K211194 · Molekule, Inc. · FRA · Aug 20, 2021 · General Hospital
Device Facts
Record ID
K211194
Device Name
Molekule Air Pro
Applicant
Molekule, Inc.
Product Code
FRA · General Hospital
Decision Date
Aug 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Molekule Air Pro air purifier is a device intended for medical purposes that is used to capture 95% of particulate matter and destroy bacteria, mold, and viruses by exposure to ultraviolet radiation when operated in Auto Mode Standard or manual mode at fan speed 2 or higher. The Molekule Air Pro air purifier has been demonstrated to entrain and destroy the following bioaerosols under the following exposure/working conditions: Test Item Average Net Log Reduction / Time @ Fan Speed 6. Room Temperature Test Bacteria Escherichia Coli 4.20 ± 0.11 / 90 minutes Bacteria Endospores Bacillus Subtilis 4.02 ± 0.23 / 30 minutes Mold Spores Aspergillus Brasiliensis 4.15 ± 0.06 / 60 minutes Virus MS2 bacteriophage 4.38 ± 0.15 / 30 minutes Test Item Single Pass Mechanical Filtration Efficiency Particulate Matter 0.3 to 1.0 micron size particles 95% or greater according to ASHRAE 52.2
Device Story
Molekule Air Pro is a freestanding ultraviolet air purifier for medical and home use. Device draws air via internal fan through a proprietary multi-layer PECO (photo electrochemical oxidation) filter. PECO technology uses UV-A light (320-400 nm) on a catalytic filter surface to generate hydroxyl radicals, which oxidize and destroy captured microorganisms (bacteria, mold, viruses). Device features an LCD capacitive touchscreen and integrated WLAN for mobile app control. Includes a particulate matter sensor and Auto Protect Mode to adjust fan speed based on detected air quality. Safety features include NFC-based filter authentication to prevent UV-A exposure and mechanical grates to prevent fan access. Output is purified air; clinical benefit is the reduction of airborne bioaerosols and particulate matter in the environment.
Clinical Evidence
No clinical data. Bench testing only. Efficacy demonstrated in sealed environmental bioaerosol chambers against E. coli, B. subtilis, A. brasiliensis, and MS2 bacteriophage, achieving >4-log reduction. Filtration efficiency verified per ASHRAE 52.2 (95% for 0.3-1.0 micron particles).
Indicated for use in medical facilities and home environments to capture 95% of particulate matter and destroy bacteria, mold, and viruses via UV-A exposure. Operates in Auto Mode Standard or manual mode at fan speed 2 or higher. Not for use in surgical suites or rooms with air separation devices.
Regulatory Classification
Identification
A medical ultraviolet air purifier is a device intended for medical purposes that is used to destroy bacteria in the air by exposure to ultraviolet radiation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 20, 2021
Molekule, Inc. % Adrienne Lenz Senior Medical Device Regulation Hymann, Phelps, & McNamara, P.C. 700 Thirteenth Street. N.W.. Suite 1200 Washington, District of Columbia 20005
Re: K211194
Trade/Device Name: Molekule Air Pro Regulation Number: 21 CFR 880.6500 Regulation Name: Medical Ultraviolet Air Purifier Regulatory Class: Class II Product Code: FRA Dated: July 16, 2021 Received: July 19, 2021
Dear Adrienne Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211194
#### Device Name Molekule Air Pro
#### Indications for Use (Describe)
The Molekule Air Pro air purifier is a device intended for medical purposes that is used to capture 95% of particulate matter and destroy bacteria, mold, and viruses by exposure to ultraviolet radiation when operated in Auto Mode Standard or manual mode at fan speed 2 or higher.
