← Product Code [FXX](/productcode/FXX) · K211260

# Sheng Jing Tang Surgical Mask (K211260)

_Liaoning Shengjingtang Biotechnology Co., Ltd. · FXX · Jan 10, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211260

## Device Facts

- **Applicant:** Liaoning Shengjingtang Biotechnology Co., Ltd.
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Jan 10, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Sheng Jing Tang Surgical Mask is intended for use in healthcare settings, procedures during which a face mask is necessary to protect both patient and healthcare personnel from transfer of body fluids, microorganisms, and particulate material. The device is indicated to be over-the-counter use. Adult only. The device is disposable and is indicated for single use. The device is not provided sterile.

## Device Story

Sheng Jing Tang Surgical Mask is a 3-layer, flat-pleated, disposable face mask for adult use in healthcare settings. It consists of an outer and inner layer of spun-bond polypropylene and a middle melt-blown polypropylene filter layer. It features spandex/nylon ear loops and a polyethylene/galvanized iron wire nosepiece for fit. The device acts as a physical barrier to reduce exposure to body fluids, microorganisms, and particulates. It is intended for over-the-counter use by individuals to protect themselves and others. The mask is non-sterile and for single use only.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM F1862-17 (fluid resistance), ASTM F2299-17 (particulate filtration), ASTM F2101-19 (bacterial filtration), and EN 14683:2019/ASTM F2100-20 (differential pressure). Biocompatibility confirmed via ISO 10993-5 and ISO 10993-10.

## Technological Characteristics

3-layer flat-pleated mask. Materials: Spun-bond polypropylene (inner/outer), melt-blown polypropylene (filter), spandex/nylon (ear loops), polyethylene/galvanized iron wire (nose piece). Non-sterile. Dimensions: 17.5cm x 9.5cm. Complies with ASTM F1862, ASTM F2299, ASTM F2101, and EN 14683 standards.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Avianz® Surgical Face Mask ([K200847](/device/K200847.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

January 10, 2022

Liaoning Shengjingtang Biotechnology Co., Ltd. Hai Luan CEO Huamei (Dalian) Health Industry Co., Ltd RM. 956 NO. 60-2 Haixing St, Dalian Tax-Free Zone Dalian, 116001 China

Re: K211260

Trade/Device Name: Sheng Jing Tang Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: October 28, 2021 Received: December 1, 2021

Dear Hai Luan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211260

Device Name Sheng Jing Tang Surgical Mask

#### Indications for Use (Describe)

Sheng Jing Tang Surgical Mask is intended for use in healthcare settings, procedures during which a face mask is necessary to protect both patient and healthcare personnel from transfer of body fluids, microorganisms, and particulate material. The device is indicated to be over the-counter use. Adult only. The device is disposable and is indicated for single use. The device is not provided sterile.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

__ Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of Title 21. CFR Section 807.92.

The assigned 510(k) number: K211260 Date of Preparation: 11/02/2021

## 1. Contact Information

## 1.1 Applicant

Applicant Name: Liaoning Shengjingtang Biotechnology Co., Ltd. Address: 123-11 Huishan St, Shenbei New District, Shenyang, Liaoning, China 110141 Contact Person: Li Ke Title: General Manager Telephone: +86-135-914-61018 Email: 174866792@qq.com

## 1.2 Consultant

Company: Huamei (Dalian) Health Industry Co., Ltd Address: RM. 956 NO.60-2 Haixing St, Dalian Tax-Free Zone, China, 116001 Contact Person: Luan Hai Telephone: +1 503-559-0930 Email: usfdahuamei@hotmail.com

## 2. Device Information

Trade Name: Sheng Jing Tang Surgical Mask Common Name: Surgical Face Mask Regulatory information: Classification Name: Mask, Surgical Classification: II Product Code: FXX Regulation Number: 21 CFR 878.4040 Review Panel: General Hospital

## 3. Legally Marketed Primary Predicate Device

Product name: Avianz® Surgical Face Mask 510(k) Number: K200847 Product Code: FXX Manufacture: MEXPO INTERNATIONAL INC.

