Single Use Electrosurgical Knife with Fluid Pump System
K211172 · Micro-Tech (Nanjing) Co., Ltd. · GEI · Dec 17, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211172
Device Name
Single Use Electrosurgical Knife with Fluid Pump System
Applicant
Micro-Tech (Nanjing) Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 17, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Single Use Electrosurgical Knife: These instruments have been designed to be used with endoscopes and electrosurgical for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract. Fluid Pump System: The Fluid Pump System is a fluid delivery device that can inject fluid with adjustable flow rate and can be used with the Single Use Electrosurgical Knife for elevation of tissue layers by injection into the submucosa.
Device Story
Device comprises a single-use electrosurgical knife and a fluid pump system. Knife is an endoscopic tool used with an electrosurgical unit (ESU) to deliver high-frequency energy for tissue cutting and coagulation. Fluid pump system is an accessory that delivers sterile saline to the submucosa via a reciprocating pump mechanism to facilitate tissue elevation. System includes a console, foot pedal, disposable pump cartridge, tubing, and control button. Physician operates the device through an endoscope working channel (≥2.8mm). Fluid flow rate is adjustable (25-100 mL/min) via console touch screen or knob. Activation of cautery or fluid delivery is controlled by foot pedal or hand-held button. Device assists in endoscopic mucosal procedures by providing controlled tissue lifting and electrosurgical cutting, potentially improving procedural precision and patient outcomes.
Clinical Evidence
No clinical data or animal studies were included in this submission. Substantial equivalence is supported by bench testing, including dimension, actuation, sealing, flow, mucosa lift, cutting, and connection force performance, alongside biocompatibility (ISO 10993-1), sterilization validation (ISO 11135), and electrical safety/EMC (IEC 60601 series).
Technological Characteristics
System includes a reusable console/foot pedal and disposable components (pump cartridge, tubing, control button). Materials include metal, glass, and plastics. Energy source: ESU (HF energy) and electrical pump (100-240V, 50/60Hz). Sterilization: EO (SAL 10^-6) for disposables. Connectivity: Standalone system. Software: Complies with IEC 62304. Packaging: ASTM F1980-16 and ISO 11607-1/2 compliant.
Indications for Use
Indicated for patients undergoing endoscopic procedures in the digestive tract requiring dissection, elevation, irrigation, and preparation of tissue layers, including submucosal injection for lesion elevation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
ERBE WaterJet Model ERBEJET® 2 System with HybridAPC Probe (K143306)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
December 17, 2021
Micro-Tech (Nanjing) Co., Ltd. Sally He RA Engineer No.10 Gaoke Third Road, Nanjing National Hi-Tech Industrial Development Zone Nanjing, Jiangsu Province China
Re: K211172
Trade/Device Name: Single Use Electrosurgical Knife with Fluid Pump System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, FQH Dated: November 3, 2021 Received: November 8, 2021
Dear Sally He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K211172
Device Name
Single Use Electrosurgical Knife with Fluid Pump System
Indications for Use (Describe)
Single Use Electrosurgical Knife:
Tyne of Use (Select one or hoth, as annlicable)
These instruments have been designed to be used with endoscopes and electrosurgical for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
Fluid Pump System:
The Fluid Pump System is a fluid delivery device that can inject fluid with adjustable flow rate and can be used with the Single Use Electrosurgical Knife for elevation of tissue layers by injection into the submucosa.
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional block font. The letters are interconnected, with the "M" on the left and the "T" on the right. A registered trademark symbol (®) is positioned in the upper right corner of the logo. The logo has a clean and modern design.
## 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K211172
### 1. Date of Preparation: 2021-10-28
### 2. Sponsor Identification
## Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing, Jiangsu Province, PRC
Establishment Registration Number: 3004837686
Contact Person: Sally He
Position: RA Engineer
Tel: +86-25-58646395
Fax: +86-25-58350006
Email: RA.Micro-Tech@outlook.com
### 3. Identification of Proposed Device
Product Name: Single Use Electrosurgical Knife with Fluid Pump System
Trade Name: Single Use Electrosurgical Knife, Fluid Pump System
Common Name: Electrosurgical Knife; Jet
## Regulatory Information
Classification Name: Electrosurgical cutting and coagulation device and accessories, Jet lavage.
Classification: 2
Product Code: GEI and FQH
Regulation Number: 878.4400 and 880.5475
Review Panel: General & Plastic Surgery, General Hospital
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are bold and appear to be interconnected, creating a sense of unity and strength. A registered trademark symbol is visible in the upper right corner of the logo.
