K211277 · Clear Choice Therapeutics, Inc. · OMP · Dec 16, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211277
Device Name
Theia NPWT Foam Wound Dressing Kit
Applicant
Clear Choice Therapeutics, Inc.
Product Code
OMP · General, Plastic Surgery
Decision Date
Dec 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Theia NPWT Foam Wound Dressing Kit is intended to be used in conjunction with the CCT Mini or CCT1 Negative Pressure Wound Drainage Pumps for the application of negative pressure wound therapy to the wound. When used in conjunction with the CCT Mini or CCT1 Negative Pressure Wound Drainage Pumps, the Theia NPWT Foam Wound Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris. The Theia NPWT Foam Wound Dressing Kit is appropriate for use on the following wounds: Pressure Ulcers Diabetic/Neuropathic Ulcers Venous Insufficiency Ulcers Traumatic Wounds Post-Operative and Dehisced Surgical Wounds Skin Flap and Grafts
Device Story
Sterile, single-use foam dressing kit for negative pressure wound therapy (NPWT). Components: reticular polyether-based polyurethane hydrophobic foam, occlusive drape, dome assembly with tubing/luer lock. Used with CCT Mini or CCT1 pumps (K082311) in institutional settings by healthcare professionals. Foam packs wound; drape creates seal; dome/tubing connects to pump canister. Pump delivers negative pressure to suction exudate. Facilitates wound healing by removing debris/exudates. Bench testing confirms uniform pressure delivery and fluid removal.
Clinical Evidence
Bench testing only. No clinical or animal studies conducted. Bench performance testing demonstrated effective removal of wound fluids and uniform pressure delivery across the dressing in simulated wound setups.
Technological Characteristics
Materials: black reticular polyether-based polyurethane foam, semipermeable polyurethane film drape, thermoplastic elastomer dome, PVC tubing. Dimensions: 1.5cm or 3.0cm thickness. Sterilization: Gamma radiation. Packaging: Tyvek pouch. Connectivity: mechanical connection to external NPWT pump.
Indications for Use
Indicated for patients requiring negative pressure wound therapy (NPWT) to promote healing via removal of excess exudates, infectious material, and tissue debris. Appropriate for pressure, diabetic/neuropathic, venous insufficiency, traumatic, post-operative/dehisced surgical wounds, and skin flaps/grafts.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
December 16, 2021
Clear Choice Therapeutics, Inc. % Kathy Herzog Regulatory Consultant DuVal & Associates 1820 Medical Arts Building, 820 Nicollet Mall Minneapolis, Minnesota 55402
Re: K211277
Trade/Device Name: Theia NPWT Foam Wound Dressing Kit Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: September 17, 2021 Received: September 20, 2021
Dear Kathy Herzog:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known) K211277
Device Name Theia NPWT Foam Wound Dressing Kit
### Indications for Use (Describe)
The Theia NPWT Foam Wound Dressing Kit is intended to be used in conjunction with the CCT Mini or CCTI Negative Pressure Wound Drainage Pumps for the application of negative pressure wound therapy to the wound. When used in conjunction with the CCT Mini or CCT1 Negative Pressure Wound Drainage Pumps, the Theia NPWT Foam Wound Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.
The Theia NPWT Foam Wound Dressing Kit is appropriate for use on the following wounds:
- Pressure Ulcers
- · Diabetic/Neuropathic Ulcers
- · Venous Insufficiency Ulcers
- · Traumatic Wounds
- · Post-Operative and Dehisced Surgical Wounds
- · Skin Flap and Grafts
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K211277
Page 1 of 5
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
#### I. SUBMITTER
Clear Choice Therapeutics, Inc. 530 W. Cloverhurst Ave. Athens, GA 30606-4216
Phone: 217.377.4834
Contact Person: John Prince Date Prepared: November 18, 2021
#### II. DEVICE
| Trade/Proprietary Names: | Theia NPWT Foam Wound Dressing Kit, Models<br>UN1083S, UN20133M, and UN25163L, FWTS21,<br>FWTM36, and FWTL44 |
|--------------------------|--------------------------------------------------------------------------------------------------------------|
| Common Name: | Foam Dressing Kit |
| Regulation Number: | 21 CFR 878.4780 |
| Regulation Name: | Powered Suction Pump |
| Device Class: | II |
| Product Code: | OMP |
| Panel: | General & Plastic Surgery |
#### PREDICATE DEVICE III.
Theia NPWT Foam Wound Dressing Kit, K161570.
