← Product Code [FRN](/productcode/FRN) · K211125

# Novum IQ Syringe Pump (K211125)

_Baxter Healthcare Corporation · FRN · Aug 30, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K211125

## Device Facts

- **Applicant:** Baxter Healthcare Corporation
- **Product Code:** [FRN](/productcode/FRN.md)
- **Decision Date:** Aug 30, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Novum IQ Syringe Pump is intended to be used for the controlled administration of fluids. These include pharmaceutical drugs, parenteral nutrition, blood and blood products, and enteral nutrition. The Novum IQ Syringe Pump is intended to deliver an infusion through the following clinically accepted routes of administration: intravenous, arterial, enteral and subcutaneous. The Novum IQ Syringe Pump is intended to be used in conjunction with legally marketed and compatible administration sets, syringes, and medications provided by the user. The Novum IQ Syringe Pump is suitable for patient care in hospitals and outpatient health care facilities. The Novum IQ Syringe Pump is intended for use on adults, pediatrics and neonates. The Novum IQ Syringe Pump is intended to aid in the reduction of operator interaction through guided programming, including a way to automate the programming of infusion parameters and documentation of infusion therapies when integrated with an Electronic Medical Record (EMR) system. This automation is intended to aid in the reduction of programming errors. The Novum IQ Syringe Pump is intended to be used by trained healthcare professionals.

## Device Story

Novum IQ Syringe Pump is a smart, electromechanical infusion pump; delivers fluids via stepper motor driving a lead screw to depress syringe plunger. Accepts 1-60 mL syringes; barrel clamp holds syringe; sensors identify syringe size via look-up table. Used in hospitals/outpatient settings by trained clinicians. Features guided programming and EMR integration for auto-programming/auto-documentation to reduce errors. Operates on AC power or internal rechargeable Li-ion battery for transport. Provides visual/audible alarms for safety. Clinician reviews pump output; device facilitates controlled therapy delivery; benefits include reduced manual programming errors and improved documentation accuracy.

## Clinical Evidence

No clinical data. Bench testing only, including software verification/validation, cybersecurity, interoperability, electrical safety (IEC 60601-1, 60601-2-24), EMC, and human factors evaluations in simulated clinical environments.

## Technological Characteristics

Stepper motor/leadscrew pumping mechanism; 1-60 mL syringe capacity; AC/battery powered (Li-ion); IPX2 ingress protection; wireless connectivity; multiple microprocessors for safety; LCD display. Software is major level of concern. Complies with IEC 60601-1, 60601-2-24, IEC 60601-1-8, and battery safety standards (UL 2054, UL 1642).

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Medfusion Model 4000 Syringe Pump ([K111386](/device/K111386.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Baxter Healthcare Corporation Dhiraj Bizzul Sr. Manager, Regulatory Affairs One Baxter Parkway Deerfield, Illinois 60015

August 30, 2022

### Re: K211125

Trade/Device Name: Novum IQ Syringe Pump Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: FRN, LZH Dated: May 18, 2022 Received: June 1, 2022

#### Dear Dhiraj Bizzul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Courtney H. Lias, Ph.D. Office Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) TBD

Device Name Novum IQ Syringe Pump

#### Indications for Use (Describe)

The Novum IQ Syringe Pump is intended to be used for the controlled administration of fluids. These include pharmaceutical drugs, parenteral nutrition, blood and blood products, and enteral nutrition.

The Novum IQ Syringe Pump is intended to deliver an infusion through the following clinically accepted routes of administration: intravenous, arterial, enteral and subcutaneous.

The Novum IQ Syringe Pump is intended to be used in conjunction with legally marketed and compatible administration sets, syringes, and medications provided by the user.

The Novum IQ Syringe Pump is suitable for patient care in hospitals and outpatient health care facilities.

The Novum IQ Syringe Pump is intended for use on adults, pediatrics and neonates.

The Novum IO Syringe Pump is intended to aid in the reduction of operator interaction through guided programming. including a way to automate the programming of infusion parameters and documentation of infusion therapies when integrated with an Electronic Medical Record (EMR) system. This automation is intended to aid in the reduction of programming errors.

