Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
Device Story
Reprocessed pulse oximeter sensor; utilizes red and infrared LEDs and a photodetector; measures light absorption across vascular bed (finger/toe/foot) to estimate arterial oxygen saturation (SpO2) and pulse rate; device connects to pulse oximeter console; used in clinical settings for continuous monitoring; reprocessing involves replacing patient-contacting materials with medical-grade adhesive tape and sterilization via vaporized hydrogen peroxide; output displayed on console for clinician assessment of patient oxygenation status; benefits include noninvasive monitoring without blood sampling.
Clinical Evidence
Clinical validation performed on healthy adult volunteers (70-100% SpO2 range) under no-motion conditions. ARMS for SpO2 was 1.48% for woven tape and 1.72% for non-woven tape. Bench testing included continuity and sensitivity testing to verify sensor functionality post-reprocessing.
Technological Characteristics
Optical sensor with red/infrared LEDs and photodetector; medical-grade adhesive tape (biocompatible per ANSI/AAMI/ISO 10993-1:2009); sterilization via vaporized hydrogen peroxide; form factor designed for finger/toe/foot application; connects to pulse oximeter console.
Indications for Use
Indicated for continuous noninvasive arterial oxygen saturation and pulse rate monitoring in patients requiring such monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 8, 2022
Stryker Sustainability Solutions Moira Barton-Varty Senior Principal Regulatory Affairs 1810 West Drake Drive Tempe, Arizona 85283
# Re: K211140
Trade/Device Name: Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer), Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer), Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer), Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer) Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: NLF Dated: February 10, 2022 Received: February 11, 2022
Dear Moira Barton-Varty:
# (NOTE: Reprocessed SUD device types require a separate attachment of the list of all models cleared in the submission. A corrected SE letter will be required if the attachment is omitted.)
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and
{1}------------------------------------------------
adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211140
Device Name Reprocessed Masimo Pulse Oximeter
Indications for Use (Describe)
Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" underneath. The Stryker logo is in bold, black font. Above the logo, the text "Page 1 of 3" is visible. The words "Sustainability Solutions" are also in bold, black font but are smaller than the Stryker logo.
# 510(k) Summary
## Contact Details
| Applicant Name: | Stryker Sustainability Solutions |
|---------------------------------|---------------------------------------------------------------|
| Applicant Address: | 1810 West Drake Drive Tempe AZ 85283 United States of America |
| Applicant Contact<br>Telephone: | 6024100204 |
| Applicant Contact: | Mrs. Moira Barton-Varty |
| Applicant Contact Email: | moira.barton@stryker.com |
### Device Name
Device Trade Name:
Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer) Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer)
Common Name: Oximeter Classification Name: Oximeter, Reprocessed Regulation Number: 870.2700 Product Code: NLF
#### Legally Marketed Predicate Devices
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
|-------------|----------------------------------------------------------|--------------|
| K081238 | Reprocessed Masimo Pulse Oximeter Sensors | NLF |
| K060143 | LNCS and SPO2.COM Sensors | DQA |
#### Device Description Summary
In a clinical setting, a pulse oximeter sensor measures the oxygen saturation of arterial blood (SpO2). A pulse oximeter sensor is composed of a light emitting diode (LED) and a sensor that are placed on opposite sides of a patient's finger or foot. The LED contains a red light and an infrared light that are differentially absorbed by oxygenated hemoglobin. Based on the relative absorption of the two wavelengths that is determined by the sensor, the POX determines the relative amount of oxygenated and deoxygenated hemoglobin, which is calculated as SpO2. In order to make the SpO2 calculation independent of skin color, finger size, etc., the pulse oximeter sensor uses only the time varying light absorption component generated by the patient's pulse. The sensor also uses the period of pulsation to measure patient pulse rate. The pulse oximeter can estimate the amount of oxygen in the blood without having to draw a blood sample.
The primary components of an oxygen transducer, or Pulse Oximeter (POX) Sensor, are light-emitting diodes (red and infrared LED) and a photo sensor. These components (with their wiring system) are embedded within a taping system designed for wrapping the POX Sensor around a patient's finger, foot, or hand so that the LED and photo sensor are directly opposite to each other. As the lights are emitted and received across a vascular bed, the rates of absorption at the two wavelengths vary depending upon the ratios of oxygenated hemoglobin within the blood.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" written below it. The Stryker logo is in bold, black letters. Above the logo, the text "Page 2 of 3" is visible. The text is smaller than the Stryker logo and is located in the upper right corner of the image.
The proposed devices of this submission do not differ from the predicate device. The only difference is that the proposed devices will be exposed to vaporized hydrogen peroxide instead of ethylene oxide.
# Intended Use/Indications for Use
Reprocessed Masimo Pulse Oximeter sensor is indicated for use in continuous noninvasive arterial oxygen saturation and pulse rate monitoring.
