ViperCath XC Peripheral Exchange Catheter

K211251 · Cardiovascular Systems, Inc. · DQY · May 26, 2021 · Cardiovascular

Device Facts

Record IDK211251
Device NameViperCath XC Peripheral Exchange Catheter
ApplicantCardiovascular Systems, Inc.
Product CodeDQY · Cardiovascular
Decision DateMay 26, 2021
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The ViperCath™ XC Peripheral Exchange Catheter is intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges, and provide a conduit for the delivery of saline or diagnostic contrast agents.

Device Story

ViperCath XC Peripheral Exchange Catheter is a percutaneous catheter used during peripheral vascular interventions. Device facilitates vascular access; supports guidewire; enables guidewire exchanges; provides conduit for saline or diagnostic contrast agent delivery. Operated by physicians in clinical settings. Device features braid-reinforced shaft; compatible with 0.035" guidewires; includes radiopaque marker bands at 1 cm and 5 cm from distal tip. Available in 65, 90, 135, and 150 cm lengths. Supplied sterile for single use. Output is physical guidance and fluid delivery path; assists physician in navigating vasculature; benefits patient by enabling efficient interventional procedures.

Clinical Evidence

Bench testing only. Dimensional verification and torque-to-failure testing performed to validate device modifications.

Technological Characteristics

Braid-reinforced shaft; 0.035" guidewire compatibility; radiopaque marker bands at 1 cm and 5 cm; 600 psi maximum injection pressure; sterile; single-use; non-hydrophilic coating.

Indications for Use

Indicated for patients undergoing peripheral vascular intervention procedures requiring vascular access, guidewire support, guidewire exchange, or delivery of saline/diagnostic contrast agents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ May 26, 2021 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Cardiovascular Systems, Inc. Erika Huffman Manager, Regulatory Affairs 1225 Old Hwy 8 NW St. Paul, Minnesota 55112 Re: K211251 Trade/Device Name: ViperCath™ XC Peripheral Exchange Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: April 23, 2021 Received: April 26, 2021 Dear Erika Huffman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211251 Device Name ViperCath™ XC Peripheral Exchange Catheter Indications for Use (Describe) The ViperCath™ XC Peripheral Exchange Catheter is intended to guide and support a guide wire during access of the vasculature, allow for wire exchanges, and provide a conduit for the delivery of saline or diagnostic contrast agents. Type of Use (Select one or both, as applicable) |--|--| X Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(K) SUMMARY | Submitter: | Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>Saint Paul, MN 55112 | |---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Erika Huffman, MS, RAC<br>Regulatory Affairs Manager<br>Cardiovascular Systems, Inc.<br>1225 Old Highway 8 NW<br>St. Paul, MN 55112<br>Ph: 651-259-1608<br>ehuffman@csi360.com | | Date Prepared: | April 23, 2021 | | Trade Name: | ViperCath™ XC Peripheral Exchange Catheter | | Common Name: | Percutaneous Catheter, 21 CFR 870.1250 | | Classification: | Class II | | Product Code: | DQY | | Predicate Device: | ViperCath XC Peripheral Exchange Catheter - K183000 | | Device Description: | The ViperCath XC Exchange Catheter is intended to be used to<br>facilitate vascular access and guidewire exchanges during<br>peripheral vascular intervention procedures. It is available in the<br>following lengths with a straight tip: 65, 90, 135, and 150 cm as<br>described in this submission. A 200 cm length is also available<br>and was previously cleared under K183000 with a straight tip<br>and an angled tip. There are no accessories for this device. The<br>catheter is compatible with .035" guidewires and has radiopaque<br>marker bands at distances of 1 cm and 5 cm from the distal tip.<br>The ViperCath XC is supplied sterile with one catheter per box. | | Intended Use: | The ViperCath™ XC Peripheral Exchange Catheter is intended<br>to guide and support a guide wire during access of the<br>vasculature, allow for wire exchanges, and provide a conduit for<br>the delivery of saline or diagnostic contrast agents. | {4}------------------------------------------------ | Comparison to<br>Predicate: | | ViperCath™ XC<br>Peripheral<br>Exchange<br>Catheter<br>(Subject Device) | ViperCath™ XC<br>Peripheral<br>Exchange<br>Catheter K183000<br>(Predicate Device) | |-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | | Effective<br>Length(s) | 65, 90, 135, 150 cm | 200 cm | | | Recommended<br>Guide Catheter | 5 Fr | 5 Fr | | | Shaft Design | Braid-reinforced | Braid-reinforced | | | Tip Shape | Straight | Straight, Angled | | | Maximum<br>Injection<br>Pressure | 600 psi | 600 psi | | | Guidewire<br>Compatibility | 0.035" | 0.035" | | | Hydrophilic<br>Coating | No | No | | | Sterile | Yes | Yes | | | Single Use | Yes | Yes | | Functional and<br>Safety Testing: | Dimensional checks and torque to failure testing were performed<br>to verify the modifications are acceptable and meet<br>specifications over the life of the device. | | | | Conclusion: | The ViperCath XC Exchange Catheter (femoral lengths) is<br>substantially equivalent to the predicate device and the<br>modifications do not introduce new questions of safety or<br>effectiveness. | | |
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...