K211168 · Philips Healthcare (Suzhou) Co., Ltd. · JAK · Nov 22, 2021 · Radiology
Device Facts
Record ID
K211168
Device Name
Philips Incisive CT on Trailer
Applicant
Philips Healthcare (Suzhou) Co., Ltd.
Product Code
JAK · Radiology
Decision Date
Nov 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Incisive CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipments and accessories. The Incisive CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society. Besides installed in hospital, the proposed Philips Incisive CT may also be installed on trailer and be transported to designated locations for use. And Incisive CT installed on trailer has the same intended use as installed in hospital.
Device Story
Incisive CT on Trailer is a whole-body CT system featuring a continuously rotating X-ray tube, detector array, and gantry; reconstructs X-ray transmission data into cross-sectional images. System includes signal analysis, display equipment, and patient support. Designed for hospital use or installation on a trailer for transport to designated locations. Trailer kit includes locking mechanisms to secure gantry and couch during transit; unlocked for operation. Operated by clinicians for diagnostic imaging and lung cancer screening. Output consists of reconstructed images for clinical review, aiding in diagnosis and decision-making. Benefits include diagnostic imaging capabilities in mobile/transportable settings.
Clinical Evidence
No clinical data. Safety and effectiveness demonstrated via bench testing and compliance with consensus standards, including vibration testing to simulate 10-year lifetime transport.
Indicated for head, whole body, cardiac, and vascular CT imaging in patients of all ages, including low-dose CT lung cancer screening for early detection of lung nodules.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
November 22, 2021
Philips Healthcare (Suzhou) Co., Ltd. % Shiguang An Regulatory Affairs Engineer No. 258, ZhongYuan Road, Suzhou Industrial Park Suzhou. Jiangsu 215024 CHINA
## Re: K211168
Trade/Device Name: Philips Incisive CT on Trailer Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: October 19, 2021 Received: October 19, 2021
## Dear Shiguang An:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K21168
Device Name Philips Incisive CT on Trailer
#### Indications for Use (Describe)
The Incisive CT is a Computed Tomography X-Ray System intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient and equipments and accessories. The Incisive CT is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
These scanners are intended to be used for diagnostic imaging and for low dose CT lung cancer screening for the early detection of lung nodules that may represent cancer*. The screening must be performed within the established inclusion criteria of programs / protocols that have been approved and published by either a governmental body or professional medical society.
Besides installed in hospital, the proposed Philips Incisive CT may also be installed on trailer and be transported to designated locations for use. And Incisive CT installed on trailer has the same intended use as installed in hospital.
*Please refer to clinical literature, including the results of the National Lung Screening Trial (N Engl J Med 2011; 365:395-409) and subsequent literature, for further information.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
# 510(k) Summary or Statement
This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.
| Date Prepared: | March 30, 2021 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Philips Healthcare (Suzhou) Co., Ltd.<br>No. 258, ZhongYuan Road, Suzhou Industrial Park, 215024<br>Suzhou, Jiangsu Province, PEOPLE'S REPUBLIC OF CHINA<br>Establishment Registration Number: 3009529630 |
| Contact Person: | Shiguang An<br>Advanced Regulatory Engineer<br>Phone cell: +86-139-40106467<br>Fax: +86-512-68018677<br>E-mail: shiguang.an@philips.com |
| Device Name: | Philips Incisive CT on Trailer |
| Classification: | Classification Name: Computed tomography x-ray system<br>Classification Regulation: 21CFR §892.1750<br>Classification Panel: Radiology<br>Device Class: Class II<br>Primary product code: JAK |
| Predicate Device: | Trade Name: Philips Incisive CT<br>Manufacturer: Philips Healthcare (Suzhou) Co., Ltd.<br>510(k) Clearance: K180015-March 20, 2018<br>Classification Regulation: 21 CFR, Part 892.1750<br>Classification Name: Computed tomography x-ray system<br>Classification Panel: Radiology<br>Device Class: Class II<br>Product Code: JAK |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out.