The Novum IO Syringe Pump is intended to be used by trained healthcare professionals.

Type of Use (*Select one or both, as applicable*)

|  | <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--|-------------------------------------------------------------|
|  | <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "Baxter" in a bold, blue font. The word is slightly italicized, giving it a dynamic appearance. The font is sans-serif, and the letters are closely spaced together.

#### 510(k) Summary

Date: August 25, 2022

#### Owner:

Baxter Healthcare Corporation One Baxter Parkway Deerfield, IL 60015

#### Contact Person:

Dhiraj Bizzul Sr. Manager, Global Regulatory Affairs 35212 W. Illinois Route 120 Round Lake, IL 60073 Telephone: (847) 909-6828 Fax: (224) 270-4119

#### IDENTIFICATION OF THE DEVICE:

Trade/Device Name: Novum IQ Syringe Pump Classification Panel: 80 General Hospital Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN, LZH

#### Table 1. Catalogue Code for Novum IO Syringe Pump

| Code Number | Name                  |
|-------------|-----------------------|
| 40800BAXUS  | Novum IQ Syringe Pump |

### PREDICATE DEVICE:

The Novum IQ Syringe Pump is substantially equivalent to the following predicate device:

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and has a clean, modern look.

| Device                               | Company                     | Predicate 510(k) | Clearance Date  |
|--------------------------------------|-----------------------------|------------------|-----------------|
| Medfusion Model 4000<br>Syringe Pump | Smith's Medical, MD<br>INC. | K111386          | August 29, 2011 |

### Table 2. Predicate Device

### DESCRIPTION OF THE DEVICE:

The Novum IQ Syringe Pump is a general purpose, syringe-based, smart pump within the Novum IQ infusion platform. The Novum IQ syringe pump offers various programmable delivery modes to address specific patient care needs. The delivery modes available to support the patient are determined by how the Novum IQ Syringe Pump and its associated drug library are configured. The Novum IQ syringe pump is capable of delivering fluids in syringes ranging in size from 1 mL - 60 mL.

The Novum IQ Syringe Pump is a system that provides delivery of fluids into a patient in a controlled manner, as identified in 21 CFR 880.5725. The system includes a software controlled, electromechanical syringe pump used for the controlled administration of fluids including pharmaceutical drugs, parenteral nutrition, blood and blood products, and enteral nutrition through compatible syringes and administration sets at user selectable rates and volumes.

The Novum IQ Syringe Pump is intended to be used for the controlled administration of fluids. These include pharmaceutical drugs, parenteral nutrition, blood and blood products, and enteral nutrition. The Novum IQ Syringe Pump is intended to deliver an infusion through the following clinically accepted routes of administration: intravenous, arterial, enteral, and subcutaneous. The Novum IQ Syringe Pump is intended to be used in conjunction with legally marketed and compatible administration sets, syringes, and medications provided by the user.

The Novum IO Syringe Pump is suitable for patient care in hospitals and outpatient health care facilities. The Novum IQ Syringe Pump is intended for use on adults, pediatrics, and neonates.

The Novum IQ Syringe Pump is intended to aid in the reduction of operator interaction through guided programming, including a way to automate the programming of infusion parameters and documentation of infusion therapies when integrated with an Electronic Medical Record (EMR) system. This automation is intended to aid in the reduction of

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font appears to be sans-serif. The word is slightly slanted to the right, giving it a dynamic appearance. The background is plain white.

programming errors. The Novum IQ Syringe Pump is intended to be used by trained healthcare professionals.

The feedback-controlled, motorized pumping mechanism of the Novum IQ Syringe Pump is a stepper motor acting with a lead screw to depress the plunger of a syringe. The syringe pump accepts a horizontally loaded syringe, held in place by a barrel clamp. The outer diameter of the loaded syringe is sensed, and compared against a stored look-up table to enable user assisted identification of the loaded syringe.