The Molekule Air Pro air purifier has been demonstrated to entrain and destroy the following bioaerosols under the following exposure/working conditions:
| Test Item | | Average Net Log Reduction / Time @ Fan Speed 6. Room Temperature Test |
|---------------------|----------------------------------|-----------------------------------------------------------------------|
| Bacteria | Escherichia Coli | $4.20 \u00b1 0.11$ / 90 minutes |
| Bacteria Endospores | Bacillus Subtilis | $4.02 \u00b1 0.23$ / 30 minutes |
| Mold Spores | Aspergillus Brasiliensis | $4.15 \u00b1 0.06$ / 60 minutes |
| Virus | MS2 bacteriophage | $4.38 \u00b1 0.15$ / 30 minutes |
| Test Item | | Single Pass Mechanical Filtration Efficiency |
| Particulate Matter | 0.3 to 1.0 micron size particles | 95% or greater according to ASHRAE 52.2 |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K211194 510(k) Summary
In accordance with 21 C.F.R. § 807.92 the following summary of information is provided:
DATE: July 16, 2021
#### SUBMITTER:
Molekule, Inc. 1301 Folsom St San Francisco, CA 94103 T 855-999-9069
#### PRIMARY CONTACT PERSON:
Adrienne R. Lenz Senior Medical Device Regulation Expert Hyman, Phelps, & McNamara, P.C. T 202-737-4292
### SECONDARY CONTACT PERSON:
Frank Bianco FDA Compliance Molekule, Inc. T 925-404-7724
### DEVICE:
TRADE NAME: Molekule Air Pro COMMON/USUAL NAME: Air Purifier CLASSIFICATION NAMES: Purifier, Air, Ultraviolet, Medical REVIEW PANEL: General Hospital PRODUCT CODE: FRA
### PREDICATE DEVICE(S):
Molekule Air Mini + Air Purifier, K202339
This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
## DEVICE DESCRIPTION:
The Molekule Air Pro air purifier is an ultraviolet air purifier that employs a photo electrochemical oxidation (PECO) ultraviolet air purification technology that destroys bacteria and viruses in air. The Molekule Air Pro may be used in medical facilities and in the home. A fan within the unit draws air up from vents in the bottom of the unit, up through the PECO filter (where it is exposed to low energy UV-A 320 - 400 nm light) and out through the vents on the top of the unit. PECO is an air purification technology that oxidizes microorganisms, including bacteria, molds and viruses. PECO works by shining UV-A light on the surface of the catalytic filter to initiate a chemical reaction that generates hydroxyl radicals. These radicals combine with microbiological contaminants that are captured on the filter. Once combined, a chemical reaction takes place destroying the contaminants.
The Molekule Air Pro is a freestanding device. It is a standalone device with an LCD display and a capacitive touchscreen user interface. Integrated WLAN provides a secondary means for controlling the device from the Molekule Android or iOS application.
Air Pro also includes a particulate matter sensor, particulate matter indicator, and Auto Protect Mode, which controls fan speed based on particulate levels detected in the use environment.
## INTENDED USE:
The Molekule Air Pro air purifier is a device intended for medical purposes that is used to capture 95% of particulate matter and destroy bacteria, mold, and viruses by exposure to ultraviolet radiation when operated in Auto Mode Standard or manual mode at fan speed 2 or higher.
The Molekule Air Pro air purifier has been demonstrated to entrain and destroy the following bioaerosols under the following exposure/working conditions:
| | Test Item | Average Net Log Reduction / Time @<br>Fan Speed 6. Room Temperature Test |
|------------------------|-------------------|--------------------------------------------------------------------------|
| Bacteria | Escherichia Coli | $4.20 \pm 0.11$ / 90 minutes |
| Bacteria<br>Endospores | Bacillus Subtilis | $4.02 \pm 0.23$ / 30 minutes |
{5}------------------------------------------------
| Test Item | | Average Net Log Reduction / Time @<br>Fan Speed 6. Room Temperature Test |
|--------------------|-------------------------------------|--------------------------------------------------------------------------|
| Mold Spores | Aspergillus<br>Brasiliensis | $4.15 \pm 0.06$ / 60 minutes |
| Virus | MS2 bacteriophage | $4.38 \pm 0.15$ / 30 minutes |
| Test Item | | Single Pass Mechanical Filtration<br>Efficiency |
| Particulate Matter | 0.3 to 1.0 micron size<br>particles | 95% or greater according to ASHRAE<br>52.2 |
# TECHNOLOGICAL CHARACTERISTIC COMPARISON:
The following table summarizes the similarities and differences between the subject and predicate devices.
