K211293 · Promisemed Hangzhou Meditech Co., Ltd. · FMI · Jun 25, 2021 · General Hospital
Device Facts
Record ID
K211293
Device Name
Safety Winged Blood Collection Sets
Applicant
Promisemed Hangzhou Meditech Co., Ltd.
Product Code
FMI · General Hospital
Decision Date
Jun 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The safety winged blood collection set is single-use, sterile, winged venipuncture needle bonded to a flexible tubing with or without a luer adaper and/or tube holder. The device is used for blood collection and/or the short-term infusion of intravenous fluids (up to 2 hours under direct clinical supervision). The blood-collection needle is designed to be covered with a safety mechanism, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury.
Device Story
Safety Winged Blood Collection Set; single-use, sterile venipuncture needle with flexible tubing, optional luer adapter, and optional tube holder. Used for blood collection or short-term IV infusion (up to 2 hours) under clinical supervision. Features integrated safety mechanism to prevent needlestick injuries; activated by pressing curved design with index finger while holding wings with thumb and middle finger, then withdrawing until audible 'click' confirms needle coverage. Operated by healthcare professionals in clinical settings. Output is successful blood draw or fluid delivery; safety mechanism protects user post-procedure. Benefits include reduced risk of accidental needlestick injury.
Clinical Evidence
No clinical data. Substantial equivalence based on non-clinical bench testing, including physical, chemical, and biological performance verification against ISO standards (e.g., ISO 7864, ISO 8536-4, ISO 9626, ISO 23908) and biocompatibility testing (ISO 10993).
Indicated for blood collection and short-term (up to 2 hours) intravenous fluid infusion in patients requiring venipuncture, under direct clinical supervision.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Gemtier Medical (Shanghai) Inc. Safety Blood Collection Device for Single Use (K170276)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 25, 2021
Promisemed Hangzhou Meditech Co., Ltd. % John Beasley Official Correspondent MedTech Review, LLC 257 Garnet Garden Street Henderson, Nevada 89015
Re: K211293
Trade/Device Name: Safety Winged Blood Collection Sets Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI, FPA Dated: April 16, 2021 Received: April 26, 2021
Dear John Beasley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known)
## K211293
### Device Name
Safety Winged Blood Collection Set - Needle Diameter (21, 23, 25 Guage) - Tube Length (10, 19, 30 cm) - HOLDER, SBC-xxxxH, Safety Winged Blood Collection Set - Needle Diameter (21, 23, 25 Guage) - Tube Length (10, 19, 30 cm) - ADAPTER, SBC-xxxx A: Safety Winged Blood Collection Set - Needle Diameter (21, 23, 25 Guage) - Tube Length (10, 19, 30 cm) - W/O Puncture, SBC-
#### Indications for Use (Describe)
The safety winged blood collection set is single-use, sterile, winged venipuncture needle bonded to a flexible tubing with or without a luer adaper and/or tube holder. The device is used for blood collection and/or the short-term infusion of intravenous fluids (up to 2 hours under direct clinical supervision). The blood-collection needle is designed to be covered with a safety mechanism, which can be activated to cover the needle immediately following venipuncture to aid in the protection against accidental needlestick injury.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human ServicesDepartment of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for "Promisemed". The word "Promise" is in green and the word "med" is in blue. There is a blue molecule graphic above the word "Promise" and a registered trademark symbol to the right of the word "med". There is a black line underneath the logo.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District,
Hangzhou City, 311121 Zhejiang, China.