The Novum IO Syringe Pump is designed to detect and react to situations that could impact patient safety. When a situation that could impact patient safety is detected, the syringe infusion pump will notify the clinician with a visual alert and an audible alarm tone.

The Novum IQ Syringe Pump is designed to operate on AC power with an internally mounted rechargeable battery pack to facilitate the continuation of therapy during patient transport or AC power loss.

## INDICATIONS FOR USE:

The Novum IQ Syringe Pump is intended to be used for the controlled administration of fluids. These include pharmaceutical drugs, parenteral nutrition, blood and blood products, and enteral nutrition. The Novum IQ Syringe Pump is intended to deliver an infusion through the following clinically accepted routes of administration: intravenous, arterial, enteral, and subcutaneous. The Novum IQ Syringe Pump is intended to be used in conjunction with legally marketed and compatible administration sets, syringes, and medications provided by the user.

The Novum IQ Syringe Pump is suitable for patient care in hospitals and outpatient health care facilities. The Novum IQ Syringe Pump is intended for use on adults, pediatrics and neonates.

The Novum IQ Syringe Pump is intended to aid in the reduction of operator interaction through guided programming, including a way to automate the programming of infusion parameters and documentation of infusion therapies when integrated with an Electronic Medical Record (EMR) system. This automation is intended to aid in the reduction of programming errors. The Novum IQ Syringe Pump is intended to be used by trained healthcare professionals.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the word "Baxter" in a stylized, bold, blue font. The font appears to be italicized, with the letters slightly slanted to the right. The word is presented against a white background, making the blue text stand out.

# DEVICE COMPARISON AND SUBSTANTIAL EQUIVALENCE:

| Characteristic                                                             | Subject Device                                                                                                                                                                                                          | Predicate Device (K111386)                                                                                                                                                                                                                                                                          | Discussion of Differences                                                                                                                                                                                                         |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>(see below for full list of<br>indications for use) | The Novum IQ Syringe Pump is<br>intended to be used for the<br>controlled administration of fluids.<br>These include pharmaceutical drugs,<br>parenteral nutrition, blood and blood<br>products, and enteral nutrition. | The Medfusion Model 4000 Syringe<br>Infusion Pump is indicated for the<br>follow uses: In the administration of<br>fluids requiring precisely controlled<br>infusion rates including blood or blood<br>products, lipids, drugs, antibiotics,<br>enteral solutions, and other therapeutic<br>fluids. | Lipids are types of nutritional fluids and<br>antibiotics are types of pharmaceutical<br>drugs. These differences do not raise<br>different questions of safety and<br>effectiveness.                                             |
| Routes of<br>Administration                                                | The Novum IQ Syringe Pump is<br>intended to deliver an infusion<br>through the following clinically<br>accepted routes of administration:<br>intravenous, arterial, enteral and<br>subcutaneous.                        | By following delivery routes: arterial,<br>epidural, intravenous, intrathecal,<br>subcutaneous and enteral.                                                                                                                                                                                         | The intended use of the device is the<br>same as the predicate. The absence of one<br>route of administration in the subject<br>device does not raise new questions of<br>safety and effectiveness.                               |
| Syringe Usage                                                              | The Novum IQ Syringe Pump is<br>intended to be used in conjunction<br>with legally marketed and<br>compatible administration sets,<br>syringes, and medications provided<br>by the user.                                | Not specified as part of indication for<br>use                                                                                                                                                                                                                                                      | Although not specified, Medfusion Model<br>4000 syringe infusion pump is to be used<br>with legally marketed and compatible<br>administration sets, syringes and<br>medications as defined in the predicate<br>operator's manual. |

## Table 3. Comparison of Indications for Use

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a company logo. The word is horizontally oriented and takes up most of the image space.