## 4. Indication For Use

Sheng Jing Tang Surgical Mask is intended for use in healthcare settings, procedures during which a face mask is necessary to protect both patient and healthcare personnel from transfer of body fluids, microorganisms, and particulate material. The device is indicated to be over-the-counter use. Adult only. The device is disposable and is indicated for single use. The device is not provided sterile.

### 5. Device Description

The Sheng Jing Tang Surgical Masks are composed of 3-layers and are flat-pleated provided blue color, the mask materials consist of an outer layer (Spun-bond polypropylene), Inner layer (Spun-bond polypropylene),

Sheng Jing Tang Surgical Mask

{4}------------------------------------------------

Middle Layer filter (Polypropylene Melt-blown Cloth), Ear-loops (Spandex + Nylon, Ear Loops way for wearing) and Nose piece (a Polyethylene + Galvanized Iron wire nosepiece to provide a firm fit over the nose and then to secure the mask over the users' mouth and face). The Sheng Jing Tang Surgical Masks are single use, disposable device, adult only, provided non-sterile.

| Comparison Items | Sheng Jing Tang Surgical Mask                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device<br>Avianz® Surgical Face Mask<br>(K200847)                                                                                                                                                                                                                                                                                                                                                                            | Comparison |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code     | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                                   | FXX                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same       |
| Regulation       | 21 C.F.R Section 878.4040                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 C.F.R Section 878.4040                                                                                                                                                                                                                                                                                                                                                                                                              | Same       |
| Classification   | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                                                                               | Same       |
| Intended use     | Sheng Jing Tang Surgical Mask is<br>intended for use in healthcare<br>settings, procedures during which<br>a face mask is necessary to<br>protect both patient and<br>healthcare personnel from transfer<br>of body fluids, microorganisms,<br>and particulate material.<br>The device is indicated to be<br>over-the-counter use. Adult only.<br>The device is disposable and is<br>indicated for single use.<br>The device is not provided sterile. | The Disposable Surgical<br>Face Masks are intended to<br>be worn to protect both the<br>patient and healthcare<br>personnel from the transfer<br>of microorganisms, body<br>fluids, and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure to blood and body<br>fluids. This is a single-use,<br>disposable device(s)<br>provided non-sterile. | Same       |
| Mask Style       | Flat-pleated, earloop, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                                       | Flat-pleated, earloop, 3 layers                                                                                                                                                                                                                                                                                                                                                                                                        | Same       |

## 6. Substantially Equivalent (SE) Comparison

## 6.1 Material Used

| Comparison Items | Sheng Jing Tang Surgical<br>Mask       | Predicate Device<br>Avianz® Surgical Face Mask<br>(K200847) | Comparison  |
|------------------|----------------------------------------|-------------------------------------------------------------|-------------|
| Inner layer      | Spun-bond polypropylene                | Spun-bond polypropylene                                     | Same        |
| Middle layer     | Polypropylene Melt-blown<br>cloth      | Melt blown polypropylene filter                             | Same        |
| Outer layer      | Spun-bond polypropylene                | Spun-bond polypropylene                                     | Same        |
| Earloop          | Spandex + Nylon                        | Polyester                                                   | *Different¹ |
| Nose piece       | Polyethylene +<br>Galvanized Iron wire | Malleable aluminum wire                                     | *Different¹ |

{5}------------------------------------------------

# 6.2 Design Features

| Comparison Items   |           | Predicate Device                                         | Comparison                             |      |
|--------------------|-----------|----------------------------------------------------------|----------------------------------------|------|
|                    |           | Sheng Jing Tang Surgical Mask Avianz® Surgical Face Mask |                                        |      |
|                    |           | (K200847)                                                |                                        |      |
| Use Environment    |           | OTC, or Healthcare or related settings                   | OTC, or Healthcare or related settings | Same |
| Single-use/<br>use | Multiple- | Single-use                                               | Single-use                             | Same |
| Color              |           | Blue                                                     | Blue                                   | Same |

