AtriCure Isolator® Synergy™ Surgical Ablation System
K211311 · AtriCure, Inc. · OCL · May 28, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211311
Device Name
AtriCure Isolator® Synergy™ Surgical Ablation System
Applicant
AtriCure, Inc.
Product Code
OCL · General, Plastic Surgery
Decision Date
May 28, 2021
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AtriCure Isolator Synergy Surgical Ablation System is intended to ablate cardiac tissue during surgery.
Device Story
System comprises Ablation and Sensing Unit (ASU), Switch Matrix (ASB), footswitch, and single-patient use electrosurgical clamp. Clamp features opposing dual electrodes on insulated jaws; various lengths/curvatures available. Operator uses footswitch to trigger ASU, which delivers RF energy to electrodes to create transmural cardiac lesions. Used in surgical settings by clinicians. Device facilitates cardiac tissue ablation; modifications include overmolded end effector design. Benefits include precise, confined lesion creation during cardiac surgery.
Clinical Evidence
Bench testing only. No clinical data presented. Biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity) performed per ISO 10993-1. Performance testing confirmed transmurality and lesion width specifications remain consistent with the predicate.
Technological Characteristics
Electrosurgical bipolar RF clamp. Materials: ABS resin (MG37EPX), Santoprene 8000 thermoplastic rubber, Ultem pins. Energy source: RF generator (ASU2/ASB3). Sterilization: Ethylene Oxide (EO) to 10^-6 SAL. Packaging: PETG blister with Tyvek lid. Complies with ISO 10993-1, IEC 60601-2-2, ASTM E1952-11, ISO 11607-2, AAMI TIR 28.
Indications for Use
Indicated for the ablation of cardiac tissue during surgery. Contraindicated for contraceptive coagulation of the fallopian tubes.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 28, 2021
AtriCure, Inc. Jon McElwee Senior Manager, Regulatory Affairs 7555 Innovation Wav Mason, Ohio 45040
Re: K211311
Trade/Device Name: AtriCure Isolator® Synergy™ Surgical Ablation System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: OCL Dated: April 27, 2021 Received: April 30, 2021
Dear Jon McElwee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Aneesh Deoras Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211311
### Device Name
AtriCure Isolator® Synergy™ Surgical Ablation System
Indications for Use (Describe)
The AtriCure Isolator Synergy Surgical Ablation System is intended to ablate cardiac tissue during surgery.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "AtriCure" in a stylized font. The letters "A," "t," "r," and "i" are in blue, while the "C," "u," "r," and "e" are in orange. There is a blue dot above the "i" and inside the "C." There is a registered trademark symbol next to the "e".
### 510(k) Summary
#### l. Applicant Information
| Manufacturer: | AtriCure, Inc.<br>7555 Innovation Way<br>Mason, Ohio 45040<br>P: 513-498-5067<br>F: 513-895-9013 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jonathan McElwee, RAC<br>Sr. Manager, Regulatory Affairs |
| Alternate Contact: | Mary Galeano<br>Sr Regulatory Affairs Specialist |
| Date Prepared: | 04/29/2021 |
| II.<br>Device Information | |
| Proprietary Name: | AtriCure Isolator® Synergy™ Surgical Ablation System |
| Common Name: | Cardiac Radio Frequency Ablation System Clamps |
| Classification: | Electrosurgical cutting and coagulation device and accessories<br>Regulatory Class: Class II; per 21 CFR 878.4400<br>Product Code: OCL<br>Classification Panel: Card ovascular |
#### lll. Device Description
The AtriCure Isolator Synergy Surgical Ablation System is comprised of the Ablation and Sensing Unit (ASU), AtriCure Switch Matrix (ASB), an AtriCure Isolator Synergy device, and a footswitch. The AtriCure Isolator Synergy device is a single-patient use, electrosurgical instrument designed for use only with the ASU.
The AtriCure Isolator Synergy clamp is used for cardiac tissue ablation. When activated, the ASU delivers radiofrequency (RF) energy to the linear electrodes on the insulated jaws of the AtriCure Isolator Synergy Clamp. The Operator controls the application of this RF energy by pressing the Footswitch.
All AtriCure Isolator Synerqy devices are configured as vascular clamps and feature clamping jaws of various lengths, curvatures and apertures; rounded jaw tips, and cables that plugs into the ASB switch matrix and ASU RF generator.
The AtriCure Isolator Synergy Clamps features two pairs of opposing dual electrodes, an in-line handle with syringetype grip handle/ actuation and button release mechanisms.
There is a Glidepath™ Tape Guide that is a single-patient use, detachable, surgical device the guidance of surgical instruments through soft tissue during general surgical procedures. The Guide is package with ISOLATOR devices that have the Attachment Tip.
The Glidepath Tape Instrument Guide is a single-patient use, surgical device designed to facilitate the guidance of surgical instruments through soft tissue during general surgical procedures.
{4}------------------------------------------------
NOTE: Please refer to the AtriCure ASU2 and ASB3 Instructions for information specific to the ASU2 and ASB3.
### IV. Indications for Use
The AtriCure Isolator Synergy Surgical Ablation System is intended to ablate cardiac tissue during surgery.
### V. Comparison of Technological Characteristics (Isolator Synergy Clamp -K110117)
This submission included modifications to allow for an alternate end effector design change to an overmold configuration. The science and fundamental technology of the Isolator Synergy Clamps (EMR2) in comparison to the predicate Isolator Synergy Clamp (EMR2, EML2), cleared per K110117, remain the same. Performance testing confirmed that the Isolator Synergy Clamps meets the same requirements for safety and efficacy as the predicate Isolator Synergy Clamp.