| Characteristic                             | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate Device (K111386)                                                                                                                                                                           | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Use<br>Environment/Treatment<br>Population | The Novum IQ Syringe Pump is<br>suitable for patient care in hospitals<br>and outpatient health care facilities.<br>The Novum IQ Syringe Pump is<br>intended for use on adults, pediatrics<br>and neonates.                                                                                                                                                                                                                                                                                          | In critical care, anesthesia, neonatal and<br>pediatric applications or other healthcare<br>settings where the use of the syringe<br>infusion pump can be monitored or<br>supervised by a clinician. | The clinical care areas for the predicate<br>device are all found within hospitals and<br>outpatient health care facilities. These<br>differences do not raise different<br>questions of safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                        |
| Other indications for<br>Use               | The Novum IQ Syringe Pump is<br>intended to aid in the reduction of<br>operator interaction through guided<br>programming, including a way to<br>automate the programming of<br>infusion parameters and<br>documentation of infusion therapies<br>when integrated with an Electronic<br>Medical Record (EMR) system. This<br>automation is intended to aid in the<br>reduction of programming errors.<br>The Novum IQ Syringe Pump is<br>intended to be used by trained<br>healthcare professionals. | By following delivery modes:<br>continuous, volume/time, mass, body<br>weight, custom dilution, intermittent and<br>bolus.                                                                           | Both devices provide guided<br>programming workflows and transfer<br>infusion information and configured drug<br>libraries using established bidirectional<br>communication protocols. The subject<br>device also utilizes auto-programming<br>and auto-documentation features to<br>transfer information to/from an EMR<br>using wireless technology. Although the<br>subject device's indications do not list<br>specific delivery modes, the subject<br>device supports the same delivery modes<br>as the predicate.  The differences do not<br>raise different issues of safety and<br>effectiveness. |

# Table 3. Comparison of Indications for Use

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is centered and takes up most of the image.

| Characteristic           | Subject Device                                     | Predicate Device (K111386) | Discussion of Differences                                                                                                                                                                        |
|--------------------------|----------------------------------------------------|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design Attributes        |                                                    |                            |                                                                                                                                                                                                  |
| Pumping Mechanism        | Stepper Motor Driving Leadscrew                    | Same as subject device     | N/A                                                                                                                                                                                              |
| Modes of Delivery        | Continuous, Intermittent                           | Same as subject device     | N/A                                                                                                                                                                                              |
| Microprocessors          | Multiple processors                                | Dual Processor             | The difference in allocated functions for the<br>subject device is based on a Hardware safety<br>feature to initiate corrective action and/or safe<br>states.                                    |
| Fluid Ingress Protection | IPX2                                               | IPX3                       | Difference in IPX2 and IPX3 does not have a<br>significant impact based on the use environment<br>indicated for the devices and do not raise different<br>questions of safety and effectiveness. |
| User Interface Display   | LCD display with adjustable backlit<br>capability. | Same                       | Displays provide pump operating and status<br>information. Differences do not raise different<br>questions of safety and effectiveness.                                                          |
| External Interfaces      | USB 2.0, Type-A receptacle                         | Ethernet                   | Both provide a means of transferring data,<br>differences do not raise different questions of<br>safety and effectiveness.                                                                       |

# Table 4. Comparison of Design Attributes

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image contains the word "Baxter" in a stylized, sans-serif font. The word is rendered in a vibrant blue color, which contrasts with the white background. The font appears to be bold and slightly italicized, giving the word a dynamic and modern look. The overall impression is clean and professional, suggesting a corporate logo or brand name.

| Table 5. Comparison of Features/Functions and Performance Specifications |  |
|--------------------------------------------------------------------------|--|
|                                                                          |  |