{6}------------------------------------------------
| | Molekule Air Pro<br>air purifier | Molekule Air Mini+<br>air purifier (K202339) | Discussion of Differences |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Holder | Molekule | Molekule | Identical |
| Device Type | Medical Ultraviolet Air purifier | Medical Ultraviolet Air purifier | Identical |
| Product Code | FRA | FRA | Identical |
| Classification<br>Regulation | 21 C.F.R. § 880.6500 | 21 C.F.R. § 880.6500 | Identical |
| Class | II | II | Identical |
| Patient Population | Not specified | Not specified | Identical |
| Rx/OTC | OTC | OTC | Identical |
| User | Healthcare Professional<br>Lay User | Healthcare Professional<br>Lay User | Identical |
| | Molekule Air Pro | Molekule Air Mini+ | Discussion of Differences |
| | air purifier | air purifier (K202339) | |
| Indications for<br>Use | The Molekule Air Pro air purifier is a<br>device intended for medical purposes that<br>is used to capture 95% of particulate<br>matter and destroy bacteria, mold, and<br>viruses by exposure to ultraviolet radiation<br>when operated in Auto Mode Standard or<br>manual mode at fan speed 2 or higher.<br>The Molekule Air Pro air purifier has been<br>demonstrated to entrain and destroy the<br>following bioaerosols under the following<br>exposure/working conditions:<br>Average Net Log Reduction / Time @ Fan<br>Speed 6. Room Temperature Test<br>Escherichia Coli<br>4.20 +/- 0.11 / 90 mins<br>Bacillus Subtilis<br>4.02 +/- 0.23 / 30 mins<br>Aspergillus Brasiliensis<br>4.15 ± 0.06 / 60 mins<br>MS2 Bacteriophage<br>4.38 ± 0.15 / 30 mins<br>Single Pass Mechanical Filtration<br>Efficiency<br>Particulate Matter 0.3 to 1.0 micron size<br>particles 95% or greater according to<br>ASHRAE 52.2 | The Molekule Air Mini + air purifier is a<br>device intended for medical purposes that<br>is used to capture 95% of particulate<br>matter and destroy bacteria and viruses by<br>exposure to ultraviolet radiation when<br>operated in manual mode at fan speed 3 or<br>higher.<br>The Molekule Air Mini + air purifier has<br>been demonstrated to destroy the<br>following MS2 bacteriophage bioaerosol<br>entrained on the filter of the subject device<br>under the following exposure/working<br>conditions:<br>Average Maximum Log Reduction /<br>Entrainment Time @ Fan Speed 5. Room<br>Temperature Test<br>MS2 Bacteriophage<br>4.38 ± 0.15 / 30 mins<br>Single Pass Mechanical Filtration<br>Efficiency<br>Particulate Matter 0.3 to 1.0 micron size<br>particles 95% or greater according to<br>ASHRAE 52.2 | Based on additional testing, the indications<br>for use for the Air Pro were updated to<br>include molds, with specific results added<br>for bacteria and mold. Air Pro also has<br>additional testing to support a greater than<br>4-log reduction of MS2 bacteriophage at<br>fan speed 2, which is also the lowest<br>possible fan speed in Auto Mode<br>Standard. The indications were also<br>revised to clarify that the reductions<br>presented in the table are average net log<br>reductions from the combination of both<br>entrainment and destruction. These<br>differences in indications do not change<br>the intended use of the device. Both<br>devices are used to destroy<br>microorganisms in the air by the same<br>mechanism. |
| Environment of<br>Use | Hospital and other healthcare<br>setting.<br>Home healthcare. | Hospital and other healthcare<br>setting.<br>Home healthcare. | Identical |
| | Molekule Air Pro | Molekule Air Mini+ | Discussion of Differences |
| | air purifier | air purifier (K202339) | |