## 6.3 Sterility

| Comparison Items | Sheng Jing Tang Surgical Mask | Predicate Device<br>Avianz® Surgical Face Mask (K200847) | Comparison |
|------------------|-------------------------------|----------------------------------------------------------|------------|
| Sterilization    | No-Sterile                    | No-Sterile                                               | Same       |

## 6.4 Dimension

| Comparison Items | Sheng Jing Tang Surgical Mask | Predicate Device<br>Avianz® Surgical Face Mask<br>(K200847) | Comparison  |
|------------------|-------------------------------|-------------------------------------------------------------|-------------|
| Length           | $17.5\pm0.875$ cm             | $17.5\pm1$ cm                                               | *Different1 |
| Width            | $9.5\pm0.475$ cm              | $9.5\pm1$ cm                                                | *Different1 |

# 6.5 Effectiveness / Performance

| Comparison Items                                                                                                                                                                                                                                         | Sheng Jing Tang Surgical Mask                                                                                                                                                                                   | Predicate Device<br>Avianz® Surgical Face Mask<br>(K200847)                                                                    | Comparison  |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------|
| Resistance to Penetration<br>by Synthetic blood<br>ASTM F1862/F1862M-<br>17 Standard Test Method<br>for Resistance of<br>Medical Face Masks to<br>Penetration by Synthetic<br>Blood (Horizontal<br>Projection of Fixed<br>Volume at a Known<br>Velocity) | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots<br><br>Lot 1: 32 Out of 32 pass at 120 mmHg<br><br>Lot 2: 32 Out of 32 pass at 120 mmHg<br><br>Lot 3: 32 Out of 32 pass at 120 mmHg                 | ≥ 30 Out of 32 pass at 120 mmHg                                                                                                | Same        |
| Particulate Filtration<br>Efficiency ASTM<br>F2299/F2299M-03<br>(2017) Standard Test<br>Method for Determining<br>the Initial Efficiency of                                                                                                              | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots<br><br>Lot 1: ≥ 98.24%                                                                                                                              | ≥ 99.9%                                                                                                                        | *Different² |
| Material Used in medical<br>Face Masks to<br>Penetration by<br>Particulates using Latex<br>Spheres                                                                                                                                                       | Lot 2: ≥ 98.78%<br>Lot 3: ≥ 98.82%                                                                                                                                                                              |                                                                                                                                |             |
| Bacterial Filtration<br>Efficiency ASTM<br>F2101-2019 Standard<br>Test Method for<br>Evaluating the Bacterial<br>Filtration Efficiency<br>(BFE) of Medical Face<br>Mask Materials, Using a<br>Biological Aerosol of<br>Staphylococcus aureus             | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots<br>Lot 1: ≥ 99.9%<br>Lot 2: ≥ 99.9%<br>Lot 3: ≥ 99.9%                                                                                               | ≥98%                                                                                                                           | *Different³ |
| Differential Pressure EN<br>14683:2019 Medical face<br>masks- Requirements<br>and test methods                                                                                                                                                           | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots<br>Lot 1: Average 5.56 mmH20/cm²<br>Lot 2: Average 5.51 mmH20/cm²<br>Lot 3: Average 4.97 mmH20/cm²<br>(EN 14683:2019, Annex C and<br>ASTM F2100-20) | 3.0 mmH20/cm² (MILM-<br>36954C)                                                                                                | *Different⁴ |
| Flammability 16 CFR<br>1610 Standard for the<br>Flammability of Clothing<br>Textiles Corrections                                                                                                                                                         | Class 1 Pass (96/96) 32 Samples<br>Each from 3 non-consecutive<br>lots<br>Lot 1: Class 1 Pass<br>Lot 2: Class 1 Pass<br>Lot 3: Class 1 Pass                                                                     | Class 1                                                                                                                        | Same        |
| 6.6 Safety                                                                                                                                                                                                                                               |                                                                                                                                                                                                                 |                                                                                                                                |             |
| Comparison<br>Items                                                                                                                                                                                                                                      | Sheng Jing Tang Surgical Mask                                                                                                                                                                                   | Predicate Device<br>Avianz® Surgical Face Mask<br>(K200847)                                                                    | Comparison  |
| In Vitro Cytotoxicity<br>Test<br>ISO 10993-5: 2009                                                                                                                                                                                                       | ISO 10993-5; Under the<br>conditions of the study, the<br>proposed device extract was<br>determined to be non-<br>cytotoxic.                                                                                    | ISO 10993-5; Under the<br>conditions of the study, the<br>proposed device extract was<br>determined to be non-<br>cytotoxic.   | Same        |
| Skin Irritation Test<br>ISO 10993-10:2010                                                                                                                                                                                                                | ISO 10993-10;<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to be<br>non-irritating.                                                                                   | ISO 10993-10;<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to be<br>non-irritating.  | Same        |
| Skin Sensitization Test<br>ISO 10993-10:2010                                                                                                                                                                                                             | ISO 10993-10;<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to be<br>non-sensitizing.                                                                                  | ISO 10993-10;<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to be<br>non-sensitizing. | Same        |
| Flammability                                                                                                                                                                                                                                             | Class 1                                                                                                                                                                                                         | Class 1                                                                                                                        | Same        |