| Characteristic                                    | Subject Device                                                                                                                                                                                                                                                                                                                                         | Predicate Device (K111386)                      | Discussion of Differences                                                                                                                                                                                                                                                                                       |  |  |  |  |                                                                                         |                                                                                                                                                                 |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Features/Functions and Performance Specifications |                                                                                                                                                                                                                                                                                                                                                        |                                                 |                                                                                                                                                                                                                                                                                                                 |  |  |  |  |                                                                                         |                                                                                                                                                                 |
| AC Power                                          | Input: 100-240 V~, 50-60 Hz, 0.5A (0.5A-<br>0.35 A)<br>Output: 16 VDC/1.25 A, short circuit<br>protected                                                                                                                                                                                                                                               | Input: 100-240 VAC 50/60Hz<br>Output: Not Known | The differences in power ratings do<br>not raise different questions of safety<br>and effectiveness.                                                                                                                                                                                                            |  |  |  |  |                                                                                         |                                                                                                                                                                 |
| Battery                                           | Smart Battery Pack<br>- Lithium Ion (internal battery unit), 10.8<br>VDC nominal<br>-Maximum 16 hour recharge time at 23 ±2° C<br>(73.4 ±3.6° F)*<br>-Battery capacity for a new fully charged<br>Smart Battery Pack, at the medium backlight<br>setting and wireless enabled use is as follows:<br>Flow Rate (mL/hr) Capacity (hrs)   10 ≥8   1200 ≥6 |                                                 |                                                                                                                                                                                                                                                                                                                 |  |  |  |  | Rechargeable Lithium ion<br>10 hrs at 5 mL/hr with 60 mL syringe<br>Charge time: 10 hrs | The pump is able to operate<br>independent from AC power duringthe voltage supply. Differences do not<br>raise different questions of safety and effectiveness. |
| Alarms                                            | Per IEC 60601-1-8                                                                                                                                                                                                                                                                                                                                      | Not known                                       | The subject device is designed<br>according to current FDA recognized<br>standards. Both the subject and<br>predicate devices have similar alarms<br>and alarm priorities, as such, any<br>difference in alarm setting to the<br>predicate do not raise different<br>questions of the safety and effectiveness. |  |  |  |  |                                                                                         |                                                                                                                                                                 |

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image contains the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be italicized. The word is centered horizontally in the image and is the only element present.

|  |  | Table 5. Comparison of Features/Functions and Performance Specifications |  |  |
|--|--|--------------------------------------------------------------------------|--|--|
|  |  |                                                                          |  |  |

| Characteristic                | Subject Device             | Predicate Device (K111386)         | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|-------------------------------|----------------------------|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Logging<br>Capability         | Yes                        | Same as subject device             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Bolus<br>Capability           | Yes                        | Same as subject device             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Bolus Volume<br>Accuracy      | ±10% under most conditions | Information not publicly available | Differences do not raise different<br>questions of safety and effectiveness.<br>Detailed bolus accuracy disclosed in<br>the labeling.                                                                                                                                                                                                                                                                                                                                  |
| Syringe size                  | 1 - 60 mL                  | Same as subject device             | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Flow Rate<br>range<br>(mL/hr) | 0.01- 1200 mL/hr           | 0.01 - 1130 mL/hr                  | Equivalent resolution over the flow<br>rate range. Both the subject and<br>predicate devices have implemented<br>Minimum programable and minimum<br>recommended flowrates.<br>The subject device has implemented<br>equivalent minimum programmable<br>flow rate as the predicate and<br>established higher minimum<br>recommended flow rate as compared<br>to the predicate.<br>These differences do not raise<br>different questions of safety and<br>effectiveness. |

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a custom typeface. The word is horizontally oriented and centered in the image. The background is plain white.

| Characteristic           | Subject Device                                                                                                                   | Predicate Device (K111386)                                                                                                                             | Discussion of Differences                                                                                                                                                           |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flow Rate<br>accuracy    | For most of the Compatible Syringes:<br>$\pm$ 5% under most conditions<br>Certain syringes:<br>-10% to +5% under most conditions | Information not publicly available                                                                                                                     | Differences do not raise different<br>questions of safety and effectiveness.<br>Detailed flow rate accuracy disclosed<br>in the labeling.                                           |
| Drug Library<br>Capacity | Number of Care Areas: 32<br>Number of Drugs: 5000<br>Number of Clinical Advisories: 800<br>Number of Modifiers: 1000             | Number of Care Areas: 16<br>Number of Drugs: 4608<br>Number of Clinical Advisories (Drug Alerts):<br>Information Not Known<br>Number of Modifiers: N/A | While the devices have different<br>capacity, both allow for user defined<br>differentiation. This difference does<br>not raise different questions of safety<br>and effectiveness. |

