K211186 · Erchonia Corporation · NHN · Oct 22, 2021 · Physical Medicine
Device Facts
Record ID
K211186
Device Name
Erchonia XLR8
Applicant
Erchonia Corporation
Product Code
NHN · Physical Medicine
Decision Date
Oct 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Erchonia® XLR8 is generally indicated: a. as adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin b. as an adjunct to liposuction procedures of the thighs, hips and stomach for reduction of pain associated with the recovery process c. for the temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery d. as an adjunctive treatment of postoperative pain
Device Story
Erchonia® XLR8 is a low-level laser system (Class II) utilizing two semiconductor diodes emitting 640nm (±10nm) visible red light. Device features variable hertz pulsed wave output. Operated by healthcare professionals, the handheld device is manually scanned 3-6 inches over the treatment area. Powered by an internal lithium-ion polymer battery, recharged via an inductive base. Mechanism of action involves stimulating mitochondria to increase ATP production. Used in clinical settings to provide temporary pain relief; output affects clinical decision-making by serving as an adjunctive therapy for post-surgical and chronic musculoskeletal pain management, potentially reducing patient pain levels.
Clinical Evidence
No new clinical data submitted. Evidence relies on seven randomized controlled clinical trials from peer-reviewed literature evaluating LLLT for post-operative pain reduction in procedures including coronary artery bypass, tibial fracture, cesarean section, endodontic surgery, and tonsillectomy.
Technological Characteristics
Class II low-level laser system. Two semiconductor diodes; 640nm (±10nm) wavelength; 7.5mW (±2mW) output. Pulsed wave output. Powered by 3.7V 1800mAh Li-ion polymer battery with inductive charging. Handheld form factor. Complies with IEC 60601-1, IEC 60601-2, IEC 60825-1, and 21 CFR 1040.10/11.
Indications for Use
Indicated for patients requiring adjunctive treatment for minor chronic neck/shoulder musculoskeletal pain, pain associated with liposuction recovery (thighs, hips, stomach), post-surgery pain 24 hours after bilateral breast augmentation, and general adjunctive treatment of postoperative pain.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
October 22, 2021
Erchonia Corporation Travis Sammons Clinical Affairs Manager 650 Atlantis Road Melbourne, Florida 32904
Re: K211186
Trade/Device Name: Erchonia® XLR8 Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared Lamp Regulatory Class: Class II Product Code: NHN Dated: September 14, 2021 Received: September 21, 2021
Dear Travis Sammons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek Pinto, PhD Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211186
Device Name
Erchonia® XLR8
Indications for Use (Describe)
The Erchonia® XLR8 is generally indicated:
a. as adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin b. as an adjunct to liposuction procedures of the thighs, hips and stomach for reduction of pain associated with the recovery process
c. for the temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery
d. as an adjunctive treatment of postoperative pain
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Owner Information
### Name and Address of Sponsor / Manufacturer
Erchonia Corporation 650 Atlantis Rd. Melbourne, FL. 32904 Telephone: 321-473-1251 Fax: 321-473-1608
# Establishment Registration Number
2032513
## Name and Address of Official Correspondent
Erchonia Corporation 650 Atlantis Road Melbourne, FL 32904 Contact: Mr. Steven Shanks Telephone: 321-473-1251 Fax: 321-473-1608 Email: sshanks@erchonia.com
## Date Prepared
10/22/2021
## Device Information
Trade Name: Erchonia® XLR8 Laser Model#: HLS Common Name: Infrared Lamp Classification Name: Powered Light-Based Laser Non-Thermal with Non-Heating Effect for Adjunctive Use in Pain Therapy (21 CFR 890.5500) Classification: Class II Panel: Physical Medicine Product Code: NHN
## Predicate Device
The Erchonia® XLR8 is the exact same model as the Erchonia® XLR8 which has previously received 510(k) market clearance under K130996 for the intended use: The Erchonia® XLR8™ laser is indicated for the following three indications:
a. adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin b. as an adjunct to liposuction procedures of the thighs. hips and stomach for reduction of pain associated with the recovery process
c. temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery
There have been no technical or design modifications made to the Erchonia® XLR8 since clearance under K130996.
{4}------------------------------------------------
## Device Description
The Erchonia® XLR8 (Model# HLS) is a low-level laser system that uses two semi-conductor diodes (visible red light), red: 640nm ±10. The Erchonia® XLR8 (Model# HLS) is a variable hertz device. The variable hertz feature of the Erchonia® XLR8 (Model# HLS) is a pulsed wave, defined as containing a selected series of breaks, variances. The Erchonia® XLR8 (Model# HLS) has been classified by the FDA/EC as a Class II/Ila device and a Class II/2 laser.