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Theia NPWT Foam Wound Dressing Kits are sterile, single use, disposable dressings for use with the CCT Mini or CCT1 Negative Pressure Wound Drainage pumps for the application of negative pressure wound therapy (NPWT) to the wound. The kits include a foam dressing made of black reticular polyether based polyurethane hydrophobic foam material, occlusive drape(s), and a dome assembly consisting of a single lumen dome, dome skirt, and drainage tubing with clamp and female luer lock connector. The foam
{4}------------------------------------------------
# K211277
dressing covers the wound and the transparent occlusive drape covers the foam dressing and creates a seal around the wound. The drainage tubing connects to the exudate canister of the negative pressure pump. The powered negative pressure pump delivers negative pressure to the dressing to suction wound exudate into the exudate canister of the pump. The dressing kits are available in three sizes (small, medium, large) based on foam dressing size (length x width) and in two foam dressing thicknesses (1.5 and 3.0 cm), for a total of six dressing kits.
The Theia NPWT Foam Wound Dressing Kits are for institutional use only with the CCT Mini and CCT1 pumps cleared through K082311.
#### V. INDICATIONS FOR USE
The Theia NPWT Foam Dressing Kit is intended to be used in conjunction with the CCT Mini or CCT1 Negative Pressure Wound Drainage Pumps for the application of negative pressure wound therapy to the wound. When used in conjunction with the CCT Mini or CCT1 Negative Pressure Wound Drainage Pumps, the Theia NPWT Foam Dressing Kit is indicated for patients who would benefit from a suction device, particularly as the device may promote wound healing by removal of excess exudates, infectious material and tissue debris.
The Theia NPWT Foam Dressing Kit is appropriate for use on the following wounds:
- Pressure Ulcers ●
- Diabetic/Neuropathic Ulcers .
- Venous Insufficiency Ulcers ●
- o Traumatic Wounds
- Post-Operative and Dehisced Surgical Wounds ●
- Skin Flap and Grafts o
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
Both the subject (1.5cm thick foam dressing) and predicate (3.0cm thick foam dressing) Theia NPWT Foam Wound Dressing Kits have the same Intended Use for the application of NPWT to the wound when used in conjunction with a negative pressure pump. The 1.5cm and 3.0cm thick foam dressing kits are identical except for the foam thickness, kit model numbers, and indicated negative pressure pumps as shown in Table 1.
{5}------------------------------------------------
# K211277
| Feature | Theia NPWT<br>Dressing Kit<br>(Subject Device) | Theia NPWT<br>Dressing Kit<br>(Predicate Device) |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | K211277 | K161570 |
| Classification | 21 CFR 878.4780 | 21 CFR 878.4780 |
| Product Code | OMP | OMP |
| Class | II | II |
| Intended Use | The Theia NPWT Dressing Kit is<br>intended to be used in conjunction<br>with a negative pressure pump for the<br>application of negative pressure<br>wound therapy to the wound. | The Theia NPWT Dressing Kit is<br>intended to be used in conjunction<br>with a negative pressure pump for the<br>application of negative pressure<br>wound therapy to the wound. |
| Indications For<br>Use | The Theia NPWT Foam Wound<br>Dressing Kit is intended to be used in<br>conjunction with the CCT Mini or<br>CCT1 Negative Pressure Wound<br>Drainage Pumps for the application of<br>negative pressure wound therapy to<br>the wound. When used in<br>conjunction with the CCT Mini or<br>CCT1 Negative Pressure Wound<br>Drainage Pumps, the Theia NPWT<br>Foam Wound Dressing Kit is<br>indicated for patients who would<br>benefit from a suction device,<br>particularly as the device may<br>promote wound healing by removal<br>of excess exudates, infectious<br>material and tissue debris. | The Theia NPWT Foam Wound<br>Dressing Kit is intended to be used in<br>conjunction with the SIMEX<br>Negative Pressure Wound Therapy<br>Pumps (K113291) for the application<br>of negative pressure wound therapy to<br>the wound. When used in conjunction<br>with the SIMEX Negative Pressure<br>Wound Therapy Pumps, the Theia<br>NPWT Foam Wound Dressing Kit is<br>indicated for patients who would<br>benefit from a suction device,<br>particularly as the device may<br>promote wound healing by removal of<br>excess exudates, infectious material<br>and tissue debris. |