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and the letters are slightly slanted to the right. The word is centered and takes up most of the image.

| Characteristic                   | Subject Device                                              | Predicate Device (K111386) | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------|-------------------------------------------------------------|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Keep Vein<br>Open (KVO)<br>Rate  | Range 0.01-5 mL/hr based on the drug in the<br>drug library | Range 0 - 9.9 mL/hr        | KVO can be enabled in the drug<br>library or configured at the pump and<br>will deliver at the KVO rate at the end<br>of an infusion for both the predicate<br>and subject device. The rate range for<br>KVO is intended to-differentiate the<br>KVO rate from the infusion rate to<br>maintain site patency. The different<br>range does not raise different<br>questions of safety and effectiveness.<br>Restrictions regarding maximum<br>KVO flow rate per syringe size is<br>provided as an additional safety<br>mitigation. |
| Occlusion<br>Pressure<br>Setting | User selectable downstream occlusion<br>pressure settings.  | Same as subject device     | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the word "Baxter" in a blue, sans-serif font. The font is bold and slightly italicized, giving the word a sense of movement. The word is centered and takes up most of the frame, with a white background.

| Characteristic         | Subject Device                                                                                                                                                                                                                                                                                                                                               | Predicate Device (K111386)                                                                                                                                                                                                                                                                                                                  | Discussion of Differences                                                                                                                                  |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dose Modes             | Continuous Delivery Modes:<br>mL/hr, mL/kg/min, mL/kg/hr, g/hr, mg/hr,<br>mg/kg/hr, mg/min, mg/kg/min, mg/kg/day,<br>mcg/hr, mcg/kg/hr, mcg/min, mcg/kg/min,<br>mcg/kg/day, ng/min, ng/kg/min, Units/hr,<br>Units/kg/hr, Units/min, Units/kg/min,<br>mUnits/min, mUnits/kg/hr, mUnits/kg/min,<br>mEq/hr, mEq/kg/hr, mmol/hr, mmol/kg/hr,<br>MillionUnits/day | Continuous<br>mL/hr, mL/kg/min, mL/kg/hr, g/hr, mg/hr,<br>mg/kg/hr, mg/min, mg/kg/min, mg/kg/day<br>mcg/hr, mcg/kg/hr, mcg/min, mcg/kg/min,<br>mcg/kg/day, ng/min, ng/kg/min, Units/hr,<br>Units/kg/hr, Units/min, Units/kg/min,<br>MillionUnits/day, mUnits/min, mUnits/kg/hr,<br>mUnits/kg/min, mEq/hr, mEq/kg/hr, mmol/hr,<br>mmol/kg/hr | The subject device also provides<br>continuous and intermittent infusions.<br>Differences do not raise different<br>questions of safety and effectiveness. |
|                        |                                                                                                                                                                                                                                                                                                                                                              | Amount / Time                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                            |
|                        | Intermittent Delivery Modes:<br>mL/kg, g, g/kg, g/m2, mg, mg/kg, mg/m2,<br>mcg, mcg/kg, mcg/m2, Units, Units/kg,<br>Units/m2, mEq, mEq/kg, mmol, mmol/kg,<br>MillionUnits, MillionUnits/kg,<br>MillionUnits/m2                                                                                                                                               | Non-weight-based:<br>mg, Units, g, mcg, mEq, mmol, MillionUnits                                                                                                                                                                                                                                                                             |                                                                                                                                                            |
|                        |                                                                                                                                                                                                                                                                                                                                                              | Weight based:<br>mL/kg, mg/kg, Units/kg, g/kg, mcg/kg,<br>mEq/kg, mmol/kg, MillionUnits/kg                                                                                                                                                                                                                                                  |                                                                                                                                                            |
|                        |                                                                                                                                                                                                                                                                                                                                                              | BSA based:<br>mg/m², Units/m², g/m², mcg/m²,<br>MillionUnits/m2                                                                                                                                                                                                                                                                             |                                                                                                                                                            |
|                        |                                                                                                                                                                                                                                                                                                                                                              | Volume / Time<br>Total Volume / Total Time                                                                                                                                                                                                                                                                                                  |                                                                                                                                                            |
| Rate or Dose<br>Limits | 0.01 - 1200 mL/hr (depending on syringe size<br>and rate).                                                                                                                                                                                                                                                                                                   | 0.01 mL/hr to 1130 mL/hr in increments of<br>0.01 or 0.1 mL/hr (depending on syringe size<br>and rate).                                                                                                                                                                                                                                     | Equivalent resolution over the flow<br>range. Differences do not raise<br>different issues of safety and<br>effectiveness.                                 |

