OCS™ Lung Solution is intended for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
Device Story
OCS Lung Solution is a sterile, colorless, pyrogen-free, colloid-based extracellular low-potassium solution. It is used by transplant clinicians to flush and store donor lungs after removal from the donor. The solution is administered at a recommended temperature to cool the organ, thereby reducing its metabolic requirements during cold storage and transport. The device is a stand-alone medical product supplied in a polypropylene bag. It facilitates the preservation of donor lungs for up to 24 hours, supporting the clinical objective of maintaining organ viability for eventual transplantation into a recipient.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, including chemical analysis and biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity, and in vitro hemolysis) per ISO-10993.
Technological Characteristics
Colloid-based extracellular low-potassium solution containing electrolytes and Dextran 40. pH adjusted with THAM/TRIS or bicarbonate. Supplied in a polypropylene bag. Terminally sterilized via steam sterilization (SAL ≤10⁻⁶). No electrical components or software.
Indications for Use
Indicated for flushing, storage, and transportation of isolated donor lungs in preparation for transplantation.
Regulatory Classification
Identification
An isolated kidney perfusion and transport system and accessories is a device that is used to support a donated or a cadaver kidney and to maintain the organ in a near-normal physiologic state until it is transplanted into a recipient patient. This generic type of device may include tubing, catheters, connectors, an ice storage or freezing container with or without bag or preservatives, pulsatile or nonpulsatile hypothermic isolated organ perfusion apparatus with or without oxygenator, and disposable perfusion set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2021
TransMedics, Inc. % Christy Foreman, M.B.E. Senior Consultant, Medical Devices Biologics Consulting Group, Inc. 1555 King Street. Suite 300 Alexandria, Virginia 22314
Re: K211314
> Trade/Device Name: OCSTM Lung Solution Regulation Number: 21 CFR 876.5880 Regulation Name: Isolated kidney perfusion and transport system and accessories Regulatory Class: II Product Code: MSB Dated: April 29, 2021 Received: April 30, 2021
Dear Christy Foreman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211314
Device Name OCS™ Lung Solution
Indications for Use (Describe)
OCS™ Lung Solution is intended for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K211314 Page 1 of 4
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the TransMedics, Inc. OCS™ Lung Solution is provided below.
#### 1. SUBMITTER
| Applicant: | TransMedics, Inc.<br>200 Minuteman Road, Suite 302<br>Andover, MA 01810 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Miriam Provost, Ph.D.<br>VP, Global Regulatory Affairs<br>978-494-7897<br>mprovost@transmedics.com |
| Submission Correspondent: | Christy Foreman, M.B.E.<br>Senior Consultant, Medical Devices<br>Biologics Consulting Group, Inc.<br>1555 King Street, Suite 300<br>Alexandria, VA 22314<br>Phone: 301-325-4245<br>Email: cforeman@biologicsconsulting.com |
| Date Prepared: | April 29, 2021 |
#### 2. DEVICE
| Device Trade Name: | OCS TM Lung Solution |
|--------------------|----------------------------------------------------------------|
| Common Name | Organ perfusion and preservation solution |
| Regulation Name: | Isolated kidney perfusion and transport system and accessories |
| Regulation: | 21 CFR 876.5880 |
| Regulatory Class: | Class II |
| Product Code: | MSB |
### PREDICATE DEVICE 3.
The subject device claims equivalence to the following legally marketed predicate device:
| 510(k) Number: | K000881 |
|----------------|-------------------------------------------------------------------------------|
| Device Name: | Perfadex® |
| Submitter: | VitroLife Sweden AB (now distributed by the subsidiary Xvive<br>Perfusion AB) |
{4}------------------------------------------------
| Regulation Name: | Isolated kidney perfusion and transport system and accessories |
|-------------------|----------------------------------------------------------------|
| Regulation: | 21 CFR 876.5880 |
| Regulatory Class: | Class II |
| Product Code: | MSB |
#### DEVICE DESCRIPTION 4.
The OCS Lung Solution is a colorless, sterile, pyrogen-free solution for cold flushing, storage and transport of donor lungs for transplantation. The OCS Lung Solution is a colloid-based extracellular low potassium solution for cold flushing and storage of donor lungs for transplantation. The OCS Lung Solution is used to flush and store a lung after removal from the donor. Administration of the solution at the recommended temperature will effectively cool the organ to reduce its metabolic requirements.
The subject 510(k) involves the class II use of the OCS Lung Solution. More specifically, the OCS Lung Solution is indicated for hypothermic flushing, storage and transportation of donor lungs for transplantation (21 CFR 876.5880, Class II Product Code MSB). The OCS Lung Solution is identical to the device approved as a component of the OCS Lung System, approved under P160013. However, as a class II device, the OCS Lung Solution is a stand-alone medical device.
#### INTENDED USE/INDICATIONS FOR USE 5.
OCS™ Lung Solution is intended for flushing, storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
### 6. SUBSTANTIAL EQUIVALENCE
# Comparison of Indications
The OCS Lung Solution has an indication for use statement that is identical to the predicate device (Perfadex). Like the predicate device, the OCS Lung Solution is indicated for flushing. storage and transportation of isolated lungs after removal from the donor in preparation for eventual transplantation into a recipient.
## Technological Comparisons
The OCS Lung Solution is substantially equivalent to the Perfadex Solution cleared under K000881. The chemical composition is nearly identical between the two products in terms of electrolytes and Dextran 40. Both products are adjusted to the same pH prior to use by 1 mmol THAM/TRIS or bicarbonate. The differences are the addition of one more gram of glucose in the OCS Lung solution, and the use of a polypropylene bag compared to a PVC bag for the predicate device.
{5}------------------------------------------------
### 7. PERFORMANCE DATA
# Biocompatibility Testing
Biocompatibility testing for the OCS Lung Solution was performed in accordance with the guidance, "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing." The duration of contact with the lung for the subject device is less than 24 hours. The level of contact was considered direct since the OCS Lung Solution is in direct contact with the donor lung. The testing to support the substantial equivalence included cytotoxicity testing, sensitization, intracutaneous reactivity, acute systemic toxicity, pyrogenicity (material mediated and bacterial endotoxin) and in vitro hemolysis. As the OCS Lung Solution in its final finished form (both solution and solution bag) is identical to the previously marketed device in formulation, processing, sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents), biocompatibility of the OCS Lung Solution has been established.
# Sterilization and Shelf Life
After formulation and filling, the OCS Lung Solution bags are terminally sterilized using a steam sterilization process to a SAL of ≤106. Each lot (or batch) of the OCS Lung Solution is tested for endotoxins prior to lot release and the product is labeled as pyrogen-free.
This OCS Lung Solution is labeled with a 24-month shelf life.
## Electrical safety and electromagnetic compatibility (EMC)
Not applicable. The device contains no electric components, generates no electrical emissions, and uses no electrical energy of any type.
# Software Verification and Validation Testing
Not applicable. The device contains no software.
# Bench Testing
Bench testing consisted of chemical analyses of the OCS Lung Solution.
# Animal Testing
Not applicable. Animal studies are not necessary to establish the substantial equivalence of this device.
## Clinical Data
Not applicable. Clinical studies are not necessary to establish the substantial equivalence of this device.
{6}------------------------------------------------
### CONCLUSION 8.
The substantial equivalence information provided in this submission demonstrates that the subject device is substantially equivalent to the predicate device in both indications for use and technological characteristics. The minor technological differences do not raise new or different questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.