# 510(k) Summary
Updated: 06/18/2021
| Contact Details | | 21 CFR 807.92(a)(1) |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|---------------------|
| Applicant Name | Promisemed Hangzhou Meditech Co., Ltd. | |
| Applicant Address | No. 12, Longtan Road, Cangqian Street, Yuhang District, Hangzhou City,<br>Zhejiang, 311121, CH | |
| Applicant Telephone Number | 865-718-8772985 | |
| Applicant Contact | Mr. Zearou Yang | |
| Applicant Contact Email | zearou.yang@promisemed.ca | |
| Correspondent Name | MedTech Review, LLC | |
| Correspondent Address | 257 Garnet Garden Street, Henderson, NV, 89015, US | |
| Correspondent Telephone<br>Number | 1-612-889-5168 | |
| Correspondent Contact | Mr. John Beasley, RAC (US) | |
| Correspondent Contact Email | john@medtechreview.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Safety Winged Blood Collection Sets | |
| Common Name | Safety Blood Collection Device for Single Use | |
| Classification Name | NEEDLE, HYPODERMIC, SINGLE LUMEN | |
| Regulation Number | 880.5570 | |
| Product Code | FMI/FPA | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate [510(k)] # | Predicate Trade Name | Product Code |
| K170276 | Gemtier Medical (Shanghai) Inc. Safety Blood Collection Device<br>for Single Use | FMI/FPA |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| The Promisemed Safety Blood Collection sets are single-use, sterile, venipuncture needles used for<br>blood collection or short-term infusion of intravenous fluids (up to 2 hours under direct clinical<br>supervision). The blood-collection needle is designed to be covered with a safety mechanism, which can<br>be activated to cover the needle immediately following venipuncture to aid in the protection against<br>accidental needlestick injury. | | |
| Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) |
| The safety winged blood collection set is single-use, sterile, winged venipuncture needle bonded to a<br>flexible tubing with or without a luer adapter and/or tube holder. The device is used for blood collection<br>and/or the short-term infusion of intravenous fluids (up to 2 hours under direct clinical supervision). The<br>blood-collection needle is designed to be covered with a safety mechanism, which can be activated to<br>cover the needle immediately following venipuncture to aid in the protection against accidental<br>needlestick injury. | | |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "Promisemed". The word "Promise" is in green, and the word "med" is in blue. Above the word "Promise" is a green and blue atom-like structure. To the right of the word "Promisemed" is a circled R symbol.
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121 Zhejiang, China.
## Indications for Use Comparison
21 CFR 807.92(a)(5)
Other than the difference in the name of the device and the fact that they are all connected to flexible tubing (these are sets), there are no differences in the indications for use of the subject device when compared to the predicate device.
### Technological Comparison
21 CFR 807.92(a)(6)
Compared with the predicate device, the subject device is identical in mechanism of action, materials, needle point, sterilization method, and biocompatibility. The subject device has the same performance specifications except for needle length. Through performance bench testing, the subject device and the predicate device demonstrated to be substantially equivalent.
| Specification | Predicate Device | Subject Device | Discussion of<br>Differences |
|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | Safety Blood Collection<br>Device for Single Use | Safety Winged Blood<br>Collection Sets | The subject device brings<br>attention to the "wings" in<br>the device name, as well<br>as properly defines the<br>device as a "set".<br><br>The differences do not<br>introduce any new<br>concerns for safety or<br>efficacy. |
| K number | K170276 | K211293 | - |
| Product code and<br>classification name | FMI, Needle,<br>Hypodermic, Single<br>Lumen | FMI, Needle<br>Hypodermic, Single<br>Lumen | No difference |