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a corporate logo. The word is centered and takes up most of the image space.

|  |  |  |  | Table 5. Comparison of Features/Functions and Performance Specifications |
|--|--|--|--|--------------------------------------------------------------------------|
|  |  |  |  |                                                                          |

| Characteristic            | Subject Device                                                                                                                                 | Predicate Device (K111386)                                                                                                                                 | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                 |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operational<br>Conditions | Operating temperature:<br>50 to 104°F (10 to 40° C), 10 to 95% relative<br>humidity non-condensing.<br>Atmospheric Pressure: 70 kPa to 102 kPa | Temperature: 5° to 40° C (40° to 104° F)<br>Relative Humidity: 20 to 95% non-condensing<br>Ambient Pressure: 70 kPa to 106 kPa (10.2 psia<br>to 15.4 psia) | Functional and performance tests<br>support the operational conditions for<br>the subject device. These differences<br>do not raise different questions of<br>safety and effectiveness.                                                                                                                                                                                                                   |
| Storage<br>Temperature    | Storage temperature:<br>14 to 120°F (-10 to 49°C), 10 to 95% relative<br>humidity non-condensing                                               | Store at controlled room temperature 15° to<br>30°C (59° to 86°F).                                                                                         | Functional and performance tests<br>support the storage conditions for the<br>subject device. These differences do<br>not raise different questions of safety<br>and effectiveness.                                                                                                                                                                                                                       |
| Approximate<br>Size       | 102mm H x 240mm W x 161mm D                                                                                                                    | 160x270x145 mm                                                                                                                                             | Human Factors studies support the<br>physical dimensions and Form Factor<br>for the subject device. These<br>differences do not raise different<br>questions of safety and effectiveness.                                                                                                                                                                                                                 |
| Approximate<br>Weight     | 2.8 kg (6.2lb)                                                                                                                                 | 4.5 lb                                                                                                                                                     | Human Factors studies support the<br>weight of the subject device. This<br>difference does not raise different<br>questions of safety and effectiveness.                                                                                                                                                                                                                                                  |
| Auto-<br>programming      | Yes                                                                                                                                            | No                                                                                                                                                         | The subject device allows automated<br>population of infusion delivery<br>information from an EMR in addition<br>to manually programming infusions<br>through the pump interface.<br>Automated data transfer using<br>establish protocols does not raise<br>different questions of safety and<br>effectiveness as infusion programming<br>is confirmed by the clinician prior to<br>starting an infusion. |
| Characteristic            | Subject Device                                                                                                                                 | Predicate Device (K111386)                                                                                                                                 | Discussion of Differences                                                                                                                                                                                                                                                                                                                                                                                 |
| Wireless                  | Yes                                                                                                                                            | Same as subject device                                                                                                                                     | N/A                                                                                                                                                                                                                                                                                                                                                                                                       |

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The font is sans-serif and appears to be a custom font. The word is the logo for the Baxter International Inc. company, a global medical technology company.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif and appears to be a custom typeface designed for the brand.

## DISCUSSION OF NON-CLINICAL TESTS

Non-Clinical testing of the Novum IO Syringe Pump has been performed against requirements for performance, physical attributes, environmental conditions and safety, and to provide objective evidence that the device's intended use is met.

As recommended by FDA guidance, "Infusion Pumps Total Product Life Cycle" issued December 2, 2014, Baxter has developed a Safety Assurance Case (SAC) to demonstrate that hazardous situations resulting from the design, intended use, and reasonably foreseeable misuse of the device have been appropriately mitigated.