#### 4. Identification of Predicate Device/Reference Device
### Predicate Device
510(k) Number: K193601
Product Name: Single Use Electrosurgical Knife
Manufacturer: Micro-Tech (Nanjing) Co., Ltd
Reference Device
510(k) Number: K143306
Product Name: ERBE WaterJet Model ERBEJET® 2 System with HybridAPC Probe
Manufacturer: ERBE USA, Inc
#### ട്. Indications for Use
### Single Use Electrosurgical Knife:
These instruments have been designed to be used with endoscopes and electrosurgical for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
### Fluid Pump System:
The Fluid Pump System is a fluid delivery device that can inject fluid with adjustable flow rate and can be used with the Single Use Electrosurgical Knife for elevation of tissue layers by injection into the submucosa.
#### 6. Device Description
The proposed device Single Use Electrosurgical Knife with Fluid Pump System includes Single Use Electrosurgical Knife and Fluid Pump System
The Single Use Electrosurgical Knife is a sterile, single-use endoscopic device, intended to be used with endoscopes and electrosurgical units for dissection, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract. The Fluid Pump System is a fluid delivery device that can inject fluid with adjustable flow rate and can be used
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected and appear to be angled slightly, giving them a dynamic look. A registered trademark symbol is visible in the upper right corner of the logo.
with the Single Use Electrosurgical Knife for elevation of tissue layers by injection into the submucosa. The Fluid Pump System is included as an accessory to the Single Use Electrosurgical Knife.
The Single Use Electrosurgical Knife is used with the Electrosurgical Unit (ESU), the ESU supplies high frequency (HF) energy through a retractable electrode of the Single Use Electrosurgical Knife for the cutting and coagulation of tissue. The Single Use Electrosurgical Knife consists of Cutting Knife, outer tube assembly, Middle Tube assembly and handle assembly. The Single Use Electrosurgical Knife is divided into I type, T type, O type and IT type according to the Cutting Knife. The Single Use Electrosurgical Knife has twenty-two (22) specifications; the main differences of these specifications are Cutting Knife Shape, Cutting Knife Length, and Effective Working Length and with injection function or without injection function.
Physicians would attach it to the ESU. For endoscopic procedures, the Single Use Electrosurgical Knife is placed down the channel of an endoscope that has a working channel equal to or greater than 2.8mm. Upon the setup of the ESU, the Single Use Electrosurgical Knife is ready for use. To activate cautery, the Cutting Knife is extended out and the ESU's footswitch is depressed. The subject devices with injection models can supply fluid into the submucosa to supply liquid, use a syringe or pump to connect with luer tap on the hand components.
The Fluid Pump System included in the proposed device is a fluid delivery device that can inject fluid with adjustable flow rate and can be used with the Single Use Electrosurgical Knife for elevation of tissue layers by injection into the submucosa.
The Fluid Pump System consists of a Foot Pedal, a Fluid Pump Console (with a Power Cord and a Control Button Cord), a Disposable Pump Cartridge, a Disposable Tubing Set, and a Disposable Control Button. The Disposable Pump Cartridge adopts the structure of a reciprocating pump, and the purpose of conveying sterile fluid is achieved through the piston's reciprocating movement. The Disposable Tubing Set is used to connect the Disposable Pump Cartridge's fluid outlet and Single Use Electrosurgical Knife's fluid injection port. The Fluid Pump Console provides a drive the Disposable Pump Cartridge to achieve reciprocating motion. The user can set the Effect value via the touch button on the touch screen or the physical knob. The Effect setting value range is 10 to 40, where Effect 10 corresponds to the minimum flow rate of 25mL/min, and Effect 40 corresponds to the
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are connected and appear to be leaning slightly to the left, giving the logo a dynamic feel. A circled "R" trademark symbol is positioned in the upper right corner of the logo.
maximum flow rate of 100mL/min. When Fluid Pump System used with Single Use Electrosurgical Knife, the recommended Effect is 10-30. The user can control the fluid delivery through the Foot Pedal or the Disposable Control Button, stepping on the Foot Pedal or pressing the Disposable Control Button to start fluid delivery, releasing the Foot Pedal or the Disposable Control Button, and immediately stop the fluid delivery.
The Single Use Electrosurgical Knife are EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of one year.
The Disposable Pump Cartridge, Disposable Tubing Set and Disposable Control Button of Fluid Pump System are EO sterilized to achieve the Sterility Assurance Level (SAL) of 10-6 and placed in a sterility maintenance package to ensure a shelf life of one year. The Foot Pedal and Fluid Pump Console of Fluid Pump System are non-sterile and reusable, the service life of the Foot Pedal and Fluid Pump Console are 8 years.