| Indications for use | The SAFETY BLOOD<br>COLLECTION DEVICE<br>FOR SINGLE USE is a<br>single-use, sterile,<br>winged venipuncture<br>needle bonded to a<br>flexible tubing with or<br>without a luer adapter<br>and/or tube holder.<br>The SAFETY BLOOD<br>COLLECTION DEVICE<br>FOR SINGLE USE is<br>used for blood<br>collection and/or the<br>short-term infusion of<br>intravenous fluids (up<br>to 2 hours under direct<br>clinical supervision).<br>The venipuncture<br>needle is designed to | The safety winged<br>blood collection set is<br>single-use, sterile,<br>winged venipuncture<br>needle bonded to a<br>flexible tubing with or<br>without a luer adapter<br>and/or tube holder.<br>The device is used for<br>blood collection and/or<br>the short-term infusion<br>of intravenous fluids<br>(up to 2 hours under<br>direct clinical<br>supervision). The<br>blood-collection needle<br>is designed to be<br>covered with a safety<br>mechanism, which can<br>be activated to cover | Difference in the name of<br>the device and the fact<br>that they are all<br>connected to flexible<br>tubing (these are sets).<br>The differences do not<br>introduce any new<br>concerns for safety or<br>efficacy. |
| Specification | Predicate Device | Subject Device | Discussion of<br>Differences |
| | be covered with a<br>safety mechanism,<br>which can be activated<br>to cover needle<br>immediately following<br>venipuncture to aid in<br>the protection against<br>accidental needlestick<br>injury. | the needle immediately<br>following venipuncture<br>to aid in the protection<br>against accidental<br>needlestick injury. | |
| Indicated for infusion | Yes | Yes | No difference |
| Single use | Yes | Yes | No difference |
| Mechanism of safety<br>blood collection | Latch mechanism:<br>press both sides of the<br>safety mechanism to<br>release the lock first<br>and then slide the<br>safety mechanism to<br>activate. | Latch mechanism:<br>Place the index finger<br>on the curved design to<br>press gently, hold the<br>two wings by the thumb<br>finger and middle finger<br>and withdraw the wings<br>back to activate the<br>safety mechanism until<br>hearing sound of<br>CLICK. The click is a<br>sign that the Safety<br>mechanism has been<br>correctly activated | The subject device<br>requires use of one finger<br>to press the release and<br>incorporates the wings for<br>activating the safety<br>mechanism.<br>The differences do not<br>introduce any new<br>concerns for safety or<br>efficacy. |
| Materials: | | | |
| Venipuncture Needle | Stainless Steel | Stainless Steel | No difference |
| Tubing | PVC | PVC | No difference |
| Wing | PVC | PVC | No difference |
| Sleeve | PE | ABS | No difference |
| Retractable Cartridge | PE | PE | No difference |
| Luer adapter | PP | PVC | No difference |
| Tube Holder | PP | ABS | No difference |
| Blood collection needle<br>protector | PP | PE | No difference |
| Rubber cover | Synthetic Rubber | Nitrile Rubber | Predicate device<br>unspecified.<br>The use of nitrile rubber<br>does not introduce any<br>new concerns for safety<br>or efficacy. |
| Specification | Predicate Device | Subject Device | Discussion of<br>Differences |
| Components: | | | |
| Tubing, Wing, Luer<br>connectors | Yes | Yes | No difference |
| Plastic sleeve,<br>venipuncture needle<br>stand | Yes | Yes | No difference |
| Luer adapter, tube<br>holder, blood collection<br>needle, protection<br>mechanism | Yes | Yes | No difference |
| Performance: | | | |
| Needle Gauge Sizes | 21G, 23G, 25G | 21G, 23G, 25G | No difference |
| Tubing Length | 10 cm to 35 cm | 10 cm, 19cm, 30 cm | No difference |
| Needle Length | 20mm | 19mm | The needle length of the<br>subject device is shorter<br>in length (19mm) as<br>compared to the<br>predicate device (20mm).<br>The short needle length<br>allows the phlebotomist to<br>insert it at a shallow angle<br>that can increase the<br>ease of use and<br>therefore, does not<br>introduce any new<br>concerns for safety or<br>efficacy. |
| Needle Point | 3-bevel | 3-bevel | No difference |
| Sterile | Ethylene Oxide | Ethylene Oxide | No difference |
| Non-pyrogen | Yes | Yes | No difference |
| Shelf life | 3 years | 3 years | No difference |
| Biocompatibility | Cytotoxicity, Skin<br>Sensitization, Irritation,<br>Acute Systemic<br>Toxicity, Hemolysis,<br>Pyrogenicity, Bacterial<br>Endotoxins testing<br>have passing results | Cytotoxicity, Skin<br>Sensitization, Irritation,<br>Acute Systemic<br>Toxicity, Hemolysis,<br>Pyrogenicity, Bacterial<br>Endotoxins testing<br>have passing results | No difference |
| Physical<br>performance | Needle appearance and cleanliness | ISO 7864:2016, clause 4.3 | |
| | Needle Size color code | ISO 6009:2016 | |
| | Needle Particulate contamination | USP<788>, METHOD 1, Test 1.B | |
| | Bonding<br>strength: | Needle and hub | ISO 7864:2016, clause 4.12 |
| |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.