| Device description: | The proposed Philips Incisive CT on Trailer is a whole-<br>body computed tomography (CT) X-Ray System featuring a<br>continuously rotating x-ray tube, detectors, and gantry with<br>multi-slice capability. The acquired x-ray transmission data<br>is reconstructed by computer into cross-sectional images of<br>the body taken at different angles and planes. This system<br>also includes signal analysis and display equipment, patient<br>and equipment support, components, and accessories.<br>The Philips Incisive CT has a 72cm bore and includes a<br>detector array that provides 50cm scan field of view (FOV). |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Besides installed in hospital, Philips Incisive CT can also<br>be installed on trailer and be transported to designated<br>locations. |
| | The main components (detection system, the reconstruction<br>algorithm, and the x-ray system) that are used in the<br>proposed Philips Incisive CT on trailer are identical to the<br>currently marketed and predicate Philips Incisive CT<br>(K180015, 20/March/2018). |
| | The components of the proposed Philips Incisive CT on<br>trailer include the following: |
| | 1. Gantry. The Gantry consists of 4 main internal units:<br>a. Stator – a fixed mechanical frame that carries HW and SW<br>b. Rotor – A rotating circular stiff frame that is mounted in<br>and supported by the stator<br>c. X-Ray Tube (XRT) and Generator, - fixed to the Rotor<br>frame |
| | d. Data Measurement System (DMS) - a detectors array,<br>fixed to the Rotor frame<br>2. Patient Support (Couch) - carries the patient in and out<br>through the Gantry bore synchronized with the scan.<br>3. Console - Containing a Host computer and display that is<br>the primary user interface. |
| | 4. CT on Trailer Kit - Modified Incisive CT installed and<br>secured on trailer requires locking motion parts during<br>trailer transportation and unlocking motion parts before<br>CT operations. |
| | In addition to the above components and the software<br>operating them, each system includes hardware and software<br>for data acquisition, display, manipulation, storage and |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The word is centered and takes up most of the frame.
| | Traditional 510(k) |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use: | filming as well as post-processing into views other than the<br>original axial images. Patient supports (positioning aids) are<br>used to position the patient. |
| | The Incisive CT is a Computed Tomography X-Ray System<br>intended to produce images of the head and body by<br>computer reconstruction of x-ray transmission data taken at<br>different angles and planes. These devices may include<br>signal analysis and display equipment, patient and equipment<br>supports, components and accessories. The Incisive CT is<br>indicated for head, whole body, cardiac and vascular X-ray<br>Computed Tomography applications in patients of all ages. |
| | These scanners are intended to be used for diagnostic<br>imaging and for low dose CT lung cancer screening for the<br>early detection of lung nodules that may represent cancer*.<br>The screening must be performed within the established<br>inclusion criteria of programs / protocols that have been<br>approved and published by either a governmental body or<br>professional medical society. |
| | Besides installed in hospital, the proposed Philips Incisive<br>CT may also be installed on trailer and be transported to<br>designated locations for use. And Incisive CT installed on<br>trailer has the same intended use as installed in hospital. |
| | *Please refer to clinical literature, including the results of the<br>National Lung Screening Trial (N Engl J Med 2011;365:395-<br>409) and subsequent literature, for further information. |
| Fundamental scientific<br>technology: | The proposed Philips Incisive CT on Trailer is advanced<br>continuous rotation computed tomography systems suitable<br>for a wide range of computed tomographic (CT)<br>applications. |
| | The proposed Philips Incisive CT on Trailer can be<br>transported by the trailer to designated locations and is used<br>clinically as a diagnostic patient imaging device that<br>produces images that correspond to tissue density. The<br>quality of the images depends on the level and amount of X-<br>ray energy delivered to the tissue. CT imaging displays both<br>high-density tissue, such as bone, and soft tissue. |
| | The principal technological components (rotating x-ray tube,<br>detector, gantry, patient support and console) of the proposed<br>Philips Incisive CT are identical to the currently marketed |
| | Traditional 510(k) |
| | predicate device Philips Incisive CT (K180015, 20/March/2018). |
| | In addition, the Proposed <b>Philips Incisive CT on Trailer</b> provides CT on trailer kit which is used to install and secure Philips Incisive CT (K180015, 20/March/2018) on a trailer. |
| | Based on the information provided above, the proposed <b>Philips Incisive CT on Trailer</b> does not raise different questions of safety and effectiveness compare to the currently marketed predicate device Philips Incisive CT (K180015, 20/March/2018). |
