any other factors. The imaging modality was selected according to the clinician's decision for each patient.
For infrared and color images, eyes with and without diagnosed pathology were included in the comparison. For Infrared images, 10 eyes which presented without pathology and the 8 eyes which presented with pathology were provided in a side-by-side comparison.
{9}------------------------------------------------
## centervue
For Color images, 11 eyes which presented without pathology and 10 eyes which presented with pathology were provided in a side-by-side comparison.
Since red-free images are obtained in EIDON FA as the green channel of the color image, comparison of color images also vields for red-free imaging.
For autofluorescence and fluorescein angiography 10 eyes respectively, with diagnosed pathology which presented were provided.
As images included have a different field of view between the two devices, namely 60° for the CenterVue EIDON FA and 80° for CenterVue EIDON FA with UWFL, only the inner 60° of the 80° images should be used for comparison.
For fluorescein angiography, direct comparison of images captured during the early phase is not possible as it would require simultaneous capture with and without the UWFL on the same device, so only intermediate and late phase images were compared. It should also be noted that, due to the need to remove the lens and resume the acquisition, FA images with and without the UWFL were taken at different times from fluorescein injection.
The comparison shows that EIDON FA with software version 4.0 and UWFL provides images that are similar in the central 60° to those of the cleared device, for all imaging modalities.
#### Conclusion
Based on the information contained within this submission, it is concluded that the CenterVue EIDON FA with software version 4.0 and Ultra Widefield Lens is substantially equivalent to the identified predicate devices already in interstate commerce within the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.