{6}------------------------------------------------

{7}------------------------------------------------

"Different! The difference in the materials, colors, and sizes do not raise additional questions for the device's safety and effectiveness. The subject devices' biocompatibility evaluation test and performed on the final finished device, including all construction materials, color and sizes. The test results indicated that pass the requirements.

*Different? The test was performed in accordance with ASTM F2299-03 (Reapproved 2017) Standard Test Method for Determining the Initial Efficiency of Materials Used in Medical Face Masks to Penetration by Particulates Using Latex Spheres, to determine the Particle Filtration Efficiency. Here, Sheng Jing Tang Surgical Mask compares to Predicate Device that results are very closed by statistics. Thus, this difference does not substantially affect equivalence between the Sheng Surgical Mask and predicate devices.

*Different The test was performed in accordance with ASTM F2101-19 Standard Test Method for Evaluating the Bacterial Filtration Efficiency (BFE) of Medical Face Materials, using a Biological Aerosol of Staphylococcus aureus the bacterial filtration efficiency (BFE) of the test article. Here, Sheng Jing Tang Surgical Mask compares to Precisate Device that results are very closed by statistics. Thus, this difference does not substantially affect equivalence between the Sheng Jing Tang Surgical Mask and predicate devices.

*Different* The test result for different pressure for the proposed device is different from predicate devices. The differential pressure test standard has been updated to EN14683:2019, Annex C in ASTM F2100:2020. The differential pressure test of the proposed device was conducted according to EN14683:2019, Annex C, and the test result demonstrates that the differential pressure of the proposed device meets the requirements of level 2 specified in ASTM F2100:2020. Thus, this difference does not substantially affect equivalence between the Sheng Jing Tang Surgical Mask and predicate devices.

## 7. Nonclinical Test

Nonclinical tests were conducted to verify that the proposed device met all design specifications, as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

## 7.1 Performance Test:

A bench test was conducted on Sheng Jing Tang Surgical Mask for the proposed device to determine