| Summary of Non-Clinical<br>Performance data: | The currently marketed <b>Philips Incisive CT</b> (K180015) complies with the following international and FDA Recognized Consensus standards:<br>• AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R) 2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, MOD).<br>FDA/CDRH recognition number 19-4<br>• IEC 60601-1-2:2014, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests<br>FDA/CDRH recognition number 19-8<br>• IEC 60601-1-3:2013 Medical electrical equipment -- Part 1-3: General requirements for basic safety - Collateral standard: Radiation protection in diagnostic X-ray equipment<br>FDA/CDRH recognition number 12-269<br>• IEC 60601-1-6:2013 Medical electrical equipment -- Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability<br>FDA/CDRH recognition number 5-89<br>• IEC 60601-2-44:2012 Medical electrical equipment - Part 2-44: Particular requirements for the safety of X-ray equipment<br>FDA/CDRH recognition number 12-256<br>• IEC 62304:2015 Medical device software -- Software life cycle processes<br>FDA/CDRH recognition number 13-79<br>• IEC 62366:2014 Medical Device-Application of Usability Engineering of Medical Devices |
| • ISO14971 Medical devices – Application of risk<br>management to medical devices (Ed. 2.0, 2007)<br>FDA/CDRH recognition number 5-40 | |
| • NEMA XR 25-2010 Computed Tomography Dose Check<br>FDA/CDRH recognition number 12-225 | |
| • NEMA XR 28-2013 Supplemental Requirements for User<br>Information and System Function Related to Dose in CT<br>FDA/CDRH recognition number 12-287 | |
| • NEMA XR 29-2013 Standard Attributes on CT Equipment<br>Related to Dose Optimization and Management | |
| • Guidance for Industry and FDA Staff – Guidance for the<br>Content of Premarket Submissions for Software Contained<br>in Medical Devices (issued May 11, 2005, document number<br>337). | |
| • Guidance for Industry and FDA Staff – Content of<br>Premarket Submissions for Management of Cybersecurity in<br>Medical Devices (issued October 2, 2014) | |
| • Guidance for Industry and FDA Staff – Use of<br>International Standard ISO 10993-1, "Biological evaluation<br>of medical devices – Part 1: Evaluation and testing within a<br>risk management process" (issued June 16, 2016) | |
| • Guidance for Industry and FDA Staff – Information to<br>Support a Claim of Electromagnetic Compatibility (EMC) of<br>Electrically-Powered Medical Devices (issued July 11, 2016) | |
| Additionally, the Philips Incisive CT complies with<br>performance standards for Computed Tomography (CT)<br>Equipment and Laser products (21 CFR 1020.33 and 21 CFR<br>1040.10, respectively). | |
| The proposed Philips Incisive CT on Trailer is identical to<br>the currently marketed and predicate Philips Incisive CT<br>(K180015) with regards to intended use, design and<br>fundamental scientific technology except for trailer kit that is<br>utilized to securely mount the CT system on the trailer. The<br>trailer kit consists of locking mechanism that lock the gantry<br>and couch parts of the CT System to the trailer. | |
| With regards to transportation testing the proposed Philips<br>Incisive CT on Trailer complies with Department of<br>Defense Test Method Standard for Environmental<br>Engineering Considerations and Laboratory Tests (MIL-<br>STD-810F). Method 514.5: Composite wheeled vehicle | |
| | Traditional 510(k) |
| | vibration exposures, as described in the main body, Annex<br>A, Annex B and Annex C of this method.<br>Philips Incisive CT on Trailer has passed the vibration test<br>of 300 hours to simulate a 10 years or 120,690 kM lifetime<br>with QA test including MeanCT, Uniformity, Noise, Spatial<br>Resolution, Slice Thickness, Linearity and Low Contrast<br>Resolution.<br>The QA test process and acceptance criteria are same as the<br>predicate device, this can demonstrate the Philips Incisive |
| | CT on Trailer performs as well as the predicate device. |
| Summary of Clinical<br>Data: | Clinical data is not warranted to demonstrate safety and<br>effectiveness of the proposed Philips Incisive CT on<br>Trailer since its design, intended use and fundamental<br>scientific technology is identical to the currently marketed<br>and predicate Philips Incisive CT (K180015,<br>20/March/2018). |
| Substantial Equivalence<br>Conclusion: | The design, intended use, fundamental scientific technology<br>and principal technological components (Tube, Generator,<br>Detector, gantry, patient support and console) of the<br>proposed Philips Incisive CT on Trailer are identical to the<br>currently marketed predicate Philips Incisive CT (K180015,<br>20/March/2018) except for the addition of a trailer Kit to<br>secure the CT System in a trailer. |
| | Based on the information provided above, the proposed<br>Philips Incisive CT on Trailer does not raise different<br>questions of safety and effectiveness as compared to the<br>currently marketed predicate Philips Incisive CT (K180015,<br>20/March/2018). The proposed Philips Incisive CT on<br>Trailer is therefore substantially equivalent to the currently<br>marketed and predicate Philips Incisive CT (K180015,<br>20/March/2018). |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The font is sans-serif and the letters are evenly spaced. The background is white, which makes the blue letters stand out. The word is centered in the image.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue letters.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "PHILIPS" in large, bold, blue letters. The letters are evenly spaced and fill most of the frame. The background is white, providing a strong contrast to the blue text.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.