The Erchonia® XLR8™ laser is currently indicated for the following three indications:
a. adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin c. as an adjunct to liposuction procedures of the thighs. hips and stomach for reduction of pain associated with the recovery process
c. temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery
This Premarket Notification is requesting expansion of the indications for use to include: d. as an adjunctive treatment of postoperative pain
The Erchonia® Laser is applied externally and has proven through clinical trials to treat post-surgical pain associated with liposuction and breast augmentation surgery. The Erchonia ® XLR8 laser emits a 640-nanometer wavelength with a tolerance of ±10 nanometer, from each of the two laser are powered by an internal battery that is recharged using a separate inductive charging base powered by an external class II medical power supply. This configuration offers portability as well as consistency of power.
The internal battery powers the two specially created and patented electronic diodes with an output of <10mW red laser beam.
The Erchonia® (HLS) laser is manufactured in accordance to the Good Manufacturing Procedures consistent with national regulatory agencies; such as FDA, EU, HC, TGA, and Anvisa. Per ISO and FDA standards the device and laser are classified as Class II.
Each of these governing agencies requires specific labeling. All required labels are affixed according to the relevant codes. Each label is pictured and described in this manual. Additionally, the placement of each label, on the Erchonia ® device, is communicated.
The associated accessories include:
- Charging Base ●
- Power Supply
- Patient protective eyewear ●
{5}------------------------------------------------
## Intended Use
The Erchonia® XLR8 is generally indicated:
a. as adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin b. as an adjunct to liposuction procedures of the thighs, hips and stomach for reduction of pain associated with the recovery process
c. for the temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery
d. as an adjunctive treatment of postoperative pain
| Comparison of Technological Characteristics with the Predicate Device | | | |
|-----------------------------------------------------------------------|--|--|--|
| Device | Erchonia® XLR8 (Model# HLS) | Erchonia® XLR8 Laser (Model# HLS) |
|-------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | Unknown | K130996 |
| | Subject Device | Predicate Device |
| Technology | Coherent light-based laser non-thermal instrument | Coherent light-based laser non-thermal instrument |
| Power | 7.5mW ± 2mW | 7.5mW ± 2mW |
| Wavelength | 640nm ±10 | 640nm ±10 |
| Energy Source | Multi diode collected then line dispersed (coherent) | Multi diode collected then line dispersed (coherent) |
| Recommended<br>treatment duration<br>(use time) based on<br>clinical evidence | 0 - 12 minutes | 0 - 12 minutes |
| Total Joules Per<br>Minute | 0.45 J | 0.45 J |
| Power Supply | External Power Supply - (100-240Vac, 50-60Hz, 0.5A;<br>12Vdc 1.5A) that connect to the Inductive Charging Base -<br>(1.5A 12V), to charge the Battery (Lithium-ion Polymer<br>3.7V, 1800mAh, 6.7W) | External Power Supply - (100-240Vac, 50-60Hz, 0.5A;<br>12Vdc 1.5A) that connect to the Inductive Charging Base<br>- (1.5A 12V), to charge the Battery (Lithium-ion Polymer<br>3.7V, 1800mAh, 6.7W) |
| Application | Line pattern, manually scanned 3-6 inches over area of<br>treatment | Line pattern, manually scanned 3-6 inches over area of<br>treatment |
| Mechanism of<br>Action | Stimulates the mitochondria to increase the production of<br>ATP | Stimulates the mitochondria to increase the production of<br>ATP |
| Indication for Use | The Erchonia® XLR8 is generally indicated:<br>a. as adjunctive use in providing temporary relief of minor<br>chronic neck and shoulder pain of musculoskeletal origin<br>b. as an adjunct to liposuction procedures of the thighs, hips<br>and stomach for reduction of pain associated with the<br>recovery process<br>c. for the temporary reduction in post-surgery pain at 24<br>hours after surgery following bilateral breast augmentation<br>surgery<br>d. as an adjunctive treatment of postoperative pain | The Erchonia® XLR8™ laser is indicated for the<br>following three indications:<br>a. adjunctive use in providing temporary relief of minor<br>chronic neck and shoulder pain of musculoskeletal origin<br>b. as an adjunct to liposuction procedures of the thighs,<br>hips and stomach for reduction of pain associated with the<br>recovery process<br>c. temporary reduction in post-surgery pain at 24 hours<br>after surgery following bilateral breast augmentation<br>surgery |
| Product Code | NHN | NHN |
## Technological Characteristics Summary
The subject device (Erchonia® XLR8 Laser, Model# HLS) is the exact same model as the predicate device (Erchonia® XLR8 Laser, Model# HLS) previously submitted under K130996. There are no differences between the subject device and predicate that render the device not substantially equivalent nor affect safety or effectiveness, as the devices have the same design, material, energy source, wavelength, and based on same mechanism of action. There have been no technical or design modifications made to the Erchonia® XLR8 since under K130996. This 510k submission is to expand the indication of use.