The stated goal of the safety assurance case is to document that the design of the Novum IQ Syringe Pump is adequately safe for its intended use. The assurance case defined the device system, including the indications for use, patient populations, use environments, and system specifications. The supporting assurance arguments confirmed that:

- potential risks have been mitigated and the residual risk is acceptable, ●
- design verification and validation of the device is acceptable, .
- device reliability is acceptable, ●
- the device meets clinically valid essential performance.

Performance testing of the Novum IQ Syringe Pump was verified against requirements for performance, physical attributes, environmental conditions, safety, and to provide objective evidence that the device intended use is met.

- . Validation demonstrated that design inputs and user needs were met. Verification involves testing of both subsystem and system level requirements against predefined and approved protocols containing validated test methods and established acceptance criteria. System verification including software verification demonstrated that design outputs meet design input requirements, connectivity requirements, interface requirements and cyber security requirements. System verification also includes testing of interfaces with other devices or accessories that are intended to be used with the system.
- . Cleaning and Disinfection validation was performed according to FDA Guidance Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling issued March 17, 2015.

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the word "Baxter" in a bold, sans-serif font. The word is blue and appears to be a logo. The font is slightly italicized, giving the word a sense of movement.

- . Wireless functionality was implemented according to FDA Guidance Radio Frequency Wireless Technology in Medical Devices, issued August 14, 2013.
- . Software verification and validation was performed according to Guidance for the Content of Premarket Submission for Software Contained in Medical Devices, issued May 11, 2005 and Infusion Pumps Total Life Cycle, issued December 2, 2014. The software is considered a major level of concern. Software testing included code review, static analysis, unit testing, integration testing, and regression testing.
- Cybersecurity verification and information was performed and provided . according to FDA's Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued October 18, 2018, and FDA's draft Guidance Postmarket Management of Cybersecurity in Medical Devices, issued December 28th 2016.
- Interoperability was assessed and testing in accordance with FDA's Guidance on . Design Considerations and Pre-Market Submission Recommendations for Interoperable Medical Device, issued September 6, 2017.

Electrical safety, EMC and essential performance testing was successfully completed according to the following standards and associated methods:

- IEC 60601-1 Medical Electrical Equipment Part 1: General Requirement for . Basic Safety and Essential Performance
- IEC 60601-2-24 Medical equipment Part 2-24: Particular requirements for the . basic safety and essential performance of infusion pumps and controllers
- IEC 60601-1-2 Medical electrical equipment Part 1-2: General requirements for . basic safety and essential performance – Collateral Standard: Electromagnetic Compatibility Requirements and Tests
- . IEC 60601-1-8 Collateral Standard: General Requirements, test and guidance for alarm systems in medical electrical equipment and medical electrical systems.
- AIM 7351731 rev 2: 2017, Medical Electrical Equipment and System . Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows the word "Baxter" in a bold, blue font. The word is slightly slanted to the right, giving it a dynamic appearance. The font is sans-serif, and the letters are closely spaced together.

- IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline . or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
- . UL 2054 2nd Edition Household and Commercial Batteries
- UL 1642 5th Edition Lithium Batteries ●

In addition to the above, and in consideration of IEC 62366-1ed. 1.0 b:2015, Medical devices - Part 1: Application of usability engineering to medical devices as well as FDA guidance, "Applying Human Factors and Usability Engineering to Medical Devices" issued February 3, 2016, Baxter conducted Human Factors evaluations in a simulated environment. The human factors study was conducted with the intended user population, use environment and use scenarios to simulate clinical conditions. Results of human factors studies show the device is suitable for its intended use.

## DISCUSSION OF CLINICAL TESTS:

No clinical testing was performed in support of this premarket notification.

### CONCLUSION:

The Novum IQ Syringe Pump has been verified and validated against design input requirements, user needs, and intended uses. The subject device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K211125](https://fda-staging.innolitics.com/device/K211125)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com