## Indications for Use Comparison
The indications for use are the same as the predicate device.
# Technological Comparison
The design of the reprocessed device is the same as the predicate device. There are no changes to the claims, clinical applications, patient population, or performance specifications for use does not change from the predicate device (K081238). The same standard mechanical design and equivalent materials are utilized. There are no changes to the claims, clinical applications, patient population, or performance specifications.
The subject device, LNCS Pulse Oximeter Sensors and the predicate device, have the following key similarities:
- intended use;
- principle of operation;
- form factor and components;
- measurement application sites;
- performance specifications;
- environmental and mechanical specifications- Both devices have the same indicated populations;
# Principle of Operation
The principle of operation of pulse oximetry is based upon the fundamental principle that hemoglobin bound to oxygen (oxyhemoglobin) and hemoglobin unbound to oxygen (deoxyhemoglobin) vary in the absorption of different wavelengths of the light and the absorptions can be used to estimate SpO2 and pulse rate. The mechanism by which this process occurs is the use of red and infrared wavelengths of light delivered by an emitter and the signal from the light absorption of oxygenated blood and deoxygenated blood to determine functional oxygen saturation of hemoglobin (SpO2).
#### Mechanism of Action for Achieving the Intended Effect
The Reprocessed Pulse Oximeter Sensor provides the intended effect equivalent to the previously cleared pulse oximeter sensor in that it utilizes an optical sensor that is applied to the patient's finger or toe through which light is transmitted to the photodetector that detects the signal transmissionis processed by the Pulse Oximeter to provide SpO2 and pulse rate.The principles of operation are still based upon the use of red and infrared wavelengths of light provided by an emitter and the detection of the signal from the light absorption of oxygenated blood and deoxygenated blood to provide functional oxygen saturation of hemoglobin (SpO2).
A functional pulse oximeter sensor when connected to a pulse oximeter console will have the ability to monitor SpO2 and Pulse Rate. Bench and laboratory testing was conducted to determine whether a pulse oximeter sensor is functional and is assessed by performing continuity and sensitivity testing. This included the following tests:
- Continuity testing to verify there are no open circuits and the current along the path of a circuit is continuous
- Sensitivity testing to detect the signal transmission between the photodiode and the LED (light emitting diode)
# Non-Clinical and/or Clinical Tests Summary & Conclusions
To support the substantial equivalence of product performance after by vaporized hydrogen peroxide to that of the predicate devices, non-clinical bench simulation testing was conducted using a stand-in device allows for SpO2 sensor verification by passing the light source of the pulse oximeter sensor (LED) thru one side of
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the Stryker logo with the words "Sustainability Solutions" underneath. The Stryker logo is in black and is a stylized version of the company name. The words "Sustainability Solutions" are in black and are in a smaller font than the Stryker logo. The words "Page 3 of 3" are in the upper right corner of the image.
the stand-in with the signal transmission measured by the photocell (photo-detector) of the pulse oximeter sensor on the opposing side of the stand-in.
A functional pulse oximeter sensor when connected to a pulse oximeter console will have the ability to monitor SpO2 and Pulse Rate. Benchand laboratory testing was conducted to determine whether a pulse oximeter sensor is functional and is assessed by performing continuity and sensitivity testing. This included the following tests:
- Continuity testing to verify there are no open circuits and the current along the path of a circuit is continuous
- Sensitivity testing to detect the signal transmission between the photodiode and the LED (light emitting diode)
Performance testing , incorporated by reference K081238, also included the following which are not altered by the use of vaporized hydrogen peroxide.
All reprocessed pulse oximeter sensors materials that are normally in contact with the patient are replaced with medical grade adhesive tape during reprocessing by Stryker Sustainability Solutions. The reprocessed pulse oximeter sensors meet biocompatibility requirements of ANSI/AAMI/ISO 10993-1: 2009: Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process for surface contacting devices with prolonged exposure (> 24 hours, but less than 30 days) with skin.
#### Clinical Testing
Stryker Sustainability Solutions performed the clinical validation testing of the SpO2 performance under no motion on healthy, adult volunteers in the range of 70% to 100%. The ARMS for SpO2 under no motion was found to be 1.48% and 1.72% for woven and non-woven tape over the range of 70-100%.
The results of the non-clinical and clinical testing demonstrate that all requirements and performance specifications were satisfied and support the subject device is substantially equivalent to its predicate.
{7}------------------------------------------------
Reprocessed Masimo Pulse Oximeter (1859 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (1860 Pediatric O2 Transducer) Reprocessed Masimo Pulse Oximeter (1861 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (1862 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2317 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2319 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2320 Adult O2 Transducer) Reprocessed Masimo Pulse Oximeter (2328 Infant O2 Transducer) Reprocessed Masimo Pulse Oximeter (2329 Adult O2 Transducer)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.