Sheng Jing Tang Surgical Mask 510(k) number: K211260 

{8}------------------------------------------------

substantial equivalence. The bench tests include the following tests:

| Performance Test                                                                | Sheng Jing Tang Surgical Mask                                                                                                                           | Acceptance Criteria              | Result |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--------|
| Fluid Resistance Performance<br>ASTM F1862-17                                   | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots                                                                                             | $\geq$ 29 of 32 pass at 120 mmHg | Passed |
|                                                                                 | Lot 1: 32 Out of 32 pass at 120 mmHg                                                                                                                    |                                  |        |
|                                                                                 | Lot 2: 32 Out of 32 pass at 120 mmHg                                                                                                                    |                                  |        |
|                                                                                 | Lot 3: 32 Out of 32 pass at 120 mmHg                                                                                                                    |                                  |        |
| Particulate Filtration<br>Efficiency ASTM F2299-17                              | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots                                                                                             | $\geq$ 98%                       | Passed |
|                                                                                 | Lot 1: $\geq$ 98.24%                                                                                                                                    |                                  |        |
|                                                                                 | Lot 2: $\geq$ 98.78%                                                                                                                                    |                                  |        |
|                                                                                 | Lot 3: $\geq$ 98.82%                                                                                                                                    |                                  |        |
| Bacterial Filtration Efficiency<br>ASTM F2101-19                                | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots                                                                                             | $\geq$ 98%                       | Passed |
|                                                                                 | Lot 1: $\geq$ 99.9%                                                                                                                                     |                                  |        |
|                                                                                 | Lot 2: $\geq$ 99.9%                                                                                                                                     |                                  |        |
|                                                                                 | Lot 3: $\geq$ 99.9%                                                                                                                                     |                                  |        |
| Differential Pressure (Delta P) EN<br>14683:2019, Annex C and ASTM<br>F2100-20. | Pass (96/96) 32 Samples Each<br>from 3 non-consecutive lots                                                                                             | <6.0 mmH20/cm²                   | Passed |
|                                                                                 | Lot 1: Average 5.56 mmH20/cm²                                                                                                                           |                                  |        |
|                                                                                 | Lot 2: Average 5.51 mmH20/cm²                                                                                                                           |                                  |        |
|                                                                                 | Lot 3: Average 4.97 mmH20/cm²                                                                                                                           |                                  |        |
|                                                                                 | ASTM F2100-20)                                                                                                                                          |                                  |        |
| Flammability 16 CFR 1610                                                        | Class 1 Pass (96/96) 32 Samples<br>Each from 3 non-consecutive<br>lots<br><br>Lot 1: Class 1 Pass<br><br>Lot 2: Class 1 Pass<br><br>Lot 3: Class 1 Pass | Class 1                          | Passed |

{9}------------------------------------------------

## 7.2 Biocompatibility Evaluation and Test

Biocompatibility evaluation conducted in accordance with the FDA's 2016 guidance supports that the subject devices are biocompatible.

The biocompatibility test includes the following tests:

- 7.2.1 In Vitro Cytotoxicity Test (ISO 10993-5: 2009 )
- 7.2.2 Skin Irritation Test(ISO 10993-10:2010)
- 7.2.3 Skin Sensitization Test (ISO 10993-10:2010)

| Biocompatibility<br>Evaluation and Test                  | Purpose                                                                                                                  | Acceptance<br>Criteria | Result |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------|--------|
| 7.2.1<br>In Vitro Cytotoxicity Test<br>ISO 10993-5: 2009 | ISO 10993-5; Under the conditions of the study, the<br>proposed device extract was determined to be non-<br>cytotoxic.   | Non-cytotoxic.         | Passed |
| 7.2.2<br>Skin Irritation Test<br>ISO 10993-10:2010       | ISO 10993-10; Under the conditions of the study, the<br>proposed device extract was determined to be non-<br>irritating. | Non-irritating.        | Passed |
| 7.2.3<br>Skin Sensitization Test<br>ISO 10993-10:2010    | ISO 10993-10.<br>Under the conditions of the study, the proposed device<br>extract was determined to be non-sensitizing. | Non-sensitizing        | Passed |

## 8. Clinical Test Conclusion

No clinical study is included in this submission.

## 9. Conclusion

The conclusions drawn from the nonclinical tests that demonstrate that Sheng Jing Tang Surgical Mask is as safe, as effective and performs as well as the legally marketed device identified in this submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K211260](https://fda-staging.innolitics.com/device/K211260)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).