| Placement | Air Pro will work in any room, but<br>giving it space in a central location,<br>is recommended. Placement near<br>the patient is key. | Air Mini + will work in any room,<br>but giving it space in a central<br>location, and on a shelf, stand, or<br>table will help maximize air intake.<br>Placement near the patient is key. | The proposed device is designed<br>for larger rooms. They are not<br>placed on shelfs or tables due to the<br>different form factor. These<br>differences do not raise different<br>questions of safety or effectiveness.<br>Both devices meet applicable<br>standards for electrical safety, |
| | It should not be used in surgical<br>suites or in rooms with air<br>separation devices. | It should not be used in surgical<br>suites or in rooms with air<br>separation devices. | |
| | It is designed for rooms under 1000<br>ft². | It is designed for rooms under 250<br>ft². | including home levels, and have<br>been demonstrated to effectively<br>remove and destroy pathogens in<br>the air. |
| | | We recommend running it in the<br>room, closer to the bed. | |
| User Control | LCD screen with capacitive<br>touchscreen interface. User<br>Interface includes several dedicated<br>screens for fan speed control, PM<br>sensor readings in addition to other<br>administrative functions.<br>Application controls mimic device<br>touch panel | Touch panel with 5 manual fan<br>settings and one auto protect<br>setting,<br>Application controls mimic device<br>touch panel | The controls on the Air Pro and Air<br>Mini+ are simple and similar to<br>user controls on many other<br>devices that healthcare<br>professionals are accustomed to<br>using.<br>The app has the same functions as<br>the firmware and does not raise<br>different questions of safety or<br>effectiveness as demonstrated by<br>software verification and<br>validation. |
| | Molekule Air Pro<br>air purifier | Molekule Air Mini+<br>air purifier (K202339) | Discussion of Differences |
| Software | Basic Firmware and App, used to<br>turn the unit on, off, and change<br>fan speed. | Basic Firmware and App, used to<br>turn the unit on, off, and change<br>fan speed. | Identical |
| Mechanism of<br>Action | UV light of sufficient energy (UV-<br>A) activates photocatalyst that<br>destroys microorganisms entrained<br>on the filter through a<br>photochemical reaction. | UV light of sufficient energy (UV-<br>A) activates photocatalyst that<br>destroys microorganisms entrained<br>on the filter through a<br>photochemical reaction. | Identical |
| Installation | Free standing | Free standing | Identical |
| Catalytic Filter | Proprietary multi-layer filter media Dimensions: 8.1 in (diameter), 11.7 in Height. Pleats per inch: 2.6 pleats per inch of outer circumference Total Filter surface area: 1779 in2 Filter coated with proprietary photocatalyst and a metal wire mesh MERV16 | Proprietary multi-layer filter media Dimensions: 6.18 in (diameter), 6.55 in Height. Pleats per inch: 3 pleats per inch of outer circumference Total Filter surface area: 616 in2 Filter coated with proprietary photocatalyst and a metal wire mesh MERV16 | Filters of both devices are identical<br>in material, coated with the same<br>photocatalyst, and feature identical<br>layers. There are differences in the<br>dimensions related to the sizing<br>requirements for each device.<br>These differences do not raise<br>differences in safety or<br>effectiveness. |
| Photocatalyst | Proprietary catalyst (same as<br>predicate) | Proprietary catalyst | Identical |
| | Molekule Air Pro<br>air purifier | Molekule Air Mini+<br>air purifier (K202339) |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.