← Product Code [KPS](/productcode/KPS) · K211247

# PET VCAR (K211247)

_GE Medical Systems SCS · KPS · Jul 27, 2021 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211247

## Device Facts

- **Applicant:** GE Medical Systems SCS
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Jul 27, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

PET VCAR (Volume Computer Assisted Reading) is a PET/CT software package which can be used by the clinician to assist in diagnosis, staging, treatment planning and monitoring treatment response.

## Device Story

PET VCAR is a PET/CT post-processing software package used by clinicians to assist in diagnosis, staging, and treatment monitoring. It takes PET/CT images as input; automatically highlights and bookmarks PET-defined regions of interest (ROI) based on user-defined thresholds; and propagates these bookmarks across longitudinal exams to facilitate comparison. The device measures Standard Uptake Value (SUV) and volume for metabolic activity. It includes automated tools for liver VOI placement, Deauville scoring, PERCIST evaluation, and bone lesion segmentation. Output is presented via a synchronized Summary Table and image display layouts, allowing clinicians to validate measurements visually. The software assists in clinical decision-making by streamlining workflow and providing quantitative data for follow-up. It does not perform automatic detection or diagnosis of pathology. Used in clinical settings on workstations or server platforms.

## Clinical Evidence

Bench testing only. Performance of the automatic reference region VOI placement algorithm was verified against defined acceptance criteria. No clinical data was required or provided.

## Technological Characteristics

Software-based post-processing package for PET/CT images. Operates on AW VolumeShare workstation and AW Server platforms. Complies with NEMA PS 3.1-3.20 (DICOM) standards. Implements automated algorithms for VOI placement, lesion segmentation, and standardized scoring (Deauville, PERCIST).

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- PET VCAR ([K063324](/device/K063324.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 27, 2021

Image /page/0/Picture/1 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

GE Medical Systems SCS % Ms. Elizabeth Mathew Senior Regulatory Affairs Manager 283, rue de la Miniere Buc, 78530 FRANCE

Re: K211247

Trade/Device Name: PET VCAR Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS. LLZ Dated: July 12, 2021 Received: July 13, 2021

Dear Ms. Mathew:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K211247

Device Name

#### PET VCAR

#### Indications for Use (Describe)

PET VCAR (Volume Computer Assisted Reading) is a PET/CT software package which can be used by the clinician to assist in diagnosis, staging, treatment plannitoring treatment response. PET VCAR automatically highlights and bookmarks PET defined regions of interest based on user-defined settings. The software can be used for visualization and analytical monitoring of disease progression or response to treatment or therapy using multi exam comparison. The software is designed to measure Standard Uptake Value (SUV) and volume for any PET defined metabolic activity. The software automatically propagates bookmarks from one time point to another for the purpose of improving analysis and workflow, which is designed to allow clinicians to make informed follow-up decisions in an efficient manner. PET VCAR does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure or function.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by a series of swirling, wave-like lines, also in blue, that give the impression of motion or energy.

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided. K211247

| Date:                      | July 22, 2021                                                                                                                  |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                 | GE Medical Systems SCS<br>Establishment Registration Number - 9611343<br>283, rue de la Minière<br>78530 Buc, France           |
| Primary Contact:           | Elizabeth Mathew<br>Senior Regulatory Affairs Manager<br>GE Healthcare<br>Tel: (262)424-7774<br>Email: Elizabeth.Mathew@ge.com |
| Secondary Contact          | Helen Peng<br>Senior Regulatory Affairs Director<br>Tel: (262)424-8222<br>Email: Hong.Peng@ge.com                              |
| Device Trade Name:         | PET VCAR                                                                                                                       |
| Common/Usual Name:         | PET VCAR                                                                                                                       |
| Primary Regulation Number: | 21 CFR 892.1200 Emission computed tomography system                                                                            |
| Primary Product Code:      | KPS                                                                                                                            |
| Secondary Product Code:    | LLZ                                                                                                                            |
| Classification:            | Class II                                                                                                                       |
| Predicate Device:          |                                                                                                                                |
| Device Name:               | PET VCAR                                                                                                                       |
| Manufacturer:              | GE Medical Systems SCS                                                                                                         |
| 510(k) number:             | K063324                                                                                                                        |
| Regulation Number:         | 21 CFR 892.1750 Computed tomography X-Ray system<br>21 CFR 892.1200 Emission computed tomography system                        |
| Product Code:              | JAK<br>KPS                                                                                                                     |
| Classification:            | Class II                                                                                                                       |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. Three curved, white shapes surround the circle, giving the impression of motion or energy.