### 7. Comparison of Technological Characteristics
The proposed device Single Use Electrosurgical Knife with Fluid Pump System includes Single Use Electrosurgical Knife and Fluid Pump System. The Single Use Electrosurgical Knife is identical to the current Single Use Electrosurgical Knife cleared under K193601. Compared with the predicate device Single Use Electrosurgical Knife cleared under K193601, the only difference of the proposed device is the addition of Fluid Pump System, which is included as accessory of Single Use Electrosurgical Knife.
For the difference, we cited a reference device ERBE WaterJet Model ERBEJET® 2 System with HybridAPC Probe (K143306) to compare with the new added accessory Fluid Pump System.
| Characteristics | Fluid Pump System | Reference Device<br>ERBEJET® 2 System with<br>HybridAPC Probe<br>(K143306) | Remark |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indications for Use | The Fluid Pump System is a<br>fluid delivery device that can<br>inject fluid with adjustable flow<br>rate and can be used with the | The ERBEJET 2 is intended<br>for lifting mucosal lesions by<br>injection into the submucosa as well<br>as the cutting and dissection of soft | The intended use<br>of the proposed<br>device Fluid<br>Pump System is |
| | | Reference Device | |
| Characteristics | Fluid Pump System | ERBEJET® 2 System with<br>HybridAPC Probe<br>(K143306) | Remark |
| | Single Use Electrosurgical<br>Knife for elevation of tissue<br>layers by injection into the<br>submucosa. | tissue in neurosurgery and soft<br>tissue such as the liver, kidney, etc.<br>within the abdomen, including Total<br>Mesorectal Excision (TME) in open<br>as well as endoscopic surgery.<br>The HybridAPC probe is indicated<br>for the induction of sterile normal<br>saline into the submucosa to lift<br>mucosal lesions using direct<br>visualization through an endoscope<br>and for HF ablation of the mucosal<br>lesion by Argon Plasma Coagulation<br>(APC). | included in the<br>reference device<br>ERBEJET® 2<br>System. |
| Energy Delivered | Pressurized sterile saline for<br>submucosa lifting | Pressurized sterile saline for cutting,<br>dissecting and submucosa lifting | The energy<br>delivered is<br>included in<br>ERBEJET® 2<br>System. |
| Power Input | 100-240 V | 100-240 V | Same |
| Frequency | 50/60 Hz | 50/60 Hz | Same |
| Nozzle Diameter | 0.15 mm | 0.12 mm | Similar |
| Flow Rate | 25~100 mL/min<br>25~75 mL/min<br>(Recommend value when used<br>with Single Use<br>Electrosurgical Knife) | 1~55 mL/min | Similar |
| Components | Fluid Pump Console<br>Connecting Cables<br>Foot Pedal<br>Disposable Pump Cartridge<br>Disposable Pump Tubing<br>Disposable Control Button | Unit<br>Connecting Cables<br>Foot Switch<br>Pump Cartridge<br>Applicators<br>Suction System | The proposed<br>device doesn't<br>include<br>Applicators and<br>Suction System |
| Main Material | Metal, Glass Display Screen,<br>Plastics, Wiring, Insulation<br>Plastic, Metal | Metal, Glass Display Screen,<br>Plastics, Wiring, Insulation Plastic,<br>Metal | SE |
| Mucosal Lift<br>Performance | Deliver flow liquid for mucosal<br>lifting | Deliver flow liquid for mucosal<br>lifting | SE |
| Characteristics | Fluid Pump System | Reference Device<br>ERBEJET® 2 System with<br>HybridAPC Probe<br>(K143306) | Remark |
| Connection<br>Performance | properly connect and detach<br>from designated<br>equipment/accessory. Once<br>connected, there are no water<br>leaks. | properly connect and detach from<br>designated equipment/accessory.<br>Once connected, there are no water<br>leaks, argon gas leakages, or leakage<br>currents above international<br>standards. | device doesn't<br>include a probe<br>subsystem,<br>without argon gas<br>or currents<br>delivered. |
| Electrical<br>Performance | Comply with<br>IEC 60601-1:2012;<br>IEC60601-1-2:2014;<br>IEC 60529 (Foot Pedal Only) | Comply with<br>IEC 60601-1:2012;<br>IEC60601-1-2:2014;<br>IEC 60529 (Foot Pedal Only) | SE |
| Usability | Comply with<br>IEC 60601-1-6:2013;<br>IEC 62366-1:2015; | Comply with<br>IEC 60601-1-6:2013;<br>IEC 62366-1:2015; | SE |
| Software | Comply with<br>IEC 62304:2015; | Comply with<br>IEC 62304:2015; | SE |
| Biocompatibility | Comply with ISO 10993-1 | Comply with ISO 10993-1 | SE |
| Shelf Life | Fluid Pump<br>Console (with<br>connecting<br>cables), Foot<br>Pedal | Unit<br>Connecting<br>cables,<br>Foot Pedal<br>Suction System | SE |
| | Reusable | Reusable | |
| | Disposable<br>Pump Cartridge,<br>Disposable<br>Pump Tubing,<br>Disposable<br>Control Button | Pump Cartridge,<br>Applicators | |
| | 1 Year<br>Comply<br>with ASTM<br>F1980-16 | 3 Years<br>Comply with<br>ASTM<br>F1980-16 | |
| | | | |
| Sterilization | Fluid Pump<br>Console (with<br>connecting<br>cables), Foot<br>Pedal | Unit<br>Connecting<br>cables,<br>Foot Pedal<br>Suction System | SE |
| | Non-Sterile | Non-Sterile | |