| | The Theia NPWT Foam Wound<br>Dressing Kit is appropriate for use on<br>the following wounds:<br>• Pressure Ulcers<br>• Diabetic/Neuropathic Ulcers<br>• Venous Insufficiency Ulcers<br>• Traumatic Wounds<br>• Post-Operative and Dehisced<br>Surgical Wounds<br>• Skin Flap and Grafts | The Theia NPWT Foam Wound<br>Dressing Kit is inappropriate for use on<br>the following wounds:<br>• Pressure Ulcers<br>• Diabetic/Neuropathic Ulcers<br>• Venous Insufficiency Ulcers<br>• Traumatic Wounds<br>• Post-Operative and Dehisced<br>Surgical Wounds<br>• Skin Flap and Grafts |
| Type of Use | Prescription Only | Prescription Only |
| User | Healthcare professional | Healthcare professional |
| Intended Use<br>Environment | Hospital or long-term care clinic<br>setting | Hospital or long-term care clinic<br>setting |
| Feature | Theia NPWT<br>Dressing Kit<br>(Subject Device) | Theia NPWT<br>Dressing Kit<br>(Predicate Device) |
| Single Use or<br>Reusable | Single Use | Single Use |
| Foam<br>Dressing<br>Dimensions<br>and Model<br>Numbers | Model Number, Size, and dimensions<br>(Length x Width x Thickness, in cm).<br>• Model FWTS21 (Small): 10x8x1.5<br>• Model FWTM36 (Medium):<br>20x13x1.5<br>• Model FWTL44 (Large):<br>25x16x1.5 | Dimensions stated as Length x Width<br>x Thickness in cm. Three sizes<br>(small, medium, and large) in one<br>thickness:<br>• Model UN1083S (Small): 10x8x3<br>• Model UN20133M (Medium):<br>20x13x3<br>• Model UN25163L (Large):<br>25x16x3 |
| Kit<br>Components | One foam dressing, occlusive drape(s)<br>(1 in the small kit and 2 in the medium<br>and large kits), and one dome<br>assembly (dome, skirt, and drainage<br>tubing with luer lock and clamp) | One foam dressing, occlusive drape(s)<br>(1 in the small kit and 2 in the medium<br>and large kits), and one dome<br>assembly (dome, skirt, and drainage<br>tubing with luer lock and clamp) |
| Kit Component<br>Function | • Foam wound dressing is used to<br>pack the wound.<br>• Occlusive drape creates and<br>maintains a seal over the peri-<br>wound area and foam dressing.<br>• Dome Assembly provides a<br>pathway for wound exudate to flow<br>from the wound/dressing to the<br>pump exudate canister. | • Foam wound dressing is used to<br>pack the wound.<br>• Occlusive drape creates and<br>maintains a seal over the peri-<br>wound area and foam dressing.<br>• Dome Assembly provides a<br>pathway for wound exudate to<br>flow from the wound/dressing to<br>the pump exudate canister. |
| Foam Dressing<br>Material | Black Polyether Polyurethane Foam | Polyether Polyurethane Foam |
| Dome Material | Thermoplastic Elastomer | Thermoplastic Elastomer |
| Occlusive<br>Drape<br>Material | Semipermeable, polyurethane<br>(polymeric) transparent film | Semipermeable, polyurethane<br>(polymeric) transparent film |
| Skirt Material | Polyurethane Medical Tape with<br>Adhesive Backing | Polyurethane Medical Tape with<br>Adhesive Backing |
| Tubing<br>Material | PVC | PVC |
| Dressing<br>Materials<br>Biocompatible | Yes | Yes |
| NPWT pump<br>provided with | No | No |
| Feature | Theia NPWT<br>Dressing Kit<br>(Subject Device) | Theia NPWT<br>Dressing Kit<br>(Predicate Device) |
| dressing kit | | |
| Dressing Kit<br>Provided<br>Sterile | Yes | Yes |
| Sterilization<br>Method | Gamma Radiation | Gamma Radiation |
| Packaging | Tyvek pouch | Tyvek pouch |
Table 1: Subject vs. Predicate Theia NPWT Foam Wound Dressing Kit Comparison
{6}------------------------------------------------
{7}------------------------------------------------
#### PERFORMANCE DATA VII.
The following performance data were provided in support of the substantial equivalence determination:
## Bench Performance Testing
Bench testing has been performed to demonstrate that the 1.5cm or 3.0cm Theia NPWT Foam Wound Dressing Kits used with the CCT Mini or CCT1 pumps are capable of effectively removing wound fluids and that pressure is delivered uniformly across the dressing and in accordance with the pump pressure setting. The subject NPWT systems performed as intended in the simulated wound and pressure maintenance test setups, and all results were acceptable.
## Animal Performance Study
No animal performance studies were conducted for the Theia NPWT Foam Wound Dressing Kit.
## Clinical Study
No clinical studies were conducted for the Theia NPWT Foam Wound Dressing Kit.
#### CONCLUSIONS VIII.
The subject and predicate Theia NPWT Foam Wound Dressing Kits have the same Intended Use. The technological characteristics comparison and results of the verification testing provide evidence that the subject Theia NPWT Foam Wound Dressing Kits are substantially equivalent to the predicate Theia NPWT Foam Wound Dressing Kits (K161570), and the performance of these dressing kits when used in conjunction with CCT pumps (K082311) is substantially equivalent to the performance of the predicate dressing kits used in conjunction with SIMEX pumps (K113291).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.