{6}------------------------------------------------
## Performance Standards
The Erchonia® XLR8 Laser complies with FDA's performance standards for light-emitting products (21CFR 1040.10 and 21 CFR 1040.11 by Laser Notice #50).
### Risk Assessment
The Erchonia® XLR8, Model#: HLS is acceptable in accordance with IEC 60601 edition 3.1, by virtue of Engineering and third-party verification. All identified risks have been mitigated to ensure the lowest acceptable risk possible using the ISO 14971 standard framework.
### Electromagnetic Compatibility and Electrical Safety
Safety and EMC testing was conducted on the Erchonia® XLR8™ (Model#:HLS) device. The device complies with the IEC 60601-1, IEC 60601-2 and IEC 60825-1 standards.
### Compliance with Voluntary Standards
The Erchonia® XLR8 complies with the following voluntary standards: IEC 60601-1-2:2014 IEC 60601-1:2005 (Third Edition) + CORR. 1:2006 + CORR. 2:2007 + A1:2012 (or IEC 60601-1: 2012 reprint) IEC 60825-1:2014 (Third Edition)
## Performance Testing- Bench
Not Applicable, as the subject device and predicate are exactly the same
### Performance Testing-Animal
No animal testing conducted
## Performance Testing-Clinical
No clinical study results are being submitted as part of this submission.
## Summary of Clinical Testing
Seven randomized controlled clinical trials published in scientific peer-reviewed journal publications were identified wherein low-level laser therapy (LLLT) was evaluated for its therapeutic ability to reduce post- operative pain following surgical procedures. These studies evaluated post-surgical pain related to:
- Coronary artery bypass graft surgery
- . Coronary bypass surgery with internal mammary artery grafts
- . Tibial fracture surgery
- Cesarean section (2 studies)
- . Endodontic surgery
- . Tonsillectomy
No new clinical data is being submitted in this 510(k) submission.
{7}------------------------------------------------
## Comparison of Erchonia® XLR8 Laser (Model# HLS) & Erchonia EML Laser
The Erchonia® XLR8 Laser, Model# HLS (predicate & subject device) has previously received substantial equivalence to the Erchonia EML Laser by the FDA, reference 510(k) #K130996. Any technological differences between the Erchonia® XLR8 Laser (subject & predicate device) and Erchonia EML Laser did not render the device not substantially equivalent, did not affect the safety or effectiveness, or raise questions regarding the safety and effectiveness due to the fact the laser wavelength and total light energy delivered per treatment is identical.
| Device | Erchonia® XLR8 Laser (Model# HLS) | Erchonia EML Laser |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| | Subject & Predicate Device | |
| 510(k) | Unknown, K130996 | K041139, K072206 |
| Output - at aperture | 7.5mW | 7.5mW |
| Wavelength | 640nm ±10 | 640nm ±10 |
| Waveform | Pulsed | Pulsed |
| Energy Source | Multi diode collected then line dispersed (coherent) | Multi diode collected then line dispersed<br>(coherent) |
| Power Supply | External Power Supply - (100-240Vac, 50-60Hz,<br>0.5A; 12Vdc 1.5A) that connect to the Inductive<br>Charging Base - (1.5A 12V), to charge the Battery<br>(Lithium-ion Polymer 3.7V, 1800mAh, 6.7W) | 115/220 V ac 50/60 Hz electrical outlet,<br>rechargeable batteries |
| Energy Delivery | Device hand-held, probe on top | Handheld treatment probe |
| Target Size | Line pattern, manually scanned over area of treatment | Line pattern, manually scanned over area<br>of treatment |
| Mechanism of Action | Stimulates the mitochondria to increase the<br>production of ATP | Stimulates the mitochondria to increase the<br>production of ATP |
## Conclusion
The subject device (Erchonia® XLR8 Laser, Model# HLS) is the exact same model as the predicate device (Erchonia® XLR8 Laser, Model# HLS) previously submitted under K130996. There are no differences between the subject device and predicate that render the device not substantially equivalent nor affect safety or effectiveness, as the devices have identical technology, provides the same outputs, and based on same mechanism of action. Therefore, this 510k submission is to expand the indication of use.
Based on Erchonia Lasers previous received FDA 510(k) market clearances for post-operative use (K072206 & K041139), and supportive published literature of low level laser therapy (LLT) effectively reducing postoperative pain across various body locations and surgical procedures, the proven physiological mechanism of action provided by the Erchonia® XLR8™ laser has substantially equivalent safety and effectiveness to support the proposed Indication for Use:
The Erchonia® XLR8 is generally indicated:
a. as adjunctive use in providing temporary relief of minor chronic neck and shoulder pain of musculoskeletal origin b. as an adjunct to liposuction procedures of the thighs, hips and stomach for reduction of pain associated with the recovery process
c. for the temporary reduction in post-surgery pain at 24 hours after surgery following bilateral breast augmentation surgery
d. as an adjunctive treatment of postoperative pain
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.