| | Disposable<br>Pump Cartridge,<br>Disposable<br>Tubing Set,<br>Disposable<br>Control Button: | Pump Cartridge,<br>Applicators | |
| | EO<br>Sterilization,<br>SAL: 10-6. | EO<br>Sterilization,<br>SAL: 10-6. | |
| | | | |
| | Section 5 510k summary | | |
| Characteristics | Fluid Pump System | Reference Device<br>ERBEJET® 2 System with<br>HybridAPC Probe<br>(K143306) | Remark |
| | Fluid Pump<br>Console (with<br>connecting<br>cables), Foot<br>Pedal<br>Reusable | Unit<br>Connecting<br>cables,<br>Foot Pedal<br>Suction System<br>Reusable | |
| Single Use | Disposable<br>Pump Cartridge,<br>Disposable<br>Pump Tubing,<br>Disposable<br>Control Button<br>Single Use | Pump Cartridge,<br>Applicators<br>Single Use | SE |
| Labeling | Comply with 21 CFR part 801 | Comply with 21 CFR part 801 | SE |
### Comparison to Reference Devices
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with the letters 'MT' in a stylized, three-dimensional design. The letters are primarily white against a blue background, giving them a bold and prominent appearance. A registered trademark symbol '®' is visible in the upper right corner of the logo, indicating that the design is legally protected.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo with the letters 'MT' in a stylized, three-dimensional block font. The letters are white against a blue background, and there is a registered trademark symbol '®' in the upper right corner of the logo. The logo has a simple, modern design.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a blue, stylized logo. The logo appears to be the letters 'M' and 'T' in a bold, sans-serif font. The letters are designed with a 3D effect, giving them a blocky, isometric appearance. The logo is simple and modern, with a clean design.
Compared with the reference device, the Fluid Pump System incorporates substantially equivalent device materials, design, configuration, packaging, sterilization process and intended use as those featured in the reference device ERBEJET® 2 System.
### 8. Performance Data
The following bench tests were performed:
- > Dimension
- A Actuation Performance
- > Compatible Performance
- > Sealing Performance
- > Flowing Performance
- > Mucosa Lift Performance
- ア Cutting Performance
- > Connected Force Performance
Shelf-life testing and packaging integrity testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices and ISO 11607-1:2019: Packaging for terminally sterilized medical devices - Part 1:
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo with the letters "MT" in a bold, sans-serif font. The letters are rendered in a dark blue color and have a three-dimensional effect, with the top and right sides appearing to extend back and to the right, creating a sense of depth. A circled "R" symbol, indicating a registered trademark, is positioned in the upper right corner of the logo.
Requirements for materials, sterile barrier systems and packaging systems and ISO 11607-2:2019: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes. One year aging test was performed to demonstrate the stability and support the results of the accelerated aging test.
Sterilization validation was carried out in accordance with ISO 11135:2014+A1:2018 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices".
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO-10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on September 4, 2020.
Electromagnetic compatibility, electric safety had been confirmed according to the following standards:
- A IEC 60601-1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- > IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
Software of Fluid Pump System had been confirmed according to FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "IEC 62304:2015 Medical device software - Software life cycle processes".
The results of all the performance testing demonstrated that the proposed device met the predetermined acceptance criteria and is substantial equivalence to the predicate device Single Use Electrosurgical Knife (K193601).
#### 9. Animal Study
No animal study is included in this submission.
### 10. Clinical Study
No clinical study is included in this submission.
{11}------------------------------------------------
K211172
Image /page/11/Picture/1 description: The image shows a logo with the letters 'MT' in a stylized, three-dimensional design. The letters are primarily white, with a blue outline and shading that gives them a sense of depth. The 'M' is positioned to the left and slightly behind the 'T', creating an overlapping effect. The overall design is simple yet modern, with a clean and professional appearance.
## 11. Substantially Equivalent (SE) Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the Single Use Electrosurgical Knife with Fluid Pump System has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Single Use Electrosurgical Knife (K193601).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.