## Table 1: Device Comparison
| # | Category | Feature/Item | Current - Isolator<br>Synergy Clamps<br>(EMR2, EML2) | Proposed Isolator<br>Synergy Clamps<br>(EMR2, EML2) | Equivalence<br>Comparison |
|----|------------------------|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------------------------|
| 1 | | Manufacturer | AtriCure, Inc. | AtriCure, Inc. | Same |
| 2 | | Product Name | Isolator Synergy Clamp<br>with Glidepath soft<br>guide | Isolator Synergy Clamp<br>with Glidepath soft<br>guide | Same |
| 3 | | 510(k) and Past<br>Submission<br>References | K110117 | K211311 | N/A |
| 4 | | Device<br>Classification /<br>Product Code | Class II,<br>21 CFR § 878.4400<br>OCL | Class II,<br>21 CFR § 878.4400<br>OCL | Same |
| 5 | | Model Number | EMR2, EML2 | EMR2, EML2 | N/A |
| 6 | General<br>Information | Indications for Use | The System is<br>intended for the<br>ablation of cardiac<br>tissue during surgery. | The System is<br>intended for the<br>ablation of cardiac<br>tissue during surgery. | Same |
| 7 | | Contraindication | The Bipolar<br>(Transpolar) System is<br>not indicated for<br>contraceptive<br>coagulation of the<br>fallopian tubes. | The Bipolar<br>(Transpolar) System is<br>not indicated for<br>contraceptive<br>coagulation of the<br>fallopian tubes. | Same |
| 8 | | Device Description | Electrosurgical Bipolar<br>RF Clamp and optional<br>accessory (Class I)<br>instrument guide | Electrosurgical Bipolar<br>RF Clamp and optional<br>accessory (Class I)<br>instrument guide | Same |
| 9 | | Generator<br>Description/<br>Energy Source | RF Generator –<br>Ablation and Sensing<br>Unit and Source Switch<br>Matrix<br>(ASU2, ASB3) | RF Generator –<br>Ablation and Sensing<br>Unit and Source Switch<br>Matrix<br>(ASU2, ASB3) | Same |
| 10 | | Accessory/<br>Positioning Guide | Product Name:<br>Glidepath Tape<br>Materials: Santoprene<br>8000 Thermoplastic<br>Rubber with Ultem pins | Product Name:<br>Glidepath Tape<br>Materials: Santoprene<br>8000 Thermoplastic<br>Rubber with Ultem pins | Same |
{5}------------------------------------------------
| # | Category | Feature/Item | Current - Isolator<br>Synergy Clamps<br>(EMR2, EML2) | Proposed Isolator<br>Synergy Clamps<br>(EMR2, EML2) | Equivalence<br>Comparison |
|----|----------------------------|--------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|---------------------------|
| | | End Effector<br>Design | Non-overmolded design<br>with ABS resin<br>MG37EP | Overmolded design<br>with ABS resin<br>MG37EPX | Equivalent |
| 11 | Performance | Ablation<br>Performance -<br>Transmurality | Transmural lesions | Transmural lesions | Same |
| 12 | | Ablation<br>Performance -<br>Lesion Width | Lesions tightly confined<br>to the region between<br>the opposing insulated<br>jaws. | Lesions tightly confined<br>to the region between<br>the opposing insulated<br>jaws. | Same |
| 13 | Sterilization<br>and Other | Sterilization Site | EO | EO | Same |
| 14 | | Packaging | PETG blister with<br>Tyvek® lid | PETG blister with<br>Tyvek® lid | Same |
| 15 | | Sterilization | Ethylene Oxide | Ethylene Oxide | Same |
| 16 | | Sterility Assurance<br>Level | 10-6 SAL | 10-6 SAL | Same |
| 17 | | Biocompatibility | Meets requirements of<br>ISO 10993 | Meets requirements of<br>ISO 10993 | Same |
| 18 | | Shelf Life | 3 year | 3 year | Same |
| 19 | | Pyrogen | Non-pyrogenic | Non-pyrogenic | Same |
| 20 | | Latex | Not made with natural<br>rubber latex | Not made with natural<br>rubber latex | Same |
| 21 | | Software | No software in the<br>clamp | No software in the<br>clamp | Same |
### VI. Performance Data
A review of the risk analysis concluded there is not overall change to the risk profile of the proposed AtriCure Isolator Synergy clamps versus the previously cleared AtriCure Isolator Synergy clamps as the modifications to the proposed AtriCure Isolator Synergy do not add or remove any features of the device or change the clinical application.
The proposed changes do not include any change to design or performance specifications of the AtriCure Isolator Synergy clamps. Additionally, the proposed changes did not modify the intended use, therefore the previously submitted bench and animal data remain valid for the AtriCure Isolator Synergy clamps.
The following consensus standards were leveraged in this submission:
- ISO 10993-1
- IEC 60601-2-2
- . ASTM E1952-11
- ISO 11607-2
- AAMI TIR 28
### Biocompatibility Testing
The biocompatibility evaluation is in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process" as recognized by FDA. Data collected from the following tests support this change:
- Cytotoxicity
- Sensitization ●
- Irritation ●
- . Acute Systemic Toxicity
- . Material Mediated Pyrogenicity
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AtriCure. The word "Atri" is in blue, and the word "Cure" is in orange. There is a blue dot in the middle of the "C" in "Cure". There is a registered trademark symbol next to the word "Cure".
### VII. Conclusions
AtriCure has demonstrated that the modifications made to the Isolator Synergy Clamp (EMR2, EML2) are substantially equivalent in intended use/indication for use as the previously cleared Isolator Synergy Clamp (EMR2, EML2) per K110117.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.