## Device Description and Marketed Devices:

PET VCAR (Volume Computer Assisted Reading) is a PET/CT post processing software package which can be used by the trained clinician to assist in diagnosis, staging, treatment planning and monitoring treatment response. PET VCAR automatically highlights and bookmarks PET defined regions of interest based on user defined threshold settings. Bookmarks are any highlighted item including 2D or 3D region of interests and their associated measurements and annotations on the images. The software can be used for visualization and analytical monitoring of disease progression or response to treatment or therapy using multi exam comparison. It is designed to measure PET Standard Uptake Value (SUV) and volume for any PET defined metabolic activity. The software automatically propagates bookmarks from one time point to another for the purpose of improving analysis and workflow. Bookmark propagating is the feature that allows all bookmarks from the baseline exam of the patient to be automatically populated to subsequent exams of the same patient to facilitate the analysis of changes between the exams.

PET VCAR offers a tool called Summary Table that compiles and manages all the analytical information in an organized and interactive design. The Summary Table is synchronized with the image display layouts offering quick measurement / image visual validation. PET VCAR's workflow is designed to allow clinicians to make informed follow-up decisions in an efficient manner. PET VCAR does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure, or function.

### Intended Use:

PET VCAR (Volume Computer Assisted Reading) is a PET/CT software package which can be used by the clinician to assist in diagnosis, staging, treatment planning and monitoring treatment response.

### Indication for Use:

PET VCAR (Volume Computer Assisted Reading) is a PET/CT software package which can be used by the clinician to assist in diagnosis, staging, treatment planning and monitoring treatment response. PET VCAR automatically highlights and bookmarks PET defined regions of interest based on user-defined threshold settings. The software can be used for visualization and analytical monitoring of disease progression or response to treatment or therapy using multi exam comparison. The software is designed to measure Standard Uptake Value (SUV) and volume for any PET defined metabolic activity. The software automatically propagates bookmarks from one time point to another for the purpose of improving analysis and workflow, which is designed to allow clinicians to make informed follow-up decisions in an efficient manner. PET VCAR does not provide or claim any automatic detection or automatic diagnosis of abnormal anatomy, structure or function.

### Technology:

The proposed device PET VCAR employs the same fundamental scientific technology as its predicate device.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The circle is surrounded by a decorative pattern of swirling lines, also in blue. The logo is simple, recognizable, and represents the General Electric brand.

# Comparison:

The table below summarizes the key feature/technological differences and similarities between the predicate device and the proposed device:

| Specification                                                     | Predicate Device:<br>PET VCAR (K063324)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Proposed Device:<br>PET VCAR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison                                                                                                                                                                                                                                                                                                                                                                                                               |
|-------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                               | PET VCAR (Volume<br>Computer Assisted<br>Reading) is a<br>PET/CT software<br>package which can<br>be used by the<br>clinician to assist in<br>diagnosis, staging,<br>treatment planning<br>and monitoring<br>treatment<br>response. PET VCAR<br>automatically<br>highlights and<br>bookmarks PET<br>defined regions of<br>interest based on<br>user-defined<br>threshold settings.<br>The software can<br>be used for<br>visualization and<br>analytical<br>monitoring of<br>disease progression<br>or response to<br>treatment or<br>therapy using multi<br>exam comparison.<br>The software is<br>designed to<br>measure Standard<br>Uptake Value (SUV)<br>and volume for any<br>PET defined<br>metabolic activity.<br>The software<br>automatically<br>propagates<br>bookmarks from<br>one time point to<br>another for the | PET VCAR (Volume<br>Computer Assisted<br>Reading) is a<br>PET/CT software<br>package which can<br>be used by the<br>clinician to assist in<br>diagnosis, staging,<br>treatment planning<br>and monitoring<br>treatment<br>response. PET VCAR<br>automatically<br>highlights and<br>bookmarks PET<br>defined regions of<br>interest based on<br>user-defined<br>threshold settings.<br>The software can<br>be used for<br>visualization and<br>analytical<br>monitoring of<br>disease progression<br>or response to<br>treatment or<br>therapy using multi<br>exam comparison.<br>The software is<br>designed to<br>measure Standard<br>Uptake Value (SUV)<br>and volume for any<br>PET defined<br>metabolic activity.<br>The software<br>automatically<br>propagates<br>bookmarks from<br>one time point to<br>another for the | Substantially Equivalent.<br><br>The Proposed Device's<br>Indications for Use is simplified<br>by removing the underlined text<br>from those of the predicate. We<br>also changed the name of<br>"Interactive Data Analysis (IDA)<br>spreadsheet "to the “summary<br>table".<br><br>These minor changes do not<br>change the clinical meaning of<br>the indications and are made for<br>improved clarity and simplicity. |
| Specification                                                     | Predicate Device:<br>PET VCAR (K063324)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Proposed Device:<br>PET VCAR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                   | improving analysis<br>and workflow. PET<br>VCAR offers a tool<br>called Interactive<br>Data Analysis (IDA)<br>spreadsheet that<br>compiles and<br>manages all the<br>analytical<br>information in an<br>organized and<br>interactive design.<br>The IDA is<br>synchronized with<br>the image display<br>layouts offering<br>quick measurement<br>/ image visual<br>validation. PET<br>VCAR's workflow is<br>designed to allow<br>clinicians to make<br>informed follow-up<br>decisions in an<br>efficient manner.<br>PET VCAR does not<br>provide or claim<br>any automatic<br>detection or<br>automatic<br>diagnosis of<br>abnormal anatomy,<br>structure or<br>function.                                                                                                                                                     | improving analysis<br>and workflow,<br>which is designed<br>to allow clinicians<br>to make informed<br>follow-up decisions<br>in an efficient<br>manner. PET VCAR<br>does not provide or<br>claim any<br>automatic<br>detection or<br>automatic<br>diagnosis of<br>abnormal anatomy,<br>structure or<br>function.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                          |
| PET lesion<br>Segmentation                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantially Equivalent<br>The subject device additionally<br>provides user flexibility to manually<br>edit the bookmarks if desired.                                                                                                                                                                                                                                                                                   |
| Automatic VOI<br>placement in the<br>liver                        | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantially Equivalent<br>Subject device places an initial VOI<br>in the liver for user convenience.                                                                                                                                                                                                                                                                                                                   |
| Deauville Score                                                   | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantially Equivalent<br>Subject device implements<br>Deauville Score                                                                                                                                                                                                                                                                                                                                                 |
| Specification                                                     | Predicate Device:<br>PET VCAR (K063324)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Proposed Device:<br>PET VCAR                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Comparison                                                                                                                                                                                                                                                                                                                                                                                                               |
| PERCIST                                                           | Not Available                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Substantially equivalent<br>Subject device implements PERCIST<br>evaluation Criteria                                                                                                                                                                                                                                                                                                                                     |
| EORTC                                                             | Yes<br>(customer defined<br>threshold)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Yes<br>(automated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Substantially Equivalent                                                                                                                                                                                                                                                                                                                                                                                                 |
| Bone lesion<br>identification and<br>lesion extent<br>calculation | Manual bone<br>segmentation &<br>PET Lesion<br>Segmentation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Automated Bone<br>Segmentation &<br>PET Lesion<br>Segmentation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Substantially Equivalent                                                                                                                                                                                                                                                                                                                                                                                                 |

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circular frame. The color of the logo is a light blue. 

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a circular frame. The logo is colored in a light blue hue, and the background is white.

## Determination of Substantial Equivalence:

### Summary of Non-Clinical, Design Control Testing

PET VCAR has successfully completed the design control testing per GE's quality system. It was designed and will be manufactured under the Quality System Regulations of 21CFR 820 and ISO 13485. No additional hazards were identified, and no unexpected test results were observed. The proposed device complies with NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard.

The following quality assurance measures were applied to the development of the device:

- · Requirements Definition
- Risk Analysis
- Technical Design Reviews
- · Formal Design Reviews
- Software Development Lifecycle
- · Performance testing (Verification, Validation)
- · Safety Testing (Verification)

In addition, bench testing was performed to evaluate the performance of the automatic reference region VOI placement algorithm and the algorithm passed the defined acceptance criteria.

The proposed PET VCAR has been successfully verified on the AW VolumeShare workstation and AW Server platforms. All the testing and results did not raise new or different questions of safety and effectiveness other than those already associated with predicate device.

Software documentation for a MODERATE level of concern.

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script, enclosed within a blue circle. The circle has small, wave-like embellishments around its perimeter, giving it a slightly decorative appearance. The logo is simple, recognizable, and associated with the multinational conglomerate corporation.

## Substantial Equivalence Conclusion:

PET VCAR has identical or equivalent technological characteristics as its predicate device.

GE's quality system's design, verification, and risk management processes did not identify any new questions of safety or effectiveness, hazards, unexpected results, or adverse effects stemming from the changes to the predicate.

Based on development under GE Healthcare's quality system, successful design verification, software documentation for a "Moderate" level of concern, along with the engineering bench testing, GE Healthcare believes that the proposed PET VCAR is substantially equivalent to, and hence as safe and as effective for its Intended Use as the legally marketed predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K211247](https://fda-staging.innolitics.com/device/K211247)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
